K Number
K183501
Date Cleared
2019-03-15

(88 days)

Product Code
Regulation Number
888.3030
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Parcus Twist AP Suture Anchors are indicated for the attachment of soft tissue to bone. This product is intended for the following indications:

ShoulderRotator Cuff Repair, Acromioclavicular Separation Repair, Bankart Lesion Repair, Biceps Tenodesis, Capsular Shift or Capsulolabral Reconstruction, Deltoid Repair, SLAP Lesion Repair.
KneeMedial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Extra Capsular Reconstruction, Iliotibial Band Tenodesis, Patellar Ligament and Tendon Avulsion Repair.
Foot/AnkleLateral Stabilization, Medial Stabilization, Midfoot Reconstruction, Achilles Tendon Repair, Hallux Valgus Reconstruction, Metatarsal Ligament Repair.
ElbowTennis Elbow Repair, Biceps Tendon Reattachment.
Hand/WristScapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction, TFCC.
Device Description

The Parcus Twist AP Suture Anchors consist of fully threaded anchor bodies in 4.5. 5.5, and 6.5mm diameters. The anchor body is comprised of a β-TCP and PLGA biocomposite that has a proven record as a safe and effective absorbable material that has been used in the industry for over 10 years. The anchor bodies feature an internal bar over which strands of UHMWPE suture and/or suture tape are suspended. This allows for free movement of the suture or suture tape in order to assist with the passing and securing of soft tissue. The anchor body material has been designed such that it will maintain the necessary strength to insert successfully and provide the necessary resistance to pull-out long enough for the body to heal and then, over time, break down, be absorbed into the body and eventually replaced by bone. While the UHMWPE suture and suture tape are non-absorbable, this does not create any problems and may remain implanted indefinitely. The presence of the osteoconductive ß-TCP will allow for boney replacement at the insertion site as the anchor is absorbed.

Suture anchors are assembled with suture and/or suture tape provided with or without needles assembled on single use drivers and provided to the end user individually packaged and sterile.

AI/ML Overview

This is a 510(k) premarket notification for the Parcus Twist AP Suture Anchors, a medical device used for the attachment of soft tissue to bone. The document focuses on demonstrating the substantial equivalence of the new device to existing predicate devices. Consequently, it does not detail a study proving the device meets acceptance criteria in the context of clinical performance or diagnostic accuracy. Instead, the "acceptance criteria" discussed relate to biocompatibility and mechanical performance for demonstrating substantial equivalence.

Here's the information extracted from the provided text regarding the testing performed:

1. A table of acceptance criteria and the reported device performance

The document does not provide a specific table of detailed acceptance criteria with numerical values and reported performance for each mechanical test. Instead, it states:

Test TypeAcceptance Criteria / Performance
Pyrogenicity Testing"The results of this testing met the acceptance criteria described in the FDA recognized standard and therefore it was concluded that the Twist AP Suture Anchors do not raise any additional concerns regarding pyrogenicity."
Biocompatibility"Extensive biocompatibility... testing has been conducted and has demonstrated substantial equivalence between the Parcus Twist AP Suture anchors and the predicate devices."
Mechanical Testing"Devices were subjected to biocompatibility and benchtop testing such as pull-out strength, cyclic loading, insertion torque testing, in-vitro degradation, and animal implantation studies. Based on the results of this testing and comparison with the performance of the predicate devices or published acceptance criteria, the Parcus Twist AP Suture Anchors are determined by Parcus Medical to be substantially equivalent to the predicate devices."
In-vitro Degradation"Devices were subjected to... in-vitro degradation... Based on the results of this testing and comparison with the performance of the predicate devices or published acceptance criteria, the Parcus Twist AP Suture Anchors are determined by Parcus Medical to be substantially equivalent to the predicate devices."
Animal Implantation"Devices were subjected to... animal implantation studies. Based on the results of this testing and comparison with the performance of the predicate devices or published acceptance criteria, the Parcus Twist AP Suture Anchors are determined by Parcus Medical to be substantially equivalent to the predicate devices."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document states, "The entire scope of the proposed Twist AP Suture Anchor product family was considered and worst-case scenarios for various concerns were identified." However, it does not specify the sample size for any of the tests (biocompatibility, pull-out strength, cyclic loading, insertion torque testing, in-vitro degradation, or animal implantation studies).
The data provenance is not specified beyond indicating that the data were generated to support a US FDA 510(k) submission. It's implicitly prospective in the sense that the testing was conducted for this specific submission to demonstrate equivalence.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This section is not applicable to this document. The testing performed is for mechanical and biological properties of a physical medical device, not related to expert interpretation of data or images to establish ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This section is not applicable to this document for the same reasons as point 3.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This section is not applicable to this document. This is not an AI/software device or an imaging device requiring human reader interpretation.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This section is not applicable to this document. This is not an algorithm or software device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" in this context refers to established standards and performance of predicate devices:

  • For Pyrogenicity: "acceptance criteria described in the FDA recognized standard"
  • For Biocompatibility, Mechanical, In-vitro Degradation, and Animal Implantation: "comparison with the performance of the predicate devices or published acceptance criteria"

8. The sample size for the training set

This section is not applicable to this document. This is not a device based on machine learning or AI that requires a training set.

9. How the ground truth for the training set was established

This section is not applicable to this document for the same reasons as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the symbol of the Department of Health & Human Services on the left and the FDA acronym followed by "U.S. FOOD & DRUG ADMINISTRATION" in blue text on the right. The FDA part of the logo is in a rectangular blue box.

March 15, 2019

Parcus Medical, LLC Paul Vagts Director of Regulatory Affairs 6423 Parkland Drive Sarasota, Florida 34231

Re: K183501

Trade/Device Name: Parcus Twist AP Suture Anchors Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: MAI Dated: December 13, 2018 Received: December 17, 2018

Dear Mr. Vagts:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm_identifies_combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laurence D. Coyne -S

For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

510(k) Number (if known) K183501

Device Name

Parcus Twist AP Suture Anchors

Indications for Use (Describe)

The Parcus Twist AP Suture Anchors are indicated for the attachment of soft tissue to bone. This product is intended for the following indications:

ShoulderRotator Cuff Repair, Acromioclavicular Separation Repair, Bankart Lesion Repair, Biceps Tenodesis, Capsular Shift or Capsulolabral Reconstruction, Deltoid Repair, SLAP Lesion Repair.
KneeMedial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Extra Capsular Reconstruction, Iliotibial Band Tenodesis, Patellar Ligament and Tendon Avulsion Repair.
Foot/AnkleLateral Stabilization, Medial Stabilization, Midfoot Reconstruction, Achilles Tendon Repair, Hallux Valgus Reconstruction, Metatarsal Ligament Repair.
ElbowTennis Elbow Repair, Biceps Tendon Reattachment.
Hand/WristScapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction, TFCC.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for PARCUS MEDICAL, LLC. The logo is primarily blue, with a red square on the left side. The text "PARCUS" is in large, bold, blue letters, and the text "MEDICAL, LLC." is in smaller, blue letters below the word "PARCUS".

Surgical Innovation | Customer Driven

510(k) Summary

Submitter:Parcus Medical, LLC6423 Parkland DrSarasota, FL 34243
Company Contact:Paul VagtsPhone: (941)755-7965Fax: (941)755-6543
Date Prepared:December 13th, 2018
Device Trade Name:Twist AP Suture Anchor
Common Name:Suture Anchor
Device Class:Class II
Classification Name:Fastener, Fixation, Biodegradable, Soft Tissue 21 CFR888.3030 - Product Code MAI
Predicate Device:The predicate devices are the:• Parcus Medical Twist PEEK Suture Anchor (K120942cleared April 20th, 2012),• DePuy Mitek HEALIX BR Anchor (K073412 clearedJanuary 17th, 2008),• DePuy Mitek Healix Knotless BR (K130917 cleared May24th, 2013), and• DePuy Mitek Lupine BR Anchor (K070925 cleared May2nd, 2007).

Device Description:

The Parcus Twist AP Suture Anchors consist of fully threaded anchor bodies in 4.5. 5.5, and 6.5mm diameters. The anchor body is comprised of a β-TCP and PLGA biocomposite that has a proven record as a safe and effective absorbable material that has been used in the industry for over 10 years. The anchor bodies feature an internal bar over which strands of UHMWPE suture and/or suture tape are suspended. This allows for free movement of the suture or suture tape in order to assist with the passing and securing of soft tissue. The anchor body material has been designed such that it will maintain the necessary strength to insert successfully and provide the necessary resistance to pull-out long enough for the body to heal and then, over time, break down, be absorbed into the body

PARCUSMEDICAL.COM

6423 Parkland Drive, Sarasota, FL 34243

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for "PARCUS MEDICAL, LLC." The word "PARCUS" is in large, blue, sans-serif font. To the left of the word is a red geometric shape. Below the word "PARCUS" is the text "MEDICAL, LLC." in a smaller, blue, sans-serif font.

Surgical Innovation | Customer Driven

and eventually replaced by bone. While the UHMWPE suture and suture tape are non-absorbable, this does not create any problems and may remain implanted indefinitely. The presence of the osteoconductive ß-TCP will allow for boney replacement at the insertion site as the anchor is absorbed.

Suture anchors are assembled with suture and/or suture tape provided with or without needles assembled on single use drivers and provided to the end user individually packaged and sterile.

Intended Use:

The Parcus Twist AP Suture Anchors are indicated for the attachment of soft tissue to bone. This product is intended for the following indications:

ShoulderRotator Cuff Repair, Acromioclavicular Separation Repair,Bankart Lesion Repair, Biceps Tenodesis, Capsular Shiftor Capsulolabral Reconstruction, Deltoid Repair, SLAPLesion Repair.
KneeMedial Collateral Ligament Repair, Lateral Collateral LigamentRepair, Posterior Oblique Ligament Repair, Extra CapsularReconstruction, Iliotibial Band Tenodesis, Patellar Ligament andTendon Avulsion Repair.
Foot/AnkleLateral Stabilization, Medial Stabilization, Midfoot Reconstruction,Achilles Tendon Repair, Hallux Valgus Reconstruction,Metatarsal Ligament Repair.
ElbowTennis Elbow Repair, Biceps Tendon Reattachment.
Hand/WristScapholunate Ligament Reconstruction, Ulnar or RadialCollateral Ligament Reconstruction, TFCC.

Substantial Equivalence Summary:

The Parcus Twist AP Suture Anchors are very similar to the predicate devices in that they are comprised of the same materials (K073412, K130917 and K070925), have identical intended use (K120942, K073412, K130917 and K070925), have identical (K120942) or very similar (K073412, K130917 and K070925) indications for use and utilize similar designs (K120942 and K073412). Pyrogenicity testing was conducted on a representative device from the Twist AP Suture Anchor product family. The results of this testing met the acceptance criteria described in the FDA recognized standard and therefore it was concluded that the Twist AP Suture Anchors do not raise any additional concerns regarding pyrogenicity. Extensive biocompatibility, mechanical and cadaveric testing has been conducted and has demonstrated substantial equivalence between the Parcus Twist AP Suture anchors and the predicate devices.

PARCUSMEDICAL.COM

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for PARCUS MEDICAL, LLC. The logo is primarily blue, with a red square on the left side. The words "PARCUS" and "MEDICAL, LLC." are written in a sans-serif font, with "PARCUS" being larger and on top of "MEDICAL, LLC."

Surgical Innovation | Customer Driven

Summary Performance Data:

The entire scope of the proposed Twist AP Suture Anchor product family was considered and worst-case scenarios for various concerns were identified. Devices were subjected to biocompatibility and benchtop testing such as pull-out strength, cyclic loading, insertion torque testing, in-vitro degradation, and animal implantation studies. Based on the results of this testing and comparison with the performance of the predicate devices or published acceptance criteria, the Parcus Twist AP Suture Anchors are determined by Parcus Medical to be substantially equivalent to the predicate devices.

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.