K Number
K182501

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-10-26

(44 days)

Product Code
Regulation Number
890.5500
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Laser Headband is indicated to treat Androgenetic Alopecia and promote hair growth in males who have Norwood Hamilton Classifications of 11a to V patterns of hair loss and treat Androgenetic Alopecia and promote hair growth in females who have Ludwig (Savin) Scale 1-4, 11-1, or frontal, both with Fitzpatrick Skin Types I to IV.

Device Description

Similar to the Hair Laser Helmet, the Hair Laser Headband is a low-level laser device and is intended/indicated for over- the-counter use. The device emits laser light with the intention to promote hair growth. The device provides distributed laser light to the scalp at 655 +/- 5 nanometers with an individual laser output of 5mW per laser diode. The lasers are configured inside a headband, designed for hands-free operation during treatment. The laser headband is for portable use with rechargeable battery.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Hair Laser Headband, based on the provided FDA 510(k) summary:

This device (Hair Laser Headband) is claiming substantial equivalence to a predicate device (Hair Laser Helmet, K160728). Therefore, the provided documentation focuses on demonstrating that the new device performs similarly to the predicate, rather than presenting a new clinical study with specific acceptance criteria related to efficacy beyond what was established for the predicate.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Implied by Substantial Equivalence)Reported Device Performance (Hair Laser Headband)
Maintain safety characteristics of predicate device (Hair Laser Helmet)- Complies with IEC 60601-1, IEC 60601-1-2, IEC 60825-1, ISO 14971 standards.- Risk Management activities performed to ISO 14971:2012; all residual risks found acceptable.- Bench testing indicates the Hair Laser Headband shares the same output parameters and dose rate characteristics as the Hair Laser Helmet.
Maintain efficacy characteristics of predicate device (Hair Laser Helmet)- Bench testing indicates the Hair Laser Headband shares the same output parameters and dose rate characteristics as the Hair Laser Helmet.- Modifications (size/dimensions, reduced diodes) have no effect on efficacy, as users are instructed to move the device in three sections of the head to achieve the same surface treatment as the Hair Laser Helmet.- "In all instances, the Hair Laser Headband performed as intended."
Meet intended use and indications for use of predicate device- Exact same indications for use as the Hair Laser Helmet.
Share fundamental scientific/technological characteristics with predicate device- Same wavelength (655 +/- 5 nm), individual laser output (5mW per diode), treatment regimen (20 min, 3x/week), materials, delivery method, helmet-like design, single on/off button, rechargeable battery.

No specific numerical acceptance criteria (e.g., "improvement in hair count by X%") are provided in this submission for the Hair Laser Headband, as its clearance is based on substantial equivalence to a previously cleared device. The implicit acceptance criterion is that its performance (safety and efficacy) is equivalent to the predicate device.

2. Sample Size Used for the Test Set and Data Provenance

The document does not describe a separate clinical "test set" in the traditional sense for the Hair Laser Headband's efficacy. Instead, it relies on:

  • Bench Testing: To verify output parameters and dose rate characteristics. No sample size for this is specified, but it would typically involve a small number of manufactured units.
  • Risk Analysis: Performed to ISO 14971:2012.
  • Design Verification and Validation: Performed per 21 CFR 820.30.

The data provenance for these activities would be internal to Pulsaderm LLC, likely in the US, and prospective as part of the device's development and validation.

For the predicate device (Hair Laser Helmet, K160728), a clinical study would have been conducted to establish its efficacy, but details about that study (sample size, provenance, etc.) are not included in this 510(k) summary for the Hair Laser Headband.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. This submission does not describe a clinical test set requiring expert ground truth for the Hair Laser Headband's efficacy. Its performance is compared to the predicate device through bench tests and design controls.

4. Adjudication Method for the Test Set

Not applicable. There is no clinical test set described that would require an adjudication method.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

Not applicable. This device is a laser treatment device for hair growth, not an AI-assisted diagnostic device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. This is a physical medical device, not an algorithm.

7. The Type of Ground Truth Used

For the Hair Laser Headband itself, the "ground truth" for its performance is established through:

  • Bench Test Results: Demonstrating equivalent output parameters and dose rate to the predicate.
  • Compliance with Standards: (IEC 60601-1, IEC 60601-1-2, IEC 60825-1, ISO 14971).
  • Design Verification and Validation: Performed according to regulatory requirements.
  • Risk Analysis: Ensuring safety.

The efficacy "ground truth" is implicitly linked to the predicate device's prior clearance, which would have been based on clinical data (likely hair counts, photographic assessment, patient self-assessment) from a study on the Hair Laser Helmet. Details of that ground truth are not provided here.

8. The Sample Size for the Training Set

Not applicable. This device is not an AI/machine learning model and therefore does not have a "training set."

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Pulsaderm LLC Gloria Avendano Regulatory Affairs Manager 12801 Commonwealth Dr. Units 2-6 Fort Myers, Florida 33913

Re: K182501

Trade/Device Name: Hair Laser Headband Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: OAP Dated: September 12, 2018 Received: September 12, 2018

October 26, 2018

Dear Gloria Avendano:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely, Neil R Ógden -S 2018.10.26 10:23:41 -04'00' For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

Device Name Hair Laser Headband

Indications for Use (Describe)

The Laser Headband is indicated to treat Androgenetic Alopecia and promote hair growth in males who have Norwood Hamilton Classifications of 11a to V patterns of hair loss and treat Androgenetic Alopecia and promote hair growth in females who have Ludwig (Savin) Scale 1-4, 11-1, or frontal, both with Fitzpatrick Skin Types I to IV.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

FDA 510(k) Summary of Safety and Effectiveness for The Hair Laser Headband

General Information

Submitter: Pulsaderm LLC 12801 Commonwealth Dr. Units 2-7 Fort Myers, Florida 33913

Contact Person:

Gloria Avendano Regulatory Affairs Manager Pulsaderm LLC 12801 Commonwealth Dr. Units 2-7 Fort Myers, Florida 33913 Gloria@pulsaderm.com

Summary Preparation Date:

Device

Device Name:Hair Laser Headband
Classification Name:Laser Surgical Instrument, for use in General and PlasticSurgery and in Dermatology
Regulation Numbers:21 CFR 890.5550
Regulatory Class:II
Product Codes:OAP

3. Predicate Devices:

The Laser Headband is substantially equivalent to the following predicate devices;

Device510(k) NumberManufacturer
NutraStim Hair Laser HelmetK160728Pulsaderm LLC

{4}------------------------------------------------

Device Description

Similar to the Hair Laser Helmet, the Hair Laser Headband is a low-level laser device and is intended/indicated for over- the-counter use. The device emits laser light with the intention to promote hair growth. The device provides distributed laser light to the scalp at 655 +/- 5 nanometers with an individual laser output of 5mW per laser diode. The lasers are configured inside a headband, designed for hands-free operation during treatment. The laser headband is for portable use with rechargeable battery.

Intended Use and Indications:

The Hair Laser Headband is indicated to treat Androgenetic Alopecia and promote hair growth in males who have Norwood Hamilton Classifications of Ila to V patterns of hair loss and treat androgenetic alopecia and promote hair growth in females who have Ludwig (Savin) Scale I-4, II-1, II-2, or frontal, both with Fitzpatrick Skin Types I to IV.

Technological Characteristics

The Hair Laser Headband shares the exact same fundamental scientific technology characteristics to the Hair Laser Helmet. The only modifications are the following:

  • Size and dimensions change
  • . Reduce number of diodes

The main technological characteristics in the Hair Laser Headband remain the same, such as wavelength and output of laser diodes, treatment regiment regimen, materials, delivery method, indications for use, helmet-like design, single on/off push button, and both devices use a rechargeable battery.

The modifications mentioned above have no effect in the safety or efficacy of the Hair Laser Headband as users are instructed to move the device in three different sections of the head to achieve the same surface treatment as the Hair Laser Helmet.

Performance Data:

Similar to the Hair Laser Helmet, the Hair Laser Headband complies with the following test standards:

IEC 60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.

IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.

IEC 60825-1 Safety of laser products - Part 1: Equipment classification and requirements.

{5}------------------------------------------------

Based on the nature of modifications implemented, the Hair Laser Headband underwent bench testing and results indicate that the Hair Laser Headband shares the same output parameters and dose rate characteristics as the Hair Laser Helmet.

Design verification and design validation testing were performed according to FDA's Design control activities per 21 CFR 820.30. Additionally, Risk Management activities were performed according to ISO 14971:2012 to evaluate new risks from design changes and all residual risks were found to be acceptable.

In all instances, the Hair Laser Headband performed as intended.

Substantial Equivalence

The Hair Laser Headband shares the same technological characteristics as the Hair Laser Helmet K160728:

DeviceHair Laser HeadbandHair Laser Helmet
510 (k) NumberpendingK160728
Product CodeOAPOAP
RegulationNumber21 CFR 890.550021 CFR 890.5500
LegalManufacturerPulsaderm LLCPulsaderm LLC
Device TypePortable headband sits on patient's headPortable helmet sits on patient's head,
Indications forUseIndicated to treat Androgenetic Alopecia andpromote hair growth in Males who haveNorwood Hamilton Classifications of IIa to Vpatterns of hair loss and treat AndrogeneticAlopecia and promote hair growth in femaleswho have Ludwig (Savin) Scale I-4, II-1, II-2, orfrontal, both with Fitzpatrick Skin Types I to IV.Indicated to treat Androgenetic Alopeciaand promote hair growth in Males who haveNorwood Hamilton Classifications of IIa toV patterns of hair loss and treatAndrogenetic Alopecia and promote hairgrowth in females who have Ludwig(Savin) Scale I-4, II-1, II-2, or frontal, bothwith Fitzpatrick Skin Types I to IV.
Type of UseOver the CounterOver the Counter
Number of LaserDiodes4082
Wavelength655 +/- 5 nanometers655 +/- 5 nanometers
LaserClassificationClass 3RClass 3R
TreatmentProtocol20 Minutes- 3 times a week20 Minutes- 3 times a week
StandardsAppliedIEC 60601-1IEC 60601-1-2IEC 60825-1ISO 14971IEC 60601-1IEC 60601-1-2IEC 60825-1ISO 14971
Device ClassClass IIClass II
SterilizationBasic cleaning, instructions provided.Basic cleaning, instructions provided.
Common NameLamp, Non-HeatingLamp, Non-Heating

{6}------------------------------------------------

The minor modifications in dimensions and number of diodes have no effect in the safety or efficacy of device, as confirmed with performance testing and risk analysis activities.

Conclusion:

Based upon the indications for use and data provided in this pre-market notification, all functional modes of the Hair Laser Headband have been shown to be substantially equivalent to the Hair Laser Helmet with respect to intended use, technological characteristics and safety characteristics. Pulsaderm LLC believes that no significant differences exist between the device and the predicate. Therefore, Substantial equivalency is requested.

§ 890.5500 Infrared lamp.

(a)
Identification. An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.(b)
Classification. Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.