K Number
K180709
Manufacturer
Date Cleared
2018-06-06

(79 days)

Product Code
Regulation Number
878.4400
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The truSculpt RF energy is intended to provide topical heating tissue temperature for the treatment of selected medical conditions, such as relief of pain and increase in local circulation.

Additionally, the 2 MHz setting for the 40 cm handpiece is indicated for reduction in circumference of the abdomen and non-invasive lipolysis (breakdown of fat) of the abdomen.

The truSculpt massage device is intended to provide a temporary reduction in the appearance of cellulite.

Device Description

The truSculpt RF Device consists of a console, one or more RF handpieces that connect to the console with an umbilical cable, and a truGlide massage roller. All system functions are controlled through the console. The handpieces deliver RF energy to generate a heating profile that produces a moderate temperature rise in the subcutaneous tissue, while monitoring epidermal temperature. The truGlide is a separate mechanical roller that can be used as a massager.

AI/ML Overview

Here's a summary of the acceptance criteria and study information for the truSculpt RF Device, based on the provided text:

Device: truSculpt RF Device

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly defined by the clinical study endpoints and the "Summary Results" in the study tables.

Acceptance Criteria (Implicit from Study Endpoints)Reported Device Performance
Non-invasive Lipolysis (Study C-17-TS13)
Efficacy: Histological evaluation of tissue for selective fat necrosis, with sparing of the dermis and epidermis, following truSculpt treatment vs. untreated contralateral control.Met endpoints. Epidermal and dermal tissues were unaffected by RF treatment and were equivalent and indistinguishable in all subject biopsies. Biopsies from the control areas showed normal subcutaneous adipose tissue structures with no indication of changes from the contralateral RF treatment area. Biopsies from treated areas showed adipocyte necrosis and/or inflammatory immune cell response. Acute inflammation was present immediately and out to 20 days. Peak adipocyte death and fat necrosis observed 30 days post-treatment.
Safety: Incidence and severity of adverse device effects during the study period, including subject pain level during RF treatment.No adverse events were observed. Immediate responses included mild to moderate erythema and edema, which resolved with no intervention.
Circumferential Reduction (Study C-17-TS15, follow-up to C-16-TS11)
Efficacy: Difference in circumferential measurement at the 18-month follow-up visit when compared with the baseline and 12-week follow-up visits, as measured and reported in C-16-TS11 (K162512).The difference between the circumferential measurements recorded during the 18-month follow-up visit when compared with the subjects' baseline circumferential measurements showed a sustained reduction in the circumference of the abdomen of 2.30 ± 0.51 cm (p < 0.001). The 18-month circumferential-measurement data when compared with the 12-week circumferential-measurement data showed a 0.17 ± 0.49 cm increase in circumference of the abdomen for this cohort of 15 subjects (p = 0.36), suggesting long-term maintenance of reduction.
Safety: Incidence and severity of treatment-related adverse effects developing after the final 12-week follow-up visit of C-16-TS11. (At which point all treatment-related expected side effects had resolved without medical intervention).No treatment-related adverse effects were observed or reported developing after the 12-week follow-up visit of C-16-TS11.

2. Sample Size Used for the Test Set and Data Provenance

  • Non-invasive Lipolysis Study (C-17-TS13):

    • Sample Size: 12 subjects (92% females, 8% males; aged 24-58 years).
    • Data Provenance: Single-center, prospective, non-randomized, open-label study conducted in an unspecified country (likely USA, given FDA submission). The study involved subjects scheduled for abdominoplasty surgery, and biopsies were harvested during these surgeries.
  • Long-term Follow-up Circumferential Reduction Study (C-17-TS15):

    • Sample Size: 15 subjects were enrolled for the 18-month follow-up (out of 46 potential subjects from the previous study C-16-TS11).
    • Data Provenance: Single-center, observational, prospective study. Follow-up data collected 18 months after initial treatment in a previous study (C-16-TS11). Country of origin is unspecified, but likely USA.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Non-invasive Lipolysis Study (C-17-TS13):

    • Number of Experts: One
    • Qualifications: "board-certified dermatopathologist"
  • Circumferential Reduction Study (C-17-TS15): The ground truth was based on physical measurements (circumferential measurements) taken by study personnel, not interpreted by experts in the same way as histological slides.

4. Adjudication Method for the Test Set

  • Non-invasive Lipolysis Study (C-17-TS13): Not explicitly stated. Given that a single "board-certified dermatopathologist" performed the examination, it implies direct assessment rather than a formal adjudication process between multiple readers. However, the study design included treated vs. contralateral (untreated) control areas on the same subject, effectively allowing for internal comparison.
  • Circumferential Reduction Study (C-17-TS15): Not applicable for measurements. The initial C-16-TS11 study (data previously submitted) included comparative outcomes from treatment and sham treatment arms.

5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

  • No MRMC comparative effectiveness study was done to evaluate human readers' improvement with AI vs. without AI assistance. The truSculpt RF device is not an AI-assisted diagnostic device, but rather a therapeutic device.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • This question is not applicable as the truSculpt RF is a physical device used by a human practitioner for treatment, not an AI algorithm. The performance studies evaluate the device's effect on patients.

7. The Type of Ground Truth Used

  • Non-invasive Lipolysis Study (C-17-TS13):
    • Ground Truth: Histological evaluation (biopsies examined by a board-certified dermatopathologist) for evidence of adipocyte necrosis and inflammatory response. This is a form of pathology ground truth.
  • Circumferential Reduction Study (C-17-TS15):
    • Ground Truth: Quantitative physical measurements (circumferential measurements of the abdomen). This is a form of outcomes data.

8. The Sample Size for the Training Set

  • The provided text describes clinical studies for validation, not the development of an AI algorithm with a distinct training set. Therefore, this information is not applicable in the context of this device and report.

9. How the Ground Truth for the Training Set Was Established

  • As this is not an AI/algorithm-based device with a separate training set, this information is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 6, 2018

Cutera, Inc. Raymond Lee Vice President, Regulatory Affairs and Quality Assurance 3240 Bayshore Blvd. Brisbane, California 94005

Re: K180709

Trade/Device Name: truSculpt RF Device; truSculpt; truSculpt 3D Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: PBX, GEI Dated: March 16, 2018 Received: March 19, 2018

Dear Raymond Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely, Jennifer R. Stevenson -S For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices

Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180709

Device Name(s) truSculpt RF Device; truSculpt; truSculpt 3D

Indications for Use (Describe)

The truSculpt RF energy is intended to provide topical heating tissue temperature for the treatment of selected medical conditions, such as relief of pain and increase in local circulation.

Additionally, the 2 MHz setting for the 40 cm handpiece is indicated for reduction in circumference of the abdomen and non-invasive lipolysis (breakdown of fat) of the abdomen.

The truSculpt massage device is intended to provide a temporary reduction in the appearance of cellulite.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments reqarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section 5 510(k) Summary

This 510(k) Summary of safety and effectiveness for the truSculpt RF Device is submitted in accordance with the requirements of the SMDA 1990 and following guidance concerning the organization and content of a 510(k) summary.

Applicant:Cutera, Inc.
Address:3240 Bayshore Blvd.Brisbane, CA 94005
Contact Person:Raymond Lee
Telephone:415-657-5539
Fax:415-715-3539
Email:rlee@cutera.com
Preparation Date:March 16, 2018
Device TradeNames:truSculpt RF Device;truSculpt; truSculpt 3D
Common Name:Massager, Vacuum, Radio Frequency Induced Heat
ClassificationName:Electrosurgical cutting and coagulation device and accessoriesPBX, GEI, 21 CFR 878.4400
Legally MarketedPredicate Devices:truSculpt RF Device (K162512)Syneron SlimShape System (K163415)
Intended Use:The truSculpt RF Device is intended to generate heat within body tissues forthe treatment of selected medical conditions, such as the relief of minoraches, pain, and muscle spasms; an increase in local circulation; a reductionin circumference of the abdomen; and non-invasive lipolysis (breakdown offat) of the abdomen. The truGlide roller is intended to provide temporaryreduction in the appearance of cellulite.
Indications for Use:The truSculpt RF energy is intended to provide topical heating for the purposeof elevating tissue temperature for the treatment of selected medicalconditions, such as relief of pain and muscle spasms and increase in localcirculation.Additionally, the 2 MHz setting for the 40 cm² handpiece is indicated forreduction in circumference of the abdomen and non-invasive lipolysis(breakdown of fat) of the abdomen.The truSculpt massage device is intended to provide a temporary reductionin the appearance of cellulite.
Device Description:The truSculpt RF Device consists of a console, one or more RF handpiecesthat connect to the console with an umbilical cable, and a truGlide massageroller. All system functions are controlled through the console. Thehandpieces deliver RF energy to generate a heating profile that produces amoderate temperature rise in the subcutaneous tissue, while monitoringepidermal temperature. The truGlide is a separate mechanical roller that canbe used as a massager.

{4}------------------------------------------------

Summary of The truSculpt RF Device that is the subject of this submission has identical Technological technological characteristics as the previously cleared truSculpt RF Device Characteristics: and very similar technological characteristics as the Syneron SlimShape System. All three devices are comprised of a console and RF applicator(s). The consoles for all devices consist of a mechanical enclosure, an RF generator, control electronics, a touch-screen user interface, and a control microprocessor. Although the truSculpt RF Device no longer includes an optional vacuum feature (removed in K133739), this difference does not alter the therapeutic effect for the intended use of the device, nor does it raise new types of safety or effectiveness questions, as demonstrated by the clinical study results. The shape and dimensions of the RF applicators of the truSculpt RF Device that is the subject of this submission are identical to those of the previously cleared truSculpt RF Device. Differences in the shape and dimensions of the RF applicators of the truSculpt RF Device and the Syneron SlimShape System do not raise any new or different safety or effectiveness questions, since the device RF applicators have the same or similar energy parameters, are intended for use in the same or similar treatment areas, and are intended for use on the same or similar population. The truSculpt RF Device that is the subject of this submission heats tissue through delivery of RF energy at 1 MHz and 2 MHz, which is identical to the previously cleared truSculpt RF Device. The Syneron SlimShape System also heats tissue through delivery of RF energy at 1 MHz but does not have a 2 MHz option. All three devices use temperature control mechanisms with redundant temperature sensors in the RF applicators to dynamically adjust the RF power delivered to first reach and then maintain therapeutic temperatures to achieve equivalent therapeutic effects, as demonstrated by the clinical study results. The clinical study results confirm that any technological differences between the truSculpt RF Device that is the subject of this submission and its predicates do not raise new types of safety or effectiveness questions. Performance Data: IEC 60601-1 Medical Electrical Equipment – Part 1: General Requirements for Safety (Edition 3.0, 2005 + AMD1:2012) including: IEC 60601-2-2 Medical Electrical Equipment – Part 2-2: Particular ● Requirements for the Basic Safety and Essential Performance of High Frequency Surgical Equipment and High Frequency Surgical Accessories - Test Report Attachment 5 IEC 60601-1-2 Medical Electrical Equipment – Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Disturbances (Edition 3, 2007)

Software Verification and Validation Testing Report: truSculpt 3D

V0005 rT-Addendum, truSculpt SW Release

This document describes the specific tests that were used to verify and validate the software used in the truSculpt 3D. Testing included verification of correct Startup/Flash screens. Operating screens and

{5}------------------------------------------------

screen selection; Software Version confirmation; Checksum verification;clearing of log files; frequency verification for all handpieces andoperating modes; correct function of displays and indicators; correctfunction of controls, buttons, and inputs; temperature limits; treatmentcountdown indicator; calibrated delivery; screen data accuracy; dateand time entry; data pop-up screens; error conditions; induced faultconditions; shot and error log capture and printing; and anomalies, ifany, encountered during testing. No anomalies were encounteredduring testing.
Results of ClinicalStudies:Two prospective IRB-approved clinical studies were conducted with thetruSculpt RF Device on the intended population to gather clinical data neededto support the revised indications for use.

Non-invasive Lipolysis Study, C-17-TS13

C-17-TS13 was a single-center, prospective, non-randomized, open-label study conducted to evaluate the safety and efficacy of treatment with the Cutera truSculpt RF Device for non-invasive lipolysis in the abdominal region.

Twelve subjects, including Caucasian, Asian and African females (92%) and males (8%), aged 24-58 years (median age: 42), with Fitzpatrick skin types II-VI, who were scheduled to undergo an abdominoplasty surgery were enrolled and received one treatment with the truSculpt RF Device on one abdominal subarea, with the contralateral side abdominal subarea being left untreated to serve as a control. Subjects were assigned to 6 subgroups based upon the timing of their abdominoplasty surgery with respect to the RF treatment date. Up to 2 subjects were enrolled into each subgroup: D0, D10, D20, D30, D60 and D90, where the number refers to the days between RF treatment and the scheduled abdominoplasty surgery date. Biopsies were harvested from the treatment and control abdominal subareas during the abdominoplasty surgery. These biopsies were cultured, stained and examined by a board-certified dermatopathologist leading to the findings summarized below.

Epidermal and dermal tissues were unaffected by RF treatment and were equivalent and indistinguishable in all subject biopsies.

Biopsies from the control areas showed normal subcutaneous adipose tissue structures in all biopsies with no indication of changes from the contralateral RF treatment area.

In contrast, biopsies from the truSculpt RF treatment abdominal areas showed adipocyte necrosis and/or inflammatory immune cell response to the subcutaneous fat. Necrosis was observed immediately and out to sixty days following treatment for six subjects. Necrosis was not present for five subjects, but four of the five subjects not showing necrosis did present inflammatory immune cell response spanning zero to ninety days following treatment. One subject at ten days post treatment did not show any response to treatment.

Acute inflammation was present immediately and out to 20 days following treatment. Peak adipocyte death and fat necrosis was observed 30 days post treatment with approximately 33% of adipocytes affected from just beneath

{6}------------------------------------------------

the dermis up to approximately 1.5 cm beneath the skin surface with no damage seen at the base of the tissue sample (3 cm). Sixty days following treatment, fat necrosis is still present although confined to small, focal areas with less than 20% of the subcutaneous affected. At the final ninety day time point, there was minimal inflammation comprised of few lymphocytes and neutrophils.

All subjects experienced transient mild to moderate erythema and edema post treatment. For eleven subjects, the mild to moderate erythema and edema resolved in 1-4 hours. One subject reported moderate erythema. mild edema and mild post-treatment pain, which persisted for 1 day.

The study results demonstrate that RF treatment with the truSculpt RF Device consistently leads to discrete areas of fat necrosis within the subcutaneous tissue. The effect is present solely in the subcutaneous tissue, while the dermal and epidermal layers remain unaffected by treatment.

Study DesignA single-center, prospective, non-randomized, open-label study
Sample Size12 subjects enrolled
Main Criteria for InclusionMale or Female, 24 to 58 years of age (inclusive); Fitzpatrick Skin Type I - VI; who are scheduled for abdominoplasty and willing to provide histology samples during the surgery from the intended to be harvested areas; has visible fat bulges on the abdomen and palpable fat pockets superior to iliac crest located bilaterally in the lower back flank region; has a Body Mass Index (BMI) $\geq$ 20 and $\leq$ 30
Follow-up Intervals1 follow-up on the day of planned abdominoplasty

The study design and results are summarized in the table below.

{7}------------------------------------------------

EndpointsEfficacy
Histological evaluation of tissue for selective fat necrosis, with sparing of the dermis and epidermis, following truSculpt treatment vs. untreated contralateral control
Safety
Incidence and severity of adverse device effects during the study period, including subject pain level during RF treatment
Summary ResultsEfficacy
Met endpoints. Epidermal and dermal tissues were unaffected by RF treatment and were equivalent and indistinguishable in all subject biopsies.Biopsies from the control areas showed normal subcutaneous adipose tissue structures in all biopsies with no indication of changes from the contralateral RF treatment area.
Safety
No adverse events were observed. The immediate responses included mild to moderate erythema and edema, which resolved with no intervention.

Long-term Follow-up Circumferential Reduction Study, C-17-TS15

C-17-TS15 was a single-center, observational, prospective study of subjects, aged 24-60 years at the time of RF treatment, who completed participation in the Treatment Group of Protocol C-16-TS11 "Pivotal Study of the truSculpt Radiofrequency Device for Circumferential Reduction" (46 subjects in total; and the maximum potential enrollment for this study) for sustained long-term circumferential reduction.

Data from protocol C-16-TS11 comparing subject outcomes from the treatment and sham treatment study arms were previously submitted under K162512, resulting in FDA clearance of a new indication for use for the truSculpt RF Device: "temporary reduction in circumference of the abdomen".

{8}------------------------------------------------

All potential subjects were contacted in no fewer than five attempts by phone calls, emails, and/or texts and were invited to participate in a new study involving one study visit: 18 months ± 2 months following their original treatment under study Protocol C-16-TS11. Of the forty-six potential subjects: two were not able to be contacted; four were not interested in participating; six were unable to participate because of relocation, travel or work constraints; nine were excluded as they had elected to have additional truSculpt RF treatments of the abdomen and flanks for further circumferential reduction after being released from study C-16-TS11 restrictions; three were excluded for various exclusion criteria (pregnancy, medical condition, and abdominal surgerv. respectively): and the remaining twenty-two subiects were scheduled for screening visits. Of the twenty-two subjects screened, seven had weight changes beyond ± 5% of baseline weight and were therefore deemed to have not met an inclusion criterion and excused prior to enrollment. The remaining fifteen subjects (32.6% of potential subjects) were enrolled.

The circumferential measurements recorded during the 18-month follow-up visit for the 15 subjects enrolled in C-17-TS15, when compared with the subjects' baseline circumferential measurements recorded and reported in C-16-TS11, demonstrated a statistical-significant, sustained reduction in the circumference of the abdomen of 2.30 ± 0.51 cm (SEM; p < 0.001).

No treatment-related adverse effects were observed or reported developing after the 12-week follow-up visit of C-16-TS11, at which point all treatmentrelated expected side effects had resolved without medical intervention.

Study Designa single-center, observational, prospective study
Sample Size46 subjects in total; 15 subjects enrolled for long-term circumferential reduction
Main Criteria for InclusionCompleted participation in the Treatment Group of Protocol C-16-TS11; and maintained the same weight within ± 5% of the baseline
Follow-up Intervals18 months post treatment in C-16-TS11
EndpointsEfficacyDifference in circumferential measurement at the 18-month follow-up visit when compared with the baseline and 12-week follow-up visits, as measured and reported in C-16-TS11 (K162512).
Safety
Incidence and severity of treatment-related adverse effects developingafter the final 12-week follow-up visitof C-16-TS11.
Summary ResultsEfficacy
The difference between thecircumferential measurementsrecorded during the 18-monthfollow-up visit when compared withthe subjects' baselinecircumferential measurementsshowed sustained reduction in thecircumference of the abdomen of$2.30 \pm 0.51$ cm.
The 18-month circumferential-measurement data when comparedwith the 12-week circumferential-measurement data showed a $0.17 \pm$0.49 cm increase in circumferenceof the abdomen for this cohort of 15subjects (p = 0.36).
Safety
No treatment-related adverseeffects were observed or reporteddeveloping after the 12-week follow-up visit of C-16-TS11, at which pointin time all treatment-relatedexpected side effects had resolvedwithout medical intervention.
Conclusion:The truSculpt RF Device that is the subject of this submission has identicaltechnological characteristics and principles of operation as the previouslycleared truSculpt RF Device and very similar technological characteristicsand principles of operation as the Syneron SlimShape System.
Performance data and clinical study results demonstrate that anydifferences between the truSculpt RF Device and its predicate devices donot raise new types of safety or effectiveness questions.
Clinical studies conducted with the truSculpt RF Device that is the subject ofthis submission have demonstrated the safety and effectiveness profile ofthe device in the intended population for the requested indications for useheld by the Syneron SlimShape System.
The truSculpt RF Device is substantially equivalent to the predicate devices.

The study design and results are summarized in the table below.

{9}------------------------------------------------

{10}------------------------------------------------

Section 5 510(k) Summary

K180709

Table 5A—Technical Specification Comparison

Cutera truSculpt RF Device(current submission)Cutera truSculpt RF Device(K162512)Syneron SlimShape System(K163415)
Energy TypeRadiofrequencyRadiofrequencyRadiofrequency + Infrared
MassageYesYesYes
Vacuum (suction)NoNoYes
Temperature sensingYesYesYes
Temperature sensing activecontrolYesYesYes
Treatment activationFingerswitchFingerswitchControl Screen Button
Area treated16 cm² and 40 cm²16 cm² and 40 cm²Unknown
Electrode shapeSquare or RectangleSquare or RectangleRectangle
RF frequency1 MHz and 2 MHz1 MHz and 2 MHz1 MHz
RF typeBipolar / MonopolarBipolar / MonopolarBipolar
Max RF power300 W300 WUnknown
Patient contact materialPolyethylene(3M Tegaderm) and 316 SSPolyethylene(3M Tegaderm) and 316 SSUnknown

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.