(211 days)
Not Found
No
The description of the device's operation relies on the oscillometric method and a fixed threshold (25% deviation) for the irregular heartbeat indicator, with no mention of AI or ML techniques.
No
The device is described as a "non-invasive blood pressure measurement device" used for "measuring systolic and diastolic blood pressure and pulse rate." Its primary function is diagnostic (measurement), not therapeutic (treatment or direct amelioration of a condition).
Yes
The device measures systolic and diastolic blood pressure and pulse rate, and includes an irregular heartbeat indicator, all of which are used to assess physiological states and detect potential abnormalities.
No
The device description explicitly states it is a "non-invasive blood pressure measurement device" that uses an "electronic pressure sensor" and "cuff pressure," indicating it includes hardware components beyond just software.
No, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostics are medical devices used to examine specimens taken from the human body, such as blood, urine, or tissue, to provide information for diagnosis, monitoring, or screening.
- Device Function: The GMC Wrist Automatic Blood Pressure Monitor, Model BQ705, measures blood pressure and pulse rate directly from the body using a non-invasive method (oscillometric). It does not analyze any specimens taken from the body.
The device is a non-invasive medical device used for physiological measurement.
N/A
Intended Use / Indications for Use
The GMC Wrist Automatic Blood Pressure Monitor, Model BQ705, is a non-invasive blood pressure measurement device that is used for measuring systolic and diastolic blood pressure and pulse rate using the oscillometric method for adults at home.
The device has an irregular heartbeat (IHB) indicator. The appearance of an irregular heartbeat during measurement, and displays an IHB symbol on the LCD with the reading once the irregular heartbeat is detected.
Product codes (comma separated list FDA assigned to the subject device)
DXN
Device Description
The GMC Wrist Automatic Blood Pressure Monitor, Model BO705, measures systolic and diastolic blood pressure and pulse rate of an adult individual. The method used to define blood pressure is similar to the auscultatory method but uses an electronic pressure sensor rather than a stethoscope and mercury manometer. The device utilizes the oscillometric method whereby the electronic pressure sensor converts variation in cuff pressure to electrical signals. The electrical signals are analyzed and used to define the systolic and diastolic measurements and determine the pulse rate. An irregular heart beat (IHB) indicator is displayed when there is more than a 25% deviation from the averaged pulse rate measurements.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Wrist
Indicated Patient Age Range
Adults
Intended User / Care Setting
Home use
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Bench testing was conducted to demonstrate that the device meets its requirements and specification. The following performance tests were completed:
- ANSI AAMI ES60601-1
- IEC 60601-1-2
- IEC 60601-1-11
- IEC 80601-2-30
The GMC Wrist Automatic Blood Pressure Monitor, Model BQ705 meets all applicable requirements.
Clinical testing was performed to insure clinical accuracy of the device in accordance with ISO 81060-2 as documented in Clinical Investigation Report. 85 patients (38 males and 47 females) were invited for the study. Standard auscultation method was used as the reference blood pressure monitor measuring in the left arm. Blood pressure measurements were repeated alternatively with the device and auscultation in the same arm according to the sequence in ISO 81060-2.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Omron BP652N (K142917)
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 870.1130 Noninvasive blood pressure measurement system.
(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).
0
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the letters "FDA" in a blue square, followed by "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
February 14, 2019
GMC Inc. Yolanda Li Director of Quality Management Div. No. 686, Su Chu Rd. Chuzhou, 239000 CHINA
Re: K181930
Trade/Device Name: GMC Wrist Automatic Blood Pressure Monitor, Model B0705 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: January 11, 2019 Received: January 11, 2019
Dear Yolanda Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
1
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Stephen C. Browning -S5
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K181930
Device Name
GMC Wrist Automatic Blood Pressure Monitor, Model BQ705
Indications for Use (Describe)
The GMC Wrist Automatic Blood Pressure Monitor, Model BQ705, is a non-invasive blood pressure measurement device that is used for measuring systolic and diastolic blood pressure and pulse rate using the oscillometric method for adults at home.
The device has an irregular heartbeat (IHB) indicator. The appearance of an irregular heartbeat during measurement, and displays an IHB symbol on the LCD with the reading once the irregular heartbeat is detected.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) |
---|
Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
Section 5- 510(k) Summary
Image /page/3/Picture/4 description: The image shows the logo for GMC 尚 钉 医疗. The logo consists of the letters "GMC" in gray, followed by a red circle with a white plus sign in the center. To the right of the circle are the Chinese characters 尚 钉 医疗, also in gray. The logo appears to be for a medical company.
GMC Inc. No. 686, Su Chu Rd., Chuzhou, Anhui, China
510(k) Summary
| 510(k) owner's information: | GMC Inc.
No. 686, Su Chu Rd., Chuzhou,
Anhui, China
Tel: +86-550-2172888
Fax: +86-550-2172889 |
|-----------------------------|-----------------------------------------------------------------------------------------------------------|
| Official Contact: | Yolanda Lin
Bus. Phone: 86-550-2172888
Fax: 86-550-2172889
E-mail: linyolanda23@gmail.com |
| Proprietary or Trade Name: | GMC Wrist Automatic Blood Pressure
Monitor, Model BQ705 |
| Common/Usual
Name: | Non-Invasive Blood Pressure Monitor |
| Classification Name: | System, Measurement, Blood-Pressure,
Non-Invasive |
| Regulation | CFR 870.1130 |
| Product Code | DXN |
| Device Class: | II |
| Predicate Device: | Omron BP652N (K142917) |
| Date Summary Prepared: | 07/16/2018 |
4
Section 5- 510(k) Summary
No. 686, Su Chu Rd., Chuzhou, Anhui, China
Device Description:
The GMC Wrist Automatic Blood Pressure Monitor, Model BO705, measures systolic and diastolic blood pressure and pulse rate of an adult individual. The method used to define blood pressure is similar to the auscultatory method but uses an electronic pressure sensor rather than a stethoscope and mercury manometer. The device utilizes the oscillometric method whereby the electronic pressure sensor converts variation in cuff pressure to electrical signals. The electrical signals are analyzed and used to define the systolic and diastolic measurements and determine the pulse rate. An irregular heart beat (IHB) indicator is displayed when there is more than a 25% deviation from the averaged pulse rate measurements.
Intended Use:
The GMC Wrist Automatic Blood Pressure Monitor, Model BQ705, is a non-invasive blood pressure measurement device that is used for measuring systolic and diastolic blood pressure and pulse rate using the oscillometric method for adults at home.
The device has an irregular heartbeat (IHB) indicator. The device detects the appearance of an irregular heartbeat during measurement, and displays an IHB symbol on the LCD with the reading once the irregular heartbeat is detected.
Substantially Equivalent (SE) Comparison
The GMC Wrist Automatic Blood Pressure Monitor, Model BO705 is substantially equivalent to the Omron BP652N (K142917).
5
BQ705
GMC0尚钧医疗
Model name | Predicate Device | Subject Device | |
---|---|---|---|
Omron BP652N (K142917) | GMC BQ705 | Comparison | |
510(k) submitter/holder | Omron Healthcare, Inc. | GMC Inc. | |
Intended use and Indications | |||
for Use | The device is a digital monitor intended | The device is a non-invasive blood pressure | Similar |
for use in measuring blood pressure and | measurement device that is used for | No difference of intended | |
pulse rate in adult patient population | measuring systolic and diastolic blood | purpose. The proposed | |
with wrist circumference ranging from | pressure and pulse rate using the | indication for use is the same as | |
5 1/4 inches to 8 1/2 inches (13.5cm to | oscillometric method for adults at home. | the cleared indication for use of | |
21.5 cm). The device detects the | The device has an irregular heartbeat (IHB) | the predicate device. The wrist | |
appearance of irregular heartbeats | indicator. The device detects the appearance | circumference is slightly | |
during measurement and gives a | of an irregular heartbeat during | different between the two | |
warning signal with readings. | measurement, and displays an IHB symbol | devices, but this difference does | |
on the LCD with the reading once the | not affect safety or | ||
irregular heartbeat is detected. | effectiveness. | ||
Use Environment | Home use | Home use | Identical |
Patient Population | Adult | Adult | Identical |
6
GMC0尚钧医疗
GMC Inc. No. 686, Su Chu Rd., Chuzhou, Anhui, China
Specifications / Features
| Model name | Predicate Device
Omron BP652N (K142917) | Subject Device
GMC BQ705 | Comparison |
|-------------------------------|--------------------------------------------|-----------------------------|--------------------------------------------------------------------------------------------|
| Measurement method | Oscillometric methods | Oscillometric methods | Identical |
| Cuff fits wrist circumference | 13.5 to 21.5cm | 13.522cm(5.3"260 mmHg | Similar8.7") | Similar299 mmHg | 30
This minor difference does not impact
safety and effectiveness of the device |
| Measurement range | Pressure 0
This minor difference does not impact
safety and effectiveness of the device |
| | Pulse 40180 beats/min | 40199 beats/min | Similar
This minor difference does not impact
safety and effectiveness of the device |
| Accuracy | Pressure ±3 mmHg or 2 % of reading | ±3 mmHg | Similar
There is no difference in the
accuracy range based on mmHg
scale |
| | Pulse ±5% of reading | ±5% of reading | Identical |
7
BQ705
GMC0尚钧医疗
| Model name | Predicate Device
Omron BP652N (K142917) | Subject Device
GMC BQ705 | Comparison |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| Reusable | Yes | Yes | Identical |
| Display Type | LCD | LCD | Identical |
| Display Parameter | Current cuff pressure, Systolic blood pressure,
Diastolic blood pressure,
Pulse rate, Error message,
measurement results in the memory | Systolic pressure, Diastolic pressure,
Pulse rate, Memory sequence number,
Date/Time, memory zone icons, pulse
mark, weak battery mark, AVG,
Hypertension Classification Indicator,
IHB, Movement indicator | Similar
The different display parameters do
not impact safety and effectiveness of
the device. |
| Inflation | Pump driven | Pump driven | Identical |
| Deflation | Automatic Air Release Valve | Automatic Air Release Valve | Identical |
| Pressure Sensor | Semi-conductor | Semi-conductor | Identical |
| Irregular Heart Beat detection | Yes | Yes | Identical |
| Body movement detection | Yes | Yes | Identical |
| Hypertension indicator | Yes | Yes | Identical |
| Advanced positioning sensor | Yes | No | Different
This feature does not impact safety |
8
Section 5- 510(k) Summary
BQ705
GMC0尚钧医疗
| Model name | Predicate Device
Omron BP652N (K142917) | Subject Device
GMC BQ705 | Comparison |
|-----------------------------------------|-----------------------------------------------|---------------------------------|-------------------------------------------------------------------------------------|
| | | | and effectiveness of the device |
| Average of latest three
measurements | Yes | Yes | Identical |
| Memory | | | Similar |
| | 100 readings | 120 reading(60memories2 zones) | This feature does not impact safety
and effectiveness of the device |
| Multiple Users | No | Yes(2) | Different
This feature does not impact safety
and effectiveness of the device |
| Power Source | AAA Battery2 pcs | AAA Battery*2 pcs | Identical |
| Dimension | 89 mm × 61 mm × 13 mm (L x W x H) | 67×75.7×30mm (L x W x H) | Similar
Size is not a factor in function of the
device |
| Weight | Approx. 80g (2 7/8 oz)
(without batteries) | 81g
(without batteries) | Similar
Weight is not a factor in function of
the device |
9
BQ705
GMC0尚钧医疗
| Model name | | Predicate Device
Omron BP652N (K142917) | Subject Device
GMC BQ705 | Comparison |
|--------------------------|---------------------|--------------------------------------------|-----------------------------|--------------------------------------------------------------------------------------------|
| Operating
Environment | Temperature | 1040℃ (50-104°F) | 1040℃ (50-104°F) | Identical |
| Environment | Humidity | 15%85% RH | 15%60 ℃(14-140°F) | Identical |85% RH | Identical |60℃ (14-140°F) | -10
| | Ambient
pressure | N/A | 700-1060 hPa | Different
This feature does not impact safety
and effectiveness of the device |
| Storage
Environment | Temperature | -20
| Environment | Humidity | 10%~95% RH | 10%~90% RH | Similar
This minor difference does not impact
safety and effectiveness of the device |
| | Ambient
pressure | 700-1060 hpa | 700-1060 hPa | Identical |
10
GMC Inc. No. 686, Su Chu Rd., Chuzhou, Anhui, China
| Model name | Predicate Device
Omron BP652N (K142917) | Subject Device
GMC BQ705 | Comparison |
|----------------------|----------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| Compliance Standards | IEC 60601-1
IEC 60601-1-2
IEC 60601-1-11
AAMI SP10
ANSI/AAMI/ISO81060-2
ISO 10993-1 | ANSI/AAMI ES60601-1
IEC 60601-1-2
IEC 60601-1-11
IEC 80601-2-30
ISO 81060-2
ISO 10993-1 | Similar
BQ705 complied with the latest
Recognized Consensus Standards |
The subject device is viewed as substantially equivalent to the predicate devices Omron BP652N because intended use and use the well known oscillometric method within to determine the systolic and diastolic blood pressure and pulse rate. There are no significant diffect the safety or effectiveness of the intended device as compared to the predicate devices.
11
Image /page/11/Picture/4 description: The image shows the logo for GMC 尚钧医疗. The logo consists of the letters "GMC" in gray, followed by a red circle with a white plus sign inside. To the right of the circle are four Chinese characters in gray. The characters are 尚钧医疗.
No. 686, Su Chu Rd., Chuzhou, Anhui, China
Performance Testing:
Bench testing was conducted to demonstrate that the device meets its requirements and specification.
The following performance tests were completed:
- ANSI AAMI ES60601-1
- IEC 60601-1-2
- · IEC 60601-1-11
- IEC 80601-2-30
The GMC Wrist Automatic Blood Pressure Monitor, Model BQ705 meets all applicable requirements.
Biocompatibility
The materials of construction are widely used in the same or similar applications; therefore, demonstrating the material's use in other medical devices and supporting existing biocompatibility.
Clinical testing
Testing to insure clinical accuracy of the device in accordance with ISO 81060-2 as documented in Clinical Investigation Report. 85 patients (38 males and 47 females) were invited for the study. Standard auscultation method was used as the reference blood pressure monitor measuring in the left arm. Blood pressure measurements were repeated alternatively with the device and auscultation in the same arm according to the sequence in ISO 81060-2.
Conclusion
Based on the evaluations completed, there are no new safety or effectiveness issues introduced with this new medical device. Therefore, the GMC Wrist Automatic Blood Pressure Monitor, Model BQ705, is substantially equivalent to the predicate devices Omron BP652N (K142917)