K Number
K182434
Device Name
LaparoGuard
Date Cleared
2018-12-13

(98 days)

Product Code
Regulation Number
876.1500
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
LaparoGuard is indicated for use with compatible laparoscopic devices. LaparoGuard can be used in general laparoscopy wherever a laparoscope is indicated for use.
Device Description
LaparoGuard allows surgeons to virtually annotate a safe anatomical volume ("safe zone") inside the body cavity of a patient during a laparoscopic surgery. A compatible laparoscope, such as the ConMed IM3300 (K031098), captures video feed that is processed by LaparoGuard, which tracks surgical instruments and the safe zone boundaries to display a real-time overlay on a compatible laparoscope monitor, such as the Stryker Vision Elect 21" (K081995). Surgical teams receive visual and auditory notifications throughout the procedure whenever a tracked instrument has exited the safe zone.
More Information

Unknown
The description mentions "processing" of video feed and "tracking" of instruments and boundaries, which could potentially involve AI/ML algorithms for image analysis and object tracking. However, the summary does not explicitly state the use of AI, ML, or deep learning, nor does it provide details about training or test sets, which are common indicators of AI/ML development. Without more information, it's impossible to definitively confirm or deny the presence of AI/ML.

No
LaparoGuard is a surgical aid for visualization and tracking, which enhances the safety of laparoscopic procedures by guiding instrument placement within a defined safe zone but does not directly treat or restore a bodily function.

No

The device is described as tracking surgical instruments and safe zone boundaries for real-time overlay during surgery, and providing notifications. This is an intra-operative guidance tool, not a diagnostic device which identifies or characterizes a disease or condition.

No

The device description explicitly states that LaparoGuard processes video feed from a compatible laparoscope and displays an overlay on a compatible monitor. While LaparoGuard itself might be software, its functionality is intrinsically tied to and relies on specific hardware components (laparoscope and monitor) for both input and output. The performance studies also include system integration testing using anatomical phantoms, further indicating a reliance on hardware interaction.

Based on the provided information, LaparoGuard is NOT an IVD (In Vitro Diagnostic) device.

Here's why:

  • IVD devices are used to examine specimens derived from the human body (like blood, urine, tissue) to provide information for diagnosis, monitoring, or screening.
  • LaparoGuard operates on real-time video feed from a laparoscope during surgery. It processes this video to track instruments and a "safe zone" within the patient's body cavity.
  • Its purpose is to provide visual and auditory guidance to the surgical team during the procedure, not to analyze biological specimens.

The device description and intended use clearly indicate that LaparoGuard is a surgical guidance system that works with laparoscopic equipment to enhance safety during surgery. It does not perform any diagnostic tests on biological samples.

N/A

Intended Use / Indications for Use

LaparoGuard is indicated for use with compatible laparoscopic devices. LaparoGuard can be used in general laparoscopy wherever a laparoscope is indicated for use.

Product codes

GCJ

Device Description

LaparoGuard allows surgeons to virtually annotate a safe anatomical volume ("safe zone") inside the body cavity of a patient during a laparoscopic surgery. A compatible laparoscope, such as the ConMed IM3300 (K031098), captures video feed that is processed by LaparoGuard, which tracks surgical instruments and the safe zone boundaries to display a real-time overlay on a compatible laparoscope monitor, such as the Stryker Vision Elect 21" (K081995). Surgical teams receive visual and auditory notifications throughout the procedure whenever a tracked instrument has exited the safe zone.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Abdomen

Indicated Patient Age Range

Any patients undergoing laparoscopic procedures

Intended User / Care Setting

Intended Users: General surgeons, gynecologists, plastic surgeons and urologists.
Care Setting: Surgical Suite

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Non-Clinical testing was conducted on the LaparoGuard system including:

  • Characterization of system accuracy using an accuracy measurement phantom
  • Characterization of laparoscopic image fidelity
  • Characterization of system display latency
  • Software verification and validation for each requirement
  • System integration testing using anatomical phantoms

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K121893

Reference Device(s)

K101979

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

0

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym followed by the full name of the agency on the right. The FDA acronym and the words "U.S. FOOD & DRUG ADMINISTRATION" are in blue.

December 13, 2018

Mariner Endosurgery Inc. % Joel Ironstone President Ironstone Product Development Inc. Suite 222, 276 Carlaw Ave Toronto, M4M 3L1 Canada

Re: K182434

Trade/Device Name: LaparoGuard Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ Dated: August 31, 2018 Received: September 6, 2018

Dear Joel Ironstone:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

1

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Image /page/1/Picture/5 description: The image shows a digital signature from Long H. Chen. The signature includes the name Long H. Chen, followed by "-S". The date of the signature is December 13, 2018, at 08:54:03, with a time zone offset of -05'00'. The signature is indicated to be "for" something, but the specific context is not provided in the image.

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K182434

Device Name LaparoGuard

Indications for Use (Describe)

LaparoGuard is indicated for use with compatible laparoscopic devices. LaparoGuard can be used in general laparoscopy wherever a laparoscope is indicated for use.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

510(k) SUMMARY

Mariner Endosurgery's LaparoGuard

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

Mariner Endosurgery Inc. Suite 305A, 175 Longwood Road South Hamilton, ON, Canada L8P 0A1

Phone: +1 905-921-8755 Contact Person: Mitchell Wilson

Date Prepared: August 31, 2018

Name of Device

LaparoGuard

Device Classification and Product Code

Endoscope and accessories, 21 CFR 876.1500, Class II, GCJ

Predicate Devices

Stryker Endoscopy's SDC3 HD Information Management System (K121893)

Indications for Use

LaparoGuard is indicated for use with compatible laparoscopic devices. LaparoGuard can be used in general laparoscopy wherever a laparoscope is indicated for use.

Device Description

LaparoGuard allows surgeons to virtually annotate a safe anatomical volume ("safe zone") inside the body cavity of a patient during a laparoscopic surgery. A compatible laparoscope, such as the ConMed IM3300 (K031098), captures video feed that is processed by LaparoGuard, which tracks surgical instruments and the safe zone boundaries to display a real-time overlay on a compatible laparoscope monitor, such as the Stryker Vision Elect 21" (K081995). Surgical teams receive visual and auditory notifications throughout the procedure whenever a tracked instrument has exited the safe zone. LaparoGuard consists of the components listed in the table below:

4

LaparoGuard ComponentPart NumberSingle Use / Reusable
TrackersType #1: C00261Single Use
Type #2: C00262Single Use
TrackersType #3: C00263Single Use
Type #4: C00264Single Use
Type #5: C00265Single Use
ProbeC00234Single Use
CalibratorC00233Single Use
NDI CameraM00034Reusable
Controlling PCX00021Reusable
Power SupplyX00026Reusable
Video Capture BoxX00022Reusable
DVI-D Bypass CableX00023Reusable
Support CartX00019Reusable

Table 1 - LaparoGuard Components

Non-Clinical Testing

Non-Clinical testing was conducted on the LaparoGuard system including:

  • . Characterization of system accuracy using an accuracy measurement phantom
  • Characterization of laparoscopic image fidelity ●
  • . Characterization of system display latency
  • . Software verification and validation for each requirement
  • . System integration testing using anatomical phantoms

Standards Compliance

LaparoGuard has been tested to be in compliance with the following FDA recognized standards:

  • Electrical safety testing per IEC 60601-1:2005 (Third Edition) + C1:2006 + C2:2007 + ● A1:2012
  • EMC per IEC 60601-1-2:2014 (4th Edition). .
  • . Positional accuracy of computer assisted surgical systems, as per ASTM F2554-10
  • Software validation per FDA's Guidance for the Content of Premarket Submission for Software Contained in Medical Devices (May 11, 2005)

5

Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "minor" level of concern, as failures or latent design flaws are unlikely to cause any injury to the patient or operator.

Substantial Equivalence

LaparoGuard is substantially equivalent to Stryker Endoscopy's SDC3 HD Information Management System. LaparoGuard has the same intended use as the predicate device. The differences in technological characteristics do not raise new questions of safety or effectiveness. LaparoGuard has been tested to FDA-recognized standards to demonstrate its safety and effectiveness. Thus, LaparoGuard is substantially equivalent to the predicate device. A summary of the substantial equivalence comparison is provided in the table below:

CHARACTERISTICSUBJECT DEVICEPREDICATE DEVICEREFERENCE DEVICE
Device NameLaparoGuardSDC3 HD Information
Management SystemExplorer Liver --- Passive
Tracking
ManufacturerMariner EndosurgeryStryker EndoscopyPathfinder Therapeutics
510(k) Number-K121893K101979
Regulatory ClassClass IIClass IIClass II
Regulation Number21 CFR 876.150021 CFR 876.150021 CFR 882.4560
Product
ClassificationGCJGCJOEW
Classification NameEndoscope and
accessoriesEndoscope and accessoriesStereotaxic instrument
Intended UseAdjunctive aid for the
monitoring of compatible
instruments during
laparoscopic surgeryAdjunctive aid for the monitoring
and control of compatible
instruments during endoscopic,
laparoscopic, or arthroscopic
procedures. Additionally, has
operating room documentation
functionality and ability to export
recorded video/images.Adjunctive aid for monitoring of
compatible instruments during
open liver surgical procedures,
along with image-guidance using
preoperative, patient-specific
MRI or CT medical images.
Indications for Use
StatementLaparoGuard device is
indicated for use with
compatible laparoscopic
devices. LaparoGuard can
be used in general
laparoscopy wherever a
laparoscope is indicated for
use.The Stryker SDC3 HD
Information Management System
(SDC3) is intended for use with
compatible endoscopic and
general surgery devices. SDC3
can be used in general
laparoscopy,
nasopharyngoscopy, ear
endoscopy, sinuscopy, and
plastic surgery wherever a
laparoscope, endoscope, or an
arthroscope is indicated for use.The Explorer™ Liver --- Passive
Tracking device is indicated for
open liver surgical procedures
where image-guidance may be
appropriate and where the
patient can tolerate long apneic
periods under general
anesthesia.
CHARACTERISTICSUBJECT DEVICEPREDICATE DEVICEREFERENCE DEVICE
A few examples of the more
common endoscopic surgeries
are laparoscopic
cholecystectomy, laparoscopic
hernia repair, laparoscopic
appendectomy, laparoscopic
pelvic lymph node dissection,
laparoscopically assisted
hysterectomy, laparoscopic &
thorascopic anterior spinal
fusion, anterior cruciate ligament
reconstruction, knee arthroscopy,
shoulder arthroscopy, small joint
arthroscopy, decompression
fixation, wedge resection, lung
biopsy, pleural biopsy, dorsal
sympathectomy, pleurodesis,
internal mammary artery
dissection for coronary artery
bypass, coronary artery bypass
grafting where endoscopic
visualization is indicated and
examination of the evacuated
cardiac chamber during
performance of valve
replacement. SDC3 users are
general surgeons, gynecologists,
cardiac surgeons, thoracic
surgeons, plastic surgeons,
orthopedic surgeons, ENT
surgeons, and urologists.
EnvironmentSurgical SuiteSurgical SuiteSurgical Suite
UsersGeneral surgeons,
gynecologists, plastic
surgeons and urologists.General surgeons,
gynecologists, cardiac surgeons,
thoracic surgeons, plastic
surgeons,
orthopedic surgeons, ENT
surgeons and urologists.General surgeons
Patient PopulationAny patients undergoing
laparoscopic proceduresAny patients undergoing
endoscopic proceduresPatients undergoing open liver
surgical procedures where
image-guidance may be
appropriate and who can tolerate
long apneic periods under
general anesthesia
Anatomical SitesAbdomenWhole Body (anywhere where anLiver
endoscope may be used)
CHARACTERISTICSUBJECT DEVICEPREDICATE DEVICEREFERENCE DEVICE
Procedural
OverviewAllows the surgeon to
monitor the position of
tracked compatible
instruments relative to a
virtual three-dimensional
"safe zone" and target
location established by the
surgeon at the initiation of
the procedure, using a
tracked probe. Using a
digital representation,
tracked instruments are
displayed relative to each
other and to the safe zone.Allows the surgeon to monitor
and control the status/settings of
any compatible instrument
attached to it.Allows the surgeon to monitor
the position of tracked
compatible instruments relative
to tumors and surrounding
anatomic structures, based on
preoperative, patient-specific
MRI or CT medical images.
Using a digital representation,
tracked instruments are
displayed relative to each other
and to the surrounding anatomy.
Electrical SafetyCompliance to IEC 60601-1Compliance to IEC 60601-1Compliance to IEC 60601-1
Energy delivered or
controlledNone delivered to patientNone directly delivered to
patient; however, device can
control the delivery of energy
from compatible instrumentsSome laser energy delivered to
patient through the use of a
Laser Range Scanner
component
NotificationsProvide visual and auditory
notification when tracked
instruments exit the
surgeon-defined "safe
zone"Provides visual and auditory
notification of surgeon-defined
"time outs" during which all
compatible instruments are
turned offN/A
Display of
Monitored
InstrumentsBoth integrated into
laparoscopic camera video
feed output and digitally
represented on a
standalone monitorIntegrated into endoscopic
camera video feed outputDigitally represented on a
standalone monitor
Tracking
TechnologyPolaris Spectra optical
position sensorN/APolaris Spectra or Vicra optical
position sensors
Method of Optical
TrackingRetroreflective passively
tracked spheres (NDI
Passive Spheres -
K033621) incorporated
onto tracked adaptersN/ARetroreflective passively tracked
spheres (NDI Passive Spheres -
K033621) incorporated onto
tracked adapters
Number of Passive
Spheres per
Tracked AdapterFourN/AFour
Compatible
InstrumentsRigid instruments onlyPowered instrumentsRigid instruments only
Number of
Compatible
Instruments
Monitored or
Controlled at a TimeFiveFiveFive
Connection to
Compatible
InstrumentsPassively optically tracked
adaptersWired connectionPassively optically tracked
adapters
CHARACTERISTICSUBJECT DEVICEPREDICATE DEVICEREFERENCE DEVICE
Compatible
Instrument
CalibrationSupports calibration of an
instrument's tip position
and orientation
simultaneously
using a calibration fixtureN/ASupports calibration of an
instruments's tip position and
orientation simultaneously
using a calibration fixture
Method for User
Indication of
Targeted
Anatomical
StructuresTracked probeN/ATracked Probe
Patient Contacting
ComponentsTracked probeNo patient contacting
componentsTracked probe and tracked
abdominal reference
BiocompatibilityBiocompatible stainless
steelN/ABiocompatible stainless steel and
other materials tested per ISO
10993-1
SterilityManufacturer-sterilized,
single use onlyN/AEnd-user sterilized, reusable

Table 2 - Substantial Equivalence Comparison Table

6

7

8

Conclusions

LaparoGuard has the same intended use and indications for use as its predicate device. Differences in technological characteristics do not raise new questions of safety or effectiveness. Therefore, LaparoGuard is substantially equivalent to its predicate.