K Number
K182587
Date Cleared
2018-11-29

(70 days)

Product Code
Regulation Number
878.4400
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The electrosurgical generator, in conjunction with electrosurgical accessories and ancillary equipment, is intended for cutting and coagulating tissue in endoscopic surgery.

Device Description

The ESG-150 is a reusable, non-sterile electrosurgical generator that features different monopolar and bipolar cutting and coagulation modes. The maximum output power is 120 W. It is intended for cutting and coagulation of tissue in endoscopic surgery in conjunction with electrosurgical accessories and ancillary equipment. The front panel features a touch screen GUI and all settings are done via the touchscreen. The ESG 150 has three contact quality indicators (CQM); green, red and grey which indicate the connection status of the neutral electrode. The generator can store selectable settings and allows the user to adjust the volume of touch screen buttons and activation. In addition to the GUI, the ESG-150 has a bipolar socket, a monopolar socket (B Type), and a neutral electrode socket. The power switch turns the generator on and off. The rear panel includes the volume control, a ventilation hole, the equipotential bonding point, the AC power socket, and a footswitch socket.

AI/ML Overview

The provided text describes the regulatory clearance of the Olympus Winter & Ibe GmbH Electrosurgical Generator ESG-150. Based on the document, here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a discrete table of performance acceptance criteria with numerical targets. Instead, the acceptance criteria are implicitly tied to compliance with recognized standards and demonstrating substantial equivalence to the predicate device (Olympus ESG-400), particularly regarding safety and tissue effects. The device is deemed to meet these criteria if it performs comparably to the predicate device in relevant aspects and adheres to established safety standards.

Acceptance Criteria CategoryReported Device Performance
Premarket Notification (510(k)) Guidance Compliance"The data was prepared in accordance with the FDA guidance. Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery Guidance for Industry and Food and Drug Administration Staff, issued on August 15, 2016. The guidance was followed for all relevant sections."
Biocompatibility"Biocompatibility testing according to ISO 10993 is not required." (because components do not come directly or indirectly into patient contact)
Electrical Safety and EMC"The design of the ESG-150 complies with recognized standards as listed in section 2.8.8 of this summary." (e.g., AAMI/ANSI ES 60601-1, IEC 60601-1-2, ANSI AAMI IEC 60601-2-2). "The FDA guidance 'Information to Support a Claim of Electromagnetic Compatibility (EMC) of Electrically-Powered Medical Devices', CDRH July 11, 2016 has been followed."
Thermal Safety"The design of the ESG-150 complies with recognized standards as listed in section 2.8.8 of this summary."
Software Validation"The software validation activities were performed in accordance with the FDA Guidance, 'Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices' (May 11, 2005), and in accordance with the FDA Guidance: 'Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery' (Aug. 15, 2016). This device type software presents a 'Major Level of Concern'." Compliance with IEC 62304.
Functional Performance (Bench Testing)"Performance and validation tests incorporated the same range of waveform outputs and power levels." "During the validation testing the waveforms and test results were compared directly between the subject and predicate device." "Basic safety and performance testing was performed in accordance with IEC standards." "Preclinical: Evidence obtained from preclinical simulated use studies demonstrate that the device performs substantially equivalent to the predicate device in relevant aspects associated with usability, tissue effects, and thermal effects." "For simulated use testing, three clinically relevant tissue types were evaluated in all applicable modes: porcine muscle, liver, and kidney." "The thermal damage of the tissue due to the HF current was measured in terms of size (length, width and depth) of the thermal coagulation zone." "Furthermore, each test included the minimum, default and maximum intensity settings. All modes were compared to the predicate." "Testing confirmed that comparable tissue effects could be achieved for applicable modes of operation with three tissue types."
Usability"Usability and user interface was also assessed according to the risk management plan. The assessment was based on Olympus predecessor product. Use-related hazardous situations were assessed and risk mitigation measures in terms of usability design for safety were defined. The residual risk was evaluated as acceptable." Compliance with IEC 60601-1-6 and IEC 62366-1.
Risk Analysis"Risk analysis was carried out in accordance with established internal acceptance criteria based on ISO 14971:2007."

2. Sample Size Used for the Test Set and the Data Provenance

  • Test Set Sample Size: The document specifies that for simulated use testing, "three clinically relevant tissue types were evaluated in all applicable modes: porcine muscle, liver, and kidney." It doesn't provide a specific numerical count for samples within each tissue type for "length, width and depth" measurements.
  • Data Provenance: The studies were preclinical simulated use studies using "porcine muscle, liver, and kidney" tissue. This indicates an in vitro or ex vivo setting, not human data. The country of origin of the data is not explicitly stated, but the manufacturer is based in Germany.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

The document does not mention the use of experts to establish ground truth for the test set in the traditional sense of clinical assessment. The "ground truth" for the performance testing was based on direct measurements of tissue effects (thermal coagulation zone size) and comparison to the predicate device's performance according to established scientific and engineering principles.

4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

No adjudication method involving multiple human reviewers is mentioned or applicable, as the performance evaluation was based on objective measurements in preclinical bench and simulated use studies.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, If So, What was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

No MRMC study was conducted. This device is an electrosurgical generator, not an AI-powered diagnostic tool that assists human readers.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was Done

This question is not applicable. The ESG-150 is a medical device, not an algorithm. Its performance is evaluated through its functional output and interaction with tissue, which is inherently standalone in its operation (though intended for human use with instruments).

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc)

For the preclinical simulated use studies, the "ground truth" was based on direct quantitative and qualitative assessment of tissue effects, specifically "thermal damage... measured in terms of size (length, width and depth) of the thermal coagulation zone" on animal tissue (porcine muscle, liver, and kidney). This is an objective, measurable physical effect, not subjective expert consensus or pathology on human samples. For other tests (electrical safety, EMC), the ground truth was conformance to recognized international standards.

8. The Sample Size for the Training Set

No training set is mentioned in the context of this device. This device is an electrosurgical generator, which does not typically involve machine learning or AI models that require training data. Its design and performance are based on engineering principles and electrical output characteristics.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 29, 2018

Olympus Winter & Ibe GmbH % Dolan Mills Senior Specialist, Regulatory Affairs Gyrus ACMI, Inc. 136 Turnpike Road Southborough, Massachusetts 01772

Re: K182587

Trade/Device Name: Device Electrosurgical Generator ESG-150 Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: September 19, 2018 Received: September 20, 2018

Dear Dolan Mills:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Long H. Chen Digitally sig Digitally signed by Long Date: 2018.11.29 10:22:54 -S -05'00' for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

K182587

Device Name

Electrosurgical generator ESG-150

Indications for Use (Describe)

The electrosurgical generator, in conjunction with electrosurgical accessories and ancillary equipment, is intended for cutting and coagulating tissue in endoscopic surgery.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

2.1 General Information

Manufacturer:Olympus Winter & Ibe GmbHKuehnstr. 6122045 HamburgGermany
Establishment Registration Number: 9610773
Official Correspondent:Dolan MillsSenior Specialist, Regulatory AffairsGyrus ACMI, Inc.136 Turnpike Rd.Southborough, MA 01772Phone: (901) 373-0236Email: dolan.mills@olympus-osta.comEstablishment Registration Number: 3003790304
Date Prepared:Sep 19, 2018

2.2 Device Identification

Proprietary name:Electrosurgical Generator ESG-150
Device Classification name:Electrosurgical cutting and coagulation device andaccessories
Regulation Medical Specialty:General & Plastic Surgery
Regulations Number:21 CFR 878.4400
Regulatory class:Class II
Product code:GEI

2.3 Predicate Device

The subject device, ESG-150, is considered substantially equivalent to the legally marketed device of K141225, GEI, 21CFR 878.4400:

Olympus ESG-400

No reference devices were used in this submission.

{4}------------------------------------------------

2.4 Product Description

The ESG-150 is a reusable, non-sterile electrosurgical generator that features different monopolar and bipolar cutting and coagulation modes. The maximum output power is 120 W. It is intended for cutting and coagulation of tissue in endoscopic surgery in conjunction with electrosurgical accessories and ancillary equipment.

The generator is a class II medical device under the regulation number 878.4400 and the product code GEI - "Electrosurgical cutting and coagulation device and accessories" Regulation Medical Specialty: General & Plastic Surgery. It is compliant with FDA recognized consensus safety standards as listed in section III, Appendix III - Standard Conformity and Test Reports.

The front panel of the ESG-150 features a touch screen GUI (graphical user interface) and all settings of the device (e.g. selecting the socket, selecting the mode, increasing/decreasing power and effect, recalling user settings, etc.) are done via the touchscreen on the front panel. The mode is selected according to the type of procedure that will be performed and according to the HF instruments that will be used.

The ESG 150 has three contact quality indicators (CQM); green, red and grey which indicate the connection status of the neutral electrode. Green highlights that a split neutral electrode is correctly connected. red highlights that the split neutral electrodes is not connected sufficiently, and grev indicates that a non-split neutral electrode is connected, or that a split type neutral electrode is connected but is not recognized by the COM, or is performing incorrectly (short circuit). The generator can store selectable settings, e.g. Power and Effect, and allows the user to adjust the volume of touch screen buttons and activation either with the volume control on the rear panel or via touchscreen (this does not apply to alarm signals).

In addition to the GUI, the ESG-150 has a bipolar socket, a monopolar socket (B Type), and a neutral electrode socket. The power switch turns the generator on and off. The rear panel includes the volume control, a ventilation hole, the equipotential bonding point, the AC power socket, and a footswitch socket.

2.5 Indications for Use

"The electrosurgical generator, in conjunction with electrosurgical accessories and ancillary equipment, is intended for cutting and coagulating tissue in endoscopic surgery."

{5}------------------------------------------------

2.6 Technological Characteristics

The ESG-150 has the same intended use and technological characteristics as the predicate device ESG-400, however, the ESG 400 is indicated for both open and laparoscopic surgery in addition to endoscopic.

While various instruments can be connected to the monopolar and bipolar sockets, because the PK modes of the predicate device (K141225) are not integrated in the ESG-150, the flare out detection is not applicable. The basic design philosophy of the User Interface (UI) and GUI flow chart concept is identical, except for the special ESG-400 amendment in regard to the PK instruments.

In comparison to the primary predicate device the following output modes are available:

Subject Device:ESG-150Predicate Device:ESG-400
PureCutPureCut
PulseCut SlowPulseCut slow
PulseCut FastPulseCut fast
N/ABlend Cut
N/APureCut
N/AFineCut

2.6.1 Output modes in comparison to the primary predicate device ESG-400

Table 2.1: Monopolar Cut Modes

Subject Device:Predicate Device:
ESG-150ESG-400
SoftCoagSoftCoag
ForcedCoagForcedCoag
N/APowerCoag
N/ASprayCoag

Table 2.2: Monopolar Coagulation Modes

{6}------------------------------------------------

Subject Device:Predicate Device:
ESG-150ESG-400
BipolarCutBipolarCut
N/ASalineCut
N/APK PureCut
N/APK SoftCut
N/APK LoopCut
N/APK MorceCut

Table 2.3: Bipolar Cut Modes

Subject Device:Predicate Device:
ESG-150ESG-400
BiSoftCoagBiSoftCoag
N/ARFCoag
N/AAutoCoag
N/ASalineCoag
N/AHardCoag
N/AFineCoag
N/APK Coag
N/APK Softcoag
N/APK AutoCoag

Table 2.4: Bipolar Coagulation Modes

The range of output waveforms is identical but the power levels are decreased in comparison to the FDA cleared ESG-400 electrosurgical generator, K141225.

2.7 Substantial Equivalence

Substantial equivalence is demonstrated by acknowledged verification/validation methodologies. The subject device has equivalent technology, performance, dimensions and materials. The differences to the primary predicate device ESG-400 are:

  • only one monopolar socket, no universal socket, no docking station interface on the bottom
  • to be used only in conjunction with a double pedal footswitch ●
  • new enclosure ●
  • no PK instruments incorporation
  • GUI modifications (additional strings, languages, sounds, screens, windows) ●
  • more effect levels available for most ESG-150 Modes than for the predicate .

2.8 Performance Data

The following performance data were provided in support of the substantial equivalence determination. All standards applied are FDA recognized international standards.

{7}------------------------------------------------

The data was prepared in accordance with the FDA guidance. Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery Guidance for Industry and Food and Drug Administration Staff, issued on August 15, 2016. The guidance was followed for all relevant sections.

2.8.1 Biocompatibility testing

The ESG-150 and its accessories do not contain components that come directly or indirectly into patient contact. Biocompatibility testing according to ISO 10993 is not required.

2.8.2 Electrical safety and electromagnetic compatibility (EMC)

The design of the ESG-150 complies with recognized standards as listed in section 2.8.8 of this summary.

The FDA guidance "Information to Support a Claim of Electromagnetic Compatibility (EMC) of Electrically-Powered Medical Devices", CDRH July 11, 2016 has been followed.

2.8.3 Thermal Safety

The design of the ESG-150 complies with recognized standards as listed in section 2.8.8 of this summary.

Animal and Clinical Studies 2.8.4

Clinical and animal studies were not necessary.

2.8.5 Software

The software validation activities were performed in accordance with the FDA Guidance, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (May 11, 2005), and in accordance with the FDA Guidance: "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery" (Aug. 15, 2016). This device type software presents a "Major Level of Concern".

Performance Testing Bench 2.8.6

To demonstrate substantial equivalence. the following aspects were considered within the validation versus the predicate devices.

    1. Performance and validation tests incorporated the same range of waveform outputs and power levels.
    1. During the validation testing the waveforms and test results were compared directly between the subject and predicate device

Basic safety and performance testing was performed in accordance with IEC standards. In addition, verification and comparison bench studies were conducted to evaluate the functional performance.

{8}------------------------------------------------

Preclinical: Evidence obtained from preclinical simulated use studies demonstrate that the device performs substantially equivalent to the predicate device in relevant aspects associated with usability, tissue effects, and thermal effects. For simulated use testing, three clinically relevant tissue types were evaluated in all applicable modes: porcine muscle, liver, and kidney. The tissue effects testing included quantitative and qualitative assessment.

In accordance with FDA Guidance. Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery, the thermal damage of the tissue due to the HF current was measured in terms of size (length, width and depth) of the thermal coagulation zone. Furthermore, each test included the minimum, default and maximum intensity settings. All modes were compared to the predicate. Where the minimum or maximum intensity settings of the test and predicate device differed in terms of output wattage by specification the closest applicable settings were chosen.

These comprehensive validation bench tests support equivalence to the predicate device. Testing confirmed that comparable tissue effects could be achieved for applicable modes of operation with three tissue types.

Usability and user interface was also assessed according to the risk management plan. The assessment was based on Olympus predecessor product. Use-related hazardous situations were assessed and risk mitigation measures in terms of usability design for safety were defined. The residual risk was evaluated as acceptable.

Risk analysis was carried out in accordance with established internal acceptance criteria based on ISO 14971:2007.

2.8.7 Reprocessing

Required cleaning, disinfecting and drying procedures are described in the instructions for use.

{9}------------------------------------------------

2.8.8 Applied standards

Standard No.Standard TitleFDA-Recognitionno + date
AAMI/ANSI ES60601-1:2005/(R)2012and C1:2009/(R)2012and, A2:2010/(R)2012Medical electrical equipment - Part 1: Generalrequirements for basic safety and essential performance(IEC 60601-1:2005, MOD)19-407/09/2014
IEC 60601-1-2Edition 4.02014Medical electrical equipment - Part 1-2: Generalrequirements for basic safety and essential performance -Collateral Standard: Electromagnetic disturbances -Requirements and tests19-806/27/2016
IEC60601-1-8Ed. 2.1:2012Medical electrical equipment - Part 1-8: Generalrequirements for basic safety and essential performance -Collateral Standard: General requirements, tests andguidance for alarm systems in medical electricalequipment and medical electrical systems5-7608/05/2013
ANSI AAMI IEC60601-2-2:2009Medical Electrical Equipment - Part 2-2: ParticularRequirements For The Basic Safety And EssentialPerformance Of High Frequency Surgical EquipmentAnd High Frequency Surgical Accessories6-38908/21/2017
IEC6230462304 Edition 1.12015-06Medical device software - Software life cycle processes13-7904/04/2016
IEC60601-1-6 Edition 3.12013-10Medical electrical equipment Part 1-6: Generalrequirements for basic safety and essential performance -Collateral standard: Usability5-8906/27/2016
ISO14971Second edition2007Medical devices - Application of risk management tomedical devices5-4006/27/2016
IEC62366-1 Edition 1.02015-02Medical Devices - Part 1: Application Of UsabilityEngineering To Medical Devices [IncludingCORRIGENDUM 1 (2016)]5-11412/23/2016

Table 2.5: Applied standards

2.9 Conclusion

The performance data support the safety of the device and demonstrate that the subject device complies with the recognized standards as specified.

In summary, we believe the ESG-150 is substantially equivalent with the predicate device with respect to the general design approach, function, and the intended use. The ESG-150 raises no new concerns of safety or effectiveness when compared to the predicate device.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.