K Number
K181269
Device Name
iCertainty
Manufacturer
Date Cleared
2018-12-14

(214 days)

Product Code
Regulation Number
870.2100
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The iCertainty System is intended for non-invasive two-dimensional imaging as an adjunctive method for evaluation of blood flow distribution (flow and perfusion) in tissues, up to a depth of 4-5mm.
Device Description
The iCertainty System is a device used to image blood flow distribution (flow in vessels and perfusion in tissues) in real time. The device employs a mobile cart so that it can be configured in the appropriate orientation to properly image the tissue in an operating room environment without restrictions. When a practitioner elects to use imaging as an adjunct to evaluate blood flow distribution the device is moved into the appropriate position and an extension arm with an optical head unit (OHU, containing the lasers/camera equipment) is positioned over the patient as described in the instructions for use. The OHU is moved to a specified height (focused) over the target image area and the imaging is performed. The imaging can be performed as often as deemed necessary by the surgeon during the procedure to visualize blood flow distribution (blood flow in vessels and perfusion in tissues) in the target tissues being operated upon. The device employs a dual laser-based system with a camera imaging system. The two lasers (450nm and 785nm) illuminate the surface and blood flow distribution up to a depth of 4-5mm of the target tissue. A digital camera records the image in real time. All real-time blood flow distribution analyses are performed on a central computer within the device. Realtime images are displayed to the surgeon on an imaging display. Data entry is performed on a separate display and data entry keyboard. A sterile drape is used as an accessory to protect the sterile operating field from contamination during use.
More Information

No
The description focuses on laser illumination, camera imaging, and real-time processing of reflected patterns using optical reflectance and laser speckle contrast analysis. There is no mention of AI or ML algorithms being used for image analysis or interpretation.

No.
The device is described as an imaging system intended for non-invasive two-dimensional imaging of blood flow distribution, used as an adjunctive method for evaluation. Its function is to visualize and display information, not to actively treat or modify physiological processes.

Yes

The device is intended for "non-invasive two-dimensional imaging as an adjunctive method for evaluation of blood flow distribution (flow and perfusion) in tissues," which is a diagnostic purpose. It helps surgeons visualize and evaluate blood flow distribution in real-time during procedures, aiding in their assessment of the patient's condition.

No

The device description explicitly details hardware components including a mobile cart, extension arm, optical head unit (OHU) containing lasers and camera equipment, a central computer, imaging display, and a separate data entry display and keyboard. While it performs image processing, it is not solely software.

Based on the provided information, the iCertainty System is not an In Vitro Diagnostic (IVD).

Here's why:

  • IVDs are used to examine specimens derived from the human body. The iCertainty System is used for non-invasive imaging of tissues directly on the patient. It does not involve the collection or analysis of samples like blood, urine, or tissue biopsies.
  • The intended use is for imaging blood flow distribution in tissues. This is a direct measurement on the living patient, not an analysis of a biological specimen in a laboratory setting.
  • The device description clearly outlines its use in an operating room environment, positioned over the patient. This is consistent with an in vivo imaging device, not an in vitro diagnostic device.

Therefore, the iCertainty System falls under the category of an in vivo imaging device, not an In Vitro Diagnostic.

N/A

Intended Use / Indications for Use

The iCertainty System is intended for non-invasive two-dimensional imaging as an adjunctive method for evaluation of blood flow distribution (flow and perfusion) in tissues, up to a depth of 4-5mm.

Product codes

DPW

Device Description

The iCertainty System is a device used to image blood flow distribution (flow in vessels and perfusion in tissues) in real time. The device employs a mobile cart so that it can be configured in the appropriate orientation to properly image the tissue in an operating room environment without restrictions. When a practitioner elects to use imaging as an adjunct to evaluate blood flow distribution the device is moved into the appropriate position and an extension arm with an optical head unit (OHU, containing the lasers/camera equipment) is positioned over the patient as described in the instructions for use. The OHU is moved to a specified height (focused) over the target image area and the imaging is performed. The imaging can be performed as often as deemed necessary by the surgeon during the procedure to visualize blood flow distribution (blood flow in vessels and perfusion in tissues) in the target tissues being operated upon. The device employs a dual laser-based system with a camera imaging system. The two lasers (450nm and 785nm) illuminate the surface and blood flow distribution up to a depth of 4-5mm of the target tissue. A digital camera records the image in real time. All real-time blood flow distribution analyses are performed on a central computer within the device. Real-time images are displayed to the surgeon on an imaging display. Data entry is performed on a separate display and data entry keyboard. A sterile drape is used as an accessory to protect the sterile operating field from contamination during use.

Mentions image processing

Yes

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Laser light illumination with digital camera recording

Anatomical Site

Targeted surgical tissues exposed to the device field of view, including both central organ and peripheral tissues

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Surgeons, physicians, nurses and other trained medical professionals, operating room environment

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Performance testing was performed to verify that the performance of the iCertainty system is substantially equivalent to currently marketed perfusion flow imaging systems, and substantially equivalent to the PeriCam PSI (K120884). Testing included:

  • use testing
  • data entry and integrity
  • electrical safety
  • electromagnetic compatibility
  • sterile barrier/drape compatibility
  • cleaning

In vivo blood flow and perfusion was evaluated through animal studies investigating skin/epidermis, static solid organ, vascular and moving solid organs in normal and diseased state.

Assessment of Clinical Data: Evaluation of human anatomy (hand, forearm and ankle from 40 patients) demonstrated that the iCertainty images blood flow distribution (flow and perfusion).
Overall Conclusions: Based on the indications for use, technological characteristics, and comparison to predicate devices, the iCertainty System has been shown to be substantially equivalent to the predicate.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

K120884 PeriCam PSI

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

K063345 Novadaq SPY

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 870.2100 Cardiovascular blood flowmeter.

(a)
Identification. A cardiovascular blood flowmeter is a device that is connected to a flow transducer that energizes the transducer and processes and displays the blood flow signal.(b)
Classification. Class II (performance standards).

0

Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized depiction of a human figure. To the right of it is the FDA logo, with the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in bold blue letters, and "ADMINISTRATION" in smaller blue letters below.

Rfpi % Jerzy Wojcik Sr Director RA/QA EdgeOne Medical 455 N Campbell Ave Chicago, Illinois 60612

December 14, 2018

Re: K181269

Trade/Device Name: iCertainty Regulation Number: 21 CFR 870.2100 Regulation Name: Cardiovascular Blood Flowmeter Regulatory Class: Class II Product Code: DPW Dated: November 20, 2018 Received: November 26, 2018

Dear Jerzy Wojcik:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

1

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Digitally signed by Neil R Neil R Ogden -S Date: 2018.12.14 Ogden -S Date: 2018.12.1-For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known)

K181269

Device Name iCertainty

Indications for Use (Describe)

The iCertainty System is intended for non-invasive two-dimensional imaging as an adjunctive method for evaluation of blood flow distribution (flow and perfusion) in tissues, up to a depth of 4-5mm.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

Date Summary Prepared:December 12, 2018
510(k) Owner:RFPi
Contact Person:Jeffery Basham
CEO, RFPi
1800 N Greene Street, Suite K
Greenville, NC 27834
919-280-2953
Jeff.basham@rfpi-co.com
510(k) Consultant Contact:
Jerzy Wojcik
Sr. Director RA/QA, EdgeOne Medical
455 N Campbell Ave, Suite 2N
Chicago, IL 60612
312-300-6643
Jerzy.wojcik@edgeonemedical.com
Device Name:iCertainty
Trade Name:iCertainty
Common Name:Flowmeter, Blood,
Cardiovascular
Regulation:870.2100
Class:II
Product Code:DPW
Predicate Device(s):
Primary Predicate K120884 PeriCam PSI
Reference Device K063345 Novadaq SPY
Device Description:The iCertainty System is a device used to image blood
flow distribution (flow in vessels and perfusion in
tissues) in real time. The device employs a mobile cart so
that it can be configured in the appropriate orientation to
properly image the tissue in an operating room
environment without restrictions. When a practitioner
elects to use imaging as an adjunct to evaluate blood flow
distribution the device is moved into the appropriate

510(k) Summary

4

position and an extension arm with an optical head unit (OHU, containing the lasers/camera equipment) is positioned over the patient as described in the instructions for use. The OHU is moved to a specified height (focused) over the target image area and the imaging is performed. The imaging can be performed as often as deemed necessary by the surgeon during the procedure to visualize blood flow distribution (blood flow in vessels and perfusion in tissues) in the target tissues being operated upon. The device employs a dual laser-based system with a camera imaging system. The two lasers (450nm and 785nm) illuminate the surface and blood flow distribution up to a depth of 4-5mm of the target tissue. A digital camera records the image in real time. All real-time blood flow distribution analyses are performed on a central computer within the device. Realtime images are displayed to the surgeon on an imaging display. Data entry is performed on a separate display and data entry keyboard. A sterile drape is used as an accessory to protect the sterile operating field from contamination during use.

  • Statement of Intended Use: The iCertainty System is intended for non-invasive twodimensional imaging as an adjunctive method for evaluation of blood flow distribution (flow and perfusion) in tissues, up to a depth of 4-5mm.
    • Accessory Device: Premier Guard sterile drape (K041501)

5

Device Comparisons
FeatureiCertainty Proposed
DevicePrimary Predicate
PeriCam PSI (K120884)Reference Device
Novadaq Spy (K063345)
Intended Use /
Indications...for non-invasive two-
dimensional imaging as
an adjunctive method for
evaluation of blood flow
distribution (flow and
perfusion) in tissues, up to
a depth of 4-5mm....for non-invasive two-
dimensional imaging of
peripheral tissue blood
perfusion.... for capturing and
viewing fluorescent
images for the visual
assessment of blood flow
as an adjunctive method
for the evaluation of
tissue perfusion, and
related tissue transfer
circulation in tissue and
free flaps used in plastic,
micro- and reconstructive
surgical procedures.
Conclusion – Both the subject and predicate devices have the same intended use as
an adjunctive method for evaluation of blood flow and perfusion.
Target UsersSurgeons, physicians,
nurses and other trained
medical professionalsSurgeons, physicians,
nurses and other trained
medical professionalsSurgeons, physicians,
nurses and other trained
medical professionals
Conclusion – Substantially equivalent. All three devices have the same target users.
Targeted
anatomical sitesTargeted surgical tissues
exposed to the device field
of view, including both
central organ and
peripheral tissuesTargeted peripheral
tissues exposed to the
device field of viewTargeted surgical tissues
exposed to the device field
of view, including both
central organ and
peripheral tissues
Conclusion - Substantially equivalent. All three devices image tissue that are in the
field of view of the device, including central organ and peripheral tissues in the subject
and primary predicate devices.
Mechanism of
Action /
TechnologyLaser light illuminates the
target surface tissue.
Using both optical
reflectance and laser
speckle contrast analysis,
the pattern reflected from
the blood in the tissue is
processed and imaged to
generate images of blood
perfusion in target tissues
up to a depth of 4-5 mm.Laser light illuminates the
target surface tissue.
Using laser speckle
contrast analysis, the
pattern reflected from the
blood in the tissue is
processed and imaged to
generate images of blood
perfusion (depth not
determined, should be
similar to iCertainty by
optical physics
parameters).Laser light illuminates the
target surface tissue
which has a fluorescent
dye coursing through
blood vessels introduced
through an intravenous
injection. The laser
excites the dye, which
emits an infrared light.
The infrared pattern
reflected is processed and
imaged to generate
images of blood perfusion
Device Comparisons
FeatureiCertainty Proposed
DevicePrimary Predicate
PeriCam PSI (K120884)Reference Device
Novadaq Spy (K063345)
Conclusion - Substantially equivalent to the predicate, PeriCam PSI and similar to
the reference, Novadaq SPY. All three devices:
use laser light illumination of target tissues, then analyze the reflected light
to generate images of blood perfusion;
have a laser illumination system, camera, computer software for analysis and
display capabilities when in use;
have the capability to view, record and replay images of blood flow and
perfusion in target tissues.
Differences between the subject and predicate and reference devices are:
the reference device requires an invasive injection of a fluorophobe ICG dye;
the subject and predicate device do not require an invasive injection;
the subject and predicate device use laser speckle contrast imaging analysis;
the reference device relatively quantifies the intensity of fluorescence as an
indicator of flow and perfusion;
the subject device displays both the visible light and perfusion image side by
side for user reference and orientation;
the predicates and reference present only the perfusion image.
The iCertainty technology used to achieve the intended use as compared to the
predicate does not raise new questions of safety or effectiveness. Refer to Section 12.2
for further substantial equivalence discussion.
Method of UseAfter preparing the device
for use, the iCertainty is
positioned above the
targeted imaging site. The
device is activated, and
images are collected and
viewed by the operator. If
desired, the images can be
viewed again at a later
time.After preparing the device
for use, the PeriCam PSI is
positioned above the
targeted imaging site. The
device is activated, and
images are collected and
viewed by the operator. If
desired, the images can be
viewed again at a later
time.After preparing the device
for use, the patient is
injected with a fluorescent
dye. Novadaq SPY is
positioned above the
targeted imaging site. The
dye must be delivered to
the target tissue by the
circulatory system. The
device is activated, and
images are collected and
viewed by the operator. If
desired, the images can be
viewed again at a later
time.
Conclusion - Substantially equivalent. All three devices utilize the same method of
use, in that the device is positioned over the target area and activated for collection of
images for immediate viewing and/or again at a later time. Refer to Section 12.3 for
further substantial equivalence discussion.
Electrical Safety
and
Electromagnetic
CompatibilityIEC 60601-1 and
60601-1-2 compliantIEC 60601-1 and
60601-1-2 compliantIEC 60601-1 and
60601-1-2 compliant
Conclusion - Substantially equivalent. All three devices are compliant with
standards for electrical safety and compatibility.
Imaging
Distance (Optical
Head to Target
Tissue)$32 +/- 3$ cm30 cm10 cm
Conclusion – substantially equivalent. The subject device and SPY predicate device
have comparable imaging distances.
Field of View9 cm x 9 cmUp to 24 cm x 24 cm5 cm x 7 cm
FeatureiCertainty Proposed DevicePrimary Predicate PeriCam PSI (K120884)Reference Device Novadaq Spy (K063345)
Conclusion - The subject device and the predicate have comparable fields of view. The SPY reference device has a smaller field of view compared to the subject device. The larger FOV provides for a larger area of the target tissue to be captured in the imaging acquisition, from the same imaging distance.
Depth of Penetration and detection4-5 mm Documented PenetrationPenetration and Detection not Reported5-6 mm Reported Penetration (Detection not reported)
4-5 mm Documented Detection
Conclusion - similar to the reference device and not comparable to the Predicate. Only iCertainty has documented both depth of penetration and depth of detection.
Resolution and ContrastImage Resolution:
112µm/pixelImage Resolution:
100µm/pixelImage Resolution:
not specified
Max:
1600 x 800 Data Points
Resolution:
1600 x x800 pixels
Speckle Contrast:
0-1Max:
1386 x 1038 Data Points
Resolution:
1386 x 1038 pixels
Speckle Contrast:
0-1Max:
1024 x 768 Data Points
Resolution:
1024 x 768 pixels
Conclusion - Substantially equivalent. All three devices have the similar resolution and contrast.
Image PresentationVisible color and MSPVBlack and White NIRBlack and White NIR
Conclusion - Substantially equivalent. All three devices display imaging that the user can interpret. However, the subject device has the added benefit of a visible light image of anatomic detail for user reference and orientation
Imaging DepictionBlood flow and perfusionBlood flow and perfusionBlood flow and perfusion
Conclusion - Substantially equivalent. Imaging from all three devices depict blood flow and perfusion.
User InterfaceKeyboard, Graphical User Interface, Buttons on Camera headButtons on Camera headKeyboard, Graphical User Interface, Buttons on Camera head
Conclusion - Substantially equivalent. The reference device and iCertainty use a keyboard, graphical user interface and buttons on the camera head. Even though the predicate does not include a keyboard and Graphical User Interface, both components are required to be provided by the user in order to operate the predicate.
Mode of UseMobile cart. When needed for imaging, the cart is positioned such that the arm can be extended into a position over the patient for imaging.The device arm is mobile. When needed for imaging, the arm is mounted to a table or other surface such that the arm can be extended into a position over the patient for imaging.Mobile cart. When needed for imaging, the cart is positioned such that the arm can be extended into a position over the patient for imaging.
Conclusion - Substantially equivalent. All three devices are configurable to achieve orientation. The reference device and iCertainty are mobile carts that can be configured in the appropriate orientation with an extension arm that is positioned to image tissue without restrictions. The predicate is fixed to a surface but has an extension arm as well.
Device Comparisons
FeatureiCertainty Proposed
DevicePrimary Predicate
PeriCam PSI (K120884)Reference Device
Novadaq Spy (K063345)
SetupThe iCertainty contains an
adjustable arm which
contains the laser and
camera componentry and
is mounted to a mobile
cart which contained
video screens, keyboard
and computer.The device has an
adjustable arm which
contains the laser and
camera componentry and
must be mounted to a
table or mobile cart (not
part of the device). The
PeriCam PSI is then
connected via firewall
cable to a firewall port on
a computer.The device has an
adjustable extension arm
which contains the
camera. The laser
illumination system is
mounted in the cart and
connected to the Camera
head by fiberoptic cable.
The mobile cart also
contains video screens,
keyboard and computer.
Conclusion - Substantially equivalent. All three devices have adjustable arms that
are stabilized for use, and utilize a computer to operate the device. While the
iCertainty and reference devices include a cart to which the arm is mounted and a
computer to process imaging, the predicate requires the computer to be supplied by
the user and the device to be mounted to a table or mobile cart. All three devices are
substantially equivalent in that all require the same componentry in order to achieve
their intended use.
Imaging Cycles10 imaging cycles, limited
by softwareMultiple imaging episodes
per evaluation, limited by
software1 to 6 imaging cycles,
limited by ICG Dye
Conclusion - Substantially equivalent. Multiple imaging episodes can be obtained
per case/evaluation.
Size81.3 x 81.3 x 198.1 cm
221 kgPeriCam PSI Head:
22 x 15 x 20 cm
PeriCam PSI Head
2.3/2.4 kg
Arm and Stand
7.1/10.1 kg
Total weight
9.4/12.5 kg
(Standard/High resolution
models)81.2 x 81.2 x 180.4 cm
165 kg
Conclusion - Substantially equivalent. The iCertainty and the reference devices are
all-in-one devices which include the computer and stand/table/cart, whereas the
predicate does not include the computer and stand/table/cart and this is smaller in size
specification. All three devices have an imaging arm. The iCertainty and reference
devices are of similar size, but the iCertainty device contains a larger counterweight,
and thus weighs more. All three camera head units are of relatively similar size (but
moderately different configurations).
Maintenance
CleaningiCertainty is used with a
sterile drape, which
protects the patient from
device contamination.
Additionally, if cleaning
or disinfection is required,
the user manual provides
instructions for cleaning.The PeriCam PSI user
manual provides
instructions for cleaning.The Novadaq SPY is used
with a sterile drape, which
protects the patient from
device contamination.
Additionally, if cleaning
or disinfection is required,
the user manual provides
instructions for cleaning.
Device Comparisons
FeatureiCertainty Proposed
DevicePrimary Predicate
PeriCam PSI (K120884)Reference Device
Novadaq Spy (K063345)
Conclusion - Substantially equivalent. Both the iCertainty and the reference device
use a sterile drape to isolate the device arm and camera head from the sterile operative
field, and both include the same instructions for cleaning or disinfection of non-sterile
surfaces. The drapes do not alter the imaging characteristics. The predicate is not
used in a true sterile environment, and thus does not require a drape. All three devices
can be cleaned by the user/operator.
BiocompatibilityThe iCertainty does not
come into contact with the
patient. Materials of
construction for the arm,
camera head and cart,
which are in contact with
the user/operator for the
purpose of movement and
manipulation, are
composed of plastic and
metal commonly found in
every-day materials.
The device does not
require the use on an
injected fluorescent dye toThe PeriCam PSI does not
come into contact with the
patient. Materials of
construction for the arm
and camera head, which
are in contact with the
user/operator for the
purpose of movement and
manipulation, are
composed of plastic and
metal commonly found in
every-day materials.
The device does not
require the use on an
injected fluorescent dye toThe Novadaq SPY does
not come into contact with
the patient. Materials of
construction for the arm,
camera head and cart,
which are in contact with
the user/operator for the
purpose of movement and
manipulation, are
composed of plastic and
metal commonly found in
every-day materials.
The device requires the
use of a biocompatible
fluorescent dye
(Indocyanine Green or
ICG) which has a low
toxicity profile.
achieve its intended use.
achieve its intended use.
Conclusion - Substantially equivalent. All three devices are not in contact with the
patient. User contact materials are considered safe common plastic and metal
materials. The reference device requires use of a low-toxicity profile fluorescent dye
to achieve intended use, while the iCertainty and predicate do not. The difference in
the biocompatibility comparison with the reference device does not raise new
questions of safety or effectiveness, and in fact reduces the safety risk for the iCertainty because the patient is not exposed to the fluorescent dye.
Operating
ConditionsTemperature:
10°C to 30°C
Relative Humidity:
30-80% non-condensing
Atmospheric Pressure
Range:
700hPA to 1060hPA.Temperature:
15°C to 30°C
Relative Humidity:
10-80% non-condensing
Atmospheric Pressure
Range:
70 kPa to 160 kPa.Temperature:
10°C to 30°C
Relative Humidity:
10% to 90%, non-
condensing
Atmospheric Pressure
Range:
94 kPa to 102 kPa
Conclusion - Substantially equivalent. The operating conditions for all three devices
are similar.
Power SupplyOperates on standard
Mains power supply in the
United States.Operates on standard
Mains power supply in the
United States.Operates on standard
Mains power supply in the
United States.
Conclusion - Substantially equivalent. All three devices operate using a standard
Mains power supply in the United States.

6

7

8

9

Performance Data: Performance testing was performed to verify that the performance of the iCertainty system is substantially equivalent to currently marketed perfusion flow imaging

10

systems, and substantially equivalent to the PeriCam PSI (K120884). Testing included:

  • . use testing
  • . data entry and integrity
  • electrical safety ●
  • . electromagnetic compatibility
  • sterile barrier/drape compatibility
  • . cleaning

In vivo blood flow and perfusion was evaluated through animal studies investigating skin/epidermis, static solid organ, vascular and moving solid organs in normal and diseased state.

  • Assessment of Clinical Data: Evaluation of human anatomy (hand, forearm and ankle from 40 patients) demonstrated that the iCertainty images blood flow distribution (flow and perfusion).
    • Overall Conclusions: Based on the indications for use, technological characteristics, and comparison to predicate devices, the iCertainty System has been shown to be substantially equivalent to the predicate.