(121 days)
The Reprocessed Torqr Catheter is intended for use in diagnostic electrophysiologic procedures. The catheter is designed for recording intracardiac electrograms and temporary pacing associated with electrophysiology studies.
The Reprocessed Torqr Diagnostic EP Catheter is a flexible, radiopaque catheter constructed of extruded polyurethane over stainless steel braid and platinum electrodes. The catheter is of high torque construction to aid in precise catheter placement.
Here's an analysis of the provided text regarding the acceptance criteria and study for the Reprocessed Torqr Diagnostic EP Catheter, formatted to address your specific questions.
1. Table of Acceptance Criteria and Reported Device Performance
The provided FDA 510(k) summary does not explicitly list acceptance criteria in a quantitative, pass/fail manner with specific numerical targets. Instead, it details the types of functional and safety testing performed to demonstrate that the reprocessed catheter is "as safe and effective as the predicate devices." The "reported device performance" is described qualitatively as having successfully undergone these tests.
| Acceptance Criteria Category (Implied) | Reported Device Performance |
|---|---|
| Biocompatibility | Tested and validated (successful) |
| Cleaning Validation | Tested and validated (successful) |
| Sterilization Validation | Tested and validated (successful) |
| Functional Testing: | |
| * Visual Inspection | Tested and validated (successful) |
| * Dimensional Verification | Tested and validated (successful) |
| * Electrical Continuity and Resistance | Tested and validated (successful) |
| * Simulated Use | Tested and validated (successful) |
| * Mechanical Characteristics | Tested and validated (successful) |
| Electrical Safety Testing: | |
| * Dielectric and Current Leakage | Tested and validated (successful) |
| Packaging Validation | Tested and validated (successful) |
| Reprocessing Cycle Limit | Reprocessed no more than one (1) time; devices rejected after maximum cycles. |
| Exclusion of previously reprocessed devices | Reprocessing restricted to exclude devices previously reprocessed by other reprocessors. |
Study Proving Device Meets Acceptance Criteria:
The document states that "Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of Reprocessed Catheter." This "Functional and Safety Testing" section (on page 6) lists the tests performed, which collectively constitute the study. The conclusion is that these tests showed the reprocessed catheter to be "as safe and effective as the predicate devices."
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the exact sample sizes used for each of the functional and safety tests. It also does not mention the country of origin of the data or whether the data was retrospective or prospective. Given the nature of the device (reprocessed catheter) and the tests listed (bench and laboratory), the data would inherently be prospective in the sense that the tests were conducted specifically for this submission.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This information is not provided in the document. The "ground truth" for this type of medical device (a reprocessed catheter) would primarily be established through objective engineering and scientific measurements and analyses against established performance standards and predicate device specifications rather than expert consensus on diagnostic interpretations.
4. Adjudication Method for the Test Set
This information is not applicable and therefore not provided for this type of device submission. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies involving interpretation (e.g., imaging) where multiple human readers' opinions need to be reconciled to form a ground truth. For bench and laboratory testing of a physical medical device, the criteria are usually objective (e.g., does it meet a specific resistance value, does it withstand a certain pull force).
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
An MRMC comparative effectiveness study was not performed. This type of study is relevant for AI-powered diagnostic tools or systems where human performance is enhanced by AI assistance. The Reprocessed Torqr Diagnostic EP Catheter is a physical medical device, not an AI diagnostic tool.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
A standalone performance study of an algorithm was not done. Again, this device is a physical medical device, not an algorithm or AI system.
7. The Type of Ground Truth Used
The ground truth for the performance of the Reprocessed Torqr Diagnostic EP Catheter is based on a combination of:
- Engineering Specifications: Adherence to design requirements, such as dimensional accuracy, electrical properties (continuity, resistance, dielectric, leakage), and mechanical characteristics.
- Predicate Device Equivalence: The reprocessed device is deemed safe and effective if its performance during these tests is comparable to the original, legally marketed predicate device (Torqr Series Diagnostic EP Catheters, K923915).
- Validation Standards: Compliance with established validation protocols for biocompatibility, cleaning, sterilization, and packaging.
8. The Sample Size for the Training Set
The concept of a "training set" is not applicable to this device. Training sets are used for machine learning algorithms. The studies performed were bench and laboratory tests on physical units of the reprocessed catheter.
9. How the Ground Truth for the Training Set Was Established
Since there was no training set for a machine learning algorithm, the question of how its ground truth was established is not applicable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 18, 2018
Innovative Health, LLC. Amanda Babcock Principal Regulatory Affairs Specialist 1435 North Hayden Road. Suite 100 Scottsdale, Arizona 85257
Re: K181618
Trade/Device Name: Reprocessed Torqr Diagnostic EP Catheter Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter Or Electrode Recording Probe Regulatory Class: Class II Product Code: NLH Dated: September 7, 2018 Received: September 10, 2018
Dear Amanda Babcock:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products, and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely
Mada Jellman
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| Description | ItemNumber | Length(cm) | Curve | FrenchSize | No. ofElectrodes | Spacing(mm) |
|---|---|---|---|---|---|---|
| TorqrDiagnostic EPCatheter | 04122JM | 110 | Josephson | 5F | 4 | 2-5-2 |
| 04125JM | 110 | Josephson | 5F | 4 | 5-5-5 | |
| 04122UM | 110 | Cournand | 5F | 4 | 2-5-2 | |
| 04125UM | 110 | Cournand | 5F | 4 | 5-5-5 | |
| 04130DS | 110 | Damato | 5F | 4 | 2-2-2 | |
| 041002JM | 110 | Josephson | 6F | 4 | 2-5-2 | |
| 041005JM | 110 | Josephson | 6F | 4 | 5-5-5 | |
| 041002UM | 110 | Cournand | 6F | 4 | 2-5-2 | |
| 041005UM | 110 | Cournand | 6F | 4 | 5-5-5 | |
| 041005DM | 110 | Damato | 6F | 4 | 5-5-5 | |
| 041565CS | 65 | CS | 5F | 10 | 2-5-2 | |
| 041865CS | 65 | CS | 5F | 10 | 2-8-2 | |
| 041590CS | 90 | CS | 5F | 10 | 2-5-2 |
The item numbers included in the scope of this submission K181618 are as follows:
{3}------------------------------------------------
Indications for Use
510(k) Number (if known) K181618
Device Name
Reprocessed Torqr Diagnostic Electrophysiology (EP) Catheter
Indications for Use (Describe)
The Reprocessed Torqr Catheter is intended for use in diagnostic electrophysiologic procedures. The catheter is designed for recording intracardiac electrograms and temporary pacing associated with electrophysiology studies.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
SECTION 5: 510(k) SUMMARY
As required by 21 CFR 807.92(c)
Submitter's Name and Address:
Innovative Health, LLC. 1435 N. Hayden Road, Suite 100 Scottsdale, AZ 85257
Contact Name and Information:
Amanda Babcock Principal Regulatory Affairs Specialist Innovative Health, LLC. (480) 525-5911 (office) (888) 965-7705 (fax) ababcock@innovative-health.com
Date prepared:
June 18, 2018
Device Information:
| Trade/Proprietary Name: | Reprocessed Torqr Diagnostic EP Catheter |
|---|---|
| Common or Usual Name: | Diagnostic Electrophysiology Catheter |
| Classification Name: | Electrode Recording Catheter or Electrode Recording Probe |
| Classification: | Class II, 21 CFR 870.1220 |
| Product Code: | NLH |
Predicate Device:
| 510(k) Number | 510(k) Title | Manufacturer |
|---|---|---|
| K923915 | Torqr Series Diagnostic EP Catheters | Medtronic (CardioRhythm) |
Device Description:
The Reprocessed Torqr Diagnostic EP Catheter is a flexible, radiopaque catheter constructed of extruded polyurethane over stainless steel braid and platinum electrodes. The catheter is of high torque construction to aid in precise catheter placement.
The item numbers in scope of this submission are as follows:
| Description | ItemNumber | Length(cm) | Curve | FrenchSize | No. ofElectrodes | Spacing(mm) |
|---|---|---|---|---|---|---|
| TorqrDiagnostic EPCatheter | 04122JM | 110 | Josephson | 5F | 4 | 2-5-2 |
| 04125JM | 110 | Josephson | 5F | 4 | 5-5-5 | |
| 04122UM | 110 | Cournand | 5F | 4 | 2-5-2 | |
| 04125UM | 110 | Cournand | 5F | 4 | 5-5-5 | |
| 04130DS | 110 | Damato | 5F | 4 | 2-2-2 |
{5}------------------------------------------------
| 041002JM | 110 | Josephson | 6F | 4 | 2-5-2 |
|---|---|---|---|---|---|
| 041005JM | 110 | Josephson | 6F | 4 | 5-5-5 |
| 041002UM | 110 | Cournand | 6F | 4 | 2-5-2 |
| 041005UM | 110 | Cournand | 6F | 4 | 5-5-5 |
| 041005DM | 110 | Damato | 6F | 4 | 5-5-5 |
| 041565CS | 65 | CS | 5F | 10 | 2-5-2 |
| 041865CS | 65 | CS | 5F | 10 | 2-8-2 |
| 041590CS | 90 | CS | 5F | 10 | 2-5-2 |
Table 5.1: Device Scope
Indications for Use:
The Reprocessed Torgr Catheter is intended for use in diagnostic electrophysiologic procedures. The catheter is designed for recording intracardiac electrograms and temporary pacing associated with electrophysiology studies.
Technological Characteristics:
The purpose, design, materials, function, and intended use of the Reprocessed Torqr Catheter are identical to the predicate devices. There are no changes to the claims, clinical applications, patient populations, performance specifications, or method of operation. In addition, Innovative Health's reprocessing of the Catheter includes removal of visible soil and decontamination. Each device is inspected and function tested prior to packaging and labeling.
Functional and Safety Testing:
Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of Reprocessed Catheter. This included the following:
- Biocompatibilitv .
- Cleaning Validation .
- Sterilization Validation 0
- Functional testing 0
- . Visual Inspection
- = Dimensional Verification
- . Electrical Continuity and Resistance
- Simulated Use
- . Mechanical Characteristics
- Electrical Safety Testing ●
- . Dielectric and Current Leakage
- o Packaging Validation
The Reprocessed Torqr Catheter is reprocessed no more than one (1) time. Each device is marked and tracked. After the device has reached the maximum number of reprocessing cycles, the device is rejected from further reprocessing. Reprocessing is performed only by Innovative Health. Innovative Health restricts its reprocessing to exclude devices previously reprocessed by other reprocessors.
{6}------------------------------------------------
Conclusion:
Innovative Health concludes that the Reprocessed Torqr Catheter is as safe and effective as the predicate devices described herein.
§ 870.1220 Electrode recording catheter or electrode recording probe.
(a)
Identification. An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.(b)
Classification. Class II (performance standards).