(53 days)
The ILICON (ocufilcon D) SPHERICAL contact lenses for daily wear are indicated for the correction of refractive ametropia (myopia and hyperopia) in aphakic or not-aphakic persons with non-diseased that may exhibit astigmatism up to 2.0 diopters that does not interfere with visual acuity. The lens is available clear or tinted and may be used to enhance or alter the apparent color of the eve.
The ILICON (ocufilcon D) TORIC Soft Contact Lenses for daily wear are indicated for the correction of refractive error in aphakic and not-aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 5.00 diopters. The lens is available clear or tinted and may be used to enhance or alter the apparent color of the eye.
The ILICON (ocufilcon D) MULTIFOCAL Soft Contact Lenses for daily wear are indicated for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic persons with non-diseased eyes that may exhibit astigmatism up to 2.0 diopters that does not interfere with visual acuity. The lens is available clear or tinted and may be used to enhance or alter the apparent color of the eye.
The ILICON (ocufilcon D) MULTIFOCAL TORIC Soft Contact Lenses for daily wear are indicated for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic persons with non-diseased eves that may exhibit astigmatism up to 5.0 diopters that does not interfere with visual acuity. The lens is available clear or tinted and may be used to enhance or alter the apparent color of the eye.
Daily wear replacement schedules may vary from patient and should be decided by eye care practitioners in consultation with their patients.
Eye care practitioners may prescribe any of the above lenses for frequent/planned replacement wear, with cleaning disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfecting system.
Eye care practitioners may prescribe any of the above lenses for single use daily disposable wear. When prescribed for daily disposable wear the lens is to be discarded after each removal.
The ILICON (ocufilcon D) Soft (Hydrophilic) Daily Wear Contact Lenses (Tinted, Color) are hemispherical shells with molded spherical base curves and molded front surfaces. The hydrophilic characteristics allow aqueous solutions to enter the lenses are fabricated from ocufilcon D, which is a hydrophilic co-polymer of 2-Hydroxyethyl methacrylate (2-HEMA) and methacrylic acid (MAA), cross-linked with ethylene glycol dimethacrylate (EGDMA), plus an initiator. The copolymer consists of 45% ocufiocon D and 55% water by weight when immersed in saline solution. The (ocufiocon D) name has been adopted by the United States Adopted Names Council (USAN).
The ILICON (ocufilcon D) Soft (Hydrophilic) Daily Wear Contact Lenses (Tinted, Color) are available tinted for visibility, and tinted to enhance or alter the apparent color of the eye. The lenses are processed to incorporate the 'listed' color additives and contain only the amount of the additive needed to accomplish the intended coloring effect. The lenses contain one or a combination of one or more of the following 'listed' color additives: Rutile TiO2, Iron Oxide, (Pathalocyaninto(2-))Copper, Carbazole Violet, Phthalocyanine Green.
When producing the color lenses, the manufacturing process changes the specifications to the clear contact lens by pad-printing the color pigment(s)-entrapping the colorants in the interpenetrating network of the contact lens material in a location that corresponds to the iris. The color pigments used are not removed by lens handling and cleaning/disinfecting procedures. Except for affecting the amount of light transmittance through the lens, the coloring process does not alter the original characteristics of the pre-tinted lens. The tinting pattern has a clear pupil diameter of 6.0 mm.
The ILICON (ocufilcon D) Soft (Hydrophilic) Daily Wear Contact Lenses (Tinted, Color) incorporate a UV absorbing monomer. The lenses block >95% in the UVB range (280nm - 315nm), and >50% in the UVA range (316nm - 380nm).
The ILICON (ocufilcon D) Soft (Hydrophilic) Daily Wear Contact Lenses (Tinted, Color) are manufactured in the sphere, toric, and multifocal design configurations.
The provided document is a 510(k) Summary for the ILICON (ocufilcon D) Soft (Hydrophilic) Daily Wear Contact Lenses (Tinted, Color). This document describes the device, its intended use, and a comparison to predicate devices to establish substantial equivalence. It does not describe a clinical study of the type you are asking about, which typically involve AI/ML performance metrics, expert adjudication, or MRMC studies.
Instead, this submission focuses on demonstrating the safety and effectiveness of the contact lenses through non-clinical testing (toxicology, shelf-life, physicochemical, and mechanical properties). Clinical testing was not deemed necessary as the clinical performance of soft contact lenses made from ocufilcon D material has been previously established.
Therefore, I cannot fulfill all parts of your request as the provided text does not contain the specific information about acceptance criteria for an AI/ML device, a study proving device meets acceptance criteria as typically understood in AI/ML performance studies, sample sizes for test sets in an AI/ML context, data provenance, expert qualifications, adjudication methods, MRMC studies, standalone performance with AI, or ground truth establishment for such a study.
However, I can extract the relevant information regarding the device's technical specifications and the non-clinical tests performed:
1. A table of acceptance criteria and the reported device performance
The document provides a table of "Parameter" and "Tolerance*" which can be considered acceptance criteria, and the "Range" or specific value as the reported performance, indicating the device meets these specifications.
| Parameter | Acceptance Criteria (Tolerance*) | Reported Device Performance (Range/Value) |
|---|---|---|
| Chord Diameter | ±0.20 mm | 11.00 mm to 15.00 mm |
| Center Thickness | When ≤ 0.10 mm → ±0.010 mm + 10%; When > 0.10 mm → ±0.015 mm + 5% | 0.05 mm to 0.15 mm |
| Base Curve | ±0.20 mm | 7.00 mm to 10.0 mm |
| Back Vertex Power (F'v) | When 0.00 < F'v ≤ 10.00 D → ±0.25 D; When 10.00 < F'v ≤ 20.00 D → ±0.50 D | +20.00D to -20.00D (in 0.25D steps) |
| Cylinder Power (F'c) | When 0.00 < F'c ≤ 2.00 D → ±0.25 D; When 2.00 < F'c ≤ 4.00 D → ±0.37 D | -0.25D to -4.00D (in 0.25D steps) |
| Cylinder Axis | When 0.00 < F'c ≤ 1.50 D → ± 8°; When F'c > 1.50 D → ± 5° | 10° to 180° (in 10° steps) |
| Multifocal Add Power | ±0.37 D | +0.25D to +4.00D (0.50D steps) |
| Surface Appearance | Lenses should be clear with no surface defect | - |
| Oxygen Permeability (x 10-11(cm²/sec)(mlO2)/(ml x mmHg)) | ±20% | 19.6 |
| Light Transmission - Tinted (@ 380-780nm) | ±5% | 95% |
| Ultraviolet Radiation Transmittance | TUVB (280 to 315 nm) < 0.05TV; TUVA (316 to 380 nm) < 0.50TV | < 5 % TUVB; < 50 %TUVA |
| Water Content | ±2% | 55% |
| Refractive Index | +0.005 | 1.410 (hydrated) |
Non-clinical Performance Data:
- Toxicology: Non-toxic (In-Vitro Cytotoxicity, Systemic Toxicity, Acute Ocular Irritation).
- Shelf Life: Stability, sterility, and package integrity demonstrated over the labeled expiration date.
- Physicochemical & Mechanical Properties: Consistent material properties (refractive index, water content, Dk, % transmission (visible & UV), tensile strength, modulus, % elongation to break, specific gravity and polymerization residuals) with predicate devices.
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not specify sample sizes for non-clinical tests (e.g., number of lenses tested for physicochemical properties, number of animals for toxicology). Data provenance is not detailed in terms of country of origin of the data for these non-clinical tests, nor is it categorized as retrospective or prospective in the context of clinical studies. However, the toxicity tests were conducted in accordance with GLP (Good Laboratory Practice) regulations.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Not applicable. This device is a medical device (contact lens), not an AI/ML diagnostic tool, and its evaluation did not involve human expert interpretation or ground truth establishment in that manner.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable, as no such clinical test set requiring adjudication was performed.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable, as this is not an AI-assisted device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable, as this is not an AI/ML algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" equivalent for this device is the established physical, chemical, and biological properties and safety profiles for contact lenses made from ocufilcon D material, demonstrated through conformance to established standards (e.g., ISO, ANSI Z80.20) and non-clinical testing results.
8. The sample size for the training set
Not applicable, as this device does not involve a "training set" in the context of AI/ML.
9. How the ground truth for the training set was established
Not applicable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
October 12, 2018
VizionFocus, Inc. % Bret Andre, MS, ABOc Principal Consultant EyeReg Consulting, Inc. 6119 Canter Lane West Linn, OR 97068
Re: K182247
Trade/Device Name: ILICON (ocufilcon D) Soft (Hydrophilic) Daily Wear Contact Lenses (Tinted, Color) Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: August 15, 2018 Received: August 20, 2018
Dear Bret Andre:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely yours.
J. Angelo Green
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K182247
Device Name
ILICON (ocufilcon D) Soft (Hydrophilic) Daily Wear Contact Lenses (Tinted, Color)
Indications for Use (Describe)
The ILICON (ocufilcon D) SPHERICAL contact lenses for daily wear are indicated for the correction of refractive ametropia (myopia and hyperopia) in aphakic or not-aphakic persons with non-diseased that may exhibit astigmatism up to 2.0 diopters that does not interfere with visual acuity. The lens is available clear or tinted and may be used to enhance or alter the apparent color of the eve.
The ILICON (ocufilcon D) TORIC Soft Contact Lenses for daily wear are indicated for the correction of refractive error in aphakic and not-aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 5.00 diopters. The lens is available clear or tinted and may be used to enhance or alter the apparent color of the eye.
The ILICON (ocufilcon D) MULTIFOCAL Soft Contact Lenses for daily wear are indicated for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic persons with non-diseased eyes that may exhibit astigmatism up to 2.0 diopters that does not interfere with visual acuity. The lens is available clear or tinted and may be used to enhance or alter the apparent color of the eye.
The ILICON (ocufilcon D) MULTIFOCAL TORIC Soft Contact Lenses for daily wear are indicated for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic persons with non-diseased eves that may exhibit astigmatism up to 5.0 diopters that does not interfere with visual acuity. The lens is available clear or tinted and may be used to enhance or alter the apparent color of the eye.
Daily wear replacement schedules may vary from patient and should be decided by eye care practitioners in consultation with their patients.
Eye care practitioners may prescribe any of the above lenses for frequent/planned replacement wear, with cleaning disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfecting system.
Eye care practitioners may prescribe any of the above lenses for single use daily disposable wear. When prescribed for daily disposable wear the lens is to be discarded after each removal.
Type of Use (Select one or both, as applicable)Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is:
SUBMITTER I.
| Date Prepared: | August 15th, 2018 |
|---|---|
| Name:Address: | VIZIONFOCUS INC.No.66, Youyi Rd., Zhunan Township.Miaoli County 350.Taiwan (R.O.C.) |
| Contact Person:Phone number: | Angus ShihGeneral Manager037-582900 |
| Consultant:Phone number: | Bret AndreEyeReg Consulting, Inc.6119 Canter Ln.West Linn, OR 97068(503) 372-5226 |
II. DEVICE
| Trade Name: | ILICON (ocufilcon D) Soft (Hydrophilic) Daily Wear Contact Lenses(Tinted, Color) |
|---|---|
| CommonName: | Contact Lens, Daily Wear |
| ClassificationName: | Soft (hydrophilic) Contact Lens (21 CFR 886.5925) |
| RegulatoryClass: | Class II |
| Product Code: | LPL; MVN |
Purpose of 510(k) Submission:
~ New Device ~
{5}------------------------------------------------
III. PREDICATE DEVICE
The ILICON (ocufilcon D) Soft (Hydrophilic) Daily Wear Contact Lenses are substantially equivalent to the following predicate device(s):
- "EZvue color (ocufilcon D)" By I-SEE VISION TECHNOLOGY INC. 510(k) number: K150293
- "BIOMEDICS (ocufilcon D)" By Coopervision, Inc. 510(k) number; K091339
IV. DEVICE DESCRIPTION
The ILICON (ocufilcon D) Soft (Hydrophilic) Daily Wear Contact Lenses (Tinted, Color) are hemispherical shells with molded spherical base curves and molded front surfaces. The hydrophilic characteristics allow aqueous solutions to enter the lenses are fabricated from ocufilcon D, which is a hydrophilic co-polymer of 2-Hydroxyethyl methacrylate (2-HEMA) and methacrylic acid (MAA), cross-linked with ethylene glycol dimethacrylate (EGDMA), plus an initiator. The copolymer consists of 45% ocufiocon D and 55% water by weight when immersed in saline solution. The (ocufiocon D) name has been adopted by the United States Adopted Names Council (USAN).
The ILICON (ocufiocon D) Soft (Hydrophilic) Daily Wear Contact Lenses (Tinted, Color) are available tinted for visibility, and tinted to enhance or alter the apparent color of the eye. The lenses are processed to incorporate the 'listed' color additives and contain only the amount of the additive needed to accomplish the intended coloring effect. The lenses contain one or a combination of one or more of the following 'listed' color additives:
| Color Additive | Listing |
|---|---|
| Rutile TiO2 | 21 CFR § 73.3126 |
| Iron Oxide | 21 CFR § 73.3125 |
| (Pathalocyaninto(2-))Copper | 21 CFR § 74.3045 |
| Carbazole Violet | 21 CFR § 73.3107 |
| Phthalocyanine Green | 21 CFR § 73.3124 |
When producing the color lenses, the manufacturing process changes the specifications to the clear contact lens by pad-printing the color pigment(s)-entrapping the colorants in the interpenetrating network of the contact lens material in a location that corresponds to the iris. The color pigments used are not removed by lens handling and cleaning/disinfecting procedures. Except for affecting the amount of light transmittance through the lens, the coloring process does not alter the original characteristics of the pre-tinted lens. The tinting pattern has a clear pupil diameter of 6.0 mm.
The ILICON (ocufilcon D) Soft (Hydrophilic) Daily Wear Contact Lenses (Tinted, Color) incorporate a UV absorbing monomer. The lenses block >95% in the UVB range (280nm - 315nm), and >50% in the UVA range (316nm - 380nm).
{6}------------------------------------------------
The ILICON (ocufilcon D) Soft (Hydrophilic) Daily Wear Contact Lenses (Tinted, Color) are manufactured in the sphere, toric, and multifocal design configurations. The material properties and available parameters of the finished lenses are as follows:
| Parameter | Range | Tolerance* |
|---|---|---|
| Chord Diameter | 11.00 mm to 15.00 mm | ±0.20 mm |
| Center Thickness | 0.05 mm to 0.15 mm | When ≤ 0.10 mm → ±0.010 mm + 10%When > 0.10 mm → ±0.015 mm + 5% |
| Base Curve | 7.00 mm to 10.0 mm | ±0.20 mm |
| Back Vertex Power (F'v) | +20.00D to -20.00D(in 0.25D steps) | When 0.00 < $ F'v $ ≤ 10.00 D → ±0.25 DWhen 10.00 < $ F'v $ ≤ 20.00 D → ±0.50 D |
| Cylinder Power (F'c) | -0.25D to -4.00D(in 0.25D steps) | When 0.00 < $ F'c $ ≤ 2.00 D → ±0.25 DWhen 2.00 < $ F'c $ ≤ 4.00 D → ±0.37 D |
| Cylinder Axis | 10° to 180°(in 10° steps) | When 0.00 < $ F'c $ ≤ 1.50 D → ± 8°When $ F'c $ > 1.50 D → ± 5° |
| Multifocal Add Power | +0.25D to +4.00D (0.50D steps) | ±0.37 D |
| Surface Appearance | - | Lenses should be clear with no surface defect |
| Oxygen Permeability(x 10-11(cm²/sec)(mlO2)/(mlx mmHg)) | 19.6 | ±20% |
| Light Transmission - Tinted(@ 380-780nm) | 95% | ±5% |
| Ultraviolet RadiationTransmittance | < 5 % TUVB< 50 %TUVA | TUVB (280 to 315 nm) < 0.05TVTUVA (316 to 380 nm) < 0.50TV |
| Water Content | 55% | ±2% |
| Refractive Index | 1.410 (hydrated) | +0.005 |
- ANSI Z80.20, Ophthalmics - Contact Lenses - Standard Terminology, Tolerances, Measurements And Physicochemical Properties (2010)
INDICATIONS FOR USE V.
The ILICON (ocufilcon D) SPHERICAL contact lenses for daily wear are indicated for the correction of refractive ametropia (myopia and hyperopia) in aphakic or not-aphakic persons with non-diseased that may exhibit astigmatism up to 2.0 diopters that does not interfere with visual acuity. The lens is available clear or tinted and may be used to enhance or alter the apparent color of the eye.
{7}------------------------------------------------
The ILICON (ocufilcon D) TORIC Soft Contact Lenses for daily wear are indicated for the correction of refractive error in aphakic and not-aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 5.00 diopters. The lens is available clear or tinted and may be used to enhance or alter the apparent color of the eye.
The ILICON (ocufilcon D) MULTIFOCAL Soft Contact Lenses for daily wear are indicated for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic or notaphakic persons with non-diseased eyes that may exhibit astigmatism up to 2.0 diopters that does not interfere with visual acuity. The lens is available clear or tinted and may be used to enhance or alter the apparent color of the eye.
The ILICON (ocufilcon D) MULTIFOCAL TORIC Soft Contact Lenses for daily wear are indicated for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic or not-aphakic persons with non-diseased eyes that may exhibit astigmatism up to 5.0 diopters that does not interfere with visual acuity. The lens is available clear or tinted and may be used to enhance or alter the apparent color of the eye.
Daily wear replacement schedules may vary from patient and should be decided by eye care practitioners in consultation with their patients.
Eye care practitioners may prescribe any of the above lenses for frequent/planned replacement wear, with cleaning disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfecting system.
Eye care practitioners may prescribe any of the above lenses for single use daily disposable wear. When prescribed for daily disposable wear the lens is to be discarded after each removal.
VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE DEVICE
The ILICON (ocufilcon D) Soft (Hydrophilic) Daily Wear Contact Lenses (Tinted, Color) are substantially equivalent to the predicate devices identified (K020392 and K150293) in terms of the following:
- 트 USAN contact lens material (ocufilcon D)
- FDA Group 4 (>50% H2O, ionic polymer)
- FDA classification - Soft (hydrophilic) Contact Lens (21 CFR 886.5925)
- 트 Intended use - daily wear contact lenses
- Actions
- 트 Indications for use
- Design configurations available (spherical, toric, multifocal)
- UV absorber
- 트 Cast molded production method
- Pad-printing contact lens tinting method
{8}------------------------------------------------
The following matrix illustrates the production method, lens function and material characteristics of the ILICON (ocufilcon D) Soft (Hydrophilic) Daily Wear Contact Lenses (Tinted, Color), as well as the predicate devices.
| VizionfocusILICON(Subject Device) | CoopervisionFrequency 55 UV(K020392) | I-SEE EZvueUV and color(K150293) | |
|---|---|---|---|
| Indications for Use | Indicated for daily wearfor the correction ofrefractive error inaphakic and not aphakicpersons with non-diseased eyes withmyopia or hyperopiaand/or presbyopia. | Indicated for daily wearfor the correction ofrefractive error inaphakic and notaphakic persons withnon-diseased eyes withmyopia or hyperopia. | For the correctionof visual acuity innot aphakicpersons with non-diseased eyes withmyopia orhyperopia and mayexhibitedrefractiveastigmatism up to2.00 diopters thatdoes not interferewith visual acuity. |
| Actions | The contact lenses act asa refractive medium thatfocus light rays fromnear and distant objectson the retina | The contact lenses actas a refractive mediumthat focus light raysfrom near and distantobjects on the retina | The contact lensesact as a refractivemedium that focuslight rays fromnear and distantobjects on theretina |
| FDA Classification | Soft (hydrophilic)Contact Lens (21 CFR886.5925) | Soft (hydrophilic)Contact Lens (21 CFR886.5925) | Soft (hydrophilic)Contact Lens (21CFR 886.5925) |
| FDA Group | FDA Group 4(>50% H2O, ionicpolymer) | FDA Group 4(>50% H2O, ionicpolymer) | FDA Group 4(>50% H2O,ionic polymer) |
| Production Method | Fully molded | Fully molded | Fully molded |
| USAN name | ocufilcon D | ocufilcon D | ocufilcon D |
| Water Content (%) | 55±2 % | 55±2 % | 55±2 % |
| Oxygen Permeabilityx 10-11 (cm²/sec)(mlO2)/(ml xmmHg @ 35°C)) (revisedFatt method) | 19.6 | 22.0 | 16 |
| Refractive Index(hydrated) | 1.410 | 1.40 | 1.410 |
| Modulus (MPa) | 0.49 | 0.48 | - |
| UV Blocker | Yes | Yes | Yes |
| Pad-Printed Tinting | Yes | No | Yes |
{9}------------------------------------------------
VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
Non-clinical Testing
A series of preclinical testing was performed to demonstrate the safety and effectiveness of the ILICON (ocufilcon D) finished contact lenses. The results support the claim that the ILICON (ocufilcon D) Soft (Hydrophilic) Daily Wear Contact Lenses (Tinted, Color) are substantially equivalent to the currently marketed predicate devices. A summary of the results from the preclinical studies is presented below.
Toxicology:
All non-clinical toxicology tests were conducted in accordance with the GLP regulation.
- · In-Vitro Cytotoxicity: Cytotoxicity testing was performed in accordance with ISO 10993-5 with results indicating that the finished lenses and packaging materials are non-toxic.
- · Systemic Toxicity: The finished lenses and packaging materials meet the requirements of the systemic injection test in accordance with ISO 10993-11 and are considered non-toxic.
- · Acute Ocular Irritation: Acute ocular irritation test was performed in accordance with ISO 10993-10 and produced no ocular irritation for the finished lenses and packaging materials.
Shelf Life:
Testing was performed to evaluate the stability, sterility, and package integrity of the ILICON (ocufilcon D) finished contact lenses over the duration of the labeled expiration date. The data presented supports substantial equivalence of the contact lenses to the already marketed predicate devices.
Physicochemical & Mechanical Properties:
The following tests were completed to verify substantial equivalence: refractive index, water content, Dk, % transmission (visible & UV), tensile strength, modulus, % elongation to break, specific gravity and polymerization residuals. Results of physicochemical and mechanical property testing demonstrate consistent material properties between the ILICON (ocuffileon D) contact lenses and the predicate devices.
{10}------------------------------------------------
Clinical Testing
Clinical testing is not required. The clinical performance of soft (hydrophilic) contact lenses manufactured from ocufilcon D materials has been demonstrated previously.
VIII. CONCLUSIONS
Validity of Scientific Data
Laboratories under Good Laboratory Practice regulations conducted toxicology, microbiology, and shelf-life stability studies following scientific protocols. The data were determined to be scientifically valid under 21 CFR 860.7.
Substantial Equivalence
Information presented in this Premarket Notification establishes that the ILICON (ocufilcon D) Soft (Hydrophilic) Daily Wear Contact Lenses (Tinted, Color) are as safe and effective as the predicate device when used in accordance with the labeled directions for use and for the proposed indication.
Risks and Benefits
The risks of the subject device are the same as those normally attributed to the wearing of soft (hydrophilic) daily wear contact lenses. The benefits to the patient are the same as those for other soft (hydrophilic) daily wear contact lenses.
§ 886.5925 Soft (hydrophilic) contact lens.
(a)
Identification. A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.