K Number
K183606
Manufacturer
Date Cleared
2019-04-08

(103 days)

Product Code
Regulation Number
892.5050
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The intended use of the device is
• quality assurance of patient specific treatment delivery prior to the treatment in IMRT (including VMAT) and 4DRT (e.g. respiratory gating and tumour tracking).
• quality assurance of the radiation delivery system.

Device Description

The device consists out of matrices of semiconductors embedded in a phantom. These matrices are inserted into the radiation field of a medical linear accelerator. If radiation from a radiotherapy treatment field) hits the semiconductors a signal is created and transferred to a computer where it is analysed and among others compared with the intended dose distribution.

The new device Delta4 Phantom+ MR is equivalent in form and function to the cleared device Delta4 Phantom+, but has been verified as an MR-conditional product.

AI/ML Overview

The provided text describes a medical device called the "Delta4 Phantom+ MR", which is a pre-treatment verification system for quality assurance in radiation therapy. It is a new version of an existing device, the "Delta4 Phantom+", with the key difference being that the new device is MR-conditional.

Here's an analysis of the acceptance criteria and study information:

Acceptance Criteria and Reported Device Performance

The document does not explicitly state numerical acceptance criteria in a table format with corresponding reported device performance values. Instead, it describes equivalence to the predicate device. The core acceptance criterion for the new device is its equivalence in safety, design, and performance to the predicate device ("Delta4 Phantom+"), with the added capability of being MR-conditional.

Specifically, the document states:

  • "The new device is equivalent, in most cases identical with the predicate devices regarding safety, design and performance."
  • "Comparison tests have been performed with the predicate device in clinical and nonclinical situations. Among others pre-treatment verification measurements of numerous treatment plans have been performed with both the new device and the predicate device. The results were compared with each other and it was determined that the results had very good correlation."

Therefore, the acceptance criteria are implicitly that the Delta4 Phantom+ MR performs equivalently to the Delta4 Phantom+ in pre-treatment verification measurements for radiation therapy, specifically in IMRT, VMAT, and 4DRT QA, while additionally being MR-conditional. The reported device performance is that it achieved "very good correlation" with the predicate device.

Acceptance CriterionReported Device Performance
Equivalence in safety, design, and performance to Delta4 Phantom+ (predicate device) in clinical and nonclinical situations for QA measurements.Achieved "very good correlation" with the predicate device in pre-treatment verification measurements of numerous treatment plans.
MR-conditional capability.Successfully designed to be MR-conditional by substituting ferromagnetic parts with non-magnetic materials (ee.g., aluminum).

Study Information

The document provides limited details about a formal study with a defined test set.

2. Sample size used for the test set and the data provenance:

  • Sample Size: The document mentions "numerous treatment plans" were used for pre-treatment verification measurements. A specific numerical sample size is not provided.
  • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). It suggests "clinical and nonclinical situations," implying some real-world or simulated clinical scenarios.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • This information is not provided. The comparison is against the predicate device, not against an expert-established ground truth in the traditional sense of diagnostic AI.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • An adjudication method is not applicable/not provided as the comparison is between two devices' measurement outputs, not expert interpretations of data.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • An MRMC study was not conducted. This device is a measurement and quality assurance tool, not an AI-assisted diagnostic device requiring human reader improvement studies.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • The performance comparison described is essentially a standalone (device-only) comparison between the new device and the predicate device, both operating without human intervention for the measurement aspect directly. The software then analyzes these measurements.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • The "ground truth" for this device's performance comparison is the measurements provided by the predicate device, which is already cleared and considered reliable for QA purposes. The new device's measurements are compared to those of the predicate device.

8. The sample size for the training set:

  • This device is not an AI/machine learning model in the typical sense that would require a "training set." It is a physical measurement device. Therefore, a training set is not applicable/not provided.

9. How the ground truth for the training set was established:

  • As a training set is not applicable, this information is not provided.

{0}------------------------------------------------

ScandiDos Logo
© ScandiDos AB

K183606
Confidential

This document may not be copied without the written consent of ScandiDos AB.

Document Title:FDA 510(k) - Summary - Delta4Phantom+ MR
Subject:Delta4 Phantom+ MR
Document ID:D017 04 074 01
Page002_06_1(002_06_5)
Author:SignatureDate
Thomas Matzen

D017 04 074 01 FDA 510(k) - Summary - Delta4 Phantom+ MR.docx

VersionCommentAuthorDescription
01releasedThomas MatzenFirst Version

1 APPLICANT

510(k) owner's NameScandiDos AB
Street addressDag Hammarskjölds väg 52A
Postal Code, CitySE-75237 Uppsala
CountrySweden
Phone Number+46-18-472 30 30
Fax Number+46-18-10 74 02
Name of official correspondentThomas Matzen
Date the summarywas prepared (yyyy-mm-dd)2018-11-21

2 DEVICE

Trade NameDelta4 Phantom+ MR
Common NamePre-treatment Verification System (for Quality Assurance)
Classification NameMedical charged-particle radiation therapy system,8892.5050. Product code: IYE

3 PREDICATE DEVICE

Predicate DeviceDelta4
510(k) Premarket submission numberK151180

{1}------------------------------------------------

Document Title:FDA 510(k) - Summary - Delta4 Phantom+ MR
Document ID:D017 04 074 01
Page002_06_2(002_06_5)

4 DESCRIPTION

The device consists out of matrices of semiconductors embedded in a phantom. These matrices are inserted into the radiation field of a medical linear accelerator. If radiation firom a radiotherapy treatment field) hits the semiconductors a signal is created and transferred to a computer where it is analysed and among others compared with the intended dose distribution.

The new device Delta4 Phantom+ MR is equivalent in form and function to the cleared device Delta4 Phantom+, but has been verified as an MR-conditional product.

5

COMPARISON WITH THE PREDICATE DEVICE

AreaDelta4 Phantom+(predicate device)Delta4Phantom+ MR(new device)Comparison
IntendedUseThe intended use of the device is• quality assurance of patient specifictreatment delivery prior to thetreatment in IMRT (includingVMAT) and 4DRT (e.g.respiratory gating and tumourtracking).• quality assurance of the radiationdelivery system.Same aspredicatedevice.Both the new and thepredicate device havethe same intended use.
IntendedUserThe intended user is a health careprofessional in the field ofradiotherapy like a medical physicistor a dosimetrist who have beeneducated in the safe use of the device.Same aspredicatedevice.Both the new and thepredicate device havethe same intendedUser.

5.1 Use context

5.1.1 Identical for both devices

The use context regarding

  • patient contact (no patient contact)
  • spatial use (professional healthcare environment)
  • . social use (users are only health care professionals who have been educated in the safe use of the device)
  • Hygienic use (clean but not sterile)
  • general physical use (controlled environmental conditions, no sharp accelerations) o
  • activity use (set up, power on, treatment plan delivery for QA purpose) ●

are identical for the new and the predicate device.

5.1.2 Differences

There are differences regarding physical use context regarding strong magnetic fields: The predicate device is MR unsafe and is never used in strong magnetic fields. The new device is MR conditional and can be used in strong magnetic fields (typically between 0.35T and 1.5T). The

{2}------------------------------------------------

© ScandiDos AB
Document Title:FDA 510(k) - Summary - Delta4 Phantom+ MR
Document ID:D017 04 074 01
Page002_06_3(002 06 5)

new device uses non-ferromagnetic materials (like aluminium) instead of ferromagnetic material (as found in the predicate device).

5.2 Energy

5.2.1 Identical for both devices

  • Neither the new nor the predicate device deliver energy to the patient. .
  • Both the new and the predicate device have the same software functionality regarding power supply operation:
    • o The software recognizes if the device is powered on and displays information about the connected hardware.
    • If the connected device is powered OFF the software informs the user that there o is no working hardware connected.

5.2.2 Differences

The predicate device can be battery powered via a power supply unit. The battery is not MR conditional. Therefore the battery powered option is NOT available for the new device. The new device is always powered via the power supply unit (the same unit as used by the predicate device). Moreover the power cable has been prolonged in the new device so that the power supply unit can be placed outside the strong magnetic field of the MR-linac, e.g. outside the treatment room.

For obvious reasons battery status is not available for the new device in the software.

5.3 Design, Dimensions, Material

5.3.1 Identical for both devices

The parameters regarding

  • Measurement phantom shape and dimension (cylinder, 22cm x 40cm)
  • Phantom length including electronics (about 71cm)
  • Phantom weight (about 27 kg)
  • o Phantom material (PMMA or Plastic Water)
  • Detector placement (two 2D arrays, perpendicular to each other)
  • . Default detector array orientation (sagittal/coronal)

are identical for the new and the predicate device.

5.3.2 Differences

The new device has an additional housing with some electronic parts, which in the predicate device are found within the same housing as the rest of the additional housing in the new device contains components that work more satisfactorily when placed outside the strong magnetic field.

5.4 Performance, technology, communication

5.4.1 Identical for both devices

The following parameter are identical for both the new and the predicate device:

{3}------------------------------------------------

ScandiDos© ScandiDos AB
Document Title:FDA 510(k) - Summary - Delta4 Phantom+ MR
Document ID:D017 04 074 01
Page002_06_4(002_06_5)
  • operating principal (detectors creating a signal in the radiation field of a medical linear ● accelerator, signal is transferred to computer program where it is analysed)
  • o Detector technology (semiconductors)
  • Measured quantity (Dose)
  • . Software analysis (comparing measured with calculated dose)
  • Main analysis parameters (Dose difference, distance to agreement, gamma index)
  • communication between PC and router (ethernet cable)

5.4.2 Differences

  • The predicate device software supports the predicate device. The software of the new device supports in addition also the new device. The new device is treated in the same way as the predicate device in the software. All functionality available for the predicate hardware is now available for the new device, no software functionality had to been added. The use of the software in combination with the new hardware is as efficient and as safe as with the predicate hardware.
  • Communication between measuring hardware and accesspoint/router: The predicate device uses Wi-Fi 802.11n communication. To allow a stable communication between hardware and PC in areas with strong magnetic fields the new device relies on standard ethernet cables instead.

5.5 Standards met

5.5.1 Identical for both devices

Both devices meet IEC 61010-1 and IEC 60601-1-2.

5.5.2 Differences

Additionally the device is conform to ASTM F2052. The predicate has not been designed to be MR conditional. Small modifications (like substituting ferromagnetic parts with non-magnetic parts) made it possible to create the new (MR conditional) device.

5.6 Biocompatibility

5.6.1 Identical for both devices

Neither of the devices is in contact with patient or body fluids.

5.7 Compatibility with the environment and other devices, safety (mechanical, chemical, physical, thermal, radiation)

5.7.1 Identical for both devices

Neither of the devices is an accessory to another medical device.

Both devices work according to their intended use without being connected to other medical devices.

5.7.2 Differences

The predicate device is MR unsafe. The new device is MR conditional. This has been achieved by substituting some ferromagnetic materials in the predicate device with non-ferromagnetic materials.

{4}------------------------------------------------

ScandiDos

Document Title:

5.8 Sterility

5.8.1 Identical for both devices

Neither of the devices is intended to be sterile.

5.9 Performance: comparison with Predicate

Comparison tests have been performed with the predicate device in clinical and nonclinical situations.

Among others pre-treatment verification measurements of numerous treatment plans have been performed with both the new device and the predicate device. The results were compared with each other and it was determined that the results had very good correlation.

SUMMARY 6

The new device is equivalent, in most cases identical with the predicate devices regarding safety, design and performance.

Whereas the predicate device is MR unsafe the new device is MR conditional. This has been achieved by substituting the few existing ferromagnetic parts with non-magnetic parts (like aluminium). This change does in no way effect the safety, usability or effectiveness in a negative way.

{5}------------------------------------------------

0020401

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known)

K183606

Device Name Delta4 Phantom+ MR

Indications for Use (Describe)

The intended use of the device is

· quality assurance of patient specific treatment delivery prior to the treatment in IMRT (including VMAT) and 4DRT (e.g. respiratory gating and tumour tracking).

· quality assurance of the radiation delivery system.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995,

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

{6}------------------------------------------------

April 8, 2019

Image /page/6/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

ScandiDos AB Thomas Matzen Manager QA, RA Dag Hammarskjölds väg 52A Uppsala, 752 37 Se

Re: K183606

Trade/Device Name: Delta4 Phantom+ MR (PMMA), Delta4 Phantom+ MR (PW) Regulation Number: 21 CFR 892.5050 Regulation Name: Medical Charged-Particle Radiation Therapy System Regulatory Class: Class II Product Code: IYE Dated: February 21, 2019 Received: February 25, 2019

Dear Thomas Matzen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{7}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Thalia T. Mills, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

§ 892.5050 Medical charged-particle radiation therapy system.

(a)
Identification. A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II. When intended for use as a quality control system, the film dosimetry system (film scanning system) included as an accessory to the device described in paragraph (a) of this section, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.