K Number
K182051
Date Cleared
2018-10-02

(63 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

ED-530XT is intended for the visualization of the duodenum and upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.

Device Description

FUJIFILM Duodenoscope Model ED-530XT is comprised of three general sections: a control portion, an insertion portion and an umbilicus. The control portion controls the angulation of the endoscope. This portion also controls the flexibility of the distal end in the endoscope. The insertion portion contains glass fiber bundles, several channels and a complementary Charge-Coupled Device (CCD) image sensor in its distal end. The channels in the insertion portion assist in delivering air/suction as well as endoscope accessories, such as forceps. The glass fiber bundles allow light to travel through the endoscope and emit light from the tip of the insertion portion to illuminate the body cavity. This provides enough light to the CCD image sensor to capture an image and display it on the monitor. The umbilicus consists of electronic components needed to operate the endoscope when plugged in to the video processor and the light source. The endoscope is used in combination with FUJIFILM's video processors, light sources and peripheral devices such as monitor, printer, foot switch, and cart.

AI/ML Overview

Here's an analysis of the acceptance criteria and study information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Test/CharacteristicReported Device Performance
Electrical SafetyANSI/AAM ES 60601-1:2012Met standard
IEC 60601-1-2:2007Met standard
IEC 60601-1-6:2013Met standard
IEC 60601-2-18:2009Met standard
BiocompatibilityISO 10993-1:2009Met standard (adopted from predicate device testing)
ISO 10993-5:2009Met standard (adopted from predicate device testing)
ISO 10993-10:2010Met standard (adopted from predicate device testing)
Functional PerformanceField of viewMet pre-defined acceptance criteria
Bending capabilityMet pre-defined acceptance criteria
Air supply rateMet pre-defined acceptance criteria
Water supply rateMet pre-defined acceptance criteria
Suction rateMet pre-defined acceptance criteria
Working lengthMet pre-defined acceptance criteria
Forceps standing angleMet pre-defined acceptance criteria
Forceps standing tensionMet pre-defined acceptance criteria
Forceps channel diameterMet pre-defined acceptance criteria
Viewing directionMet pre-defined acceptance criteria
ResolutionMet pre-defined acceptance criteria
LG outputMet pre-defined acceptance criteria
Transport & Storage ConditionsTemperature range (-20°C to 60°C)Supported (expansion of conditions)
Humidity range (10 to 85% RH)Supported (expansion of conditions)

2. Sample Size Used for the Test Set and Data Provenance

The provided text does not explicitly state the sample size for the test set regarding the functional performance tests (e.g., field of view, bending capability, etc.). It only mentions that Fujifilm conducted "performance testing on the proposed device ED-530XT." Also, the provenance of the data (e.g., country of origin, retrospective or prospective) is not specified.

3. Number of Experts Used to Establish Ground Truth and Qualifications

The provided document does not mention the use of experts to establish ground truth for the performance tests. The nature of the tests (e.g., measuring field of view, bending capability) suggests objective measurements rather than expert interpretation.

4. Adjudication Method for the Test Set

Since experts are not mentioned in the establishment of ground truth for performance tests, an adjudication method is not applicable and not mentioned.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

The document does not indicate that a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done. The device in question is an endoscope, and the performance tests focus on its physical and functional attributes, not on its assistance to human readers in interpretation. Therefore, an effect size of human readers improving with AI vs. without AI assistance is not relevant or reported here.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

This is a physical medical device (an endoscope), not an AI algorithm. Therefore, a standalone performance study in the context of "algorithm only without human-in-the-loop performance" is not applicable and not mentioned. The tests performed are on the device itself.

7. Type of Ground Truth Used

The ground truth used for the performance tests appears to be objective measurements against pre-defined engineering and performance specifications. For electrical safety and biocompatibility, the ground truth is established by the relevant international consensus standards (e.g., IEC, ISO).

8. Sample Size for the Training Set

Since this is a physical medical device and not an AI algorithm, the concept of a "training set" in the context of machine learning is not applicable and not mentioned.

9. How the Ground Truth for the Training Set Was Established

As there is no training set for an AI algorithm, this information is not applicable. The "ground truth" for the device's performance is established by its adherence to specified engineering parameters and relevant safety and performance standards.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 2, 2018

FUJIFILM Corporation % Jeffrey Wan Specialist, Regulatory Affairs FUJIFILM Medical Systems U.S.A., Inc. 10 High Point Drive Wayne, NJ 07470

Re: K182051

Trade/Device Name: FUJIFILM Duodenoscope Model ED-530XT Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FDT Dated: September 19, 2018 Received: September 20, 2018

Dear Jeffrey Wan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/7 description: The image shows the name Jeffrey W. Cooper -S, followed by the date 2018.10.02 and the time 21:34:02. The image also contains the timestamp -04'00'. There is also a logo on the left side of the image.

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K182051

Device Name

FUJIFILM Duodenoscope Model ED-530XT

Indications for Use (Describe)

ED-530XT is intended for the visualization of the duodenum and upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

FUJIFILM Duodenoscope Model ED-530XT

Date: July 30, 2018

Submitter's Information:

FUJIFILM Corporation 798 Miyanodai Kaisei-Machi Ashigarakami-Gun, Kanagawa, 258-8538, Japan FDA Establishment Registration Number: 3001722928

Contact Person:

Jeffrey Wan Specialist, Regulatory Affairs Telephone: (973) 709-2219 Facsimile: (201) 995-2452 E-Mail: jeffrey.wan@fujifilm.com

Identification of the Proposed Device:

Proprietary/Trade Name:FUJIFILM Duodenoscope Model ED-530XT
Common Name:Endoscope
Device Class:Class II
Review Panel:Gastroenterology/Urology
Classification:Endoscope and accessories, 21 C.F.R. § 876.1500
Product Code:FDT

Predicate Device:

FUJIFILM Duodenoscope Model ED-530XT (K152257)

Intended Use / Indications for Use:

ED-530XT is intended for the visualization of the duodenum and upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.

Device Description:

FUJIFILM Duodenoscope Model ED-530XT is comprised of three general sections: a control portion, an insertion portion and an umbilicus. The control portion controls the angulation of the endoscope. This portion also controls the flexibility of the distal end in the endoscope. The insertion portion contains glass fiber bundles, several channels and a complementary Charge-Coupled Device (CCD) image sensor in its distal end. The channels in the insertion portion assist in delivering air/suction as well as endoscope accessories, such as forceps. The glass fiber bundles allow light to travel through the endoscope and emit light from the tip of the insertion portion to illuminate the body cavity. This provides enough light to the CCD image sensor to capture an image and display it on the monitor. The umbilicus

{4}------------------------------------------------

consists of electronic components needed to operate the endoscope when plugged in to the video processor and the light source. The endoscope is used in combination with FUJIFILM's video processors, light sources and peripheral devices such as monitor, printer, foot switch, and cart.

Technological Characteristics:

The proposed device ED-530XT differs from the predicate device in the following minor modifications:

  • . Compatibility with Video Processor VP-7000 and Light Source BL-7000 (K163675).
  • Compliance to IEC60601-1 edition 3.1 to support compatibility with VP-7000 and BL-7000. .
  • Expansion of transport and storage conditions to a temperature range of -20°C to 60°C and . a humidity range of 10 to 85% RH.
  • . Use of a new epoxy resin in the manufacturing process. Both the new and original epoxy resins can be used for repair.
  • . Compatibility with disposable cleaning brushes WB11003DV and WB7025DC.

Performance Data:

Electrical safety of the proposed device was evaluated using following standards: ANSI/AAM ES 60601-1:2012, IEC 60601-1-2:2007, IEC 60601-1-6:2013, and IEC 60601-2-18:2009.

The proposed device ED-530XT was adopted into biocompatibility testing conducted on the predicate device ED-530XT using the following consensus standards: ISO 10993-1:2009, ISO 10993-5:2009, and ISO 10993-10:2010.

Fujifilm conducted the following performance testing on the proposed device ED-530XT to ensure that the modified device performs equivalently to the predicate device:

  • . Field of view
  • . Bending capability
  • . Air supply rate
  • Water supply rate
  • Suction rate
  • . Working length
  • Forceps standing angle
  • Forceps standing tension
  • . Forceps channel diameter
  • . Viewing direction
  • Resolution
  • . LG output

{5}------------------------------------------------

In all cases, the devices met the pre-defined acceptance criteria for the test.

Substantial Equivalence:

The company's ED-530XT has the same intended use and indications for use as the previously cleared predicate ED-530XT (K152257). In addition, the proposed device has similar technological characteristics and principles of operation as the predicate. The minor differences between the proposed devices and their predicate devices do not raise new or additional questions of safety or effectiveness of the proposed devices. Thus, the proposed device ED-530XT is substantially equivalent to its predicate device.

Conclusions:

The modified ED-530XT is substantially equivalent to the predicate ED-530XT and conforms to applicable medical device safety and performance standards.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.