(83 days)
Tranquil-CTM Interbody System
The Tranquil-C™ Interbody System is intended for use for anterior cervical interbody fusion in skeletally mature patients with cervical disc degeneration and/or cervical instability, as confirmed by imaging studies (radiographs, CT, MRI), that results in radiculopathy, and/or pain at multiple contiguous levels from C2 – T1. The Tranquil-C™ Interbody System is also to be used with supplemental fixation systems that have been cleared for use in the cervical spine. The Tranquil-C™ Interbody System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion.
Tranquil-LTM Interbody System
The Tranquil-L™ Interbody System is intended for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2 – S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radios. Patients should have received 6 months of non-operative treatment prior to treatment with the devices. The device must be used with supplemental fixation. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). It is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone.
The Tranquil™ Interbody System is made of Ti-6Al-4V ELI. The implant is offered in various angles, widths, heights, and lengths to meet patient anatomy for both cervical and lumbar spine. The devices and instruments are provided clean and non-sterile for steam sterilization at the user's facility.
The purpose of this traditional 510(k) is to expand the indications for use to include use with either autograft and/or allograft bone for the Tranquil™ Interbody System, as well as update the fusion level for the Tranquil-C™ Interbody System to include use at multiple levels (e.g., up to 4 levels) from C2 – T1, and add additional footprints.
The provided document is a 510(k) clearance letter from the FDA for the Tranquil™ Interbody System. It details the device's intended use and compares it to predicate devices to establish substantial equivalence. However, this document does not contain information about acceptance criteria, a specific study proving the device meets those criteria, or any of the detailed aspects of a clinical study or performance evaluation that would typically involve human readers, ground truth, or sample sizes for testing and training sets.
The only "performance data" mentioned is:
- "Mechanical performance testing data was provided as part of the previous submission to establish substantial equivalence for its use."
- "As was demonstrated in this submission through engineering rationale, the additional footprints do not present a new worst case for the testing that was previously performed. Therefore, no new mechanical testing was performed for this 510(k) submission."
This indicates that the submission relies on bench testing and engineering rationale to demonstrate that the expanded indications and additional footprints do not alter the substantial equivalence established by previous mechanical testing. There is no mention of a study involving human subjects or AI performance.
Therefore, I cannot fulfill the request for information regarding acceptance criteria and a study that proves the device meets them because that information is not present in the provided text. The document focuses on regulatory clearance based on substantial equivalence to predicate devices, primarily through engineering and material comparisons, not clinical performance studies.
{0}------------------------------------------------
August 27, 2018
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Nexus Spine, LLC Jared Crocker Director of Quality and Regulatory Affairs 2825 East Cottonwood Parkway, Suite 330 Salt Lake City, Utah 84121
Re: K181483
Trade/Device Name: Tranquil™ Interbody System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX, ODP Dated: August 6, 2018 Received: August 8, 2018
Dear Mr. Crocker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brent Showalter -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
Device Name Tranquil™ Interbody System
Indications for Use (Describe)
Tranquil-CTM Interbody System
The Tranquil-C™ Interbody System is intended for use for anterior cervical interbody fusion in skeletally mature patients with cervical disc degeneration and/or cervical instability, as confirmed by imaging studies (radiographs, CT, MRI), that results in radiculopathy, and/or pain at multiple contiguous levels from C2 – T1. The Tranquil-C™ Interbody System is also to be used with supplemental fixation systems that have been cleared for use in the cervical spine. The Tranquil-C™ Interbody System is designed for use with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft to facilitate fusion.
Tranquil-LTM Interbody System
The Tranquil-L™ Interbody System is intended for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2 – S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radios. Patients should have received 6 months of non-operative treatment prior to treatment with the devices. The device must be used with supplemental fixation. These DDD patients may also have up to Grade I spondylolisthesis at the involved level(s). It is indicated to be used with autograft bone and/or allograft bone comprised of cancellous and/or corticocancellous bone.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary
| Submitter: | Nexus Spine, LLC |
|---|---|
| Contact Person: | Mr. Jared Crocker, Director of Quality and Regulatory Affairs2825 East Cottonwood Parkway Suite 330Salt Lake City, UT 84121Telephone: (801) 702-8592Fax: (801) 702-8585 |
| Date Prepared: | August 6, 2018 |
| Trade Name: | Tranquil™ Interbody System |
| Classification, Nameand Number: | Class IIIntervertebral body fusion device21 CFR 888.3080 |
| Product Code: | MAX. ODP |
Predicate Device(s):
| Manufacturer | Device | 510(k) Number |
|---|---|---|
| Primary Predicate | ||
| Nexus Spine LLC | Tranquil™ Interbody System | K170297 |
| Additional Predicates | ||
| HD LifeSciences LLC | HD Lumbar Interbody System | K170676 |
| EIT Emerging ImplantTechnologies GmbH | EIT Cellular Titanium Cervical Cages,EIT Cellular Titanium PLIF Cages,EIT Cellular Titanium TLIF Cages,EIT Cellular Titanium ALIF Cages | K172888 |
Reference Device(s):
The following devices have been 510(k) cleared with these same or similar technological elements: ODP product code, indications for use including autograft and/or allograft, and anatomical region (multiple contiguous levels from C2 – T1):
- NuVasive: Modulus-C Interbody System (K172676) ●
- Astura Medical: ALTA Anterior Cervical Interbody Spacer (K173324) ●
- NuVasive: Cohere Cervical Interbody Fusion Device (K173030)
- NuVasive: CoRoent Small Interbody System (K163491)
{4}------------------------------------------------
| Device Description: | The Tranquil™ Interbody System is made of Ti-6Al-4V ELI. The implant isoffered in various angles, widths, heights, and lengths to meet patientanatomy for both cervical and lumbar spine. The devices and instrumentsare provided clean and non-sterile for steam sterilization at the user'sfacility.The purpose of this traditional 510(k) is to expand the indications for use toinclude use with either autograft and/or allograft bone for the Tranquil™Interbody System, as well as update the fusion level for the Tranquil-C™Interbody System to include use at multiple levels (e.g., up to 4 levels) fromC2 – T1, and add additional footprints. |
|---|---|
| Intended Use: | Tranquil-C™ Interbody SystemThe Tranquil-C™ Interbody System is intended for use for anterior cervicalinterbody fusion in skeletally mature patients with cervical discdegeneration and/or cervical spinal instability, as confirmed by imagingstudies (radiographs, CT, MRI), that results in radiculopathy, myelopathy,and/or pain at multiple contiguous levels from C2 – T1. The Tranquil-C™Interbody System is also to be used with supplemental fixation systems thathave been cleared for use in the cervical spine. The Tranquil-C™ InterbodySystem is designed for use with autogenous and/or allogeneic bone graftcomprised of cancellous and/or corticocancellous bone graft to facilitatefusion.Tranquil-L™ Interbody SystemThe Tranquil-L™ Interbody System is intended for use in skeletally maturepatients with Degenerative Disc Disease (DDD) at one or two contiguouslevels from L2 – S1. DDD is defined as discogenic back pain withdegeneration of the disc confirmed by patient history and radiographicstudies. Patients should have received 6 months of non-operative treatmentprior to treatment with the devices. The device must be used withsupplemental fixation. These DDD patients may also have up to Grade Ispondylolisthesis or retrolisthesis at the involved level(s). It is indicated tobe used with autograft bone and/or allograft bone comprised of cancellousand/or corticocancellous bone. |
| Statement ofTechnologicalComparison: | The technological principle for both the subject and the primary predicatedevice is intervertebral body fusion of the spine in skeletally mature patientswith degenerative disc disease.The subject device, Tranquil™ Interbody System, and the primary predicatedevice, Tranquil™ Interbody System (K170297), are based on thefollowing technological elements: |
{5}------------------------------------------------
- . Same FDA Product Codes: MAX, ODP
- Same implant material: Ti-6Al-4V ●
- Same manufacturing process: Additive ●
- Similar indications for use ●
- . Equivalent multiple footprints, geometries, and heights to account for variations in patient anatomy
- Equivalent anatomical region ●
- . Same surgical approach
The difference between the subject device and the primary predicate device is the addition of new footprints, as well as the expanded indications as described in the following paragraphs.
Tranquil-LTM Interbody System
The primary predicate Tranquil-L™ Interbody System is already cleared to be used with autogenous bone graft (autograft). The proposed indication for the subject Tranquil-LTM Interbody System is to include the use of allograft.
The additional predicate device, HD Lumbar Interbody System (K170676) is equivalent to the subject device in the following technological elements:
- . Same FDA product code: MAX
- Equivalent implant material: Ti-6Al-4V ●
- Equivalent manufacturing process: Additive ●
- Same Indications for Use: including autograft and/or allograft
- . Same anatomical region: up to 2 levels from L2 - S1
- . Equivalent multiple footprints, geometries, and heights to account for variations in patient anatomy
- Same surgical approach .
As these technological characteristics of the subject Tranquil-LTM Interbody System are equivalent to those of predicate K170676, the use of cancellous and/or corticocancellous allograft in the Tranquil-L™ Interbody System may also be deemed equivalent.
Tranquil-CTM Interbody System
The primary predicate Tranquil-C™ Interbody System is already cleared to be used with autogenous bone graft (autograft) and is intended to be used for procedures at one level from C3 - C7. The proposed indication for the subject Tranquil-CTM Interbody System is to include the use of allograft for procedures at multiple levels (e.g., up to 4 levels) from C2 – T1.
{6}------------------------------------------------
The additional predicate device, EIT Cellular Titanium Cervical Cages (K172888), is equivalent to the subject device in the following technological elements:
- . Same FDA product code: ODP
- Equivalent implant material: Ti-6Al-4V
- . Equivalent manufacturing process: Additive
- . Same Indications for Use: including autograft and/or allograft
- . Same anatomical region: multiple levels from C2 - T1
- Equivalent multiple footprints, geometries, and heights to account for variations in patient anatomy
- Same surgical approach .
Reference Devices:
In addition to the additional predicate device, EIT Cellular Titanium Cervical Cages, many other 510(k) cleared devices have the same or similar technological elements:
- . Same FDA product code: ODP
- Same Indications for Use: including autograft and/or allograft
- Same anatomical region: multiple levels from C2 T1 ●
- Equivalent multiple footprints, geometries, and heights to account for . variations in patient anatomy
- . Same surgical approach
In the years 2017 and 2018, there were at least four devices with the technological elements stated above which received 510(k) clearance:
- NuVasive: Modulus-C Interbody System (K172676) ●
- Astura Medical: ALTA Anterior Cervical Interbody Spacer (K173324) ●
- NuVasive: Cohere Cervical Interbody Fusion Device (K173030) ●
- . NuVasive: CoRoent Small Interbody System (K163491)
Therefore, the expansion of the Tranquil-CTM Interbody System indications for use to include allograft at multiple levels from C2 – T1 is substantially equivalent to the predicate devices.
Performance Data: Mechanical performance testing data was provided as part of the previous submission to establish substantial equivalence for its use. As was demonstrated in this submission through engineering rationale, the additional footprints do not present a new worst case for the testing that was previously performed. Therefore, no new mechanical testing was performed for this 510(k) submission.
{7}------------------------------------------------
Documentation provided demonstrates the Tranquil™ Interbody System is Conclusion: substantially equivalent to predicate devices.
§ 888.3080 Intervertebral body fusion device.
(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.