K Number
K181920

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-12-11

(146 days)

Product Code
Regulation Number
886.5925
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CLARITI 1 DAY (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eyes that may exhibit astigmatism up to 2.00 Diopters that does not interfere with visual acuity.

The CLARITI 1 DAY TORIC (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia) in phakic or aphakic persons with non-diseased eyes that may exhibit astigmatism up to 10.00 Diopters.

The CLARITI 1 DAY MULTIFOCAL (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia and hyperopia) and/or presbyopia in phakic persons with non-diseased eyes that may require a reading addition of +3.00 Diopters or less and may exhibit astigmatism up to 1.50 Diopters or less.

The CLARITI 1 DAY MULTIFOCAL TORIC (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia and hyperopia) and/or presbyopia in phakic persons with non-diseased eyes that may exhibit astigmatism up to 10.00 Diopters and require a reading addition of +3.00 Diopters or less.

The Clarti 1 Day (somofilcon A) Soft (hydrophilic) Contact Lenses are indicated for daily wear single use only. The lenses are to be discarded upon removal; therefore, no cleaning or disinfection is required.

Device Description

The Clariti 1 day (somofilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens with UV Blocker is available as sphere lens, toric lens, multifocal lens, and multifocal toric lens.

In its hydrated state, Clariti 1 day (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV Blocker when placed on the cornea acts as a refracting media to focus light rays on the retina.

Clariti 1 day (somofilcon A) Soft (hydrophilic) Contact Lenses for Daily Wear Single Use are a hydrophilic co-polymer of silicone containing monomers and hydrophilic monomers which is cross-linked with tetraethyleneglycol dimethacrylate and di-functional methacryloxypropyl-terminated poly(dimethylsiloxane).

When hydrated the lens consists of 44.0% somofilcon A and 56.0% water by weight of saline immersed in normal saline. A benzophenone UV absorbing monomer is used in the contact lens to help protect against transmission of harmful UV radiation and Clariti 1 day (somofilcon A) Soft contact lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.

The average transmittance characteristics are less than 5% in the UVB range of 280 to 315nm and less than 50% in the UVA range of 316-380nm.

The lens has a hemispherical flexible shell, which covers the cornea and a portion of the adjacent sclera, with the following dimensions:

  • Chord Diameter: 13.0mm to 15.5mm
  • Centre Thickness: 0.03mm to 0.50mm
  • Base Curve: 7.5mm to 9.30mm
  • Powers: -20.00 DS to +20.00 DS
  • Toric Cylinder options: -0.75, -1.25, -1.75 and -2.25
  • Toric Axis options: 10° to 180° (10° steps)
  • Multifocal Add: Lens "LOW" = "low" for spectacle near ADD lens (Max +2.25 ADD) Lens "HIGH" = "high" for spectacle near ADD lens (+2.50 ADD or greater)
AI/ML Overview

Here's a summary of the acceptance criteria and study information for the Clariti 1 day (somofilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens with UV Blocker, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance & Device Description

The document states that the performance specifications/parameters for both the subject and predicate device are identical. Therefore, the acceptance criteria are implicitly those listed under "Performance Specifications/Parameters Comparison" in the Predicate Device column, and the reported device performance for the subject device is stated as "Same" for each parameter.

Performance Specifications/ParametersPredicate Device (Clariti 1 day, K130331)Subject Device (Clariti 1 day, K181920)
Water Content (%)56Same
Refractive Index1.401Same
Oxygen Permeability (Dk @ 35°C)60 x 10^-11 [(cm²/sec) x (ml O2)/(ml x mm Hg)]Same
Base Curve (mm)8.6Same
Diameter (mm)14.0Same
Light Transmittance (%)>95Same
Modulus (MPa)≥0.3Same
Tensile Strength (MPa)≥0.4Same
Elongation to Break (%)≥100Same
Surface TreatmentNoSame
Center Thickness (mm)Varies with powerSame
Power Range (D)-20.00 to +20.00Same
Additional Device Description Information (not acceptance criteria but important characteristics):
Chord Diameter:13.0mm to 15.5mmSame (implied, as device description is identical)
Centre Thickness:0.03mm to 0.50mmSame (implied)
Base Curve:7.5mm to 9.30mmSame (implied)
Powers:-20.00 DS to +20.00 DSSame (implied)
Toric Cylinder options:-0.75, -1.25, -1.75 and -2.25Same (implied)
Toric Axis options:10° to 180° (10° steps)Same (implied)
Multifocal Add:"LOW" (Max +2.25 ADD), "HIGH" (+2.50 ADD or greater)Same (implied)
%Transmittance @ 590nm:98.13Same (implied)
%Transmittance @ 280-315nm:0.71Same (implied)
%Transmittance @ 316-380nm:20.62Same (implied)
Surface Character:HydrophilicSame (implied)
Specific Gravity:1.17Same (implied)

Key Differences Between Subject and Predicate Device:
There were two technological differences between the subject device (K181920) and the predicate device (K130331) that required performance data:

  1. Addition of 4,4'-Diallyoxyl benzophenone (DAB): The subject device includes both UV416 and DAB as UV blockers, whereas the predicate only had UV416.
  2. Change from borate to phosphate buffered packaging solution: The subject device uses Phosphate Buffered Saline Solution containing 0.020% w/v Poloxamer 407, while the predicate used Borate Buffered Saline Solution containing 0.005% w/v Poloxamer 407.

2. Sample Size Used for the Test Set and Data Provenance

The document states that "Conclusive evidence was provided to demonstrate that the subject device lens material is equivalent to the currently marketed predicate device lens material through the statistical analysis of the physiochemical properties, especially water content, oxygen permeability, modulus, and toughness of the lens."

For the performance testing (bench testing), the document mentions that "Each test was conducted according to the ANSI, ISO, and/or ASTM standard indicated." However, specific sample sizes for the test set are not provided in the given text.

The data provenance is not explicitly stated as country of origin, but the testing was conducted in accordance with international standards (ISO, ANSI, ASTM) and FDA guidance documents (e.g., Use of International Standard ISO 10993-1, Premarket Notification [510(k)] Guidance Document for Daily Wear Contact Lenses). This implies that standard laboratory testing was performed, likely at the manufacturer's or a contract research organization's facilities. The studies appear to be prospective in nature, as they were conducted to support the substantial equivalence claim for the subject device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable to this submission. The "ground truth" in this context refers to the defined physical, chemical, and optical properties of the contact lenses, as measured by standardized bench tests, and biocompatibility evaluations. These measurements do not rely on expert interpretation in the way that image-based diagnostic devices might.

4. Adjudication Method for the Test Set

This is not applicable. The performance testing involves direct physical, chemical, and optical measurements against established standards, not subjective assessments requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

An MRMC comparative effectiveness study was not done. This type of study is relevant for AI-powered diagnostic devices where human readers interpret medical images. This submission is for a medical device (contact lens) where performance is assessed through bench and biocompatibility testing, not human interpretation of data. The document explicitly states: "clinical performance data was not required to be submitted in this 510(k)."

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

This is not applicable. This is not an algorithm or AI-powered device. The device's performance is assessed through its physical characteristics and biological safety.

7. The Type of Ground Truth Used

The "ground truth" for the performance criteria was established through:

  • Bench Testing: Measurements of physical, chemical, and optical properties (e.g., water content, oxygen permeability, refractive index, mechanical properties, light transmittance, specific gravity, contact angle, total extractables, non-polymeric residuals) using established ANSI, ISO, and ASTM standards.
  • Biocompatibility Testing: Evaluation according to ISO 10993 series for cytotoxicity, irritation, systemic toxicity, and sensitization.
  • Comparison to Predicate: The key ground truth was the established performance of the legally marketed predicate device (Clariti 1 day, K130331), demonstrating that the subject device's performance was "Same" or equivalent across all measured parameters despite minor material and packaging solution changes.

8. The Sample Size for the Training Set

This is not applicable. This submission is for a physical medical device (contact lens), not an AI algorithm. Therefore, there is no "training set" in the context of machine learning.

9. How the Ground Truth for the Training Set was Established

This is not applicable for the reasons stated above.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 11, 2018

Marie Dutton Senior Regulatory Affairs Specialist 5870 Stoneridge Drive Suite 1 Pleasanton, CA 94588

Re: K181920

Trade/Device Name: Clariti 1 day (somofilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens with UV Blocker Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: November 6, 2018 Received: November 7, 2018

Dear Marie Dutton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device. please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely yours.

J. Angelo Green

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K181920

Device Name

Clariti 1 day (somofilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens with UV Blocker

Indications for Use (Describe)

The CLARITI 1 DAY (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eyes that may exhibit astigmatism up to 2.00 Diopters that does not interfere with visual acuity.

The CLARITI 1 DAY TORIC (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia) in phakic or aphakic persons with non-diseased eyes that may exhibit astigmatism up to 10.00 Diopters.

The CLARITI 1 DAY MULTIFOCAL (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia and hyperopia) and/or presbyopia in phakic persons with non-diseased eyes that may require a reading addition of +3.00 Diopters or less and may exhibit astigmatism up to 1.50 Diopters or less.

The CLARITI 1 DAY MULTIFOCAL TORIC (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia and hyperopia) and/or presbyopia in phakic persons with non-diseased eyes that may exhibit astigmatism up to 10.00 Diopters and require a reading addition of +3.00 Diopters or less.

The Clarti 1 Day (somofilcon A) Soft (hydrophilic) Contact Lenses are indicated for daily wear single use only. The lenses are to be discarded upon removal; therefore, no cleaning or disinfection is required.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

I. SUBMITTER:

CooperVision, Inc. 6150 Stoneridge Mall Road, Suite 370 Pleasanton, CA 94588

Contact Person:

Marie Dutton Senior Regulatory Affairs Specialist CooperVision, Inc. 5870 Stoneridge Drive, Suite 1 Pleasanton, CA 94588 Phone: (925) 251-6645 (925) 251-6643 Fax: E-mail: MDutton@coopervision.com

Date Prepared:

November 6, 2018

II. DEVICE:

Name of Device:Clariti 1 day (somofilcon A) Soft (Hydrophilic)Daily Disposable Contact Lens with UV Blocker
Common Name:Soft (hydrophilic) Contact Lens
Classification Name:Lens, Contact, (Disposable) [21 CFR 886.5925 (b) (1)]
Regulatory Class:II
Product Code:MVN, LPL
Classification Panel:Ophthalmic

III. PREDICATE DEVICE:

Clariti 1 day (somofilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens with UV Blocker, K130331

{4}------------------------------------------------

IV. DEVICE DESCRIPTION:

The device description for the subject device and predicate Clariti 1 day (somofilcon A) device is identical. The Clariti 1 day (somofilcon A) Soft (Hydrophilic) Daily Disposable Contact Lens with UV Blocker is available as sphere lens, toric lens, multifocal lens, and multifocal toric lens.

In its hydrated state, Clariti 1 day (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV Blocker when placed on the cornea acts as a refracting media to focus light rays on the retina.

Clariti 1 day (somofilcon A) Soft (hydrophilic) Contact Lenses for Daily Wear Single Use are a hydrophilic co-polymer of silicone containing monomers and hydrophilic monomers which is cross-linked with tetraethyleneglycol dimethacrylate and di-functional methacryloxypropyl-terminated poly(dimethylsiloxane).

When hydrated the lens consists of 44.0% somofilcon A and 56.0% water by weight of saline immersed in normal saline. A benzophenone UV absorbing monomer is used in the contact lens to help protect against transmission of harmful UV radiation and Clariti 1 day (somofilcon A) Soft contact lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.

The average transmittance characteristics are less than 5% in the UVB range of 280 to 315nm and less than 50% in the UVA range of 316-380nm.

The lens has a hemispherical flexible shell, which covers the cornea and a portion of the adjacent sclera, with the following dimensions:

  • Chord Diameter: 13.0mm to 15.5mm ●
  • Centre Thickness:
  • Base Curve:
  • Powers: ●
  • Toric Cylinder options:
  • Toric Axis options: ●
  • 0.03mm to 0.50mm 7.5mm to 9.30mm -20.00 DS to +20.00 DS -0.75, -1.25, -1.75 and -2.25
  • 10° to 180° (10° steps)
  • Multifocal Add: ●

Lens "LOW" = "low" for spectacle near ADD lens (Max +2.25 ADD) Lens "HIGH" = "high" for spectacle near ADD lens (+2.50 ADD or greater)

{5}------------------------------------------------

The physical/optical properties of the lenses are:

• Refractive Index:1.4003
• %Transmittance @ 590nm:98.13
• %Transmittance @ 280-315nm:0.71
• %Transmittance @ 316-380nm:20.62
• Surface Character:Hydrophilic
• Water Content:56%
• Oxygen Permeability (DK):60 x 10-11 (cm²/sec)(ml O2/ml x mmHg) at 35°C (FattMethod for determination of oxygenpermeability)
• Specific Gravity:1.17

V. INDICATIONS FOR USE:

The indications for use statement for the subject device and predicate Clariti 1 day (somofilcon A) device is identical.

Lens DesignIndication
SphereThe CLARITI 1 DAY (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes that may exhibit astigmatism up to 2.00 Diopters that does not interfere with visual acuity.
ToricThe CLARITI 1 DAY TORIC (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes that may exhibit astigmatism up to 10.00 Diopters.
MultifocalThe CLARITI 1 DAY MULTIFOCAL (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia and hyperopia) and/or presbyopia in phakic or aphakic persons with non-diseased eyes that may require a reading addition of +3.00 Diopters or less and may exhibit astigmatism up to 1.50 Diopters or less.
Multifocal ToricThe CLARITI 1 DAY MULTIFOCAL TORIC (somofilcon A) Soft (hydrophilic) Daily Disposable Contact Lens with UV blocker is indicated for daily wear single use only for the optical correction of refractive ametropia (myopia and hyperopia) and/or presbyopia in phakic or aphakic persons with non-diseased eyes that may exhibit astigmatism up to 10.00 Diopters and require a reading addition of +3.00 Diopters or less.

{6}------------------------------------------------

The Clariti 1 day (somofilcon A) Soft (hydrophilic) Contact Lenses are indicated for daily wear single use only. The lenses are to be discarded upon removal; therefore, no cleaning or disinfection is required.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE:

The technological principle for both the subject and predicate device somofilcon A soft (hydrophilic) contact lenses with UV blocker is identical. The subject and predicate device are based on the same technological elements. The two technological differences that exist between the subject and predicate device consist of the addition of 4,4'-Diallyoxyl benzophenone (DAB) and the change from borate to phosphate buffered packaging solution. A comparison of the technological characteristics of the subject device with the predicate device is provided in the table below.

For the technological difference of the addition of DAB, performance data was provided to demonstrate that the subject device lens material is equivalent to the currently marketed predicate device lens material. Additionally, there is no change to the USAN name of the device because additives such as UV absorbers are excluded for the purpose of nomenclature.

For the technological difference of the change from borate to phosphate buffered packaging solution, performance data was provided to ensure compatibility of the contact lens with the lens storage solution that comes into direct contact with the contact lens and to ensure that the changes in the packaging material has not compromised the sterility or stability of the contact lens for the labeled expiration date. Additionally, the requirements of the procedure for implementing changes in packaging materials, specifically the types of buffers utilized in lens packaging solutions or addition of buffers to saline solutions (e.g., borate to phosphate buffer), from the Premarket Notification [510(k)] Guidance Document for Daily Wear Contact Lenses, May 12, 1994 amended June 28, 1994 were met.

{7}------------------------------------------------

Technology/Material Comparison
Predicate DeviceSubject Device
Product NameClariti 1 day (somofilconA) Soft (Hydrophilic)Daily Disposable ContactLens with UV BlockerSame
Material USAN Namesomofilcon ASame
510(k) NumberK130331K181920
FDA Category (Group)Materials having a Dkgreater than 40 Dk units(using mmHg) and havinga Dk greater than thatexpected based on thematerials' water contentalone (Group V)Same
Manufacturing MethodCast MoldingSame
Wearing andReplacement ScheduleDaily Wear Single UseSame
SterilizationMoist HeatSame
Packaging MaterialsInjection moldedpolypropylene blisterscovered by aluminum foillaminate and blister stripsare packed into printedcartonsSame
Packaging SolutionBorate Buffered SalineSolution containing0.005% w/v Poloxamer407Phosphate Buffered SalineSolution containing0.020% w/v Poloxamer407
Blue Visibility TintNoSame
TintNoneSame
UV BlockerUV416UV416 and DAB

{8}------------------------------------------------

VII. PERFORMANCE DATA:

The performance specifications/parameters for both the subject and predicate device Clariti 1 day (somofilcon A) contact lenses with UV blocker are identical and provided in the comparison table below.

Performance Specifications/Parameters Comparison
Predicate DeviceSubject Device
Clariti 1 dayClariti 1 day
(somofilcon A)(somofilcon A)
K130331K181920
Water Content (%)56Same
Refractive Index1.401Same
Oxygen Permeability(Dk @ 35°C)60 x 10-11 [(cm²/sec) x (ml O2)/(ml x mm Hg)]Same
Base Curve (mm)8.6Same
Diameter (mm)14.0Same
Light Transmittance (%)>95Same
Modulus (MPa)≥0.3Same
Tensile Strength (MPa)≥0.4Same
Elongation to Break (%)≥100Same
Surface TreatmentNoSame
Center Thickness (mm)Varies with powerSame
Power Range (D)-20.00 to + 20.00Same

The following performance data were provided in support of the substantial equivalence determination. All tests were conducted in accordance with the GLP regulation (21 CFR Part 58) or according to valid scientific protocols.

Biocompatibility testing:

The biocompatibility evaluation for the subject Clariti 1 day (somofilcon A) device was conducted in accordance with the Guidance for Industry and Food and Drug Administration Staff entitled Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process," June 16, 2016 and the Premarket Notification [510(k)] Guidance Document for Daily Wear Contact Lenses, May 12, 1994 amended June 28, 1994. The battery of testing included the following tests:

  • Cytotoxicity per ISO 10993-5:2009 ●
  • Irritation per ISO 10993-10:2010 and ISO 9394:2012 ●
  • Systemic Toxicity per ISO 10993-11:2006 and ISO 10993-1:2009 .
  • Sensitization per ISO 10993-10:2010 ●

{9}------------------------------------------------

Performance testing - bench:

The performance testing for the subject Clariti 1 day (somofilcon A) device was conducted in accordance with the Premarket Notification [510(k)] Guidance Document for Daily Wear Contact Lenses, May 12, 1994 amended June 28, 1994. The battery of testing included the following. Each test was conducted according to the ANSI, ISO, and/or ASTM standard indicated:

  • Light transmittance per ANSI Z80.20-2010 and ISO 18369-3:2006 ●
  • Oxygen permeability per ISO 18369-4:2006 ●
  • Water content per ISO 18369-4:2006 .
  • . Refractive index per ISO 16369-4:2006
  • Mechanical properties per ANSI Z80.20-2010 and ASTM D1708-13 ●
  • Lens specification per ANSI Z80.20-2010, ISO 18369-2:2012, and ISO ● 18369-3:2006
  • Specific gravity per ISO 1183-1:2012
  • Contact angle per ANSI Z80.20-2010 ●
  • Total extractables per ISO 18369-4:2006 ●
  • Non-polymeric residuals in lens and packaging solution, solvents chosen per ISO 18369-4:2006

Clinical testing:

Data provided in this 510(k) is sufficient to adequately characterize the modified subject Clariti 1 day (somofilcon A) device in terms of its physical/chemical/optical, and toxicological performance characteristics when compared to the predicate. The results are equivalent determining that additional clinical performance data was not required to complete the substantial equivalent determination. Additionally, the technical characteristics and manufacturing and sterilization processes of the subject lens are equivalent to somofilcon A contact lens currently marketed by CooperVision; therefore, it was confirmed no clinical data is required.

CONCLUSIONS: VIII.

Conclusive evidence was provided to demonstrate that the subject device lens material is equivalent to the currently marketed predicate device lens material through the statistical analysis of the physiochemical properties, especially water content, oxygen permeability, modulus, and toughness of the lens. Therefore, there is no change to the USAN name of the device. Based on this performance testing and the fact that the subject device has the same manufacturing process as the marketed predicate device lens, clinical performance data was not required to be submitted in this 510(k). There were no differences in physiochemical properties, especially water content, oxygen permeability, modulus, and toughness of the lens. The performance testing demonstrates that the subject Clariti 1 day (somofilcon A) device performs comparably to the predicate Clariti 1 day (somofilcon A) device that is currently marketed for the same intended use.

{10}------------------------------------------------

The modifications made to the subject Clariti 1 day (somofilcon A) device did not have a significant impact on the performance characteristics of the 510(k) cleared predicate device. There is sufficient data provided in this 510(k) to adequately characterize the modified lens in terms of its physical/chemical/optical, and toxicological performance characteristics when compared to the previously cleared lens. Results of the pre-clinical testing (physical/chemical/optical and toxicological data) are sufficient to support the claim and clinical performance data are not needed to assess effects of the new characteristics.

N/A