(351 days)
Not Found
No
The device description and intended use are solely focused on the physical properties and function of sterilization pouches and rolls, with no mention of any computational or analytical capabilities that would involve AI/ML.
No.
This device is a sterilization pouch/roll intended to enclose other medical devices for sterilization and maintain their sterility, not to provide therapeutic treatment itself.
No
Explanation: The device is a sterilization pouch/roll intended to enclose medical devices for sterilization and maintain their sterility. While it includes chemical indicators to show exposure to sterilization processes, these do not diagnose any condition or disease. They merely indicate a physical event (exposure to sterilizing agents).
No
The device described is a physical sterilization pouch/roll made of paper and plastic film with chemical indicators. It is a hardware device used for packaging and maintaining sterility of other medical devices.
Based on the provided text, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostic devices are used to examine specimens derived from the human body (like blood, urine, tissue) to provide information for diagnosis, monitoring, or screening.
- Device Intended Use: The intended use of the Heat Sealing Sterilization Pouch/Roll is to enclose medical devices for sterilization and maintain their sterility. It is used on medical devices, not on or with human specimens for diagnostic purposes.
- Device Description: The description focuses on the physical characteristics of the pouches/rolls, their ability to withstand sterilization processes, and their function in maintaining sterility. It does not mention any interaction with biological samples or diagnostic procedures.
- Performance Studies: The listed performance studies are related to the physical properties of the packaging, its ability to maintain sterility, and the effectiveness of the sterilization indicators. These are not studies typically associated with IVD performance (like sensitivity, specificity, etc.).
The device is clearly intended as a sterilization packaging material for other medical devices.
N/A
Intended Use / Indications for Use
The Heat Sealing Sterilization Pouch Flat are intended to enclose another medical devices that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
The Heat Sealing Sterilization Pouch Flat are intended for sterilization of dental instruments, excluding complex devices (endoscopes and instruments with lumen/channels).
The intended sterilization cycles are listed below:
Prevacuum steam: 4 minutes at 132 ℃; 10 minute dry time.
Ethylene oxide: 1 hours at 55 °C; relative humidity between 40%- 80%; 100% ethylene oxide at a concentration of 740 mg/L, 7 day aeration time at 20℃.
The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EtO sterilization process.
The Heat Sealing Sterilization Pouch Flat are not intended use for any load with lumen/channels and complex device. The maximum wrapped package weight for 500 x 600 mm size pouch is 1470 g, 200 x 300 mm size pouch is 300x, 90 x 559 mm size pouch is 250g, and 102 x 203 size pouch is 110g.
The sterilization pouch maintains the sterility of the enclosed devices for up to 6 months post sterilization.
Device Size (mm): 90x559, 102x203, 127x381, 150x300, 150x300, 152x254, 152x559, 191x330, 200x300, 203x406, 300x400, 305x381, 305x457, 400x500, 457x559, 500x600.
Heat Sealing Sterilization Pouch Gusseted
The Heat Sealing Sterilization Pouch Gusseted are intended to enclose another medical devices that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
The Heat Sealing Sterilization Pouch Gusseted are intended for sterilization of dental instruments, excluding complex devices (endoscopes and instruments with lumen/channels).
The intended sterilization cycles are listed below:
Prevacuum steam; 4 minutes at 132 ℃; 10 minute dry time. Ethylene oxide: 1 hours at 55 °C; relative humidity between 40%- 80%; 100% ethylene oxide at a concentration of 740 mg/L, 7 day aeration time at 20°C.
The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EtO sterilization process.
The Heat Sealing Sterilization Pouch Gusseted are not intended use for any load with lumen/channels and complex device. The maximum wrapped package weight for 300 x 65 x 560 mm size pouch is 1050g, 200 x 33 x 340 mm size pouch is 400g, and 65 x 40 x 254 mm size pouch is 200g.
The sterilization pouch maintains the sterility of the enclosed devices for up to 6 months post sterilization. Device Size (mm): 65x40x254.70 x 35 x 145.90 x 50 x 155.128 x 50 X 190.137x40x305, 150x50x380, 165x75x295. 200x33x340, 200x98x440,250x60x340,250x60x480,300x65x560.
Heat Sealing Sterilization Pouch Roll Flat
The Heat Sealing Sterilization Roll Flat are intended to be used to enclose another medical devices that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
The Heat Sealing Sterilization Roll Flat are intended for sterilization of dental instruments, excluding complex devices (endoscopes and instruments with lumen/channels).
The intended sterilization cycles are listed below:
Prevacuum steam; 4 minutes at 132 ℃; 10 minute dry time.
Ethylene oxide: 1 hours at 55 °C; relative humidity between 40%- 80%; 100% ethylene oxide at a concentration of 740 mg/L, 7 day aeration time at 20°C.
The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EtO sterilization process.
The Heat Sealing Sterilization Roll Flat are not intended use for any load with lumen/channels and complex device. The maximum wrapped package weight for 600 x 200 mm size pouch is 910g, 300 x 200 mm size pouch is 450g and 50 x 200 mm size pouch is 100g.
The sterilization roll maintains the sterility of the enclosed devices for up to 6 months post sterilization. Device Size (mm): 50 x 200,100 x 200,120 x 200,150 x 200,200 x 200,250 x 200,300 x 200,350 x 200, 400 x 200, 500 x 200,600 x 200.
Heat Sealing Sterilization Pouch Roll Gusseted
The Heat Sealing Sterilization Roll Gusseted are intended to enclose another medical devices that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
The Heat Sealing Sterilization Roll Gusseted are intended for sterilization of dental instruments, excluding complex devices (endoscopes and instruments with lumen/channels).
The intended sterilization cycles are listed below:
Prevacuum steam; 4 minutes at 132 ℃; 10 minute dry time.
Ethylene oxide: 1 hours at 55 °C; relative humidity between 40%- 80%; 100% ethylene oxide at a concentration of 740 mg/L, 7 day aeration time at 20°C.
The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EtO sterilization process.
The Heat Sealing Sterilization Roll Gusseted are not intended use for any load with lumen/channels and complex device. The maximum wrapped package weight for 500 x 100 mm size pouch is 750g, 250 x 100 mm size pouch is 375g and 75 x 100 mm size pouch is 100g.
The sterilization roll maintains the sterility of the enclosed devices for up to 6 months post sterilization. Device Size (mm): 75 x 100,100 x 100,200 x 100,200 x 100,300 x 100,350 x 100,400 x 100,500 x 100,
Product codes
FRG, JOJ
Device Description
The Heat Sealing Sterilization Pouch/Roll is intended to contain medical devices to be terminally sterilized by the EtO or Steam sterilization process. The recommended sterilization cycle parameter is:
Prevacuum steam: 4 minutes at 132 ℃; 10 minute dry time.
Ethylene oxide: 1 hours at 55 ℃; relative humidity between 40%- 80%; 100% ethylene oxide at a concentration of 740 mg/L, 7 day aeration time at 20℃.
The medical devices are inserted into the Pouch/Roll, sealed, and then sterilized for the EtO or Steam Sterilization Process. The heat sealed pouch/roll are heat sealed prior to sterilization processing. After completion of the sterilization process, the Pouch/Roll maintain sterility of the enclosed medical devices until the seal is opened.
The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 6 months post sterilization.
The Pouch/Roll is printed with a chemical indicator bar that changes from Pink to Yellow (EtO) or Blue to Dark Green (Steam) when exposed to EtO gas or Steam vapor during process. The EtO and Steam Chemical Indicator offers an addition way to verity processing in the sterilization cycle. The Chemical Indicator should be used in addition to, not in place of, the biological indicator. The EtO and Steam Chemical Indicators do not signify sterilization; they only indicate that the indicator has been exposed to the EtO or Steam.
The Sterilization Pouch/Roll is offered in the follow 4 types:
- Heat Sealing Sterilization Pouch Flat;
- Heat Sealing Sterilization Pouch Gusseted;
- Heat Sealing Sterilization Pouch Roll Flat;
- Heat Sealing Sterilization Pouch Roll Gusseted.
The defining characteristics of the 4 types as follows:
Heat Sealing Sterilization Pouch Flat: These pouches are made from a medical grade paper and plastic (CPP/PET) film that is heat sealed on three sides, the forth side is left opened and will be heat-sealed when using. The Process Indicators Ink printed on the Pouch will exhibit a color change after the pouch is exposed to EtO gas or Steam.
Heat Sealing Sterilization Pouch Gusseted: These rolls are the same with the Heat Sealing Sterilization Pouch Flat, except that the plastic film is folded on both longest sides instead of flat. This design is convenient to enclose the medical devices with certain height.
Heat Sealing Sterilization Pouch Roll Flat: These rolls are made from a medical grade paper and plastic (CPP/PET) film that are heat sealed on opposite two sides. It will be cut into the suitable length and the opened sides will be heat-sealed. The indicators printed on it are the same with the sterilization pouches.
Heat Sealing Sterilization Pouch Roll Gusseted: These rolls are the same with the flat sterilization roll, except that the plastic film is folded on both longest sides instead of flat. This design is convenient to enclose the medical devices with certain height.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
health care provider
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Non-clinical tests were conducted with the subject device. The test results demonstrated that the subject device met the acceptance criteria. A list on the non-clincial test are provided below:
- ISO 14937:2009 Sterilization of health care products - General requirements or characterization of a sterilizing agent and the development, validation and routine control of a sterilization process for medical device
- ASTM D638-14 Standard Test Method For Tensile Properties of Plastics;
- ASTM F2251-03 Standard Test Method for thickness measurement of flexible packaging A material;
- ASTM D1922-03 Standard Test Method for Propagation tear resistance of plastic film and thin sheeting by pendulum method;
- ISO 5636-3:2013 Paper and board - Determination of air permeance (medium range) - Part 3: Bendtsen method;
- ASTM F1140/f1140M-13 Standard Test Methods for internal pressurization failure resistance of unrestrained package;
- ASTM F1608-00 Standard test methods for Microbial Ranking of Porous packaging materials; A
- ISO 10993-7:2009 Biological Evaluation of Medical Device - Part 7: Ethylene Oxide Sterilization residuals;
- ISO 11140-1:2009 Sterilization of Health Care Products - Chemical Indicators - Part 1: General Requirements;
- ASTM F1929-12 Standard test methods for detecting seal leaks in porous medical packaging by dye penetration.
- A ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro Cytotoxicity;
- ISO 10993-10:2010 Biological Evaluation of Medical Devices Part 10: Tests for irritation A and skin sensitization;
- Shelf Life Validation Test, validate the shelf life performance to the proposed device as real time aging method.
- Sterilization Process Validation Test of Self Sealing Sterilization Pouch for EO and Steam sterilization process
- Verification Test of Self Sealing Sterilization Pouch for EO and Steam Sterilization Process
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Microorganism Penetration testing as ASTM F1608: Retention rate >99.9%
Ethylene oxide residues testing as ISO 10993-7: 30 kPa
Chemical Indicators performance testing as ISO 11140-1: EO indicator, Initial Color: Pink -> Signal Color: Yellow; Steam indicator, Initial Color: Blue -> Signal Color: Dark Green
Tensile strength Testing as ASTM D882: >7.0 KN/m
Thickness testing as ASTM F2251: (138+/-3) micrometer
Air Permeability Coefficient Testing as ISO 5636-3: > 8micrometer/Pa.s
Tear Strength testing as ASTM D1922: > 500 mN/15mm
Seal strength testing as ASTM F88 / F88M: >4 N/15mm
Dye Penetration Testing as ASTM F1929: No Dye Leakage
Dry Validation: 10 mins
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 880.6850 Sterilization wrap.
(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 9, 2019
Tianchang Jiarui Packaging Material Co., Ltd. % Ray Wang General Manager Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District Beijing, 102401 Cn
Re: K181957
Trade/Device Name: Heat Sealing Sterilization Pouch Flat, Heat Sealing Sterilization Pouch Gusseted, Heat Sealing Sterilization Pouch Roll Flat, Heat Sealing Sterilization Pouch Roll Gusseted Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG, JOJ Dated: June 10, 2019 Received: June 13, 2019
Dear Ray Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
1
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportinemdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Elizabeth Claverie, M.S. Assistant Director for THT4B2 Acting Assistant Director for THT4B1 DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
2
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K181957
Device Name
Heat Sealing Sterilization Pouch Flat, Heat Sealing Sterilization Pouch Gusseted, Heat Sealing Sterilization Pouch Roll Flat, Heat Sealing Sterilization Pouch Roll Gusseted
Indications for Use (Describe)
The Heat Sealing Sterilization Pouch Flat are intended to enclose another medical devices that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
The Heat Sealing Sterilization Pouch Flat are intended for sterilization of dental instruments, excluding complex devices (endoscopes and instruments with lumen/channels).
The intended sterilization cycles are listed below:
Prevacuum steam: 4 minutes at 132 ℃; 10 minute dry time.
Ethylene oxide: 1 hours at 55 °C; relative humidity between 40%- 80%; 100% ethylene oxide at a concentration of 740 mg/L, 7 day aeration time at 20℃.
The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EtO sterilization process.
The Heat Sealing Sterilization Pouch Flat are not intended use for any load with lumen/channels and complex device. The maximum wrapped package weight for 500 x 600 mm size pouch is 1470 g, 200 x 300 mm size pouch is 300x, 90 x 559 mm size pouch is 250g, and 102 x 203 size pouch is 110g.
The sterilization pouch maintains the sterility of the enclosed devices for up to 6 months post sterilization.
Device Size (mm): 90x559, 102x203, 127x381, 150x300, 150x300, 152x254, 152x559, 191x330, 200x300, 203x406, 300x400, 305x381, 305x457, 400x500, 457x559, 500x600.
Heat Sealing Sterilization Pouch Gusseted
The Heat Sealing Sterilization Pouch Gusseted are intended to enclose another medical devices that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
The Heat Sealing Sterilization Pouch Gusseted are intended for sterilization of dental instruments, excluding complex devices (endoscopes and instruments with lumen/channels).
The intended sterilization cycles are listed below:
Prevacuum steam; 4 minutes at 132 ℃; 10 minute dry time. Ethylene oxide: 1 hours at 55 °C; relative humidity between 40%- 80%; 100% ethylene oxide at a concentration of 740 mg/L, 7 day aeration time at 20°C.
The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EtO sterilization process.
The Heat Sealing Sterilization Pouch Gusseted are not intended use for any load with lumen/channels and complex device. The maximum wrapped package weight for 300 x 65 x 560 mm size pouch is 1050g, 200 x 33 x 340 mm size pouch is 400g, and 65 x 40 x 254 mm size pouch is 200g.
The sterilization pouch maintains the sterility of the enclosed devices for up to 6 months post sterilization. Device Size (mm): 65x40x254.70 x 35 x 145.90 x 50 x 155.128 x 50 X 190.137x40x305, 150x50x380, 165x75x295. 200x33x340, 200x98x440,250x60x340,250x60x480,300x65x560.
3
Heat Sealing Sterilization Pouch Roll Flat
The Heat Sealing Sterilization Roll Flat are intended to be used to enclose another medical devices that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
The Heat Sealing Sterilization Roll Flat are intended for sterilization of dental instruments, excluding complex devices (endoscopes and instruments with lumen/channels).
The intended sterilization cycles are listed below:
Prevacuum steam; 4 minutes at 132 ℃; 10 minute dry time.
Ethylene oxide: 1 hours at 55 °C; relative humidity between 40%- 80%; 100% ethylene oxide at a concentration of 740 mg/L, 7 day aeration time at 20°C.
The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EtO sterilization process.
The Heat Sealing Sterilization Roll Flat are not intended use for any load with lumen/channels and complex device. The maximum wrapped package weight for 600 x 200 mm size pouch is 910g, 300 x 200 mm size pouch is 450g and 50 x 200 mm size pouch is 100g.
The sterilization roll maintains the sterility of the enclosed devices for up to 6 months post sterilization. Device Size (mm): 50 x 200,100 x 200,120 x 200,150 x 200,200 x 200,250 x 200,300 x 200,350 x 200, 400 x 200, 500 x 200,600 x 200.
Heat Sealing Sterilization Pouch Roll Gusseted
The Heat Sealing Sterilization Roll Gusseted are intended to enclose another medical devices that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
The Heat Sealing Sterilization Roll Gusseted are intended for sterilization of dental instruments, excluding complex devices (endoscopes and instruments with lumen/channels).
The intended sterilization cycles are listed below:
Prevacuum steam; 4 minutes at 132 ℃; 10 minute dry time.
Ethylene oxide: 1 hours at 55 °C; relative humidity between 40%- 80%; 100% ethylene oxide at a concentration of 740 mg/L, 7 day aeration time at 20°C.
The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EtO sterilization process.
The Heat Sealing Sterilization Roll Gusseted are not intended use for any load with lumen/channels and complex device. The maximum wrapped package weight for 500 x 100 mm size pouch is 750g, 250 x 100 mm size pouch is 375g and 75 x 100 mm size pouch is 100g.
The sterilization roll maintains the sterility of the enclosed devices for up to 6 months post sterilization. Device Size (mm): 75 x 100,100 x 100,200 x 100,200 x 100,300 x 100,350 x 100,400 x 100,500 x 100,
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
4
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
5
510(k) Summary
This 510(k) Summary of 510(k) is being submitted in accordance withTitle 21, CFR Section 807.92.
The assigned 510(k) Number: K181957
-
- Date of Preparation: 07/08/2019
-
- Sponsor Identification
Tianchang Jiarui Packaging Material Co., Ltd. Yeshan Industria Zone, Tianchang City, 239300, Anhui Province of China
Contact Person: Zhang Ruiqing Position: General Manager Tel: +86 550 7981778 Fax: +86 550 7323988 Email: sales@cncarate.com
-
- Designated Submission Correspondent
Mr. Ray Wang
- Designated Submission Correspondent
Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, BeiJing, China 102401
Tel: +86-18910677558 Fax: +86-10-56335780 Email: ray.wang@believe-med.com
6
4. Identification of Proposed Device
Trade Name: Heat Sealing Sterilization Pouch Flat/Heat Sealing Sterilization Pouch Gusseted/Heat Sealing Sterilization Pouch Roll Flat/Heat Sealing Sterilization Pouch Roll Gusseted Common Name: Sterilization Pouches & Roll
Model(s):
Heat Sealing Sterilization Pouch Flat
90 x 559mm, 102 x 203 mm, 127 x 381 mm, 130 x 360 mm, 150 x 300 mm, 150 x 300 mm, 152 x 254 mm, 152 x 559 mm, 191 x 330 mm, 200 x 300 mm, 203 x 400 mm, 300 x 400 mm, 305 x 381 mm, 305 x 457 mm, 400 x 500 mm, 300 x 400 mm, 457 x 559 mm, 500 x 600 mm;
Heat Sealing Sterilization Pouch Gusseted
65 x 40 x 254 mm, 70 x 35 x 145 mm, 90 x 50 x 155 mm, 128 x 50 x 190 mm, 137 x 40 x 305 mm, 150 x 50 x 380 mm, 165 x 75 x 295 mm, 200 x 33 x 340 mm, 200 x 98 x 440 mm, 250 x 60 x 340 mm, 250 x 60 x 480 mm, 300 x 65 x 560 mm;
Heat Sealing Sterilization Pouch Roll Flat
75 x 200 mm, 100 x 200 mm, 120 x 200 mm, 150 x 200 mm, 200 x 200 mm, 300 x 200 mm, 350 x 200 mm, 400 x 200 mm, 500 x 200 mm, 600 x 200 mm;
Heat Sealing Sterilization Pouch Roll Gusseted
75 x 100 mm, 100 x 100 mm, 150 x 100 mm, 200 x 100 mm, 300 x 100 mm, 300 x 100 mm, 350 x 100 mm, 400 x 100 mm, 500 x 100 mm;
Regulatory Information
Classification Name: Wrap, Sterilization/Indicator, Physical/Chemical Sterilization Process Classification: 2 Product Code: FRG/JOJ Regulation Number: 21 CFR 880.6850/ 21 CFR 880.2800 Review Panel: General Hospital
Intended Use Statement:
Heat Sealing Sterilization Pouch Flat
The Heat Sealing Sterilization Pouch Flat are intended to enclose another medical devices that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
The Heat Sealing Sterilization Pouch Flat are intended for sterilization of dental instruments,
excluding complex devices (endoscopes and instruments with lumen/channels).
The intended sterilization cycles are listed below:
Prevacuum steam; 4 minutes at 132 ℃; 10 minute dry time.
Ethylene oxide: 1 hours at 55 ℃; relative humidity between 40%- 80%; 100% ethylene oxide at a concentration of 740 mg/L, 7 day aeration time at 20℃.
7
The pouch's external chemical ink indicators are designed to the user that the pouch has undergone either a steam or EtO sterilization process.
The Heat Sealing Sterilization Pouch Flat are not intended use for any load with lumen/channels and complex device. The maximum wrapped package weight for 500 x 600 mm size pouch is 1470 g, 200 x 300 mm size pouch is 300g, 90 x 559 mm size pouch is 250g, and 102 x 203 size pouch is 110g.
The sterilization pouch maintains the sterility of the enclosed devices for up to 6 months post sterilization.
Heat Sealing Sterilization Pouch Gusseted
The Heat Sealing Sterilization Pouch Gusseted are intended to enclose another medical devices that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
The Heat Sealing Sterilization Pouch Gusseted are intended for sterilization of dental instruments, excluding complex devices (endoscopes and instruments with lumen/channels).
The intended sterilization cycles are listed below:
Prevacuum steam: 4 minutes at 132 ℃; 10 minute dry time.
Ethylene oxide: 1 hours at 55 ℃; relative humidity between 40%- 80%; 100% ethylene oxide at a concentration of 740 mg/L, 7 day aeration time at 20℃.
The pouch's external chemical ink indicators are designed to the user that the pouch has undergone either a steam or EtO sterilization process.
The Heat Sealing Sterilization Pouch Gusseted are not intended use for any load with
lumen/channels and complex device. The maximum wrapped package weight for 300 x 65 x 560 mm size pouch is 1050g, 200 x 33 x 340 mm size pouch is 400g, and 65 x 40 x 254 mm size pouch is 200g.
The sterilization pouch maintains the sterility of the enclosed devices for up to 6 months post sterilization.
Heat Sealing Sterilization Pouch Roll Flat
The Heat Sealing Sterilization Roll Flat are intended to enclose another medical devices that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
The Heat Sealing Sterilization Roll Flat are intended for sterilization of dental instruments,
excluding complex devices (endoscopes and instruments with lumen/channels).
The intended sterilization cycles are listed below:
Prevacuum steam; 4 minutes at 132 ℃; 10 minute dry time.
Ethylene oxide: 1 hours at 55 ℃; relative humidity between 40%- 80%; 100% ethylene oxide at a concentration of 740 mg/L, 7 day aeration time at 20℃.
The pouch's external chemical ink indicators are designed to the user that the pouch has undergone either a steam or EtO sterilization process.
The Heat Sealing Sterilization Roll Flat are not intended use for any load with lumen/channels and
8
complex device. The maximum wrapped package weight for 600 x 200 mm size pouch is 910g, 300 x 200 mm size pouch is 450g and 50 x 200 mm size pouch is 100g..
The sterilization roll maintains the sterility of the enclosed devices for up to 6 months post sterilization.
Heat Sealing Sterilization Pouch Roll Gusseted
The Heat Sealing Sterilization Roll Gusseted are intended to be used to enclose another medical devices that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
The Heat Sealing Sterilization Roll Gusseted are intended for sterilization of dental instruments, excluding complex devices (endoscopes and instruments with lumen/channels).
The intended sterilization cycles are listed below:
Prevacuum steam; 4 minutes at 132 ℃; 10 minute dry time.
Ethylene oxide: 1 hours at 55 ℃; relative humidity between 40%- 80%; 100% ethylene oxide at a concentration of 740 mg/L, 7 day aeration time at 20℃.
The pouch's external chemical ink indicators are designed to the user that the pouch has undergone either a steam or EtO sterilization process.
The Heat Sealing Sterilization Roll Gusseted are not intended use for any load with lumen/channels and complex device. The maximum wrapped package weight for 500 x 100 mm size pouch is 750g, 250 x 100 mm size pouch is 375g and 75 x 100 mm size pouch is 100g.
The sterilization roll maintains the sterility of the enclosed devices for up to 6 months post sterilization.
Device Description
The Heat Sealing Sterilization Pouch/Roll is intended to contain medical devices to be terminally sterilized by the EtO or Steam sterilization process. The recommended sterilization cycle parameter is:
Prevacuum steam: 4 minutes at 132 ℃; 10 minute dry time.
Ethylene oxide: 1 hours at 55 ℃; relative humidity between 40%- 80%; 100% ethylene oxide at a concentration of 740 mg/L, 7 day aeration time at 20℃.
The medical devices are inserted into the Pouch/Roll, sealed, and then sterilized for the EtO or Steam Sterilization Process. The heat sealed pouch/roll are heat sealed prior to sterilization processing. After completion of the sterilization process, the Pouch/Roll maintain sterility of the enclosed medical devices until the seal is opened.
The device is intended to allow sterilization of enclosed devices and also to maintain sterility of the enclosed devices until used up to 6 months post sterilization.
The Pouch/Roll is printed with a chemical indicator bar that changes from Pink to Yellow (EtO) or Blue to Dark Green (Steam) when exposed to EtO gas or Steam vapor during process. The EtO and Steam Chemical Indicator offers an addition way to verity processing in the sterilization cycle. The
9
Chemical Indicator should be used in addition to, not in place of, the biological indicator. The EtO and Steam Chemical Indicators do not signify sterilization; they only indicate that the indicator has been exposed to the EtO or Steam.
The Sterilization Pouch/Roll is offered in the follow 4 types:
-
- Heat Sealing Sterilization Pouch Flat;
-
- Heat Sealing Sterilization Pouch Gusseted;
-
- Heat Sealing Sterilization Pouch Roll Flat;
-
- Heat Sealing Sterilization Pouch Roll Gusseted.
The defining characteristics of the 4 types as follows:
Heat Sealing Sterilization Pouch Flat: These pouches are made from a medical grade paper and plastic (CPP/PET) film that is heat sealed on three sides, the forth side is left opened and will be heat-sealed when using. The Process Indicators Ink printed on the Pouch will exhibit a color change after the pouch is exposed to EtO gas or Steam.
Heat Sealing Sterilization Pouch Gusseted: These rolls are the same with the Heat Sealing Sterilization Pouch Flat, except that the plastic film is folded on both longest sides instead of flat. This design is convenient to enclose the medical devices with certain height.
Heat Sealing Sterilization Pouch Roll Flat: These rolls are made from a medical grade paper and plastic (CPP/PET) film that are heat sealed on opposite two sides. It will be cut into the suitable length and the opened sides will be heat-sealed. The indicators printed on it are the same with the sterilization pouches.
Heat Sealing Sterilization Pouch Roll Gusseted: These rolls are the same with the flat sterilization roll, except that the plastic film is folded on both longest sides instead of flat. This design is convenient to enclose the medical devices with certain height.
-
- Identification of Predicate Device(s)
Predicate Device 510(k) Number: K143637 Product Name: U & U Sterilization Pouch and Roll Model Name: U&U Medical Technology Co., Ltd.
- Identification of Predicate Device(s)
10
6. Technological Characteristics
ITEM | Subject Device K181957 | Predicate Device K143637 | Comparison | |
---|---|---|---|---|
Indication | ||||
For | ||||
Use | The Heat Sealing Sterilization Pouch Flat are intended to be used to enclose | |||
another medical devices that is to be sterilized by a health care provider. It is | ||||
intended to allow sterilization of the enclosed medical device and also to | ||||
maintain sterility of the enclosed device until used. | ||||
The Heat Sealing Sterilization Pouch Flat are intended for sterilization of | ||||
dental instruments, excluding complex devices (endoscopes and instruments | ||||
with lumen/channels). | ||||
The intended sterilization cycles are listed below: | ||||
Prevacuum steam; 4 minutes at 132 °C; 10 minute dry time. | ||||
Ethylene oxide: 1 hours at 55 °C; relative humidity between 40%- 80%; 100% | ||||
ethylene oxide at a concentration of 740 mg/L, 7 day aeration time at 20°C. | ||||
The pouch's external chemical ink indicators are designed to indicate to the user | ||||
that the pouch has undergone either a steam or EtO sterilization process. | ||||
The Heat Sealing Sterilization Pouch Flat are not intended use for any load | ||||
with lumen/channels and complex device. The maximum wrapped package | ||||
weight for 500 x 600 mm size pouch is 1470 g, 200 x 300 mm size pouch is | ||||
300g,90x 559mmsize pouch is 250g, and102x203 size pouch is110g. The | ||||
sterilization pouch maintains the enclosed devices for up to 6 months post | ||||
sterilization. | The U&U sterilization pouch and roll are intended to provide health care | |||
workers with an effective method to enclose devices intended for sterilization in | ||||
steam and Ethylene Oxide (EtO). |
The recommended gravity steam sterilization cycle parameters are 30 minutes at
121 °C.
The recommended EtO sterilization cycle is 4 hours at 55 °C with a relative
humidity between 50%- 85% and a sterilant concentration of 600 mg/L.
Furthermore, the sterilization pouch and roll maintains the enclosed devices up
until 90Days post sterilization. Lastly, the pouch's external chemical ink
indicators are designed to indicate to the user that the pouch has undergone
either a steam or EtO sterilization process.
The recommended gravity steam sterilization cycle parameters
Steam sterilization temperature: 121°C (250°F)
Steam sterilization time: 30 minutes.
Drying time: 25 minutes
The recommended EtO sterilization cycle parameters
EtO sterilization temperature: 55°C (130 °F)
Eto sterilization time: 4 hour
Eto sterilization humidity: 50% to 85% RH
EtO sterilization concentration:600mg/L
Aeration time: 8 hours. | Different |
| | | Heat Sealing Sterilization Pouch Gusseted
The Heat Sealing Sterilization Pouch Gusseted are intended to be used to
enclose another medical devices that is to be sterilized by a health care provider. | | |
Table 1 General Comparison
11
| It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
The Heat Sealing Sterilization Pouch Gusseted are intended for sterilization of dental instruments, excluding complex devices (endoscopes and instruments with lumen/channels).
The intended sterilization cycles are listed below:
Prevacuum steam; 4 minutes at 132 °C; 10 minute dry time.
Ethylene oxide: 1 hours at 55 °C; relative humidity between 40%- 80%; 100% ethylene oxide at a concentration of 740 mg/L, 7 day aeration time at 20°C.
The pouch's external chemical ink indicators are designed to indicate to the user that the pouch has undergone either a steam or EtO sterilization process.
The Heat Sealing Sterilization Pouch Gusseted are not intended use for any load with lumen/channels and complex device. The maximum wrapped package weight for 300 x 65 x 560 mm size pouch is 1050g, 200 x 33 x 340 mm size pouch is400g, and65x40x254 mmsizepouch is200g.
The sterilization pouch maintains the enclosed devices for up to 6 months post sterilization.
Heat Sealing Sterilization PouchRoll Flat
The Heat Sealing Sterilization RollFlat areintended tobeused toenclose another medical devices that is to be sterilized by ahealth care provider. It is intended to allowsterilization ofthe enclosed medicaldeviceand also to maintain sterility of theencloseddevice until used.
The Heat Sealing Sterilization RollFlat areintendedforsterilizationofdental instruments, excludingcomplex devices (endoscopes andinstruments with lumen/channels).
The intended sterilizationcycles are listed below: | Aeration Temperature: 60°C
Sterilization load claim:
Two types of sterilization loads were validated.
Load A: Metal medical instruments and Hand-control pen, the total weight is 24lbs.
Load B: Tubing (Silicone) and Surgical Towels. The total weight is 18lbs. |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Prevacuum steam; 4 minutes at 132 °C; 10 minute dry time.
Ethylene oxide: 1 hours at 55 °C; relative humidity between 40%-80%; 100%
ethylene oxide at a concentration of 740 mg/L, 7 day aeration time at 20°C.
The pouch's external chemical ink indicators are designed to indicate to the user
that the pouch has undergone either a steam or EtO sterilization process.
The Heat Sealing Sterilization Roll Flat are not intended use for any load with
lumen/channels and complex device. The maximum wrapped package weight
for 600 x 200 mm size pouch is 910g, 300 x 200 mm size pouch is 450g and 50
x200mmsize pouch is 100g.
The sterilization pouch maintains the enclosed devices for up to 6 months post
sterilization. | Heat Sealing Sterilization PouchRoll Gusseted
The Heat Sealing Sterilization RollGusseted are intended to be used to enclos
another medical devices that is to be sterilized by ahealth care provider. It is
intended to allowsterilization ofthe enclosed medicaldeviceand also to maintain
sterility of theencloseddevice until used.
The Heat Sealing Sterilization RollGusseted areintended for sterilization of
dental instruments, excluding complex devices (endoscopes and instruments wit
lumen/channels).
The intended sterilizationcycles are listed below:
Prevacuum steam; 4minutes at 132 °C; 10 minutedry time.
Ethylene oxide: 1 hoursat55°C;relative humidity between 40%-80%;100%
ethylene oxide at aconcentrationof 740 mg/L, 7 day aeration time at20°C. The
pouch's external chemical ink indicatorsare designed toindicate tothe user that
the pouch has undergone either asteam or EtO sterilization process.
The Heat Sealing Sterilization RollGusseted are not intended use for any load |
12
13
| | with lumen/channels and complex device. The maximum wrapped package
weight for 500 x100mm size pouchis750g, 250x 100 mmsize pouch is 375g
and 75x100 mm size pouch is 100g.
The sterilization pouch maintains the enclosed devices for up to 6 months post
sterilization. | | |
|-------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Material | TopWeb -Medical Porous Paper | Top Web - Medical Porous Paper | Similar |
| Compostion | Bottom Web-MedicalPlastic film(CPP) | Bottom Web - Medical Plastic film(CPP) | |
| | | | |
| | EtO gas indicator ink-Process Indicators
Steam indicator ink-Process Indicators | EtO gas indicator ink-Process Indicators class 1
Steam indicator ink-Process Indicators class 1 | |
| Sterilization
Cycles | Prevacuum steam; 4 minutes at 132 °C; 10 minute dry time.
Ethylene oxide: 1 hours at 55 °C; relative humidity between 40%-80%; 100%
ethylene oxide at a concentration of 740 mg/L, 7 day aeration time at 20°C. | The recommended gravity steam sterilization cycle parameters are 30 minutes at
121 °C. The
recommended EtO sterilization cycle is 4 hours at 55 °C with a relative humidity
between 50%-85% and a sterilant concentration of 600 mg/L. | Different |
| Configuration/
Dimension | Various Size, Heat Sealing | Various Size, Heat Sealing and Self Sealing | Similar |
| Air Permeance | The maximum equivalent pore size diameter shall not exceed 50um. | The maximum equivalent pore size diameter shall not exceed 50um. | Similar |
| Microbial Barrier
Properties
(Packaging
Integrity) | Use ASTM 1608 method, the LRV is more than 4.0 Use ASTM 1929 method,
the inspection result is PASS | Use ASTM 1608 method, the LRV is more than 3.5 Use ASTM 1929 method,
the inspection result is PASS | Similar |
| Material
Compatibility | After sterilization, the materials were not degraded | After sterilization, the materials were not degraded | Similar |
| Biocompatibility | ISO10993-10, Test for Irritation ; ISO10993-10, Test for Skin sensitization; | ISO10993-10,Test forIrritation ;ISO10993-10,Testfor Skin sensitization; | Similar |
| Maintenance
of
Sterility | 6 months | 90 Days | Different |
| or
chemical
indicator endpoint
stability | | | |
| Shelf Life | 2 years | 18 months | Different |
| Drying Time | 10 minutes | 25 minutes | Different |
| Aeration Time | 7 days at 20°C | 8 hours at 60°C | Different |
| Chemical
Indicator Efficacy | Changed color EtO- Pink to Yellow;
Steam- Blue to Dark Green | Changed color EtO- YELLOW to COCOA;
Steam- GREEN to PURPLE | Different |
14
15
7. Non-Clinical Test Conclusion
Non-clinical tests were conducted with the subject device. The test results demonstrated that the subject device met the acceptance criteria. A list on the non-clincial test are provided below:
-
ISO 14937:2009 Sterilization of health care products - General requirements or characterization of a sterilizing agent and the development, validation and routine control of a sterilization process for medical device
-
ASTM D638-14 Standard Test Method For Tensile Properties of Plastics;
- ASTM F2251-03 Standard Test Method for thickness measurement of flexible packaging A material;
-
ASTM D1922-03 Standard Test Method for Propagation tear resistance of plastic film and thin sheeting by pendulum method;
-
ISO 5636-3:2013 Paper and board - Determination of air permeance (medium range) - Part 3: Bendtsen method;
-
ASTM F1140/f1140M-13 Standard Test Methods for internal pressurization failure resistance of unrestrained package;
- ASTM F1608-00 Standard test methods for Microbial Ranking of Porous packaging materials; A
-
ISO 10993-7:2009 Biological Evaluation of Medical Device - Part 7: Ethylene Oxide Sterilization residuals;
-
ISO 11140-1:2009 Sterilization of Health Care Products - Chemical Indicators - Part 1: General Requirements;
-
ASTM F1929-12 Standard test methods for detecting seal leaks in porous medical packaging by dye penetration.
- A ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro Cytotoxicity;
- ISO 10993-10:2010 Biological Evaluation of Medical Devices Part 10: Tests for irritation A and skin sensitization;
-
Shelf Life Validation Test, validate the shelf life performance to the proposed device as real time aging method.
-
Sterilization Process Validation Test of Self Sealing Sterilization Pouch for EO and Steam sterilization process
-
Verification Test of Self Sealing Sterilization Pouch for EO and Steam Sterilization Process
Name of Test | Purpose of the test | Acceptance criteria | Test Results |
---|---|---|---|
Microorganism Penetration | |||
testing as ASTM F1608 | Microorganism | ||
Penetration Test | Retention | ||
rate >99.9% | Meet the acceptance criteria | ||
Ethylene oxide residues testing as | |||
ISO 10993-7 | Ethylene oxide | ||
residues | 30 kPa | Meet the acceptance criteria |
16
F1140/F1140M | for bursting | ||
---|---|---|---|
Chemical Indicators performance | |||
testing as ISO 11140-1 | Chemical | ||
Indicators | |||
performance | EO indicator, | ||
Initial Color: Pink | |||
-> Signal Color: | |||
Yellow; | |||
Steam indicator, | |||
Initial Color: Blue | |||
-> Signal Color: | |||
Dark Green | Meet the acceptance criteria | ||
Tensile strength | |||
Testing as ASTM D882 | Tensile strength | >7.0 KN/m | Meet the acceptance criteria |
Thickness testing as ASTM | |||
F2251 | Variation in | ||
thickness | $(138±3)$ µm | Meet the acceptance criteria | |
Air Permeability Coefficient | |||
Testing as ISO 5636-3 | Air Permeability | ||
Coefficient | > 8μm/Pa.s | Meet the acceptance criteria | |
Tear Strength testing as ASTM | |||
D1922 | Tear Strength | > 500 mN/15mm | Meet the acceptance criteria |
Seal strength testing as ASTM | |||
F88 / F88M | Seal strength test | >4 N/15mm | Meet the acceptance criteria |
Dye Penetration Testing as | |||
ASTM F1929 | Dye Penetration | ||
Test | No Dye Leakage | Meet the acceptance criteria | |
Dry Validation | Dry time | 10 mins | Meet the acceptance criteria |
17
- Clinical Test Conclusion 8.
No clinical study is included in this submission.
9. Conclusion
The conclusions drawn from the non-clinical tests demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed predicate device(K143637)