(84 days)
The StatStrip Glucose Hospital Meter System is intended for point-of-care, in vitro diagnostic, multiple-patient use for the quantitative determination of glucose in capillary finger stick, venous whole blood, neonate arterial whole blood, and neonate heel stick specimens throughout all hospital and all professional healthcare settings, including patients receiving intensive medical intervention/therapy.
The system should only be used with single-use, auto-disabling lancing devices when performing a capillary finger stick or neonate heel stick.
It is not intended for use with neonate cord blood specimens.
It is not intended for the screening or diabetes mellitus but is indicated for use in determining dysglycemia.
The StatStrip Glucose Hospital Meter System was previously cleared under K150281. The purpose of this dual 510(k) and CLIA Waiver submission for the StatStrip Glucose Hospital Meter System is to expand the indication for use to include the use of capillary specimens on patients receiving intensive medical intervention/therapy and support a CLA Waived categorization for this expanded indication for use to include the quantitative determination of glucose in capillary finger stick throughout all hospital and all professional healthcare settings. No changes have been made to the software, the test strips, controls or linearity solutions.
The system contains the following:
-
- StatStrip Meter, with integrated Wi-Fi connection and antenna option
-
- Charging Station
-
- Vial of StatStrip Test Strips
-
- Battery (3.7 V Lithium)
-
- Quick Reference Guide
-
- Instructions for Use Manual
-
- StatStrip Log Book
Offered separately:
- . StatStrip Test Strips
- Quality Control Solutions, Level 1, 2, 3
- Linearity Solutions, Levels 1, 2, 3, 4, 5.
Meter: The StatStrip Glucose Hospital Meter is a hand-held testing device that works in conjunction with Nova StatStrip Glucose test strips to measure glucose in a whole blood sample. Meter operation is selfprompting using an illuminated touch-screen Graphical User Interface (GUI). In addition to measuring glucose, the meter also stores patient test data, quality control test data and other information relating to the patient, patient sample, operator, reagents and meter.
The self-prompting menu system is navigated by means of a combination of touch-panel on-screen keys, on-screen soft keyboard. The operator can use a finger or a PDA stylus to select options for the onscreen display. It offers audible feedback for user inputs, and audible and/or visual feedback for prompts and user alerts. The Meter also has a barcode scanner that automates data entry.
The rechargeable batteries provide sufficient power to operate for 8 hours before requiring recharging. A "battery fuel gauge" bitmap constantly informs the user as to the current state of charge on the battery. Battery charge state information is available on the "meter Welcome screen". The meter will shutoff (Sleep) after 90 seconds of inactivity. Test data and meter setup information will be stored in a nonvolatile form to prevent data loss.
Charging (Docking) Station: The meter charging station is a stationary accessory used to recharge the meter. The charging station has one slot for the meter to be placed for charging and a slot for an extra battery to be charged.
The charging station should be located central to the patient care area being served by the meter (e.g. a nursing station). The data charging station must remain plugged in to a wall outlet for power.
The system still allows the charging station to be used to transfer data from the meter to a central workstation, and allow meter setup information to be downloaded from the central workstation to the meter.
Test Strips: The test strips contain a reaction layer that contains a glucose-enzyme (greater than 1.0 IU) and ferricvanide as a mediator. The test strip is touched to a drop of blood to initiate the test process. The strip is designed such that when a drop of blood is touched to the strip, the blood is drawn into the reaction space via capillary action. A simple one-step process provides a blood glucose result. Test strips will be sold in vials of 25 strips.
Control Solutions: The control solutions are aqueous assayed solutions, containing buffered D-Glucose, viscosity-adjusting agent, preservatives and other non-reactive ingredients (dye). They contain no products of human origin. There are three levels of controls, (Level 1, Level 2 and Level 3). These solutions will be offered for sale separately from the meter.
Linearity Solutions: There are 5 levels of Linearity solutions containing buffered D-Glucose, viscosity-adjusting agent, preservatives and other non-reactive ingredients (dye). They contain no products of human origin. These solutions are offered separately from the system for users to verify the performance of the system.
Replacement batteries will be offered separately.
Wi-Fi Option: The StatStrip Glucose Hospital Meter provides a Wi-Fi communication method with a healthcare facility's network system. The wireless radio chipset utilizing the IEEE 802.11 alb/g communication protocols (Wi-Fi) is used as an optional means to transmit and receive test results, previously cleared under K150281.
The provided text describes the Nova Biomedical Corporation's StatStrip Glucose Hospital Meter System. Here's a breakdown of the acceptance criteria and study information:
Acceptance Criteria and Device Performance for StatStrip Glucose Hospital Meter System
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for glucose meters, particularly those for critical care, typically align with ISO 15197 standards or internal company-defined standards that demonstrate clinical agreement. While explicit "acceptance criteria" are not given in quantitative terms like "must achieve X% within Y mg/dL," the clinical study performance data serves as the reported performance against FDA's substantial equivalence requirements. Therefore, the "acceptance criteria" can be inferred from the performance metrics demonstrated.
Interpreted Acceptance Criteria (Based on Study Performance):
- For glucose concentrations < 75 mg/dL: A high percentage of results should be within ±15 mg/dL of the comparator method.
- For glucose concentrations ≥ 75 mg/dL: A high percentage of results should be within ±20% of the comparator method.
Reported Device Performance (from Clinical Study Performance section):
| Glucose Concentration Range | Performance Metric | Study #1 (Prospective) | Study #2 & #3 (Combined Real-World) |
|---|---|---|---|
| < 75 mg/dL | Within ± 5 mg/dL | 100% (1/1) | 47.9% (907/1894) |
| Within ± 10 mg/dL | 100% (1/1) | 77.6% (1470/1894) | |
| Within ± 12 mg/dL | 100% (1/1) | 85.2% (1614/1894) | |
| Within ± 15 mg/dL | 100% (1/1) | 91.7% (1737/1894) | |
| Exceeds ± 15 mg/dL | 0% (0/1) | 8.3% (157/1894) | |
| ≥ 75 mg/dL | Within ± 5% | 48.9% (277/567) | 50.2% (7473/14884) |
| Within ± 10% | 79.4% (450/567) | 79.3% (11807/14884) | |
| Within ± 12% | 85.4% (484/567) | 86.0% (12799/14884) | |
| Within ± 15% | 91.0% (516/567) | 92.1% (13712/14884) | |
| Within ± 20% | 96.8% (549/567) | 96.4% (14350/14884) | |
| Exceeds ± 20% | 3.2% (18/567) | 3.6% (534/14884) |
2. Sample size used for the test set and the data provenance:
- Prospective Clinical Study (Site #1):
- Sample Size: 568 critical care patients.
- Data Provenance: Prospective study, conducted at a single leading medical center in Minnesota, U.S. (one of the top-ranked hospitals in Diabetes & Endocrinology).
- Real-World Evidence Studies (Site #2 & #3):
- Sample Size:
- Study #2: 2,133 critical care patients.
- Study #3: 14,645 critical care patients.
- Combined for reporting: 1894 (for <75 mg/dL) and 14884 (for >=75 mg/dL). This suggests that after filtering or specific data points, these are the totals used for analysis.
- Data Provenance: Retrospective data mining ("real world evidence") from two leading medical centers in the U.S. (one in Minnesota and one in Maryland). These were described as CLIA Waived retrospective studies.
- Sample Size:
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not provided in the document. The ground truth was established by a central laboratory comparator method.
4. Adjudication method for the test set:
This information is not provided in the document.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This is not applicable. The device is a glucose meter, not an imaging device or AI diagnostic tool that assists human readers.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
Yes, the studies assessed the performance of the StatStrip Glucose Hospital Meter System (algorithm-driven device) by comparing its readings to a central laboratory comparator method. This is essentially a standalone performance evaluation of the device itself. The "CLIA Waived operators" used the device, but the performance being measured is of the device's accuracy compared to a gold standard, not the operator's diagnostic ability.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
The ground truth was established using plasma glucose specimens obtained from either an arterial or venous specimen as measured on a central laboratory comparator method, specifically an IDMS traceable hexokinase comparator method (Roche Cobas System, Roche Diagnostics, Indianapolis, IN). This is considered a highly accurate and standardized laboratory method for glucose determination.
8. The sample size for the training set:
This information is not provided in the document. The document describes clinical studies for performance evaluation for a 510(k) submission, not specifically developmental or training data for the device's underlying technology. The device itself was previously cleared (K150281) and this submission is an expansion of its indication for use, implying the core technology was already developed and validated.
9. How the ground truth for the training set was established:
This information is not provided in the document, as the document focuses on the expanded indication's validation rather than the initial device development.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
July 12, 2018
Nova Biomedical Corporation John Mchale VP of RA/QA and Technical Support 200 Prospect Street Waltham, MA 02454
Re: K181043
Trade/Device Name: StatStrip Glucose Hospital Meter System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Code: PZI Dated: April 18, 2018 Received: April 19, 2018
Dear John Mchale:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR
{1}------------------------------------------------
Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Kellie B. Kelm -S
Courtney H. Lias, Ph.D. for Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
K181043
Device Name StatStrip Glucose Hospital Meter System
Indications for Use (Describe)
The StatStrip Glucose Hospital Meter System is intended for point-of-care, in vitro diagnostic, multiple-patient use for the quantitative determination of glucose in capillary finger stick, venous whole blood, neonate arterial whole blood, and neonate heel stick specimens throughout all hospital and all professional healthcare settings, including patients receiving intensive medical intervention/therapy.
The system should only be used with single-use, auto-disabling lancing devices when performing a capillary finger stick or neonate heel stick.
It is not intended for use with neonate cord blood specimens.
It is not intended for the screening or diabetes mellitus but is indicated for use in determining dysglycemia.
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
[X] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995,
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) SUMMARY
510(k) Number: K181043
| 510(k) Owner: | Nova Biomedical Corporation |
|---|---|
| Registration Number: | 1219029 |
| Address: | 200 Prospect St.Waltham, MA 02453 |
| Fax Number: | 784-891-4806 |
| Contact Person: | John F. Mchale |
| Phone: | 781-894-0800 |
| Date Prepared: | July 3, 2018 |
Proprietary Name: StatStrip Glucose Hospital Meter System
Common or Usual Name: Blood Glucose Meter
Product Code: PZI
Classification Name:
| Classification Name | Class No. | Regulation No. | Class |
|---|---|---|---|
| Prescription Use Blood Glucose Meter ForNear-Patient Testing | 75 PZI | 862.1345 | II |
Predicate Device: StatStrip Glucose Hospital Meter System, K150281.
Device Description:
The StatStrip Glucose Hospital Meter System was previously cleared under K150281. The purpose of this dual 510(k) and CLIA Waiver submission for the StatStrip Glucose Hospital Meter System is to expand the indication for use to include the use of capillary specimens on patients receiving intensive medical intervention/therapy and support a CLA Waived categorization for this expanded indication for use to include the quantitative determination of glucose in capillary finger stick throughout all hospital and all professional healthcare settings. No changes have been made to the software, the test strips, controls or linearity solutions.
The system contains the following:
-
- StatStrip Meter, with integrated Wi-Fi connection and antenna option
-
- Charging Station
-
- Vial of StatStrip Test Strips
-
- Battery (3.7 V Lithium)
-
- Quick Reference Guide
-
- Instructions for Use Manual
-
- StatStrip Log Book
Offered separately:
- . StatStrip Test Strips
- Quality Control Solutions, Level 1, 2, 3 ●
- Linearity Solutions, Levels 1, 2, 3, 4, 5. ●
Meter
The StatStrip Glucose Hospital Meter is a hand-held testing device that works in conjunction with Nova StatStrip Glucose test strips to measure glucose in a whole blood sample. Meter operation is selfprompting using an illuminated touch-screen Graphical User Interface (GUI). In addition to measuring
{4}------------------------------------------------
glucose, the meter also stores patient test data, quality control test data and other information relating to the patient, patient sample, operator, reagents and meter.
The self-prompting menu system is navigated by means of a combination of touch-panel on-screen keys, on-screen soft keyboard. The operator can use a finger or a PDA stylus to select options for the onscreen display. It offers audible feedback for user inputs, and audible and/or visual feedback for prompts and user alerts. The Meter also has a barcode scanner that automates data entry.
The rechargeable batteries provide sufficient power to operate for 8 hours before requiring recharging. A "battery fuel gauge" bitmap constantly informs the user as to the current state of charge on the battery. Battery charge state information is available on the "meter Welcome screen". The meter will shutoff (Sleep) after 90 seconds of inactivity. Test data and meter setup information will be stored in a nonvolatile form to prevent data loss.
Charging (Docking) Station
The meter charging station is a stationary accessory used to recharge the meter. The charging station has one slot for the meter to be placed for charging and a slot for an extra battery to be charged.
The charging station should be located central to the patient care area being served by the meter (e.g. a nursing station). The data charging station must remain plugged in to a wall outlet for power.
The system still allows the charging station to be used to transfer data from the meter to a central workstation, and allow meter setup information to be downloaded from the central workstation to the meter.
Test Strips
The test strips contain a reaction layer that contains a glucose-enzyme (greater than 1.0 IU) and ferricvanide as a mediator. The test strip is touched to a drop of blood to initiate the test process. The strip is designed such that when a drop of blood is touched to the strip, the blood is drawn into the reaction space via capillary action. A simple one-step process provides a blood glucose result. Test strips will be sold in vials of 25 strips.
Control Solutions
The control solutions are aqueous assayed solutions, containing buffered D-Glucose, viscosity-adjusting agent, preservatives and other non-reactive ingredients (dye). They contain no products of human origin. There are three levels of controls, (Level 1, Level 2 and Level 3). These solutions will be offered for sale separately from the meter.
Linearity Solutions
There are 5 levels of Linearity solutions containing buffered D-Glucose, viscosity-adjusting agent, preservatives and other non-reactive ingredients (dye). They contain no products of human origin. These solutions are offered separately from the system for users to verify the performance of the system.
Replacement batteries will be offered separately.
Wi-Fi Option
The StatStrip Glucose Hospital Meter provides a Wi-Fi communication method with a healthcare facility's network system. The wireless radio chipset utilizing the IEEE 802.11 alb/g communication protocols (Wi-Fi) is used as an optional means to transmit and receive test results, previously cleared under K150281.
Intended Use:
The StatStrip Glucose Hospital Meter System is intended for point-of-care, in vitro diagnostic, multiplepatient use for the quantitative determination of glucose in capillary finger stick, venous whole blood, arterial whole blood, neonate arterial whole blood, and neonate heel stick specimens throughout all
{5}------------------------------------------------
hospital and all professional healthcare settings, including patients receiving intensive medical intervention/therapy.
The system should only be used with single-use, auto-disabling lancing devices when performing a capillary finger stick or neonate heel stick.
lt is not intended for use with neonate cord blood specimens.
It is not intended for the screening or diagnosis of diabetes mellitus but is indicated for use in determining dysqlycemia.
Summary of the Technological Characteristics:
The StatStrip Glucose Hospital Meter System has the same scientific technology as the predicate. No changes have been made to the meter, the software, the controls or the linearity solutions. The only change is to the labeling of the device to allow the product to be used on capillary samples in critical care patient settings. The StatStrip Glucose Hospital Meter System is substantially equivalent to the StatStrip Glucose Hospital Meter System.
Summary of the Clinical Study Data:
Nova Biomedical conducted a large multi-center study of the StatStrip Glucose Hospital Meter System at two (2) leading medical centers to demonstrate the clinical performance of capillary whole blood specimens obtained by finger stick within critical care hospital settings on patients receiving intensive medical intervention/therapy. StatStrip capillary glucose specimens were compared to plasma glucose specimens obtained from either an arterial or venous specimen within 15 minutes of the capillary specimen as measured on a central laboratory comparator method. Clinical Study Site #1 was a CLIA Waived prospective study and the other 2 studies conducted in Site #3 were CLIA Waived retrospective studies designed to assess real world clinical performance throughout all hospital and all professional healthcare settings. Study Site #1 and #3 is one of the top ranked hospitals in Diabetes & Endocrinology located in Minnesota and study Site #2 is one the world's preeminent healthcare institutions located in Maryland. A description of the clinical study is found below. All central laboratory comparisons were against an IDMS traceable hexokinase comparator method (Roche Cobas System, Roche Diagnostics, Indianapolis, IN).
Prospective Clinical Study
The prospective capillary critical study (Site #1) included 568 critical care patients undergoing intensive medical intervention/therapy within 3 critical care departments (CVCU, MICU and OR). The critical care patients from the CVICU and MCU included 80 unique patient conditions receiving a total of 3,785 medications representing 17 parent drug classes. Medical conditions were unavailable for critical care patients in the OR.
Real World Evidence Studies
The purpose of these studies was to retrospectively data mine real world evidence on the performance of StatStrip glucose with capillary whole blood specimens obtained by skin puncture from adult and pediatric patients in Intensive Care Units (ICUs) on critically ill patients receiving Intensive Medical Intervention/Therapy as compared to plasma glucose obtained from either an arterial or venous specimen. These real world retrospective studies were conducted by CLIA Waived operators within these healthcare settings. The CLIA Waived operators were qualified to use the StatStrip system per institutional requirements in these critical care settings.
Real World Study Site Breakdown
Clinical Study #2 was a real world clinical review of 2,133 critical care patients throughout all critical care settings that had a glucose test performed using a capillary whole blood specimen obtained by finger stick and a plasma glucose test obtained from either an arterial or venous specimen performed in the
{6}------------------------------------------------
central laboratory within 15 minutes of the capillary test. Medical conditions were unavailable for critical care patients in this study.
Clinical Study #3 was a real world clinical review of 14,645 critical care patients throughout all critical care settings that had a glucose test performed using a capillary whole blood specimen obtained by finger stick and a plasma glucose test obtained from either an arterial or venous specimen performed in the central laboratory within 15 minutes of the capillary test. Medical conditions were unavailable for critical care patients in this study.
Clinical Study Performance
Study #1 (Prospective Study)
| Fingertip capillary samples with glucose concentrations <75 mg/dL | ||||
|---|---|---|---|---|
| Within | Within | Within | Within | Exceeds |
| $\pm$ 5 mg/dL | $\pm$ 10 mg/dL | $\pm$ 12 mg/dL | $\pm$ 15 mg/dL | $\pm$ 15 mg/dL |
| 1/1 (100%) | 1/1 (100%) | 1/1 (100%) | 1/1 (100%) | 0/1 (0%) |
Fingertin canillary samples with glucose concentrations <75 mg/dl
Fingertip capillary samples with glucose concentration ≥75 mg/dL
| Within± 5 % | Within± 10 % | Within± 12 % | Within± 15 % | Within± 20 % | Exceeds± 20 % |
|---|---|---|---|---|---|
| 277/567(48.9%) | 450/567(79.4%) | 484/567(85.4%) | 516/567(91.0%) | 549/567(96.8%) | 18/567(3.2%) |
Study #2 and #3 (Combined)
Fingertip capillary samples with glucose concentrations <75 mg/dL
| Within | Within | Within | Within | Exceeds |
|---|---|---|---|---|
| ± 5 mg/dL | ± 10 mg/dL | ± 12 mg/dL | ± 15 mg/dL | ± 15 mg/dL |
| 907/1894 | 1470/1894 | 1614/1894 | 1737/1894 | 157/1894 |
| (47.9%) | (77.6%) | (85.2%) | (91.7%) | (8.3%) |
Fingertip capillary samples with glucose concentration ≥75 mg/dL
| Within | Within | Within | Within | Within | Exceeds |
|---|---|---|---|---|---|
| ± 5 % | ± 10 % | ± 12 % | ± 15 % | ± 20 % | ± 20 % |
| 7473/14884 | 11807/14884 | 12799/14884 | 13712/14884 | 14350/14884 | 534/14884 |
| (50.2%) | (79.3%) | (86.0%) | (92.1%) | (96.4%) | (3.6%) |
{7}------------------------------------------------
Comparison of Predicate and Proposed devices
| Characteristic | Predicate Device: Nova StatStripHospital Glucose Meter (K150281) | Proposed: Device |
|---|---|---|
| Indication ForUse | The StatStrip Glucose Hospital MeterSystem is intended for point-of-care, invitro diagnostic, multiple-patient use forthe quantitative determination ofglucose in capillary finger stick, venouswhole blood, arterial whole blood,neonate arterial whole blood, andneonate heel stick specimens.The StatStrip Glucose Hospital MeterSystem is also intended for use in thequantitative determination of glucose invenous whole blood, arterial wholeblood, neonate heel stick and neonatalarterial samples throughout all hospitaland all professional healthcare settings.The system should only be used withsingle-use, auto-disabling lancingdevices when performing a capillaryfinger stick or neonate heel stick.It is not intended for use with neonatecord blood specimens.It is not intended for the screening ordiagnosis of diabetes mellitus but isindicated for use in determiningdysglycemia. | The StatStrip Glucose Hospital MeterSystem is intended for point-of-care, invitro diagnostic, multiple-patient use forthe quantitative determination ofglucose in capillary finger stick, venouswhole blood, arterial whole blood,neonate arterial whole blood, andneonate heel stick specimensthroughout all hospital and allprofessional healthcare settings,including critical care settings.The system should only be used withsingle-use, auto-disabling lancingdevices when performing a capillaryfinger stick or neonate heel stick.It is not intended for use with neonatecord blood specimens.It is not intended for the screening ordiagnosis of diabetes mellitus but isindicated for use in determiningdysglycemia. |
| MeasuringRange | 10-600 mg/dL | Same as Predicate |
| HematocritRange | 20-65% | Same as Predicate |
| OperatingPrinciple | Electrochemical biosensor | Same as Predicate |
| Sample size | 1.2 µL | Same as Predicate |
| Handheldmeter? | Yes | Same as Predicate |
| Calibration | Automatic, no Calibration Code | Same as Predicate |
| Data Storage | 1,500 patient or QC test results | Same as Predicate |
| Test time toresult | 6 seconds | Same as Predicate |
| Weight | 8.8 ounces | Same as Predicate |
| Barcodescanner | Yes | Same as Predicate |
| Power source | Rechargeable 3.7 volt Lithium battery | Same as Predicate |
| Characteristic | Predicate Device: Nova StatStripHospital Glucose Meter (K150281) | Proposed: Device |
| Test Strips -Active reagent: | Glucose Oxidase | Same as Predicate |
| QualityControls | Liquid, 3 levels | Same as Predicate |
| Linearity | 5 levels | Same as Predicate |
| DockingStation? | Yes | Same as Predicate |
| Wi-Fi NetworkConnectivity | Yes | Same as Predicate |
| Software | Yes | Same as Predicate |
{8}------------------------------------------------
Conclusion:
The results of the clinical study data confirmed that the StatStrip Glucose Hospital Meter System is safe and effective for its intended purpose. StatStrip Glucose Hospital Meter System is substantially equivalent to that of the predicate StatStrip Glucose Hospital Meter System (K15028).
§ 862.1345 Glucose test system.
(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.