K Number
K181887
Date Cleared
2019-01-22

(193 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

AMD Anti-Fog Solution is a sterile and single-use laparoscopic accessory device for use prior to and during endoscopic and laparoscopic procedures. The product is intended to prevent condensation of the distal lenses of the endoscopic/ laparoscopic instruments during the surgery. The product is intended for prescription use only.

Device Description

The AMD Anti-Fog Solution is a single-use, sterile, and biocompatible laparoscopic accessory device. The product is packaged by the Tyvek" Pouch which includes one clear plastic screw top dispenser bottle with the volume of 6 mL and one polyurethane (PU) foam pad with x-ray detectable radiopaque ribbon. The solution is a clear/colorless, odorless, and aqueous solution which comprises of surfactant (≤2%), isopropyl alcohol (≤2%), and water (≥96%). The polyurethane (PU) foam pad is an adhesivebacked sponge with radiopaque ribbon that can be used to introduce the solution into the surface of the lenses.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance:

The document primarily focuses on demonstrating substantial equivalence to a predicate device, rather than defining explicit performance acceptance criteria with numerical targets for the AMD Anti-Fog Solution itself (beyond biocompatibility, sterility, and shelf life). However, the "Performance" row in Table 5.1 and the "Product Performance Test" section provide the core performance claim.

Acceptance Criteria (Implied)Reported Device Performance
Biocompatibility (no adverse biological reactions)All biocompatibility tests (Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, Material-Mediated Pyrogenicity) passed requirements.
Sterility (Sterility Assurance Level < 10⁻⁶)Routine sterilization parameter at ≥25 kGy achieved SAL < 10⁻⁶.
Shelf-life (maintenance of sterility and function for 5 years)All aging tests (Package Integrity, Seal Peel Strength, Product Sterility, Fog Resistance) passed for a proposed 5-year shelf life.
Performance (Preventing condensation of distal lenses)"AMD Anti-Fog Solution is as effective as the predicate device."

2. Sample Size for the Test Set and Data Provenance:

  • Biocompatibility Testing: The specific number of samples (e.g., number of animals for in vivo tests, number of cell cultures for in vitro tests) is not explicitly stated in the provided document for each test. The tests were performed according to ISO standards, which would typically specify sample sizes.
  • Sterilization Validation: The VDmax 25 method was used, implying a standard number of samples derived from product bioburden testing to establish the sterilization dose. The document states bioburden was < 1000 CFU/sample, but the number of samples tested for bioburden is not specified.
  • Product Aging Validation (Shelf Life Study): The number of samples for each test (Package Integrity, Seal Peel Strength, Product Sterility, Product Stability/Fog Resistance) is not explicitly stated. These tests were performed for a 5-year shelf life validation.
  • Product Performance Test (Fog Resistance Test): The sample size for this test is not explicitly stated.
  • Data Provenance: The studies were conducted by the manufacturer, Advanced Medical Design Co., Ltd., based in Taiwan. The location where the tests were performed (e.g., specific labs) is not detailed. The data is retrospective in the sense that it was completed prior to the 510(k) submission.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

  • This information is not applicable in the context of this submission. The device is a physical product (anti-fog solution), and its performance (biocompatibility, sterility, fog resistance, shelf life) is evaluated through objective laboratory tests against established standards, not through human interpretation or expert-driven ground truth. Therefore, no "experts" were required to establish ground truth in the sense of clinical assessment or image interpretation.

4. Adjudication Method for the Test Set:

  • Not applicable. As the performance is based on objective laboratory tests against specified standards, there is no need for an adjudication method as would be used for subjective assessments or disputes among human readers/interpreters.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done:

  • No. This type of study is typically used for diagnostic or screening devices where human interpretation plays a role. The AMD Anti-Fog Solution is a laparoscopic accessory intended to prevent fogging, not a diagnostic tool requiring human reader interpretation. Therefore, an MRMC study and analysis of AI vs. human reader improvement is not relevant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) study was done:

  • Not applicable. The device is a physical solution, not an algorithm or AI system.

7. The Type of Ground Truth Used:

  • Objective Laboratory Measurements and Standard Criteria:
    • Biocompatibility: Ground truth was established by recognized international standards (ISO 10993 series) which define acceptable biological responses.
    • Sterility: Ground truth was established by microbiological testing methods (ISO 11137-2, ISO 11737-1, ISO 11737-2) to achieve a specified sterility assurance level (SAL).
    • Shelf Life (Packaging Integrity, Seal Strength): Ground truth was established by recognized ASTM and ISO standards for packaging materials.
    • Shelf Life (Product Stability/Fog Resistance): Ground truth was established by an "internal testing protocol." The specifics of this protocol are not detailed, but it would involve a measurable criterion for fog resistance.
    • Overall Performance (Fog Resistance): The claim is based on equivalence to the predicate device, implying that its fog resistance is at least as good as the predicate's, which itself would have been demonstrated through specific test methods.

8. The Sample Size for the Training Set:

  • Not applicable. This device is not an AI/ML algorithm or software that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

  • Not applicable. See point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 22, 2019

Advanced Medical Design Co., Ltd Silvia Yeh Regulatory Affairs 4-5F, No 29, Wuquan 5th Rd., Wugu Dist. New Taipei City, 24888 Taiwan

Re: K181887 Trade/Device Name: AMD Anti-Fog Solution Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: OCT Dated: July 10, 2018 Received: December 12, 2018

Dear Silvia Yeh:

We reviewed your Section 510(k) premarket notification of intent to market the device referenced above and determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmb/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark R. Kreitz -S

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K181887

Device Name AMD Anti-Fog Solution

Indications for Use (Describe)

AMD Anti-Fog Solution is a sterile and single-use laparoscopic accessory device for use prior to and during endoscopic and laparoscopic procedures. The product is intended to prevent condensation of the endoscopic/ laparoscopic instruments during the surgery. The product is intended for prescription use only.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a logo for AMD Global Solutions for Medical Industry. The logo features the letters "AMD" in large, bold, blue font, with a globe graphic incorporated into the "A". Below the letters is the text "GLOBAL SOLUTIONS FOR MEDICAL INDUSTRY" in a smaller font. The logo is clean and professional, suggesting a company that is reliable and innovative.

I. SUBMITTER

510(k) Summary

Company Name:Advanced Medical Design Co., Ltd.
Mailing Address:4F~5F, No29, Wuquan 5th Rd., Wugu Dist., New Taipei city, 248 TaiwanR.O.C.
Phone:+886-2-2290-2627
Fax:+886-2-2298-1789
Contact Person:Silvia YehRegulatory AffairPhone: +886-2-2290-2627 Ext. 511E-mail: silvia.yeh@amdintl.com.tw
Date Prepared:July 10, 2018

II. DEVICE INFORMATION

Name of Device:AMD Anti-Fog Solution
Common or Usual Name:Endoscope Anti-fogging Device
Classification Name:Endoscope and accessories (21 CFR 876.1500)
Regulatory Class:II
Product Code:OCT

III. PREDICATE DEVICE

Trade/Device NameClear-It Anti-Fog Solution
Common or Usual NameEndoscope Anti-fogging Device
510(k) NumberK022826
Product CodeOCT
SubmitterPreservation Solution, Inc.

IV. DEVICE DESCRIPTION

The AMD Anti-Fog Solution is a single-use, sterile, and biocompatible laparoscopic accessory device. The product is packaged by the Tyvek" Pouch which includes one clear plastic screw top dispenser bottle with the volume of 6 mL and one polyurethane (PU) foam pad with x-ray detectable radiopaque

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image is a logo for AMD Global Solutions for Medical Industry. The logo is blue and white. The logo has a globe on the left side of the word AMD. The text "GLOBAL SOLUTIONS FOR MEDICAL INDUSTRY" is below the logo.

ribbon. The solution is a clear/colorless, odorless, and aqueous solution which comprises of surfactant (≤2%), isopropyl alcohol (≤2%), and water (≥96%). The polyurethane (PU) foam pad is an adhesivebacked sponge with radiopaque ribbon that can be used to introduce the solution into the surface of the lenses.

The device is intended to be used prior and during minimally invasive surgical operation, to prevent condensation of the distal lenses of the endoscopic instruments during the clinical operation. AMD Anti-Fog Solution is comprised of a surfactant that provides the same physical characteristics as the predicate device. The mechanism of anti-fogging function of AMD Anti-Fog Solution is to reduce the surface tension of the water, resulting in a non-scattering film of water instead of single droplets, on the lenses once the product is smeared on the lenses. Therefore, the non-scattering film of water was formed on the lenses instead of the single water droplets. This phenomenon is called "wetting".

V. INDICATIONS FOR USE

AMD Anti-Fog Solution is a sterile and single-use laparoscopic accessory device for use prior to and during endoscopic and laparoscopic procedures. The product is intended to prevent condensation of the distal lenses of the endoscopic/laparoscopic instruments during the surgery. The product is intended for prescription use only.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE DEVICE

AMD Anti-Fog Solution is designed to prevent condensation of the distal lenses, that is the same as the predicate device (K number: K022826). A comparison of the device features, technological characteristics, intended use, and other information demonstrates that AMD Anti-Fog Solution is substantially equivalent to the predicate device. The substantially equivalent comparison table was summarized in Table 5.1.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for AMD Global Solutions for the Medical Industry. The logo is in blue and features a globe on the left side of the letters AMD. The text "GLOBAL SOLUTIONS FOR MEDICAL INDUSTRY" is written in a smaller font below the letters AMD.

Device NameProposed DevicePredicate DeviceComment
Indication for UseAMD Anti-Fog Solution is a sterile and single-use laparoscopic accessory device for use prior to and during endoscopic and laparoscopic procedures. The product is intended to prevent condensation of the distal lenses of the endoscopic/laparoscopic instruments during the surgery. The product is intended for prescription use only.Clear-It Anti-Fog Solution (K022826)Clear-It Anti-Fog Solution is indicated for use in the sterile surgical arena to eliminate condensation from endoscopic lenses, goggles and other devices that are likely to fog.Same
Reusable / DisposableDisposableDisposableSame
Target Patient PopulationThe patient who treated by minimum invasive laparoscopic surgery.The patient who treated the surgery in the sterile surgical arena.Similar. The patient population of the proposed device is for minimum invasive laparoscopic surgery, the predicate device included.
Target User PopulationPrescription use onlyPrescription use onlySame
Where UsedHospital O.R. roomHospital O.R. roomSame
ContraindicationsThere are no known contraindications to the patient.N/ASame

Page 3 of 7

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for AMD Global Solutions for the Medical Industry. The logo is in blue and features a globe on the left side of the letters AMD. The text "GLOBAL SOLUTIONS FOR MEDICAL INDUSTRY" is located below the letters AMD.

Table 5.1. The Substantially Equivalent Comparison Table
DeviceNameProposed DevicePredicate Device(K022826)Comment
Elements- Surfactant (≤2%)- Isopropyl alcohol (≤2%)- Water (≥96%)b. Sponge- Polyurethane (PU) foam- Water- Isopropyl alcohol- Sodium Alcohol EtherSulfate- AmmoniumDodecylbenzeneb. Sponge- Polyurethane (PU) foambiological evaluationresults of theproposed devicemeet therequirements of theapplicablestandards.
Method ofIntroductionWiping distal end of the lensby sponge with solution.Wiping distal end of the lensby sponge with solution.Same
CompatibilityWith OtherDevicesLaparoscope/Endoscope:5 mm, 10 mm and all othersizesLaparoscope/Endoscope:5 mm, 10 mm and all othersizesSame
PerformancePreventing condensation ofthe distal lenses of theendoscopic/laparoscopicinstruments during theclinical operation.Preventing condensation ofthe distal lenses of theendoscopic/laparoscopicinstruments during theclinical operation.Same
SafetyThe followingbiocompatibility tests wereconducted.- Cytotoxicity Test- Sensitization Test- Intracutaneous ReactivityTest- Acute Systemic Toxicity Test- Material-MediatedPyrogenicity TestAll of the tests passed therequirements as indicated inthe applicable standardsThe cytotoxicity test wasconducted.Same. Weperformed all of thebiocompatibilitytests that areindicated in the FDArecognizedstandards.

Table 5,1, The Substantially Equivalent Comparison Table

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for AMD Global Solutions for the Medical Industry. The logo is in blue and features a globe on the left side. The letters AMD are in large, bold font and are the main focus of the logo. Below the letters, the words "GLOBAL SOLUTIONS FOR MEDICAL INDUSTRY" are written in a smaller font.

Proposed DevicePredicate DeviceComment
DeviceNameAMD Anti-Fog SolutionClear-It Anti-Fog Solution(K022826)
SterileBarrierSingle Layer Of Tyvek® PouchSingle Layer Of Tyvek® PouchSame
SterilizationMethodGamma-radiation Sterilization(10-6 SAL)Gamma-radiation Sterilization(10-6 SAL)Same
EnergySourceNo energy sourceNo energy sourceSame

Table 5.1. The Substantially Equivalent Comparison Table

Through the substantially equivalent comparison table, the differences do not raise any different issues on the safety or effectiveness of the product.

VII. PERFORMANCE DATA

A series of the studies were performed to evaluate the safety and effectiveness of AMD Anti-Fog Solution. The following test results were provided to confirm the product is safe and effective as indicated.

A. Biocompatibility Testing

The biocompatibility test was evaluated per the FDA recognized consensus standard named "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" issued on June 16, 2016, the biocompatibility tests include the following items since the product is classified as the classification.

Nature of Body ContactContact Duration
CategoryContact
External communicating deviceTissue/bone/dentinLimited (≤24 h)
No.Test NameApplicable StandardsComment
1Cytotoxicity TestISO 10993-5:2009 Biological Evaluation Of Medical Devices- Part 5: Tests For In Vitro CytotoxicityPass
2Sensitization TestISO 10993-10:2010 Biological Evaluation Of MedicalDevices - Part 10: Tests For Irritation And Skin SensitizationPass
3IntracutaneousISO 10993-10:2010 Biological Evaluation Of MedicalPass

Page 5 of 7

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for AMD Global Solutions for Medical Industry. The logo is in blue and features a globe on the left side. The text "AMD" is in large, bold letters, and the text "GLOBAL SOLUTIONS FOR MEDICAL INDUSTRY" is in smaller letters below the logo.

No.Test NameApplicable StandardsComment
Reactivity TestDevices - Part 10: Tests For Irritation And Skin Sensitization
4Acute SystemicToxicity TestISO 10993-11:2017 Biological Evaluation Of MedicalDevices - Part 11: Tests For Systemic ToxicityPass
5Material-MediatedPyrogenicityUSP <151> Pyrogen TestPass

All of the test studies listed above showed that AMD Anti-Fog Solution did not raise any safety issues and is biocompatible.

B. Sterilization Validation and Shelf Life Study

The product is designed to perform gamma-radiation sterilization prior to place into the market, therefore the following studies should be evaluated by the applicable standards/guidance.

a. Gamma-radiation Sterilization Validation Study

The gamma-radiation sterilization study was performed per the requirements of the FDA recognized consensus standards listed below.

  • ISO 11137-2:2013 Sterilization of Health Care Products Radiation Part 2: Establishing the Sterilization Dose
  • ISO 11737-1:2006 Sterilization of Medical Devices Microbiological Methods Part 1: Determination of A Population of Microorganisms on Products
  • ISO 11737-2:2009 Sterilization of Medical Devices Microbiological Methods Part 2: Tests of Sterility Performed in The Definition, Validation And Maintenance Of A Sterilization Process

The method used for gamma-radiation sterilization study was VDmax 25 method, since the product bioburden was less than 1000 CFU/sample. The test reports showed that the product can become sterile when the routine sterilization parameter was controlled at the dose of no less than 25 kGy, which meets the regulatory requirement of sterile condition (SAL < 10t).

b. Product Aging Validation Study (Shelf Life Study)

The product aging validation study was performed for 5 years per the FDA recognized consensus standards "ISO 11607-1:2006 Packaging for Terminally Sterilized Medical Devices - Part 1: Requirements for Materials, Sterile Barrier Systems And Packaging Systems" to determine the shelf life of the product, since the product is supplied in the sterile status.

The Aging Validation Study included the following test studies. Since the shelf life of the product is proposed to be stored for 5 years, the aging validation study is

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image is a logo for AMD Global Solutions for the Medical Industry. The logo is blue and white. The logo has a globe on the left side and the letters AMD on the right side. Below the logo is the text "GLOBAL SOLUTIONS FOR MEDICAL INDUSTRY".

performed which included the following test items.

No.Test ItemApplicable Standards/GuidanceComment
1Package Integrity Test(Dye penetration Test)ASTM F1929-15 Standard Test Method For DetectingSeal Leaks In Porous Medical Packaging By DyePenetrationPass
2Seal Peel Strength TestASTM F88/F88M-15 Standard Test Method For SealStrength Of Flexible Barrier MaterialsPass
3Product Sterility TestISO 11737-2:2009 Sterilization Of Medical Devices -Microbiological Methods - Part 2: Tests Of SterilityPerformed In The Definition, Validation AndMaintenance Of A Sterilization ProcessPass
4Product Stability Studies(Fog Resistance Test)N/A (followed by the internal testing protocol)Pass

All of the aging test data showed that the product can be safe and effective during its predetermined shelf life. Hence, the sterile assurance level and the functional specification of the product meet the firm's definition and regulatory requirement, the shelf life of the product is 5 years.

C. Product Performance Test

The test result of the fog resistance test shows that AMD Anti-Fog Solution is as effective as the predicate device.

VIII. CONCLUSIONS

Based on the previous data and comparison to the predicate device, AMD Anti-Fog Solution is as safe and effective as the predicate device and do not raise any new issues of safety and effectiveness.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.