(261 days)
No
The device description focuses on the physical properties and materials of the embolization plug and its delivery system. There is no mention of software, algorithms, or data processing that would indicate the use of AI or ML. The performance studies described are standard mechanical, animal, and biocompatibility tests for a physical implantable device.
Yes.
The device is used to obstruct or reduce the rate of blood flow in the peripheral vasculature to treat medical conditions.
No
The device is an embolization plug designed to obstruct or reduce blood flow in the peripheral vasculature. It is a therapeutic device, not one that diagnoses conditions.
No
The device description clearly states it is a permanent implant comprised of physical components (polyurethane foam plug, platinum-iridium alloy marker band, introducer sheath). It is a hardware device, not software.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use is to "obstruct or reduce the rate of blood flow in the peripheral vasculature." This is a therapeutic intervention performed in vivo (within the body) to treat a medical condition.
- Device Description: The device is a permanent implant designed to physically block blood flow. This is a mechanical action within the body, not a test performed on a sample in vitro (outside the body).
- Lack of IVD Characteristics: The description does not mention any of the typical characteristics of an IVD, such as:
- Analyzing biological samples (blood, urine, tissue, etc.)
- Detecting or measuring substances in samples
- Providing information for diagnosis, monitoring, or screening
IVD devices are used to perform tests on samples taken from the body to provide information about a person's health status. This device is used to directly treat a condition within the body.
N/A
Intended Use / Indications for Use
The IMPEDE-FX Embolization Plug is indicated for use with the IMPEDE Embolization Plug to obstruct or reduce the rate of blood flow in the peripheral vasculature.
Product codes
KRD
Device Description
The IMPEDE-FX Embolization Plug is a permanent implant, pushable embolization device comprised of a porous, degradable polyurethane thermoset foam plug and a proximal platinumiridium alloy marker band. The foam plug is crimped into a secondary shape to allow delivery through a catheter to the target vessel using a guidewire. Upon deployment in the target vessel and exposure to an aqueous environment and body temperature, the plug will self-expand to its primary shape to embolize the target vessel. The IMPEDE-FX device may only be used in conjunction with the IMPEDE® Embolization Plug System to further enhance vessel occlusion or increase the length of occlusion within the target vessel. It is indicated to obstruct or reduce the rate of blood flow in the peripheral vasculature.
The IMPEDE-FX polyurethane foam undergoes slow oxidative degradation with the majority (> 90%) of the foam remaining at 30 days in a swine intravascular model. The amount of foam degradation was also assessed in vivo in 180-day rat subcutaneous implants and 26-week rabbit intramuscular implants. Analysis of the histology slides from both studies revealed near complete degradation of the polymer foam at the majority of the implant sites in each study.
The device is pre-loaded into an introducer sheath which is made of a PEEK tube bonded to a polycarbonate luer hub to facilitate the transfer of the device into the catheter. The device implant is categorized as a permanent implant, blood contacting device, and the introducer is categorized as an external communicating device, limited (
§ 870.3300 Vascular embolization device.
(a)
Identification. A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.(b)
Classification. Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).
0
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 23, 2019
Shape Memory Medical Meghan Reu Ouality and Regulatory Manager 807 Aldo Avenue, Suite 109 Santa Clara, California 95054
Re: K182390
Trade/Device Name: IMPEDE-FX Embolization Plug Regulation Number: 21 CFR 870.3300 Regulation Name: Vascular embolization device Regulatory Class: Class II Product Code: KRD Dated: April 25, 2019 Received: April 26, 2019
Dear Meghan Reu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
1
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Kenneth Cavanaugh Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K182390
Device Name IMPEDE-FX Embolization Plug
Indications for Use (Describe)
The IMPEDE-FX Embolization Plug is indicated for use with the IMPEDE Embolization Plug to obstruct or reduce the rate of blood flow in the peripheral vasculature.
Type of Use (Select one or both, as applicable) | |
---|---|
------------------------------------------------- | -- |
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
510(k) SUMMARY
SUBMITTER
Shape Memory Medical Inc 807 Aldo Avenue, Suite 109 Santa Clara, CA 95054
Contact Person: Meghan Reu, Quality and Regulatory Sr. Manager Phone: 408-649-5175 x 711 Fax: 669-230-5932 Date Prepared: March 29, 2019
DEVICE
Name of Device: IMPEDE-FX Embolization Plug Common Name: Vascular Embolization Device Classification Name: Device, embolization, arterial Regulatory Class: Class II, 21 CFR 870.3300 Product Code: KRD
PREDICATE DEVICE IMPEDE Embolization Plug (K181051)
REFERENCE DEVICE Vascular Solutions Gel Block Embolization Pledgets (K113266)
DEVICE DESCRIPTION
The IMPEDE-FX Embolization Plug is a permanent implant, pushable embolization device comprised of a porous, degradable polyurethane thermoset foam plug and a proximal platinumiridium alloy marker band. The foam plug is crimped into a secondary shape to allow delivery through a catheter to the target vessel using a guidewire. Upon deployment in the target vessel and exposure to an aqueous environment and body temperature, the plug will self-expand to its primary shape to embolize the target vessel. The IMPEDE-FX device may only be used in conjunction with the IMPEDE® Embolization Plug System to further enhance vessel occlusion or increase the length of occlusion within the target vessel. It is indicated to obstruct or reduce the rate of blood flow in the peripheral vasculature.
The IMPEDE-FX polyurethane foam undergoes slow oxidative degradation with the majority (> 90%) of the foam remaining at 30 days in a swine intravascular model. The amount of foam degradation was also assessed in vivo in 180-day rat subcutaneous implants and 26-week rabbit intramuscular implants. Analysis of the histology slides from both studies revealed near complete degradation of the polymer foam at the majority of the implant sites in each study.
The device is pre-loaded into an introducer sheath which is made of a PEEK tube bonded to a polycarbonate luer hub to facilitate the transfer of the device into the catheter. The device implant is categorized as a permanent implant, blood contacting device, and the introducer is categorized as an external communicating device, limited (90%) of the foam remaining after 30
days in a swine intravascular occlusion
animal model. Near complete degradation
in 180-day rat subcutaneous implants and
26-week rabbit intramuscular implants | Yes - slow degradation with the majority
(>90%) of the foam remaining after 30
days in a swine intravascular occlusion
animal model. Near complete degradation
in 180-day rat subcutaneous implants and
26-week rabbit intramuscular implants |
| Radiopacity | Proximal Pt/Ir Marker Band | Proximal Pt/Ir Marker Band and distal
Pt/Ir and Nitinol Anchor Coil |
TABLE 1: PREDICATE DEVICE COMPARISON CHART
5
PERFORMANCE DATA - NEW
The following performance data were provided in support of the substantial equivalence determination.
Mechanical, Visual and Characterization testing
- Simulated use testing ●
- o Accessory Compatibility (Guidewire, Catheter and Introducer)
- Delivery Performance (Deliverability, Migration) O
- Implantation Performance (Plug Expansion, Migration, Bond Integrity) о
- Radial Force ●
- Dimensional Inspection ●
- Visual Inspection
Animal Study
The IMPEDE-FX Embolization Plug was evaluated acutely (i.e. 4 hour sacrifice) in a porcine large animal study to evaluate usability, performance, and vessel response.
PERFORMANCE DATA - LEVERAGED
The following performance data was leveraged from the predicate IMPEDE device, due to similarities between the devices.
Biocompatibility testing
The biocompatibility evaluation for the IMPEDE device was conducted in accordance with ISO10993-1 Biological Evaluation of Medical Devices - Part 1: Guidance on Selection of Tests. The implant is considered a permanent implant, blood contacting device, and the introducer is considered an external communicating device, limited (