K Number
K182595
Date Cleared
2018-12-18

(89 days)

Product Code
Regulation Number
892.1550
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ultrasound diagnostic system and probes are designed to obtain ultrasound images and analyze body fluids. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.

Device Description

The HERA W10 is general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound dasta and to display the data as B-mode. M-mode, Pulsed wave (PW) Doppler, Continuous wave (CW) Doppler, Color Doppler, Tissue Doppler Imaging (TDI), Tissue Doppler Wave (TDW), Power Amplitude Doppler, Pulse Inversion Harmonic Imaging (S-Harmonic), Directional Power Doppler (S-Flow), Color M-Mode, 3D Imaging Mode, 4D Imaging Mode, Elastoscan+ Mode, Tissue Harmonic Imaging, MV-Flow Mode or as a combination of these modes. The HERA W10 also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The HERA W10 has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

AI/ML Overview

I am sorry, but the provided text does not contain sufficient information to answer your request. The document is a 510(k) summary for the HERA W10 Diagnostic Ultrasound System, which primarily focuses on establishing substantial equivalence to predicate devices and detailing the device's intended uses and compliance with safety standards.

Specifically, the text does not include:

  • A table of acceptance criteria and reported device performance.
  • Details about sample sizes used for test sets, data provenance, number or qualifications of experts for ground truth, or adjudication methods for test sets.
  • Information about a multi-reader multi-case (MRMC) comparative effectiveness study or its effect size.
  • Information about a standalone (algorithm only) performance study.
  • The type of ground truth used (expert consensus, pathology, outcomes data, etc.).
  • The sample size for the training set or how its ground truth was established.

In fact, section 11 explicitly states: "The subject of this premarket submission, HERA W10, did not require clinical studies to support substantial equivalence." This indicates that detailed performance studies as you've requested were not part of this specific FDA submission.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 18, 2018

Samsung Medison Co., Ltd. Ji Yea Lee Regulatory Affairs Specialist 3366, Hanseo-ro, Nam-myeon, Hongcheon-gun, Gangwon-do 25108 REPUBLIC OF KOREA

Re: K182595

Trade/Device Name: HERA W10 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic Pulsed Doppler Imaging System Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: November 23, 2018 Received: November 26, 2018

Dear Ji Yea Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Hole 2. Nils

for Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K182595

Device Name HERA W10 Diagnostic Ultrasound System

Indications for Use (Describe)

The ultrasound diagnostic system and probes are designed to obtain ultrasound images and analyze body fluids. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

Form Approved: OMB No. 0910-0120

Expiration Date: 06/30/2020

See PRA Statement below.

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995,

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white, set against a blue oval background. The font is sans-serif and bold, giving the logo a clean and modern look. The blue oval is slightly tilted, adding a dynamic element to the design.

510(k) No.:

Device Name: HERA W10 Diagnostic Ultrasound System Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)NNNNNNote 1Notes 2, 3, 4, 7, 8, 9, 11, 14
Abdominal (See Note 10)NNNNNNote 1Notes 2, 4, 5, 6, 7, 8, 9, 11, 12, 14
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricNNNNNNote 1Notes 2, 7, 8, 9, 11
Small Organ (See Note 5)NNNNNNote 1Notes 2, 5, 6, 7, 8, 9, 11, 12, 14
Neonatal CephalicNNNNNNote 1Notes 8, 9, 11
Adult CephalicNNNNNNote 1Notes 7
Trans-rectalNNNNNNote 1Notes 2, 7, 8, 9, 11, 12
Trans-vaginalNNNNNNote 1Notes 2, 7, 8, 9, 11, 12
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)NNNNNNote 1Notes 2, 5, 6, 7, 8, 9, 11, 12, 14
Musculo-skel. (Superfic.)NNNNNNote 1Notes 2, 5, 6, 7, 8, 9, 11, 12, 14
Intra-luminal
Other (See Note 13)NNNNNNote 1Notes 2, 7, 8, 9, 12
Cardiac AdultNNNNNNote 1Notes 4, 7
CardiacCardiac PediatricNNNNNNote 1Notes 4, 7
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselNNNNNNote 1Notes 2, 5, 6, 7, 8, 9, 11, 14
Other (spec.)

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+PPI, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white, set against a blue oval background. The font is sans-serif and appears to be bolded.

510(k) No.:

Device Name: L3-12A for use with HERA W10

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 14
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricNNNNNote 1Note 2, 5, 6, 7, 8, 9, 11
Small Organ (See Note 5)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 12, 14
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 12, 14
Musculo-skel. (Superfic.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 12, 14
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 14
Other (spec.)

N=new indication; P= previously cleared by FDA K173204, K173513; E=added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+CW, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white, set against a blue oval background. The font is sans-serif and bold, and the letters are evenly spaced.

510(k) No.:

Device Name: LA2-9A for use with HERA W10

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal ImagingLaparoscopic
&OtherPediatricNNNNNote 1Note 2, 5, 6, 7, 8, 9, 11
Small Organ (See Note 5)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11,12
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11,12
Musculo-skel. (Superfic.)PPPPNote 1Note 2, 5, 6, 7, 8, 9, 11,12
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vesselPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11
VesselOther (spec.)

N= new indication; P= previously cleared by FDA K173204, K173513; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+CW, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 14: S-Fusion

Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109)

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white, set against a blue oval background. The font is sans-serif and bold, and the oval is tilted slightly to the right.

510(k) No.:

Device Name: LA4-18B for use with HERA W10

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)PPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 12, 14
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Note 2,5,6,7,8,9,11,12,14
Musculo-skel. (Superfic.)PPPPNote 1Note 2,5,6,7,8,9,11,12,14
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vesselPPPPPNote 1Note 2, 5, 6, 7, 8, 9, 11, 14
VesselOther (spec.)

N= new indication; P= previously cleared by FDA K173204, K173513; E=added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CV, B+C+PW, B+DD+PW, B+DPD+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis
  • Note 11: Panoramic imaging
  • Note 12: ElastoScan
  • Note 13: Includes Urology/Prostate
  • Note 14: S-Fusion

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white letters on a blue background. The background is an oval shape with rounded corners.

510(k) No.:

Device Name: CA1-7A for use with HERA W10

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPNote 1Notes 2,4,7,8,9,11,14
Abdominal (See Note 10)PPPPNote 1Notes 2,6,7,8,9,11,14
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote 1Notes 2,6,7,8,9,11,14
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPNote 1Notes 2,6,7,8,9,11,14
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)PPPPNote 1Notes 2,6,7,8,9,11,14
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPNote 1Notes 2,6,7,8,9,11,14
Other (spec.)

N= new indication; P= previously cleared by FDA K173204, K173513; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CV, B+C+PW, B+DD+PW, B+DPD+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white letters on a blue background. The background is an oval shape with rounded corners.

510(k) No.:

Device Name: CA2-9A for use with HERA W10

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal/Obstetrics (See Note 3)PPPPNote 1Notes 2,4,7,8,9,11,14
Abdominal (See Note 10)PPPPNote 1Notes 2,6,7,8,9,11,14
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K173513; E= added under Appendix E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CV, B+C+PW, B+PD+PW, B+DPD+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white, set against a blue oval background. The font is sans-serif and bold, and the letters are evenly spaced.

510(k) No.:

Device Name: CA3-10A for use with HERA W10

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Notes 2, 3, 7, 8, 9, 11, 14
Abdominal (See Note 10)PPPPPNote 1Notes 2, 7, 8, 9, 11, 14
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPPNote 1Notes 2, 7, 8, 9, 11, 14
Small Organ (See Note 5)
Neonatal CephalicNNNNNNote 1Notes 2, 7, 8, 9, 11, 14
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)PPPPPNote 1Notes 2, 7, 8, 9, 11, 14
Musculo-skel. (Superfic.)PPPPPNote 1Notes 2, 7, 8, 9, 11, 14
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPPNote 1Notes 2, 7, 8, 9, 11, 14
Other (spec.)

N= new indication; P= previously cleared by FDA K173204, K173513; E= added under Appendix E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+ID, B+CV, B+C+PW, B+PD+PW, B+PPH+N, B+PPH+N, B+ID+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

  • Note 4: Color M-mode
    Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

  • Note 6: Abdominal organs and peripheral vessel

  • Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging

  • Note 9: MultiVision (Spatial Compound Imaging)

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

  • Note 14: S-Fusion

Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109)

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white letters on a blue background. The background is an oval shape with rounded corners.

510(k) No.:

Device Name: CF4-9 for use with HERA W10

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPNote 1Notes 7, 8, 9, 11
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPNote 1Notes 7, 8, 9, 11
Small Organ (See Note 5)
Neonatal CephalicPPPPNote 1Notes 7, 8, 9, 11
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vesselPPPPPNote 1Notes 8, 9, 11
Other (spec.)

N= new indication; P= previously cleared by FDA K173204, K173513; E=added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+PPI, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPI+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white letters on a blue background. The background is an oval shape with rounded corners.

510(k) No.:

Device Name: E3-12A for use with HERA W10

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Notes 2, 7, 8, 9
Abdominal (See Note 10)PPPPPNote 1Notes 2, 7, 8, 9, 12
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPNote 1Notes 2, 7, 8, 9, 12, 14
Trans-vaginalPPPPPNote 1Notes 2, 7, 8, 9, 12
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)PPPPPNote 1Notes 2, 7, 8, 9, 12, 14
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K173204, K173513; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CW, B+C+PW, B+PD+PW, B+PPI+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsv
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging
    Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

{12}------------------------------------------------

Image /page/12/Picture/1 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white letters on a blue background. The background is an oval shape with rounded corners.

510(k) No.:

Device Name: EA2-11B for use with HERA W10

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Notes 2, 7, 8, 9
Abdominal (See Note 10)PPPPPNote 1Notes 2, 7, 8, 9, 12
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPNote 1Notes 2, 7, 8, 9, 12, 14
Trans-vaginalPPPPPNote 1Notes 2, 7, 8, 9, 12
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)PPPPPNote 1Notes 2, 7, 8, 9, 12, 14
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K173204, K173513; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 2: Includes imaging for guidance of biopsy

{13}------------------------------------------------

Image /page/13/Picture/1 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white, set against a blue oval background. The font is sans-serif and bold, and the letters are evenly spaced. The blue oval is slightly tilted.

510(k) No.:

Device Name: VR5-9 for use with HERA W10

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Notes 2, 7, 8, 9
Abdominal (See Note 10)PPPPPNote 1Notes 2, 7, 8, 9, 12
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPNote 1Notes 2, 7, 8, 9, 12
Trans-vaginalPPPPPNote 1Notes 2, 7, 8, 9, 12
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)PPPPPNote 1Notes 2, 7, 8, 9, 12,
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K173513; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CW, B+C+PW, B+PD+PW, B+PD+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

{14}------------------------------------------------

Image /page/14/Picture/1 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white letters on a blue background. The background is an oval shape with rounded corners.

510(k) No.:

Device Name: PA4-12B for use with HERA W10

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Fetal Imaging&OtherAbdominal (See Note 10)PPPPPNote 1Note 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
PediatricPPPPPNote 1Note 7
Small Organ (See Note 5)
Neonatal CephalicPPPPPNote 1Note 7
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
CardiacCardiac AdultPPPPPNote 1Note 4, 7
Cardiac PediatricPPPPPNote 1Note 4, 7
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K173204, K173513; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+PPI, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging
    Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate

{15}------------------------------------------------

Image /page/15/Picture/1 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white letters on a blue background. The background is an oval shape with rounded corners.

510(k) No.:

Device Name: PA3-8B for use with HERA W10

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPPNote 1Note 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
PediatricPPPPPNote 1Note 7
Small Organ (See Note 5)
Neonatal CephalicPPPPPNote 1Note 7
Adult CephalicPPPPPNote 1Note 7
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
CardiacCardiac AdultPPPPPNote 1Note 4, 7
Cardiac PediatricPPPPPNote 1Note 4, 7
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K173204, K173513; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+PPI, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

{16}------------------------------------------------

Image /page/16/Picture/1 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white, bold, sans-serif font. The word is placed inside a blue, rounded oval shape that is tilted slightly to the right. The blue oval provides a background for the brand name, making it stand out.

510(k) No.:

Device Name: PM1-6A for use with HERA W10

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)
Abdominal (See Note 10)PPPPPNote 1Note 7
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal Imaging& OtherLaparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult CephalicPPPPPNote 1Note 7
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
Cardiac AdultPPPPPNote 1Note 4, 7
CardiacCardiac PediatricPPPPPNote 1Note 4, 7
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K173204, K173513; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+PPI, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

{17}------------------------------------------------

Image /page/17/Picture/1 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white letters on a blue background. The background is an oval shape with rounded corners.

510(k) No.:

Device Name: CV1-8A for use with HERA W10

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
General(Track I only)Specific(Tracks I & III)BMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
OphthalmicOphthalmic
Fetal Imaging& OtherFetal/Obstetrics (See Note 3)PPPPPNote 1Note 2, 4, 7, 8, 9, 11
Abdominal (See Note 10)PPPPPNote 1Note 2, 4, 7, 8, 9
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Laparoscopic
Pediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)
CardiacCardiac Adult
Cardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralVesselPeripheral vessel
Other (spec.)

N= new indication; P= previously cleared by FDA K173204, K173513; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+PPI, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

{18}------------------------------------------------

Image /page/18/Picture/1 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white letters on a blue background. The background is an oval shape with rounded corners.

510(k) No.:

Device Name: EV3-10B for use with HERA W10

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical ApplicationMode of Operation (*includes simultaneous B-mode)
GeneralSpecificBMPWDCWDColorDoppler*Combined*(Spec.)Other(Spec.)
(Track I only)(Tracks I & III)
OphthalmicOphthalmic
Fetal/Obstetrics (See Note 3)PPPPPNote 1Note 2, 7, 8, 9, 11, 12
Abdominal (See Note 10)PPPPPNote 1Note 2, 7, 8, 9, 11, 12
Intra-operative (See Note 6)
Intra-operative (Neuro.)
Fetal ImagingLaparoscopic
&OtherPediatric
Small Organ (See Note 5)
Neonatal Cephalic
Adult Cephalic
Trans-rectalPPPPPNote 1Note 2, 7, 8, 9, 11, 12
Trans-vaginalPPPPPNote 1Note 2, 7, 8, 9, 11, 12
Trans-urethral
Trans-esoph. (non-Cardiac)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Intra-luminal
Other (See Note 13)PPPPPNote 1Note 2, 7, 8, 9, 11, 12
Cardiac Adult
CardiacCardiac Pediatric
Trans-esophageal (Cardiac)
Other (spec.)
PeripheralPeripheral vessel
VesselOther (spec.)

N= new indication; P= previously cleared by FDA K173513; E= added under Appendix E

Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+PPI, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

  • Note 2: Includes imaging for guidance of biopsy
    Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

  • Note 8: 3D imaging
    Note 9: Spatial Compound Imaging

  • Note 10: Includes Renal, Gynecology/Pelvis

  • Note 11: Panoramic imaging

  • Note 12: ElastoScan

  • Note 13: Includes Urology/Prostate Note 14: S-Fusion

{19}------------------------------------------------

6. 510(K) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

  • Date Prepared November 23, 2018 1.
    K182595

    1. Manufacturer SAMSUNG MEDISON CO., LTD. 3366, Hanseo-ro, Nam-myeon, Hongcheon-gun, Gangwon-do 25108, REPUBLIC OF KOREA
    1. Primary Contact Person Ji Yea Lee Regulatory Affairs Specialist Phone: +82.2.2194.1594 Fax: +82. 31.8017.9573 Email: jiyea722.lee@samsungmedison.com
  • Proposed Device 4.

    • Common/Usual Name: Diagnostic Ultrasound System and Accessories
    • Proprietary Name: HERA W10 Diagnostic Ultrasound System
    • Common Name: Diagnostic Ultrasound System
    • Classification Names: system, imaging, pulsed doppler, ultrasonic
    • Product Code: IYN, IYO, ITX
    • Regulation: 892.1550, 892.1560, 892.1570
  • Primary Predicate Device 5.

  • RS85 Diagnostic Ultrasound System (K173204) -

Reference Device

  • WS80A Diagnostic Ultrasound System (K173513) -
  • Voluson E10 (K172342) ।
  • Device Description 6.

The HERA W10 is general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound dasta and to display the data as B-mode. M-mode, Pulsed wave (PW) Doppler, Continuous wave (CW) Doppler, Color Doppler, Tissue Doppler Imaging (TDI), Tissue Doppler Wave (TDW), Power Amplitude Doppler, Pulse Inversion Harmonic Imaging (S-Harmonic), Directional Power Doppler (S-Flow), Color M-Mode, 3D Imaging Mode, 4D Imaging Mode, Elastoscan+ Mode, Tissue Harmonic Imaging, MV-Flow Mode or as a combination of these modes. The HERA W10 also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The HERA W10 has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

    1. Intended Uses
      The ultrasound diagnostic system and probes are designed to obtain ultrasound images and analyze body fluids.

The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.

8. Technology

  • The HERA W10 employs the same fundamental scientific technology as its predicate device(s).
  • Determination of Substantial Equivalence 9.

{20}------------------------------------------------

The proposed HERA W10 is substantially equivalent to the predicate devices with regards to intended use, imaging capabilities, technological characteristics and safety and effectiveness.

  • The systems are all intended for diagnostic ultrasound imaging and fluid flow analysis. .
  • The proposed HERA W10 and predicate RS85 system have the same clinical intended use. .
  • The proposed HERA W10 and predicate RS85 system have the same imaging modes, and ● mode of operation.
  • . The proposed HERA W10 system transducers are cleared in the predicate RS85 and WS80A.
  • The system indications for use are equivalent to the predicate devices. .
  • . The system are manufactured with materials with have been evaluated and found to be safe for the intended use of the device.
  • The systems have acoustic power levels which are below the applicable FDA limits. ●
  • The proposed HERA W10 and predicates RS85 and WS80A Systems have similar capacity . in terms of performing measurements, capturing digital images, reviewing and reporting studies.
  • The HERA W10 and predicate systems have been designed in compliance with approved ● electrical and physical safety standards.
  • The proposed HERA W10 system adds an improved version of cleared software feature MPI . and change name to MPI+.
  • . The proposed HERA W10 system adds a transducer recognition tool called QuickPrep.
  • The proposed HERA W10 system adds a simple image transferring software feature called ● SonoSync which is not intended for diagnosis.
  • The proposed HERA W10 system adds three transducers CV1-8A, EV3-10B, LA4-18B . supporting the cleared software feature HQ Vision and adds a supporting preset OB/GYN.
  • . The proposed HERA W10 system adds two transducers CV1-8A, CA1-7A supporting the cleared software feature MV-Flow and adds a supporting preset OB.
  • The proposed HERA W10 system adds two new features LumiFlow, and ShadowHDR. .
  • The proposed HERA W10 system adds pediatric application to cleared L3-12A and LA2-9A . due to Pediatric ABD and Pediatric Hip presets. The presets have already included in the L3-12A and LA2-9A of the predicate RS85 and WS80A.
  • . The proposed HERA W10 system adds Neonatal Cephalic application to cleared CA3-10A due to Neonatal preset. The preset has already included in the CA3-10A of the predicate RS85.
    1. Summary of Non-Clinical Test

The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electromagnetic and mechanical safety, and has been found to conform with applicable medical device safety standards. The HERA W10 and its applications comply with voluntary standards.

{21}------------------------------------------------

Reference No.Title
IEC 60601-1ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012Medical Electrical Equipment - Part 1: General Requirements for basic safety and essential performance.
IEC 60601-1-2IEC60601-1-2: 2014(4th Edition), Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - EMC
IEC 60601-2-37IEC60601-2-37:2007 + A1:2015, Particular requirements for the safety of ultrasonic medical diagnostic and monitoring equipment
ISO10993-1ISO 10993-1, Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process.
ISO14971ISO 14971:2007, Medical devices - Application of risk management to medical devices
NEMA UD 2-2004NEMA UD 2-2004 (R2009)Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment Revision 3
NEMA UD 3-2004NEMA UD 3-2004 (R2009)Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment, Revision 2
    1. Summary of Clinical Tests
      The subject of this premarket submission, HERA W10, did not require clinical studies to support substantial equivalence.
    1. Conclusion
      Intended uses and other key features are consistent with traditional climical practices and FDA guidelines. The design, development and quality process of the manufacturer confirms with 21 CFR 820 and ISO 13485. The device is designed to conform to applicable medical device safety standards and compliance. Therefore, SAMSUNG MEDISON CO., LTD. considers the HERA W10 to be as safe, as effective, and performance is substantially equivalent to the predicate devices.

END of 510(K) Summary

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.