K Number
K183231

Validate with FDA (Live)

Device Name
SIGNA Premier
Date Cleared
2019-01-18

(59 days)

Product Code
Regulation Number
892.1000
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SIGNA Premier system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the region of interest being imaged, contrast agents may be used.

The images produced by the SIGNA Premier system reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

Device Description

SIGNA Premier is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times, and is designed for improved patient comfort and workflow. The system features a 3.0T superconducting magnet with a 70cm bore size and can image in the sagittal, coronal, axial, oblique, and double oblique planes, using various pulse sequences, imaging techniques and reconstruction algorithms. The system is designed to conform to NEMA DICOM standards (Digital Imaging and Communications in Medicine). The modifications to this system include the AIRx software features, which allows users the flexibility to automate and standardize a number of connected steps required for an MRI examination of the brain.

AI/ML Overview

This document describes a 510(k) premarket notification for the GE Healthcare SIGNA Premier magnetic resonance diagnostic device, specifically focusing on the addition of the AIRx software feature. The AIRx feature automates and standardizes aspects of MRI brain examinations using deep learning algorithms.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of acceptance criteria with corresponding performance metrics for the AIRx feature in terms of diagnostic accuracy (e.g., sensitivity, specificity, accuracy for a specific pathology). Instead, the performance is described in terms of "productivity and consistency benefits" and maintaining the imaging performance of the predicate device.

Acceptance Criteria (Implied)Reported Device Performance
Maintains imaging performance of the predicate device (K171128)"The MR System maintains the same imaging performance results as its predicate device (K171128)." This implies that image quality and diagnostic capabilities are not degraded by the AIRx feature.
Demonstrates productivity benefits for AIRx workflow"Internal scans were conducted as part of validation for AIRx workflow to confirm the productivity... benefits of the proposed feature." (Specific metrics for productivity, such as time saved, are not provided in the summary.)
Demonstrates consistency benefits for AIRx workflow (e.g., in scan prescription automation)"Internal scans were conducted as part of validation for AIRx workflow to confirm the... consistency benefits of the proposed feature." and "The AIRx feature... automates and standardizes a number of connected steps required for an MRI examination of the brain." (Specific metrics for consistency are not provided in the summary.)
Compliance with relevant standards (IEC 62304, ANSI/AAMI 60601-1, IEC 60601-2-33) and quality assurance measures (Risk Analysis, Requirements Reviews, etc.)"The modifications to SIGNA Premier include the AIRx software only feature and complies with the following voluntary standards: • IEC 62304 • ANSI/AAMI 60601-1 • IEC 60601-2-33" and "The following quality assurance measures were applied to the development of the subject device, as they were for the predicate device: • Risk Analysis • Requirements Reviews • Design Reviews • Integration testing (System verification) • Performance testing (Verification) • Simulated use testing (Validation)" The non-clinical tests were "completed with passing results per pass/fail criteria defined in the test cases."
No new questions of safety and effectiveness"These technological differences do not raise any different questions regarding safety and effectiveness." and "performance data demonstrating that the feature is as safe and effective as the predicate, and does not raise different questions of safety and effectiveness."

2. Sample Size Used for the Test Set and the Data Provenance

The document mentions "Internal scans were conducted as part of validation for AIRx workflow." However, it does not specify the sample size (number of patients or scans) used for this clinical testing. The data provenance is also not explicitly stated in terms of country of origin or whether it was retrospective or prospective, beyond being "internal scans."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

The document does not provide any information regarding the number of experts or their qualifications used to establish ground truth for the clinical test set. The focus is on the automation and standardization capabilities of AIRx, and the claim is that the MR system's imaging performance remains the same as the predicate, which presumably means human interpretation of the resulting images is still the primary diagnostic method.

4. Adjudication Method for the Test Set

The document does not describe any adjudication method for the clinical test set.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

The document does not describe an MRMC comparative effectiveness study and therefore does not report any effect size of human readers improving with or without AI assistance. The AIRx feature appears to be a "pre-scan algorithm" designed to improve workflow and consistency before image interpretation, rather than directly assisting in the diagnostic interpretation by a human reader in an MRMC setting.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The document states, "The AIRx feature is a pre-scan algorithm," and its purpose is to "automate and standardize a number of connected steps required for an MRI examination of the brain." This suggests that the algorithm's performance is in setting up the scan, and the final interpretation is still by a "trained physician." While "bench testing" and "non-clinical tests" were performed for the software, it's not clear if a standalone performance evaluation in terms of diagnostic accuracy (e.g., for detecting specific pathologies) was conducted without a human interpreting the resulting images. The statement "The MR System maintains the same imaging performance results as its predicate device (K171128)" implies that the imaging capability is maintained, and not that AIRx performs diagnosis standalone.

7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

The document does not explicitly state the type of ground truth used for the "internal scans" validating the AIRx workflow. Given the focus on "productivity and consistency benefits" in scan prescription, the ground truth might relate to objective measures of scan parameter consistency, successful automation of steps, or adherence to a predefined scan protocol, rather than definitive diagnostic outcomes like pathology.

8. The Sample Size for the Training Set

The document states, "AIRx was developed using deep learning algorithms." However, it does not provide the sample size used for the training set for these deep learning algorithms.

9. How the Ground Truth for the Training Set Was Established

The document does not describe how the ground truth for the training set was established for the deep learning algorithms used in AIRx.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo features a stylized human figure. The FDA part of the logo has the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

GE Medical Systems, LLC % Mr. Brian R. Zielski Regulatory Affairs Leader 3200 N. Grandview Blvd WAUKESHA WI 53188

January 18, 2019

Re: K183231

Trade/Device Name: SIGNA Premier Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH, LNI and MOS Dated: November 19, 2018 Received: November 20, 2018

Dear Mr. Zielski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael D. O'Hara For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K183231

Device Name SIGNA Premier

Indications for Use (Describe)

The SIGNA Premier system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the region of interest being imaged, contrast agents may be used.

The images produced by the SIGNA Premier system reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 81 CFR 201 Subpart D) ☐ Over-The-Counter Use (81 CFR 201 Subpart C)

Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a blue circle. There are small, rounded shapes at the top and bottom of the circle, giving it a slightly decorative appearance.

GE Healthcare 510(k) Premarket Notification Submission K183231

510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

Date:November 19, 2018
Submitter:GE Medical Systems, LLC (GE Healthcare)3200 N. Grandview Blvd.,Waukesha, WI 53188USA
Primary Contact Person:Brian R. ZielskiRegulatory Affairs LeaderGE HealthcarePhone: 262-521-6609
Secondary Contact Person:James McMahonSenior Director, Regulatory AffairsGE HealthcarePhone: 508-382-2858
Device Trade Name:SIGNA Premier
Common/Usual Name:Magnetic Resonance Diagnostic Device
Classification Names:Magnetic Resonance Diagnostic Device per 21 CFR892.1000
Product Code:LNH, LNI, MOS
Predicate Device(s):SIGNA Premier (K171128)
Device Description:SIGNA Premier is a whole body magnetic resonancescanner designed to support high resolution, high signal-to-noise ratio, and short scan times, and is designed forimproved patient comfort and workflow. The systemfeatures a 3.0T superconducting magnet with a 70cm boresize and can image in the sagittal, coronal, axial, oblique,and double oblique planes, using various pulse sequences,imaging techniques and reconstruction algorithms. Thesystem is designed to conform to NEMA DICOM
standards (Digital Imaging and Communications inMedicine).The modifications to this system include the AIRxsoftware features, which allows users the flexibility toautomate and standardize a number of connected stepsrequired for an MRI examination of the brain.
Indications for UseThe SIGNA Premier system is a whole body magneticresonance scanner designed to support high resolution,high signal-to-noise ratio, and short scan times. It isindicated for use as a diagnostic imaging device toproduce axial, sagittal, coronal, and oblique images,spectroscopic images, parametric maps, and/or spectra,dynamic images of the structures and/or functions of theentire body, including, but not limited to, head, neck,TMJ, spine, breast, heart, abdomen, pelvis, joints,prostate, blood vessels, and musculoskeletal regions of thebody. Depending on the region of interest being imaged,contrast agents may be used.The images produced by the SIGNA Premier systemreflect the spatial distribution or molecular environment ofnuclei exhibiting magnetic resonance. These imagesand/or spectra when interpreted by a trained physicianyield information that may assist in diagnosis.
Comparison of Indicationsfor UseThe indications for use statement and intended use areidentical to the predicate device, in accordance with theFDA's guidance document "The 510(k) Program:Evaluating Substantial Equivalence in PremarketNotifications [510(k)]", dated 28 July 2014.
Technology:The SIGNA Premier with the proposed software featureemploy the same fundamental technology as the predicatedevice.The SIGNA Premier has been modified to include theAIRx feature, which automates and standardizes a numberof connected steps required for an MRI examination of thebrain. AIRx was developed using deep learningalgorithms.
These technological differences do not raise any differentquestions regarding safety and effectiveness. Both devicesmust allow for an effective method to setup an appropriatescan prescription. The performance data described in thissubmission include results of both bench testing andclinical testing that show the performance of the SIGNAPremier compared to the predicate device.
Determination ofSubstantial Equivalence:Summary of Non-Clinical Tests:The modifications to SIGNA Premier include the AIRxsoftware only feature and complies with the followingvoluntary standards:• IEC 62304• ANSI/AAMI 60601-1• IEC 60601-2-33The following quality assurance measures were applied tothe development of the subject device, as they were forthe predicate device:• Risk Analysis• Requirements Reviews• Design Reviews• Integration testing (System verification)• Performance testing (Verification)• Simulated use testing (Validation)The non-clinical tests have been summarized in theverification and validation testing for AIRx. The testingwas completed with passing results per pass/fail criteriadefined in the test cases. This supports substantialequivalence to its predicate because it was also developedunder quality assurance Design Controls. In addition, thesoftware complies with the same applicable Standards.
Summary of Clinical Tests:Internal scans were conducted as part of validation forAIRx workflow to confirm the productivity andconsistency benefits of the proposed feature. The AIRxfeature is a pre-scan algorithm, and the MR System
maintains the same imaging performance results as itspredicate device (K171128).
Conclusion:The SIGNA Premier with the modified software featurehas the same intended use as the predicate. This 510(k)submission includes information on the technologicalcharacteristics of the proposed software feature, as well asperformance data demonstrating that the feature is as safeand effective as the predicate, and does not raise differentquestions of safety and effectiveness.In conclusion, GE Healthcare considers the SIGNAPremier to be as safe, as effective, and performance issubstantially equivalent to the predicate devices.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of a blue circle with a stylized, cursive "GE" in white inside. The "GE" letters are interconnected and have a flowing, elegant design. The blue circle has a slightly raised, rounded edge, giving it a three-dimensional appearance.

GE Healthcare 510(k) Premarket Notification Submission

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized, cursive font, enclosed within a blue circle. The blue color is a vibrant, medium shade, and the cursive letters are white, creating a contrast that makes the logo easily recognizable.

GE Healthcare 510(k) Premarket Notification Submission

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized, cursive font, enclosed within a circular frame. The frame has decorative swirls or flourishes around the letters, and the entire logo is in a light blue color.

GE Healthcare 510(k) Premarket Notification Submission

§ 892.1000 Magnetic resonance diagnostic device.

(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.