(85 days)
MIM software is used by trained medical professionals as a tool to aid in evaluation and information management of digital medical images. The medical image modalities include, but are not limited to, CT. MRI. CR. DX. MG. US. SPECT. PET and XA as supported by ACRNEMA DICOM 3.0. MIM assists in the following indications:
- · Receive, transmit, store, retrieve, display, print, and process medical images and DICOM objects.
- · Create, display and print reports from medical images.
- · Registration, fusion display, and review of medical images for diagnosis, treatment planning,
· Evaluation of cardiac left ventricular function, including left ventricular end-diastolic volume, end-systolic volume, and ejection fraction.
· Localization and definition of objects such as tumors and normal tissues in medical images.
· Creation, transformation, and modification of contours for applications including, but not limited to, quantitative analysis, aiding adaptive therapy, transfering contours to radiation therapy treatment planning systems, and archiving contours for patient follow-up and management.
· Quantitative and statistical analysis of PET/SPECT brain scans by comparing to other registered PET/SPECT brain scans.
· Planning and evaluation of permanent implant brachytherapy procedures (not including radioactive microspheres). - · Post-treatment dose calculation of permanent Yttrium-90 (Y90) microsphere implants.
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Images that are printed to film must be printed using a FDA-approved printer for the diagnosis of digital mammography images. Mammographic images must be viewed on a display system that has been cleared by the FDA for the diagnosis of digital mammography images.
The software is not to be used for mammography CAD.
MIM - SPECTRA Quant is a feature of MIM. It is designed for use in medical imaging and operates on both Windows and Mac computer systems. MIM - SPECTRA Quant extends the functionality MIM - Y90 Dosimetry (K172218) software. The following functions have been added to allow quantitative reconstructions of Single Photon Emission Computed Tomography (SPECT) scans:
- Allows for decay correction of activity back to time of injection ●
- Allows for correction of photon attenuation ●
- . Allows for correction of photon scatter
- Allows for correction of collimator detector resolution effects ●
- Allows for camera sensitivity calibration providing output in Becquerels per ml (Bq/ml) .
Here's a breakdown of the acceptance criteria and study details for MIM - SPECTRA Quant based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
| Isotope | Collimator | Acceptance Criteria (Quantitative Error) | Reported Quantitative Error | Verification Result |
|---|---|---|---|---|
| Tc99m | LEHR | <10% | <5% | Passed |
| I-123 | LEHR | <10% | <5% | Passed |
| I-131 | HE | <10% | <5% | Passed |
| Lu-177 | ME | <10% | <5% | Passed |
Note: The acceptance criterion for MIM – SPECTRA Quant is a quantitative error of smaller or equal to 10% in regions that are large enough not to be affected by the partial volume effect.
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Description: The test set included "standard quality control phantoms, a Liver phantom, and simulated phantoms based on the NEMA IEC Body Phantom." It also involved "clinical data."
- Sample Size: The exact number of phantoms or clinical cases is not specified in the provided text.
- Data Provenance: The document does not explicitly state the country of origin for the data or whether it was retrospective or prospective. It only mentions the use of "physical and simulated phantoms as well as with clinical data." It's likely that the phantom data would be generated in-house or sourced from standard organizations.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- Number of Experts: "Board certified clinicians" performed "validation testing" and "qualified Nuclear Medicine professionals" rated reconstructions of clinical data. The exact number of clinicians or Nuclear Medicine professionals is not specified.
- Qualifications: "Board certified clinicians" and "qualified Nuclear Medicine professionals." Specific experience level (e.g., "10 years of experience") is not specified.
4. Adjudication Method for the Test Set
The document states that "All reconstructions of clinical data were rated as acceptable for clinical use by the qualified Nuclear Medicine professionals." This suggests an expert review process. However, a specific adjudication method (like 2+1, 3+1) for resolving disagreements among multiple experts, if multiple were involved, is not described.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
- Was it done? The document mentions "clinicians using MIM – SPECTRA Quant performed validation in comparison to independently implemented, industry-standard solutions." This suggests a comparison. However, it does not explicitly state that a formal Multi-Reader Multi-Case (MRMC) comparative effectiveness study was conducted with a focus on human reader improvement with vs. without AI assistance. The focus described is on the software's performance compared to standard solutions.
- Effect Size: No effect size or quantification of how much human readers improve with AI vs. without AI assistance is provided.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Yes, a standalone performance evaluation was done. The "Quantitative Accuracy Verification of MIM - SPECTRA Quant" table directly measures the "Quantitative Error" of the software itself (the algorithm) against predefined acceptance criteria using phantoms. This is a direct measure of the algorithm's standalone performance.
7. The Type of Ground Truth Used
The ground truth for the quantitative accuracy assessment (the core of the reported performance data) was established based on the known properties of the phantoms. Phantoms are designed to have known concentrations of radioactive isotopes, which serve as the ground truth for measurement accuracy. For "clinical data," the ground truth of "acceptability for clinical use" was established by "qualified Nuclear Medicine professionals" (expert review).
8. The Sample Size for the Training Set
The document does not provide any information about the sample size used for the training set. This is a key omission for understanding the AI model's development.
9. How the Ground Truth for the Training Set was Established
The document does not provide any information on how ground truth was established for a training set. Given the lack of mention of a training set, it's possible the quantitative accuracy features (decay, attenuation, scatter, resolution corrections) are based on physical models and algorithms rather than a trained machine learning model, or if a model was used, its training details are not disclosed.
{0}------------------------------------------------
June 22, 2018
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The Department of Health & Human Services logo features an abstract image of a human figure. The FDA acronym is in a blue square, and the full name "U.S. Food & Drug Administration" is in blue text.
MIM Software Inc. % Ms. Lynn Hanigan Quality Assurance Director 25800 Science Park Drive - Suite 180 CLEVELAND OH 44122
Re: K180815
Trade/Device Name: MIM - SPECTRA Quant Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: June 7, 2018 Received: June 8, 2018
Dear Ms. Hanigan:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and
Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
Device Name
MIM - SPECTRA Quant
Indications for Use (Describe)
MIM software is used by trained medical professionals as a tool to aid in evaluation and information management of digital medical images. The medical image modalities include, but are not limited to, CT. MRI. CR. DX. MG. US. SPECT. PET and XA as supported by ACRNEMA DICOM 3.0. MIM assists in the following indications:
- · Receive, transmit, store, retrieve, display, print, and process medical images and DICOM objects.
- · Create, display and print reports from medical images.
- · Registration, fusion display, and review of medical images for diagnosis, treatment planning,
· Evaluation of cardiac left ventricular function, including left ventricular end-diastolic volume, end-systolic volume, and ejection fraction.
· Localization and definition of objects such as tumors and normal tissues in medical images.
· Creation, transformation, and modification of contours for applications including, but not limited to, quantitative analysis, aiding adaptive therapy, transfering contours to radiation therapy treatment planning systems, and archiving contours for patient follow-up and management.
· Quantitative and statistical analysis of PET/SPECT brain scans by comparing to other registered PET/SPECT brain scans.
· Planning and evaluation of permanent implant brachytherapy procedures (not including radioactive microspheres).
- · Post-treatment dose calculation of permanent Yttrium-90 (Y90) microsphere implants.
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Images that are printed to film must be printed using a FDA-approved printer for the diagnosis of digital mammography images. Mammographic images must be viewed on a display system that has been cleared by the FDA for the diagnosis of digital mammography images.
The software is not to be used for mammography CAD.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for MIM Software. The logo consists of two overlapping rounded squares, one gray and one red, with a white circle cut out of the red square. To the right of the squares is the text "mim" in black, with the word "SOFTWARE" in smaller black letters below it. The logo is simple and modern, and the colors are eye-catching.
510(k) Summary of Safety and Effectiveness (The following information is in conformance with 21 CFR 807.92)
Submitter:
MIM Software Inc. 25800 Science Park Drive - Suite 180 Cleveland, OH 44122
| Phone:Fax: | 216-455-0600216-455-0601 |
|---|---|
| Contact Person: | Lynn Hanigan |
| Date Summary Prepared: | 6/7/2018 |
Device Name
| Trade Name: | MIM – SPECTRA Quant |
|---|---|
| Common Name: | Medical Imaging Software |
| Regulation Number / Product Code: | 21 CFR 892.2050 Product Code LLZ |
| Classification Name: | System, Imaging Processing, Radiological |
Predicate Devices
| K172218 | MIM – Y90 Dosimetry | MIM Software Inc. |
|---|---|---|
| K023190 | E.CAM Computer / e.soft Workstation | Siemens |
| K162483 | Symbia Intevo Bold (xSPECT Quantification) | Siemens |
Intended Use
MIM software is intended for trained medical professionals including, but not limited to, radiologists, oncologists, physicians, medical technologists, dosimetrists and physicists.
MIM is a medical image and information management system that is intended to receive, transmit, store, retrieve, display, print and process digital medical images, as well as create, display and print reports from those images. The medical modalities of these medical imaging systems include, but are not limited to, CT, MRI, CR, DX, MG, US, SPECT, PET and XA as supported by ACR/NEMA DICOM 3.0.
MIM provides the user with the means to display, register and fuse medical images from multiple modalities. Additionally, it evaluates cardiac left ventricular function and perfusion, including left ventricular end-diastolic volume, end-systolic volume, and ejection fraction.
The Region of Interest (ROI) feature reduces the time necessary for the user to define objects in medical image volumes by providing an initial definition of object contours. The objects include, but are not limited to, tumors and normal tissues.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for MIM Software. The logo consists of two overlapping rounded squares, one gray and one red, with a white circle cut out of the red square. To the right of the squares is the text "mim" in a bold, sans-serif font, with the word "SOFTWARE" underneath in a smaller font.
MIM provides tools to quickly create, transform, and modify contours for applications including, but not limited to, quantitative analysis, aiding adaptive therapy, transferring contours to radiation therapy treatment planning systems and archiving contours for patient follow-up and management.
MIM aids in the assessment of PET/SPECT brain scans. It provides automated quantitative and statistical analysis by automatically registering PET/SPECT brain scans to a standard template and comparing intensity values to a reference database or to other PET/SPECT scans on a voxel by voxel basis, within stereotactic surface projections or standardized regions of interest.
MIM allows the dose distribution of an implant to be individually shaped for each patient and is a general purpose brachytherapy planning system used for prospective and confirmation dose calculations for patients undergoing a course of brachytherapy using permanent implants of various radioisotopes.
For Yttrium-90 (Y90) microspheres, MIM should only be used for the retrospective determination of dose and should not be used to prospectively calculate dose for the case where there is a need for retreatment using Y90 microspheres.
Indications for Use
MIM software is used by trained medical professionals as a tool to aid in evaluation and information management of digital medical images. The medical image modalities include, but are not limited to, CT, MRI, CR, DX, MG, US, SPECT, PET and XA as supported by ACR/NEMA DICOM 3.0. MIM assists in the following indications:
- Receive, transmit, store, retrieve, display, print, and process medical images and DICOM obiects.
- . Create, display and print reports from medical images.
- Registration, fusion display, and review of medical images for diagnosis, treatment ● evaluation, and treatment planning.
- . Evaluation of cardiac left ventricular function and perfusion, including left ventricular enddiastolic volume, end-systolic volume, and ejection fraction.
- Localization and definition of objects such as tumors and normal tissues in medical images.
- . Creation, transformation, and modification of contours for applications including, but not limited to, quantitative analysis, aiding adaptive therapy, transferring contours to radiation therapy treatment planning systems, and archiving contours for patient follow-up and management.
- Quantitative and statistical analysis of PET/SPECT brain scans by comparing to other registered PET/SPECT brain scans.
- . Planning and evaluation of permanent implant brachytherapy procedures (not including radioactive microspheres).
- Post-treatment dose calculation of permanent Yttrium-90 (Y90) microsphere implants.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for MIM Software. The logo consists of a red square with rounded corners and a white circle in the lower left corner, partially covered by a gray square with rounded corners. To the right of the graphic is the text "mim" in a bold, sans-serif font, with the word "SOFTWARE" in a smaller font below.
Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Images that are printed to film must be printed using a FDA-approved printer for the diagnosis of digital mammography images. Mammographic images must be viewed on a display system that has been cleared by the FDA for the diagnosis of digital mammography images. The software is not to be used for mammography CAD.
Device Description
MIM - SPECTRA Quant is a feature of MIM. It is designed for use in medical imaging and operates on both Windows and Mac computer systems. MIM - SPECTRA Quant extends the functionality MIM - Y90 Dosimetry (K172218) software. The following functions have been added to allow quantitative reconstructions of Single Photon Emission Computed Tomography (SPECT) scans:
- Allows for decay correction of activity back to time of injection ●
- Allows for correction of photon attenuation ●
- . Allows for correction of photon scatter
- Allows for correction of collimator detector resolution effects ●
- Allows for camera sensitivity calibration providing output in Becquerels per ml (Bq/ml) .
Substantial Equivalence
MIM - SPECTRA Quant is substantially equivalent to a combination of the predicate devices MIM -Y90 Dosimetry (K172218), Siemens - E.CAM Computer / e.soft Workstation (K023190), and Siemens - Symbia Intevo Bold (xSPECT Quantification) (K162483).
Performance Data
Validation testing with MIM - SPECTRA Quant was performed by board certified clinicians. In all cases, the software passed its performance requirements and met specifications. Additionally, clinicians using MIM – SPECTRA Quant performed validation in comparison to independently implemented, industry-standard solutions. It was demonstrated that MIM software, when used according to operating instructions, can be used safely and effectively.
Furthermore, MIM Software Inc. has conducted performance and integration testing on MIM – SPECTRA Quant software. Standard quality control phantoms, a Liver phantom, and simulated phantoms based on the NEMA IEC Body Phantom were used for verification testing. All tests were performed using standard clinical acquisition and reconstruction protocols.
The accuracy of corrections for attenuation, scatter, and resolution effects were verified in physical and simulated phantoms as well as with clinical data. In all cases, the software passed its performance requirements and met specifications. All reconstructions of clinical data were rated as acceptable for clinical use by the qualified Nuclear Medicine professionals.
The absolute quantitative accuracy of MIM – SPECTRA Quant was verified in physical and simulated phantoms for various isotope-collimator combinations. The acceptance criterion for MIM – SPECTRA Quant is a quantitative error of smaller or equal to 10% in regions that are large enough not to be affected by the partial volume effect. The error in reconstructed activity concentration for these regions was less than 5% which was better than the predefined requirements. In small targets the
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for MIM Software. The logo consists of two overlapping rounded squares, one gray and one red, with a white circle cut out of the red square. To the right of the squares is the text "mim" in a stylized black font, with the word "SOFTWARE" in a smaller font below it.
partial volume effect lowers the accuracy as expected and was similar to the accuracy of the predicate device for these regions.
| Isotope | Collimator | Acceptance Criteria* | Quantitative Error | Verification Result |
|---|---|---|---|---|
| Tc99m | LEHR | <10% | <5% | Passed |
| I-123 | LEHR | <10% | <5% | Passed |
| I-131 | HE | <10% | <5% | Passed |
| Lu-177 | ME | <10% | <5% | Passed |
Quantitative Accuracy Verification of MIM - SPECTRA Quant
Acceptance criteria was <10% error in regions that are large enough not to be affected by partial volume effect.
The testing for half-time or half-dose imaging demonstrated that OSEM with corrections for attenuation, scatter, and resolution met the requirements of similar or superior image quality metrics compared to a full-time or full-dose FBP reconstruction.
In conclusion, MIM Software Inc. has conducted performance testing on the MIM – SPECTRA Quant software. In all cases, the software passed its performance requirements and met specifications.
§ 892.2050 Medical image management and processing system.
(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).