(37 days)
The VERIFY® V24 Self-Contained Biological Indicator is intended for routine monitoring of the following sterilizer cycles:
· Lumen, Non Lumen, Flexible, Fast and Fast Non Lumen Cycles of the V-PRO® 1, 1 Plus, maX, maX 2, 60 and s 2 Sterilizers
· STERRAD® 100S Sterilizer (default cycle)
• STERRAD® 200 Sterilizer (default cycle)
· Standard and Advanced Cycles of the STERRAD® NX Sterilizers with All Clear Technology
· Express, Flex Scope and Standard Cycles of the STERRAD® 100NX Sterilizers with All Clear technology
The VERIFY V24 BI Challenge Pack is intended for qualification testing of the V PRO 1 Plus, VPRO maX, V-PRO maX 2, V-PRO 60 and V PRO s 2 Low Temperature Sterilization Systems following installation, relocation, malfunctions or major repairs.
The challenge pack is placed in an otherwise empty sterilizer chamber; a hospital-defined challenge load is not included.
The challenge pack is not intended for routine of V-PRO Sterilizers. It has been tested and validated solely for use in periodic testing of the Sterilizers.
The VERIFY V24 Self-Contained Biological Indicator (SCBI) is used by healthcare facilites to monitor the V-PRO® Low Temperature Sterilization Systems. It is designed to accompany medical devices placed in the sterilizer. The user places the VERIFY V24 SCBI into the V-PRO Low Temperature Sterilization System and performs a sterilization cycle. After cycle completion, the SCBI can either be immediately activated or it can be held at room temperature for a maximum of 72 hours (3 days) prior to activation. The SCBI is activated by sealing the vial and rupturing the medium ampoule using the STERIS VERIFY SCBI HP activator. The activated SCBI is incubated at 55-60°C for ≥ 24 hours. The SCBI indicates a pass if the media remains orange and non-turbid. The SCBI indicates a failure of sterilization if the media changes from orange to yellow and/or if the media is turbid.
The VERIFY V24 Biological Indicator Challenge Pack is used by healthcare facilities for qualification testing of the V-PRO Low Temperature Sterilization Systems following installation, relocations, malfunctions or major repairs. The challenge pack is placed in an otherwise empty sterilizer chamber; a hospital defined challenge load is not included. The user places the VERIFY V24 Biological Indicator Challenge Pack into the V-PRO Low Temperature Sterilization System and performs a sterilization cycle. After cycle completion, the VERIFY HPU Chemical Indicator (CI) and the VERIFY V24 Self-Contained Biological Indicator (SCBI) contained in the challenge pack are retrieved. The CI is assessed for a passing color change immediately and the SCBI can either be immediately activated or it can be held at room temperature for a maximum of 72 hours (3 days) prior to activation. The SCBI is activated by sealing the vial and rupturing the medium ampoule using the STERIS VERIFY SCBI HP activator. The activated SCBI is incubated at 55-60°C for ≥ 24 hours. The SCBI indicates a pass if the media remains orange and non-turbid. The SCBI indicates a failure of sterilization if the media changes from orange to yellow and/or if the media is turbid.
Here's a breakdown of the acceptance criteria and study information for the VERIFY V24 Self-Contained Biological Indicator and VERIFY V24 Biological Indicator Challenge Pack, based on the provided document.
1. Table of Acceptance Criteria and Reported Device Performance
It's important to note that the document presents separate acceptance criteria and testing for the "Self-Contained Biological Indicator (SCBI)" and the "Challenge Pack."
For the VERIFY V24 Self-Contained Biological Indicator (SCBI):
Test | Acceptance Criteria | Reported Device Performance |
---|---|---|
½ & ¾ Cycle Performance | All test SCBIs exposed in either the ½ or ¾ cycle display passing results | PASS |
Simulated Use | All processed SCBIs exhibit negative growth results | |
All processed SCBI Label PIs exhibit a “pass” result | ||
All processed CIs exhibit a “pass” result | PASS | |
Verification of Viable Spore Population after exposure to ALLClear | The mean population of V24 SCBI exposed to the ALLClear pre-conditioning was -50% to +300% of the unexposed V24 SCBI population | PASS |
Reduced Incubation Time (RIT) Testing | Meets FDA's requirement of > 97% alignment of the 20-minute results with the conventional incubation time of 7 days | PASS |
Viable spore population | 2.0 - 3.4 x 10^6 spore/SCBI | PASS (stated to contain "greater than 10^6 spores/BI" and "Same" as predicate in Table 5-1) |
D-value | 4-8 sec | PASS (stated "Same" as predicate in Table 5-1) |
Survival Time | 4-30 sec | PASS (stated "Same" as predicate in Table 5-1) |
Kill Time | ≤ 16 min | PASS (stated "Same" as predicate in Table 5-1) |
Hold Time | Performance not affected if incubated within 72 hours of exposure to VHP sterilization | PASS |
For the VERIFY V24 Biological Indicator Challenge Pack:
Test | Acceptance Criteria | Reported Device Performance |
---|---|---|
Comparative Dose Response to Biological Model | The challenge pack shall demonstrate equal or greater resistance as compared to the worst case biological model | PASS |
Simulated Use | All processed SCBIs exhibit negative growth results | |
All processed SCBI Label PIs exhibit a “pass” result | ||
All processed CIs exhibit a “pass” result | PASS |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the numerical sample size used for the test set in any of the studies. It refers to "All test SCBIs" and "All processed SCBIs" without providing a count.
The data provenance is not specified beyond being "nonclinical tests" and "performance testing to demonstrate substantial equivalence to the predicate." There is no mention of country of origin or whether the studies were retrospective or prospective.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not provided in the document. The document describes tests related to the biological and chemical indicators themselves and their performance in sterilization cycles, not a diagnostic or interpretive task that would typically involve human expert adjudication for ground truth.
4. Adjudication Method for the Test Set
This information is not provided as the tests described are objective measures of biological indicator performance (e.g., growth/no growth, color change, spore population) rather than subjective assessments requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance
This is not applicable. The device described (Biological Indicator and Challenge Pack) is a sterilization process indicator, not an AI or imaging device that would typically be involved in human-in-the-loop diagnostic tasks or benefit from a reader study.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
This is not applicable as the device is a physical biological indicator, not an algorithm. Its performance is evaluated through laboratory and simulated use testing.
7. The Type of Ground Truth Used
The ground truth for the performance tests appears to be based on:
- Biological viability: Growth or lack thereof of Geobacillus stearothermophilus spores after exposure to sterilization cycles, as determined by a color change in the growth medium.
- Established physical and chemical properties: D-value, survival time, kill time, and viable spore population are laboratory-measured physical characteristics against established standards.
- Chemical indicator color change: A visible, objective change in color on the indicator.
- Comparative resistance: The challenge pack's resistance compared to a "worst case biological model."
Essentially, the "ground truth" is defined by the expected biological or chemical response to a sterilization process.
8. The Sample Size for the Training Set
This is not applicable. There is no concept of a "training set" for a biological indicator. The tests described are performance and validation tests, not machine learning model training.
9. How the Ground Truth for the Training Set Was Established
This is not applicable as there is no training set for this type of device.
§ 880.2800 Sterilization process indicator.
(a)
Biological sterilization process indicator —(1)Identification. A biological sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization.(2)
Classification. Class II (performance standards).(b)
Physical/chemical sterilization process indicator —(1)Identification. A physical/chemical sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.(2)
Classification. Class II (performance standards).