K Number
K181343
Date Cleared
2018-06-12

(22 days)

Product Code
Regulation Number
892.1200
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

HeartSee Software for cardiac position tomography (PET) is indicated for determining regional and global absolute rest and stress myocardial perfusion in co'min/g, Coronary Flow Reserve and their combination into the Coronary Flow Capacity (CFC) Map in patients with suspected or known coronary artery disease (CAD) in order to assist clinical interpretation of PET perfusion images by quantification of their severity.

HeartSee is intended for use by trained professionals, such as nuclear technicians, nuclear cardiology physicians, or cardiologists with appropriate training and certification. The climician remains ultimately responsible for the final assessment and diagnosis based on standard practices, clinical judgment and interpretation of PET images or quantitative data.

Device Description

HeartSee is a software tool for cardiac positron emission tomography (PET) for determining regional and global absolute rest and stress myocardial perfusion in cc/min/q. Coronary Flow Reserve and their combination into the Coronary Flow Capacity (CFC) Map for facilitating the interpretation of PET perfusion images in patients with suspected of known coronary artery disease. HeartSee is intended for use by trained professionals, such as nuclear technicians, nuclear medicine or nuclear cardiology physicians, or cardiologists with appropriate training and certification.

HeartSee contains two fundamental components. First, the software imports cardiac PET images in DICOM format from PET scanners with DICOM output. These images are reoriented to cardiac axes to produce standard tomographic and topographic displays of relative uptake. Second, the software quantifies absolute rest and stress myocardial perfusion per unit tissue (cc/min/gm), Coronary Flow Reserve (CFR) as the stress/rest perfusion ratio and the Coronary Flow Capacity combining CFR and stress perfusion, all on a pixel basis for regional and global values. Archiving output data is supported for clinical diagnostics, quality control and research.

AI/ML Overview

The provided text does not contain detailed acceptance criteria and the results of a study proving the device meets these criteria. The document is a 510(k) summary for the HeartSee Cardiac P.E.T. Processing Software (K181343).

Here's what can be extracted based on the prompt's requirements:

1. A table of acceptance criteria and the reported device performance

The document states: "The verification and validation testing of the predicate device verification is deemed to be sufficient for the proposed device as the devices are identical with the exception of branding." This indicates that no new performance data or acceptance criteria specific to K181343 are presented because its performance is considered identical to its predicate (K171303). Therefore, a table for K181343's performance against specific acceptance criteria is not provided in this document.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

Not provided for K181343. Since no new testing was performed for K181343, this information would pertain to the predicate device (K171303) and is not detailed here.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not provided for K181343. This information would relate to the testing of the predicate device (K171303) and is not detailed here.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not provided for K181343. This information would relate to the testing of the predicate device (K171303) and is not detailed here.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

The document states: "No clinical testing was performed in support of this 510(k) premarket notification." Therefore, no MRMC comparative effectiveness study was done for this specific submission (K181343).

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The document states: "No clinical testing was performed in support of this 510(k) premarket notification." Therefore, no standalone algorithm performance study was done for this specific submission (K181343).

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not provided for K181343. This information would relate to the testing of the predicate device (K171303) and is not detailed here.

8. The sample size for the training set

Not provided for K181343. This information would relate to the development and training of the algorithm for the predicate device (K171303) and is not detailed here.

9. How the ground truth for the training set was established

Not provided for K181343. This information would relate to the development and training of the algorithm for the predicate device (K171303) and is not detailed here.

Summary of the Study that Proves the Device Meets Acceptance Criteria:

The document explicitly states: "The proposed device and K171303 are software tools using identical standard, industrial computing hardware and applications. The code in the proposed software package is identical to K171303." And further: "The verification and validation testing of the predicate device verification is deemed to be sufficient for the proposed device as the devices are identical with the exception of branding."

This means that no new study was performed for K181343 to demonstrate its performance against acceptance criteria. The device is considered substantially equivalent to a previously cleared predicate device (K171303) because it is essentially the same software, differing only in branding. Therefore, the prior testing and validation of the predicate device (K171303) is the "study" referenced, and its acceptance criteria and performance data are considered applicable to K181343 by extension. The details of that predicate study are not present in this document.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 12, 2018

Bracco Diagnostics Inc. % Ms. Patrice Marchildon Senior Director Regulatory Affairs 259 Prospect Plains Road, Building H MONROE TOWNSHIP NJ 08831

Re: K181343

Trade/Device Name: HeartSee Cardiac P.E.T. Processing Software Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: KPS Dated: May 18, 2018 Received: May 21, 2018

Dear Ms. Marchildon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

Page 2 - Ms. Patrice Marchildon

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Willin C. Jy for

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use 4.

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below

510(k) Number (if known)

K181343

Device Name

HeartSee Cardiac P.E.T. Processing Software

Indications for Use (Describe)

HeartSee Software for cardiac position tomography (PET) is indicated for determining regional and global absolute rest and stress myocardial perfusion in co'min/g, Coronary Flow Reserve and their combination into the Coronary Flow Capacity (CFC) Map in patients with suspected or known coronary artery disease (CAD) in order to assist clinical interpretation of PET perfusion images by quantification of their severity.

HeartSee is intended for use by trained professionals, such as nuclear technicians, nuclear cardiology physicians, or cardiologists with appropriate training and certification. The climician remains ultimately responsible for the final assessment and diagnosis based on standard practices, clinical judgment and interpretation of PET images or quantitative data.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including thetime to review instructions, search existing data sources, gather and maintain the data needed and completeand review the collection of information. Send comments regarding this burden estimate or any other aspectof this information collection, including suggestions for reducing this burden, to:
Department of Health and Human ServicesFood and Drug AdministrationOffice of Chief Information OfficerPaperwork Reduction Act (PRA) StaffPRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection ofinformation unless it displays a currently valid OMB number."

information unless it displays a currently valid OMB number

FORM FDA 3881 (1/14)

Page of

Service (501) 441-6741

{3}------------------------------------------------

510(k) Summary 5.

Bracco Diagnostics Inc. HeartSee Cardiac P.E.T. Processing Software (per 21CFR 807.92)

1. SUBMITTER/510(K) Holder

Headquarters

Bracco Diagnostics Inc. 259 Prospect Plains Road Monroe Township, NJ 08831 USA

Contact: Patrice Marchildon, Senior Director Regulatory Affairs Contact Phone: 609-514-2284 Contact Email: Patrice.Marchildon@diaq.bracco.com

Date Prepared: May 18, 2018

2. DEVICE NAME

Proprietary Name:HeartSee Cardiac P.E.T. Processing Software
Common name:Cardiac Positron Emission Tomography (PET)Analysis Software
Classification Name:Emission computed tomography system.
Classification Regulation:21 CFR 892.1200 (Class II)
Product Code:KPS

3. PREDICATE DEVICE

  • HeartSee Cardiac P.E.T. Processing Software (K171303) ●

4. Device Description

HeartSee is a software tool for cardiac positron emission tomography (PET) for determining regional and global absolute rest and stress myocardial perfusion in cc/min/q. Coronary Flow Reserve and their combination into the Coronary Flow Capacity (CFC) Map for facilitating the interpretation of PET perfusion images in patients with suspected of known coronary artery disease. HeartSee is intended for use by trained professionals, such as nuclear technicians, nuclear medicine or nuclear cardiology physicians, or cardiologists with appropriate training and certification.

HeartSee contains two fundamental components. First, the software imports cardiac PET images in DICOM format from PET scanners with DICOM output. These images are reoriented to cardiac axes to produce standard tomographic and topographic displays of relative uptake. Second, the software quantifies absolute rest and stress myocardial perfusion per unit tissue (cc/min/gm), Coronary Flow Reserve (CFR) as the stress/rest perfusion ratio and the Coronary Flow Capacity combining CFR and stress perfusion, all on a pixel

{4}------------------------------------------------

basis for regional and global values. Archiving output data is supported for clinical diagnostics, quality control and research.

5. INDICATIONS FOR USE

HeartSee Software for cardiac positron emission tomography (PET) is indicated for determining regional and global absolute rest and stress myocardial perfusion in cc/min/q, Coronary Flow Reserve and their combination into the Coronary Flow Capacity (CFC) Map in patients with suspected or known coronary artery disease (CAD) in order to assist clinical interpretation of PET perfusion images by quantification of their severity.

HeartSee is intended for use by trained professionals, such as nuclear technicians, nuclear medicine or nuclear cardiology physicians, or cardiologists with appropriate training and certification. The clinician remains ultimately responsible for the final assessment and diagnosis based on standard practices, clinical judgment and interpretation of PET images or quantitative data.

6. Summary of Technological Characteristics Compared to the Predicate DEVICE/S

The proposed device and K171303 are software tools using identical standard, industrial computing hardware and applications. The code in the proposed software package is identical to K171303.

7. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE

The verification and validation testing of the predicate device verification is deemed to be sufficient for the proposed device as the devices are identical with the exception of branding.

8. Summary of Clinical Testing as Basis for Substantial Equivalence

No clinical testing was performed in support of this 510(k) premarket notification.

9. SUMMARY OF OTHER INFORMATION

No other information is available.

10. Conclusions Drawn From Non-clinical and Clinical Tests

Since the proposed HeartSee Cardiac P.E.T. Processing Software is identical to the K171303 version. Bracco Diagnostics Inc. has concluded that the proposed HeartSee Cardiac P.E.T. Processing Software is substantially equivalent to the predicate device K171303.

§ 892.1200 Emission computed tomography system.

(a)
Identification. An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.(b)
Classification. Class II.