K Number
K182052
Device Name
Omni Instrument Tray
Manufacturer
Date Cleared
2019-02-26

(210 days)

Product Code
Regulation Number
880.6850
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The Omni Instrument Tray is intended to enclose, protect, and organize the Omni Hysteroscope, Sheaths, and Removable Outflow Channels during sterilization and storage. The Omni Instrument Tray must be used in conjunction with a sterilization wrap that is cleared by FDA for the indicated sterilization cycles. The Omni Instrument Tray has been validated for the following sterilization cycles: Prevacuum Steam 132°C (270°F) 4 minutes exposure 35 minutes dry time Contents: - 1 Omni Hysteroscope - 1 Removable Outflow Channel - 1 XL Removable Outflow Channel - 1 Omni 3.7mm Diagnostic Sheath - 1 Omni 5.5mm Operative Sheath - 1 Omni 6mm Operative Sheath - 3 Seal Caps - 2 Light Guide Adapters STERRAD ® - STERRAD 100S Short - STERRAD NX Standard - STERRAD 100NX Standard Contents: - 1 Omni Hysteroscope - 1 Removable Outflow Channel - 1 XL Removable Outflow Channel - 1 Omni 3.7mm Diagnostic Sheath - 1 Omni 5.5mm Operative Sheath - 1 Omni 6mm Operative Sheath - 3 Seal Caps - 2 Light Guide Adapters
Device Description
The Omni Instrument Tray is intended to enclose, protect, and organize The Omni Hysteroscope, Sheaths, and Removable Outflow Channels during sterilization and storage. The Omni Instrument Tray must be used in conjunction with a sterilization wrap that is cleared by FDA for the indicated sterilization cycles. The Omni Instrument Tray is made of a stainless steel, santoprene, and silicone and is supplied non-sterile and is designed for multiple use for the sterilization of the Omni Hysteroscope set prior to each use of the Hysteroscope.
More Information

No
The device description and performance studies focus on the physical properties and sterilization capabilities of an instrument tray, with no mention of AI or ML.

No
The device is an instrument tray used for sterilization and storage of other medical instruments, not for directly treating a medical condition.

No

Explanation: The device described is an instrument tray designed for sterilization and storage of other medical instruments (like a hysteroscope and sheaths). Its function is to protect and organize these instruments, not to diagnose medical conditions itself.

No

The device description explicitly states it is made of stainless steel, santoprene, and silicone, indicating it is a physical hardware device.

Based on the provided text, the Omni Instrument Tray is not an IVD (In Vitro Diagnostic) device.

Here's why:

  • Intended Use: The intended use is to "enclose, protect, and organize the Omni Hysteroscope, Sheaths, and Removable Outflow Channels during sterilization and storage." This describes a device used for the preparation and storage of other medical devices, not for performing diagnostic tests on biological samples.
  • Device Description: The description reinforces its function as a container for sterilization and storage of surgical instruments.
  • Lack of IVD Characteristics: The text does not mention any of the typical characteristics of an IVD device, such as:
    • Analyzing biological samples (blood, urine, tissue, etc.)
    • Providing diagnostic information about a patient's health
    • Using reagents or assays

The Omni Instrument Tray is a device used in the process of preparing surgical instruments for use, which falls under the category of medical devices used in healthcare settings, but not specifically IVD devices.

N/A

Intended Use / Indications for Use

The Omni Instrument Tray is intended to enclose, protect, and organize the Omni Hysteroscope, Sheaths, and Removable Outflow Channels during sterilization and storage.
The Omni Instrument Tray must be used in conjunction with a sterilization wrap that is cleared by FDA for the indicated sterilization cycles.
The Omni Instrument Tray has been validated for the following sterilization cycles:
Prevacuum Steam: 132°C (270°F), 4 minutes exposure, 35 minutes dry time.
Contents for Prevacuum Steam: - 1 Omni Hysteroscope, - 1 Removable Outflow Channel, - 1 XL Removable Outflow Channel, - 1 Omni 3.7mm Diagnostic Sheath, - 1 Omni 5.5mm Operative Sheath, - 1 Omni 6mm Operative Sheath, - 3 Seal Caps, - 2 Light Guide Adapters.
STERRAD ®: - STERRAD 100S Short, - STERRAD NX Standard, - STERRAD 100NX Standard.
Contents for STERRAD: - 1 Omni Hysteroscope, - 1 Removable Outflow Channel, - 1 XL Removable Outflow Channel, - 1 Omni 3.7mm Diagnostic Sheath, - 1 Omni 5.5mm Operative Sheath, - 1 Omni 6mm Operative Sheath, - 3 Seal Caps, - 2 Light Guide Adapters.

Product codes (comma separated list FDA assigned to the subject device)

KCT

Device Description

The Omni Instrument Tray is intended to enclose, protect, and organize The Omni Hysteroscope, Sheaths, and Removable Outflow Channels during sterilization and storage.
The Omni Instrument Tray must be used in conjunction with a sterilization wrap that is cleared by FDA for the indicated sterilization cycles.
The Omni Instrument Tray is made of a stainless steel, santoprene, and silicone and is supplied non-sterile and is designed for multiple use for the sterilization of the Omni Hysteroscope set prior to each use of the Hysteroscope.
The Omni Instrument Tray has been validated for the following sterilization cycles:
Pre-vacuum Steam: 132°C (270°F), 4 minutes exposure, 35 minutes dry time. Contents: - 1 Omni Hysteroscope, -1 Removable Outflow Channel, -1 XL Removable Outflow Channel, -1 Omni 3.7mm Diagnostic Sheath, -1 Omni 5.5mm Operative Sheath, - 1 Omni 6mm Operative Sheath, - 3 Seal Caps, - 2 Light Guide Adapters.
STERRAD®: STERRAD NXTM Standard, STERRAD 100NXTM Standard, STERRAD 100S Short. Contents: - 1 Omni Hysteroscope, -1 Removable Outflow Channel, -1 XL Removable Outflow Channel, -1 Omni 3.7mm Diagnostic Sheath, -1 Omni 5.5mm Operative Sheath, - 1 Omni 6mm Operative Sheath, - 3 Seal Caps, - 2 Light Guide Adapters.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Clinical Testing: No clinical testing was performed, submitted, referenced or relied upon for the Omni Instrument Tray.
Bench Testing: Non-clinical bench testing and simulated use testing demonstrate the Omni Instrument Tray is similar to the predicate device. Design verification testing included:

  • Biocompatibility testing to ISO 10993-5:2009 (Biological evaluation of medicaldevices-Part 5: Test for in vitro cytotoxicity)
  • Usability testing to IEC 62366-1:2015 (Medical devices-Part 1: Application of usability engineering to medical devices)
  • Packaging testing to ISTA 3A (Packaged-Products for Parcel Delivery System Shipment70 kq (150 lb or less)
  • Design verification testing to hold the Omni hysteroscope. Sheaths, and Removable Outflow Channels (ANSI/AAMI ST77:2013 Containment devices for reusable medical device sterilization)
    • Attribute verification of component fit
  • Reliability testing including durability, drop test, and handle strength (AAMI TIR12:2010 Design, testing, and labeling reusable medical devices for reprocessing in health care facilities: A quide for medical device manufacturers)
    • Latch cycle durability
    • Simulated table height drop test
    • Handle pull strength to sustained load
      Cleaning and Sterilization: The Omni Instrument Tray is supplied non-sterile and required processing through cleaning and sterilization at the health care facility by the end user. Validation testing confirmed the methods and parameters of these processes as specified in the device labeling.
      The above referenced design verification testing and cleaning and sterilization validations demonstrate the subject device met the design specifications.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K121280

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

K171692

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).

0

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

February 26, 2019

Hologic, Inc Rachelle Fitzgerald Senior Regulatory Affairs Specialist 250 Campus Drive Marlborough, Massachusetts 01752

Re: K182052

Trade/Device Name: Omni Instrument Tray Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: January 29, 2019 Received: January 25, 2019

Dear Rachelle Fitzgerald:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

1

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Clarence W. Murray lii III -S

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

2

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
510(k) Number (if known)K182052
Device NameOmni Instrument Tray
Indications for Use (Describe)
The Omni Instrument Tray is intended to enclose, protect, and organize the Omni Hysteroscope, Sheaths, and Removable Outflow Channels during sterilization and storage.
The Omni Instrument Tray must be used in conjunction with a sterilization wrap that is cleared by FDA for the indicated sterilization cycles.
The Omni Instrument Tray has been validated for the following sterilization cycles:
Prevacuum SteamSTERRAD ®
132°C (270°F)- STERRAD 100S Short
4 minutes exposure- STERRAD NX Standard
35 minutes dry time- STERRAD 100NX Standard
Contents:Contents:
- 1 Omni Hysteroscope- 1 Omni Hysteroscope
- 1 Removable Outflow Channel- 1 Removable Outflow Channel
- 1 XL Removable Outflow Channel- 1 XL Removable Outflow Channel
- 1 Omni 3.7mm Diagnostic Sheath- 1 Omni 3.7mm Diagnostic Sheath
- 1 Omni 5.5mm Operative Sheath- 1 Omni 5.5mm Operative Sheath
- 1 Omni 6mm Operative Sheath- 1 Omni 6mm Operative Sheath
- 3 Seal Caps- 3 Seal Caps
- 2 Light Guide Adapters- 2 Light Guide Adapters
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)Page 1 of 1
K182052/S001Attachment 3M
PSC Publishing Services (301) 443-6740
EF
1 of 1

3

510(K) SUMMARY K182052

Date: February 20, 2019

510(k) Submitter:

Hologic, Inc. 250 Campus Drive Marlborough, MA 01752 Attn: Rachelle D. Fitzgerald P: 508.263.8631 F: 877.793.6434

Establishment Reqistration Number: 1222780

Trade Name: Omni Instrument Tray

Common/Usual Name: Sterilization Tray

Regulation Name: Sterilization wrap

Regulation Number: 21 CFR 880.6850

Product Code: KCT

Classification: Class II

Panel: General Hospital

PREDICATE DEVICE

Tradename: MyoSure Instrument Tray Submitter/510(k) Holder: Hologic, Inc. 510(k) #: K121280 Product Code: KCT Regulation: 21CFR 880.6850

DEVICE DESCRIPTION

The Omni Instrument Tray is intended to enclose, protect, and organize The Omni Hysteroscope, Sheaths, and Removable Outflow Channels during sterilization and storage.

The Omni Instrument Tray must be used in conjunction with a sterilization wrap that is cleared by FDA for the indicated sterilization cycles.

The Omni Instrument Tray is made of a stainless steel, santoprene, and silicone and is supplied non-sterile and is designed for multiple use for the sterilization of the Omni Hysteroscope set prior to each use of the Hysteroscope.

The Omni Instrument Tray has been validated for the following sterilization cycles:

4

Pre-vacuum SteamSTERRAD®
132°C (270°F)STERRAD NXTM Standard
4 minutes exposureSTERRAD 100NXTM Standard
35 minutes dry timeSTERRAD 100S Short
Contents:Contents:
- 1 Omni Hysteroscope- 1 Omni Hysteroscope
-1 Removable Outflow Channel-1 Removable Outflow Channel
-1 XL Removable Outflow Channel-1 XL Removable Outflow Channel
-1 Omni 3.7mm Diagnostic Sheath-1 Omni 3.7mm Diagnostic Sheath
-1 Omni 5.5mm Operative Sheath-1 Omni 5.5mm Operative Sheath
- 1 Omni 6mm Operative Sheath- 1 Omni 6mm Operative Sheath
- 3 Seal Caps- 3 Seal Caps
- 2 Light Guide Adapters- 2 Light Guide Adapters

INDICATIONS FOR USE:

The Omni Instrument Tray is intended to enclose, protect, and organize The Omni Hysteroscope, Sheaths, and Removable Outflow Channels during sterilization and storage.

The Omni Instrument Tray must be used in conjunction with a sterilization wrap that is cleared by FDA for the indicated sterilization cycles.

The Omni Instrument Tray has been validated for the following sterilization cycles:
-----------------------------------------------------------------------------------------
Pre-vacuum SteamSTERRAD®
132°C (270°F)STERRAD NX™ Standard
4 minutes exposureSTERRAD 100NXT™ Standard
35 minutes dry timeSTERRAD 100S Short
Contents:Contents:
- 1 Omni Hysteroscope- 1 Omni Hysteroscope
-1 Removable Outflow Channel-1 Removable Outflow Channel
-1 XL Removable Outflow Channel-1 XL Removable Outflow Channel
-1 Omni 3.7mm Diagnostic Sheath-1 Omni 3.7mm Diagnostic Sheath
-1 Omni 5.5mm Operative Sheath-1 Omni 5.5mm Operative Sheath
- 1 Omni 6mm Operative Sheath- 1 Omni 6mm Operative Sheath
- 3 Seal Caps- 3 Seal Caps
- 2 Light Guide Adapters- 2 Light Guide Adapters

5

Technological Omni Instrument Trav Predicate Device Reference Device Characteristic (K121280) (K171692) The tray is intended The tray is intended to The tray is intended Intended Use enclose, protect, and to enclose, protect, to enclose, protect, and organize and organize organize medical medical devices. medical devices. devices. Steam Autoclave Steam Autoclave; Pre-Steam Autoclave; Biocompatibility Vacuum 132°C (270°F), Pre-Vacuum 132°C Pre-Vacuum 132°C 4 min Exposure, 30 min (270°F), 3 min (270°F); 4 min Exposure; Exposure, 15-30 min dry time; 35 min dry time dry time; STERRAD® 100S Short STERRAD NX Standard STERRAD: NX cycle STERRAD 100NX Standard and Standard Hemolytic Test 10993-4 Express, 100S Short STERRAD 100S Short Biological evaluation of Cytotoxicity ISO 10993medical devices-Part: Cytotoxicity ISO 5:2009 Biological Selection of tests for 10993-5:2009 evaluation of medical interactions with blood; Biological evaluation devices-Part 5: Test for in Irritation 10993-10: Tests of medical devicesvitro cytotoxicity. for irritation and skin Part 5: Test for in vitro cytotoxicity. sensitization. Under the condition of the study, the device Under the conditions of Under the condition extract was found not to the study, device extract of the study, the be cytotoxic to the was found not to be device extract was testing assay. hemolytic to the testing found not to be assay. cytotoxic to the testing assay. Under the conditions of the study, the polar and non-polar device extracts were found not to be a sensitizer nor an irritant to the animal model.

TECHNOLOGICAL CHARACTERISTICS:

6

| Technological
Characteristic | Omni Instrument Tray | Predicate Device
(K121280) | Reference Device
(K171692) |
|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Prescription Only | No | No | No |
| Device Description | Stainless Steel
tray and lid
with metal lid locking
mechanism and handles
Accessories include:
silicone
grommets, and stainless
steel
posts | Fenestrated plastic tray
and
lid with metal lid locking
mechanism and handles
Accessories include:
silicone grommets, and
stainless steel posts | Rigid containment
device consisting of
a base with a lid
which can be
fastened by a
latching
mechanism. The
device is perforated
in order to enable
reprocessing of
enclosed medical
devices held in
place by silicone
retainers. |
| Materials of
construction
Tray
Lid
Handles/
Latching
Grommets | Stainless Steel/
Santoprene
Stainless Steel
Stainless Steel/
Silicone/Santoprene
Silicone | Radel®R resin
Radel®R resin
Stainless Steel
Silicone | Stainless steel
Stainless steel
Stainless Steel
Silicone |
| Dimensions
Length
Width
Height | 16.20"
10.11"
2.27" | 13.45"
8.41"
2.00" | 22.83"
3.03"
10.63" |
| Sterilization cycles | Steam Autoclave:
Pre-Vacuum 132°C
(270°F);
4 min Exposure;
35 min dry time
STERRAD NX
Standard
STERRAD 100NX
Standard STERRAD
100S Short | Steam Autoclave
Pre-Vacuum 132°C
(270°F),
4 min Exposure, 30 min
dry time;
STERRAD® 100S Short
cycle | Steam Autoclave
Pre-Vacuum 132°C
(270°F), 3 min
Exposure, 15-30 min
dry time;
STERRAD: NX
Standard and
Express, 100S
Short |
| Capacity
(Maximum total
weight) | 25 lbs | 13 lbs | 8.708 lbs |
| Reusable | Yes | Yes | Yes |
| Perforations
(Vent to Volumne
Ratio) | .435 in²/in³ | .020 in²/in³ | 72% |
| Technological
Characteristic | Omni Instrument Tray | Predicate Device
(K121280) | Reference Device
(K171692) |
| Patient Contacting
Materials | None | None | None |

7

PERFORMANCE TESTING

Clinical Testing

No clinical testing was performed, submitted, referenced or relied upon for the Omni Instrument Tray.

Bench Testing

Non-clinical bench testing and simulated use testing demonstrate the Omni Instrument Tray is similar to the predicate device. Design verification testing included:

  • . Biocompatibility testing to ISO 10993-5:2009 (Biological evaluation of medicaldevices-Part 5: Test for in vitro cytotoxicity)
  • Usability testing to IEC 62366-1:2015 (Medical devices-Part 1: Application of usability . engineering to medical devices)
  • . Packaging testing to ISTA 3A (Packaged-Products for Parcel Delivery System Shipment70 kq (150 lb or less)
  • Design verification testing to hold the Omni hysteroscope. Sheaths, and Removable ● Outflow Channels (ANSI/AAMI ST77:2013 Containment devices for reusable medical device sterilization)
    • Attribute verification of component fit ●
  • Reliability testing including durability, drop test, and handle strength (AAMI TIR12:2010 Design, testing, and labeling reusable medical devices for reprocessing in health care facilities: A quide for medical device manufacturers)
    • . Latch cycle durability
    • Simulated table height drop test ●
    • Handle pull strength to sustained load

Cleaning and Sterilization

The Omni Instrument Tray is supplied non-sterile and required processing through cleaning and sterilization at the health care facility by the end user. Validation testing confirmed the methods and parameters of these processes as specified in the device labeling.

8

The following standards were used during the validation process:

Standard ReferenceStandard Title
AAMI/ANSI ST79:2017Comprehensive guide to steam sterilization and sterility assurance in
health care facilities
ANSI/AAMI ST77:2013Containment devices for reusable medical device sterilization
ISO 17664:2017Processing Of Health Care Products - Information To Be Provided By The
Medical Device Manufacturer For The Processing Of Medical Devices
ISO 17665-1:2006Sterilization Of Health Care Products - Moist Heat - Part 1:
Requirements For The Development, Validation And Routine Control Of
A Sterilization
Process For Medical Devices
ISO TS 17665-2:2009Sterilization Of Health Care Products - Moist Heat - Part 2: Guidance On
The Application Of ISO 17665-1
ANSI/AAMI/ISO
14937:2009/(R)
2013Sterilization of Health Care Products-General Requirements for
characterization of a sterilizing agent and the development, validation and
routine control of a sterilization process for medical devices
AAMI TIR12:2010Designing, testing, and labeling reusable medical devices for reprocessing
in health care facilities: A guide for medical device manufacturers
AAMI TIR30:2011A compendium of processes, materials, test methods, acceptance criteria
for cleaning reusable medical devices

The above referenced design verification testing and cleaning and sterilization validations demonstrate the subject device met the design specifications.

CONCLUSION

The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device MyoSure Instrument Tray (K121280).