K Number
K181503
Date Cleared
2018-10-11

(126 days)

Product Code
Regulation Number
870.2700
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Fingertip Pulse Oximeter MD300CI218 is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent, child and infant patients in hospital-type facilities and homecare environment.

Device Description

The proposed device Fingertip Pulse Oximeter MD300CI218 is a battery powered device. It can detect and display the measured %SpO2 and pulse rate value and will automatically power off when there is no signal for longer than 8 seconds. The proposed device is adopted colorful color OLED screen to display SpO2, PR and waveform which can be displayed in 2 directions. And it is designed with the battery indicator function to warn the user that the battery power may be low. The proposed device is normally applied to adult, adolescent, child and infant patients in hospitals, hospital facilities and homecare environment. And it can transmit the measurements to Smart Device installed the mobile APP via Bluetooth 4.0 to help the users to organize and track their health information. The proposed device consists of power supply module, detector and emitter LED, signal collection and processor module, display module, Bluetooth module, user interface and button control. The fingertip pulse oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660nm, which is red light; the other is 905nm, which is infrared-red light. Skin, bone, tissue and venous vessels normally absorb a constant amount of light over time. The photo detector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2. The proposed device is not for life-supporting or life-sustaining, not for implant. The device is not sterile and the transducers are reusable and do not need sterilization and re-sterilization. The device is for prescription. The device does not contain drug or biological products. The device is software-driven and the software validation is provided in software.

AI/ML Overview

The provided text describes the 510(k) premarket notification for the Fingertip Pulse Oximeter MD300CI218. It includes details about the device, its intended use, comparison to a predicate device, and various tests conducted to demonstrate substantial equivalence.

Here's an analysis of the acceptance criteria and study data based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the SpO2 (oxygen saturation) and PR (Pulse Rate) accuracy are implicitly stated as the device's accuracy specifications, which are identical to the predicate device. The performance is reported in the clinical test results.

MetricAcceptance Criteria (from Predicate Device)Reported Device Performance (MD300CI218)
SpO2 Accuracy70%100%, ±2%; 069% no definitionArms of 1.66 (during steady state 70-100%)
PR Accuracy30bpm99bpm, ±2bpm; 100bpm235bpm, ±2%Not explicitly stated in the results summary, but stated as compliant with claimed accuracy.
SpO2 Display Range0~100%0~100%
SpO2 Measurement Range70%~100%70%~100%
PR Display Range0bpm~235bpm30bpm~250bpm
PR Measurement Range30bpm~235bpm30bpm~250bpm

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: 12 healthy adult subjects.
  • Data Provenance: The clinical study was conducted from March 9-11, 2018, in the Yue Bei People's Hospital, cardiovascular medicine, in China. It was a prospective study where subjects were intentionally involved for the purpose of testing the device.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

The document does not specify the number or qualifications of experts used to establish ground truth for the clinical test (e.g., co-oximeter readings for SpO2 accuracy). It only mentions that the study followed ISO 80601-2-61:2011, which outlines specific methodologies for pulse oximeter testing, implying that the ground truth would have been established using reference methods as per the standard.

4. Adjudication Method for the Test Set

The document does not describe any specific adjudication method for the test set. For pulse oximetry accuracy testing as per ISO 80601-2-61, ground truth is typically established by arterial blood gas analysis using a co-oximeter, where direct measurements serve as the reference, rather than expert adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. This study is a technical performance test of a medical device (pulse oximeter) against a reference standard, not an AI-assisted diagnostic study involving human readers.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

Yes, this was a standalone performance study. The MD300CI218 Finger Pulse Oximeter is a device that directly measures and displays SpO2 and pulse rate. The clinical test evaluated the accuracy of the device's measurements compared to a reference standard (likely arterial blood sampling and co-oximetry, as per the ISO standard). There is no "human-in-the-loop" interaction for interpreting the device's output, beyond a clinician reading the displayed values.

7. The Type of Ground Truth Used

The type of ground truth used for the clinical study was based on the methodology described in ISO 80601-2-61:2011. This standard typically mandates the use of reference co-oximeter measurements from arterial blood samples (induced hypoxia studies) as the ground truth for SpO2 accuracy. The document states, "The SpO2 accuracy performance results showed the MD300CI218 Fingertip Pulse Oximeter to have an Arms of 1.66 during steady state conditions over the range of 70-100%," which is a standard metric derived from comparing device readings to co-oximeter readings.

8. The Sample Size for the Training Set

This document describes a clinical validation study for a medical device that does not appear to involve a machine learning or AI algorithm that requires a "training set." The device is based on established optical principles (Lambert-Beer Law) for pulse oximetry. Therefore, the concept of a "training set" as it applies to AI models is not relevant here.

9. How the Ground Truth for the Training Set Was Established

As noted in point 8, there is no mention or indication of a "training set" for an AI model. The device operates on optical principles rather than an AI/ML algorithm that is 'trained' on data.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.

Beijing Choice Electronic Technology Co., Ltd. Haiying Zhao Quality Director No.9 Shuangyuan Road, Badachu Hi-tech Zone, Shijingshan District Beijing, 100041 China

Re: K181503

Trade/Device Name: Fingertip Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: September 4, 2018 Received: September 4, 2018

Dear Haiying Zhao:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice

(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Todd D. Courtney -S

Tina Kiang, Ph.D. for Acting Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Section II Indications for Use Statement

Indications for Use

510(k) Number (if known): K181503

Device Name: _Fingertip Pulse Oximeter MD300CI218

Indications for Use:

The Fingertip Pulse Oximeter MD300CI218 is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent, child and infant patients in hospital-type facilities and homecare environment.

Prescription Use √AND/OROver-The-Counter Use
(Part 21 CFR 801 Subpart D)(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE

(1) OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{3}------------------------------------------------

Section III 510(k) Summary

This summary of 510(k) is being submitted in accordance with the requirements of 21 CFR 807.92.

There is no prior submission for the device.

3.1 Submitter Information

● Manufacturer Name:

Establishment Registration Number: 3005569927 Beijing Choice Electronic Technology Co., Ltd. Room 4104, No. A12 Yuquan Road Haidian District 100143 Beijing, P.R.China

. Contact Person:

Lei Chen Beijing Choice Electronic Technology Co., Ltd. North Building 3F, No. 9 Shuangyuan Road, Badachu Hi-tech Zone, Shijingshan District Beijing China 100041 Phone: +86-10-88798300 Ext 6020 Fax:215-4052545 Email: cc@choicemmed.com

  • . Date prepared: June 04, 2018

3.2 Proposed Device Information

Device Common Name: Pulse Oximeter Device Trade/Proprietary Name: Fingertip Pulse Oximeter Model: MD300CI218 Classification Name: Oximeter Regulation Number: 870.2700 Product Code: DQA

Class: II

Panel: Anesthesiology

{4}------------------------------------------------

3.3 Predicate Device

510(k) Number: K142888 Common Name: Pulse Oximeter Device Trade/Proprietary Name: Fingertip Pulse Oximeter Model: MD300C318T2 Classification Name: Oximeter Product Code: DQA Regulation Number: 870.2700

Device Class: II

Panel: Anesthesiology

Manufacturer: Beijing Choice Electronic Technology Co., Ltd.

Intended Use: The MD300C318T2 is intended for spot checking in measuring and displaying functional arterial hemoglobin (SpO2) and pulse rate. It is intended for adult, adolescent, child and infant users in hospitals, hospital facilities and home healthcare environments.

3.4 Device Description

The proposed device Fingertip Pulse Oximeter MD300CI218 is a battery powered device. It can detect and display the measured %SpO2 and pulse rate value and will automatically power off when there is no signal for longer than 8 seconds. The proposed device is adopted colorful color OLED screen to display SpO2, PR and waveform which can be displayed in 2 directions. And it is designed with the battery indicator function to warn the user that the battery power may be low.

The proposed device is normally applied to adult, adolescent, child and infant patients in hospitals, hospital facilities and homecare environment. And it can transmit the measurements to Smart Device installed the mobile APP via Bluetooth 4.0 to help the users to organize and track their health information.

The proposed device consists of power supply module, detector and emitter LED, signal collection and processor module, display module, Bluetooth module, user interface and button control.

The fingertip pulse oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660nm, which is red light; the other is 905nm, which is infrared-red light. Skin, bone, tissue and venous vessels normally absorb a constant amount of light over time. The photo detector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an

{5}------------------------------------------------

oxygen saturation measurement. This measurement is referred to as SpO2.

The proposed device is not for life-supporting or life-sustaining, not for implant.

The device is not sterile and the transducers are reusable and do not need sterilization and re-sterilization.

The device is for prescription.

The device does not contain drug or biological products.

The device is software-driven and the software validation is provided in software.

{6}------------------------------------------------

3.5 Comparison list of the technological characteristics

Comparison ElementsProposed DevicePredicate Device
Product NameFingertip Pulse OximeterFingertip Pulse Oximeter
ModelMD300CI218MD300C318T2
Regulation No.21 CFR 870.270021 CFR 870.2700
ClassificationIIII
Classification NameOximeterOximeter
Product CodeDQADQA
Indications for UseThe MD300CI218 is a handheld non-invasivedevice intended for spot-checking of oxygensaturation of arterial hemoglobin (SpO2) andPulse Rate of adult, adolescent, child andinfantpatients in hospitals, hospital-typefacilities and homecare environment.The MD300C318T2 is intended for spot-checking in measuring and displayingfunctional arterial hemoglobin (SpO2) andpulse rate. It is intended for adult, adolescent,child and infant users in hospitals, hospitalfacilities and home healthcare environments.
Comparison StatementThe proposed device and the predicated device have the same intended use andclassification
ComponentsPower supply module, detector and emitterLED, signal collection and processor module,display module,Bluetoothmodule,userinterface and button control.Power supply module, detector and emitterLED, signal collection and process module,display module, Bluetooth module, indicatormodule, user interface.
Design PrincipleThe fingertip pulse oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660nm, which is red light; the other is 905nm, which is infrared-red light. Skin, bone, tissue and venous vessels normally absorb a constant amount of light over time. The photo detector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2.Principle of the oximeter is as follows: A mathematical formula is established making use of Lambert Beer Law according to Spectrum Absorption Characteristics of Reductive hemoglobin (RHb) and Oxyhemoglobin (HbO2) in red and near-infrared zones. Operation principle of the instrument: Photoelectric Oxyhemoglobin Inspection Technology is adopted in accordance with Capacity Pulse Scanning and Recording Technology, so that two beams of different wavelength of lights (660nm glow and 940nm near infrared light) can be focused onto a human nail tip through a clamping finger-type sensor. A measured signal obtained by a photosensitive element, will be shown on the oximeter's display through process in electronic circuits and microprocessor.
MeasurementRed660±3nm660nm
WavelengthInfrared905±10nm940nm
Comparison StatementThe proposed device and the predicate device have the same design principle and similar measurement wavelength. The only difference is the wavelength of the infrared LED emitter, and we can verify that which will not effect the essential performance of the proposed device.
Display TypeOLEDOLED
User Interface2 display directions2 display directions
Power supply2*AAA alkaline batteries2*AAA alkaline batteries
Display DataSpO2, PRSpO2, PR
SpO2 Display Range0~100%0%~100%
SpO2 Measurement Range70%~100%70%~100%
SpO2 Accuracy70%100%, ±2%;069% no definition70%100%, ±2%;069% no definition
SpO2 Resolution1%1%
PR Display Range30bpm~250bpm0bpm~235bpm
PR Measurement Range30bpm~250bpm30bpm~235bpm
PR Accuracy30bpm99bpm, ±2bpm;100bpm250bpm, ±2%30bpm99bpm, ±2bpm;100bpm235bpm, ±2%
PR Resolution1%1%
Wireless Transmission Range0~10m0~10m
Antenna TypeInternalInternal
TransmitterBluetooth Compliance: Version 4.0Bluetooth Compliance: Version 2.1

Table 3-1 Performance Specification Comparison Table between the Proposed Device and Predicate Device

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

Premarket Notification 510(k) Submission-Section III 510(k) Summary
---------------------------------------------------------------------
OperatingTemperature5°C~40°C5°C~40°C
Relative Humidity15% ~93%, no condensation in operation;≤93% no condensation in storage15% ~93%, no condensation in operation;≤93% no condensation in storage
AtmospherePressure70kPa~106kpa86kPa~106kpa
Comparison StatementThe proposed device has similar product specification as predicate device.
ContactingMaterialBattery CoverEnclosurePower ButtonABSABS
Fingertip CushionMedical Silicone GelMedical Silicone Gel
Comparison StatementPerformance TestingThe contacting materials of the proposed device are same to those of the predicate device.
Laboratory TestingThe laboratory tests include SpO2 and PRaccuracy Test, Weak Perfusion Test, High andLow Temperature and Humidity Test,Performance Test After Cleaning andISO80601-2-61The laboratory tests include SpO2 and PRaccuracy Test, Weak Perfusion Test, High andLow Temperature and Humidity Test,Performance Test After Cleaning andISO80601-2-61
Electrical SafetyConformed to IEC60601-1, IEC 60601-1-11Conformed to IEC60601-1, IEC 60601-1-11
EMC and Electrical SafetyElectromagnetic CompatibilityConformed to IEC60601-1-2Conformed to FCC certificationConformed to IEC60601-1-2Conformed to FCC certification
Software and CybersecurityModerate level of concernModerate level of concern
Compliance with:FDA Guidance for the content of PremarketSubmissions for Software Contained inMedical Devices and,FDA Guidance for Content of PremarketSubmission for Management of Cybersecurityin Medical DeviceRisk Management in Compliance withISO14971:2007Compliance with FDA Guidance for thecontent of Premarket Submissions forSoftware Contained in Medical DevicesRisk Management in Compliance withISO14971:2007
Label and LabelingCompliance with the Guidance of pulseoximeter-premarket notification issued onMarch 4.2013Compliance with the Guidance of pulseoximeter-premarket notification issued onMarch 4.2013

{10}------------------------------------------------

{11}------------------------------------------------

3.6 Intended use

The Fingertip Pulse Oximeter MD300Cl218 is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent, child and infant patients in hospital-type facilities and homecare environment.

3.7 Testing

The Fingertip Pulse Oximeter MD300CI218 was supported by both laboratory and clinical accuracy testing in order to ensure that they were appropriate performance and functional features to fully comply with recognized standards and is substantially equivalent to the predicate device.

Non-clinical Test

The Fingertip Pulse Oximeter MD300CI218 is designed and tested and will be manufactured in accordance with both mandatory and voluntary standards, including:

IEC60601-1:2005+A1:2012 Medical Electrical Equipment-Part 1: General Requirements for Safety.

IEC 60601-1-11:2010 Medical Electrical Equipment Part 1-11: General requirements for basic safety and essential performance-Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.

IEC60601-1-2:2014 Medical Electrical Equipment Part 1-2: General Requirements for Basic Safety and Essential Performance Collateral Standard: Electromagnetic disturbances - Requirements and tests.

ISO 80601-2-61:2011 Medical Electrical Equipment- Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment.

We have also conducted other performance test including SpO2 and PR Accuracy Test, Device Output Time and Finger Out Time Test. Device Response Time Test. Weak Perfusion Test, High and Low Temperature & Humidity Test Per Guidance for Industry and FDA Staff: Pulse Oximeter-Premarket Notification submission [510(k)s].

The Software Validation is in compliance with FDA Guidance to Compliance with FDA

Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.

The list of non-clinical test performed on the proposed devices.

No. Test Name

{12}------------------------------------------------

1System Performance Test
2Performance Test according to ISO 80601-2-61
3Electromagnetic Compatibility Test According to IEC 60601-1-2
4Electrical Safety Test According to IEC 60601-1
5Used in the home healthcare environment test according to IEC 60601-1-11
6Bluetooth Wireless Test according to FCC Part 15B and Part 15C
7wireless coexistence testing

The test results indicate that the effectiveness of the proposed device is identical to that of the predicate device.

Clinical Test

The clinical study of MD300C1218 was conducted following the testing described in clause 201.12.1 of ISO 80601-2-61:2011 Medical electrical equipment- Part 2-61 Particular requirements for basic safety and essential performance of pulse oximeter equipment.

The Clinical Study of MD300CI218 was conducted Mar.9-11,2018 in the Yue Bei people's Hospital, cardiovascular medicine in accordance with the study procedure. There were no deviations from the study procedure. Institutional Review Board(IRB) approval was obtained for testing through the Yue Bei People's Hostiptal-IRB. 12 healthy adult subjects, gave their informed consent and were studied. The clinical study report was presented in Performance Testing-Clinical Study Report.

Subjects information:

There were no adverse events during the study. The subject demographics included a total of 12 healthy adult volunteer subjects, 7 females and 5 males. The ages ranged from 18 to 45years. The subject weights ranged from 42 to 83kg. The subject Height ranged from 153 to 179cm. The Skin tones included in the study were as follows: 4 subjects with very dark pigmentation, 1 subject with very light pigmentation. The remaining subjects with light skin tones of China origins.

Results:

The SpO2 accuracy performance results showed the MD300CI218 Fingertip Pulse Oximeter to have an Arms of 1.66 during steady state conditions over the range of 70-100% .

Conclusion:

The results of the study provide supporting evidence that the pulse oximeter MD300CI218 Fingertip Pulse Oximeter was compliance to the accuracy specification claimed by the manufacturer.

The clinical statement was presented in Performance Testing-Clinical Statement.

{13}------------------------------------------------

3.8 Determination of substantial equivalence

The proposed device of Fingertip Pulse Oximeter MD300CI218 has the same classification information, same intended use, same design principle, similar product design and specifications, same performance effectiveness as the predicated device. The main difference is the wavelength of the infrared LED emitter, and we can verify that which will not affect the essential performance of the proposed device. So the proposed devices are Substantially Equivalent (SE) to the predicate device which is US legally market device.

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).