K Number
K180684
Manufacturer
Date Cleared
2018-04-19

(35 days)

Product Code
Regulation Number
864.7750
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CoaguChek XS Plus system is intended for use by professional healthcare providers for quantitative prothrombin time testing for the monitoring of warfarin therapy. The system uses fresh capillary or non-anticoagulated venous whole blood.

Device Description

No technological, material, performance, or design changes to the CoaguChek XS Plus System have been implemented since its clearance on K071041. Thus, the device description for the CoaguChek XS Plus System remains the same as that presented and cleared in K071041. This submission deals only with the performance of Super Sani-Cloth wipes for the effective cleaning and disinfection of the CoaguChek XS Plus System housing and components.

AI/ML Overview

The provided document describes a Special 510(k) submission for extending cleaning and disinfection claims for the CoaguChek XS Plus System to include "Super Sani-Cloth" wipes.

Here's an analysis of the acceptance criteria and study for the device based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance CriteriaReported Device Performance
Super Sani-Cloth wipes are effective for cleaning and disinfection of the device.The corresponding test results on the device components demonstrate that the Super Sani-Cloth wipes are effective for the cleaning and disinfection of the device.
The device must meet performance requirements for its intended use after being subjected to cleaning and disinfection with Super Sani-Cloth.Robustness and effectiveness cleaning and disinfecting testing on the CoaguChek XS Plus System for Super Sani-Cloths demonstrated that the device meets the performance requirements for its intended use. Overall, the data demonstrate that the CoaguChek XS Plus System operates in the same manner when cleaning and disinfecting occurs with the Super Sani-Cloth wipes.
Neither meter accuracy nor meter functionality should be impacted by the cleaning and disinfection procedure over the device's lifetime. (Implicit: the device should continue to accurately measure prothrombin time and function correctly.)Acceptance criteria were met when testing with venous blood the CoaguChek XS Plus meters that had been subjected to 10,950 Cleaning and Disinfection (C&D) cycles representing three-years lifetime of the meter based on a testing frequency of 10 tests per day. Neither meter accuracy nor meter functionality were impacted by the cleaning and disinfection procedure.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: The document mentions that the CoaguChek XS Plus meters were subjected to 10,950 Cleaning and Disinfection (C&D) cycles. This number represents the simulated lifetime (three years at 10 tests/day) of the cleaning/disinfection robustness. The number of actual meters tested (i.e., the sample size of the devices themselves) is not explicitly stated. However, it implies testing was performed on at least one or more devices to generate this data.
  • Data Provenance: Not specified, but the context is an FDA submission from Roche Diagnostics, implying internal testing conducted by the manufacturer. The document does not mention the country of origin of the data or whether it was retrospective or prospective. It is implied to be prospective testing carried out for this submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:

  • Not Applicable. This study is focused on the robustness of the device itself to a cleaning and disinfection process, not on diagnostic performance where expert interpretation of results would be required to establish ground truth. The 'ground truth' here is the continued accurate function and performance of the instrument.

4. Adjudication Method for the Test Set:

  • None. Adjudication methods like 2+1 or 3+1 are used in studies involving human interpretation or subjective assessments. This study evaluates the objective performance (accuracy and functionality) of a medical device after a specific treatment (cleaning/disinfection).

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not Applicable. This is not an AI-assisted diagnostic device, nor is it a study involving human readers.

6. If a Standalone (i.e. algorithm only, without human-in-the-loop performance) was done:

  • Yes, in the sense that the device's performance was evaluated standalone. The study assessed the inherent accuracy and functionality of the CoaguChek XS Plus System after cleaning with Super Sani-Cloth wipes. There's no human-in-the-loop variable being tested here; it's about the device's physical and operational integrity.

7. The Type of Ground Truth Used:

  • The ground truth used is the inherent accurate measurement capability and continued functionality of the CoaguChek XS Plus System before and after repeated cleaning and disinfection cycles. This is established by comparing the device's performance (specifically, meter accuracy) using venous blood measurements against established standards or baseline performance, and verifying that its functionality (e.g., ability to power on, display results, etc.) remains intact.

8. The Sample Size for the Training Set:

  • Not Applicable. There is no "training set" for this type of robustness study. The study involves demonstrating that the already cleared device, K071041, maintains its performance after being subjected to a specific cleaning agent. It's not a machine learning or AI algorithm that requires training data.

9. How the Ground Truth for the Training Set was Established:

  • Not Applicable. As there is no training set, there is no ground truth established for one.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 19, 2018

Roche Diagnostics Bin Sun Regulatory Affairs Principal 9115 Hague Road Indianapolis, Indiana 46250

Re: K180684

Trade/Device Name: CoaguChek XS Plus System Regulation Number: 21 CFR 864.7750 Regulation Name: Prothrombin time test Regulatory Class: Class II Product Code: GJS Dated: March 14, 2018 Received: March 15, 2018

Dear Bin Sun:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR

{1}------------------------------------------------

Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Leonthena R. Carrington -S

Lea Carrington Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180684

Device Name CoaguChek® XS Plus System

Indications for Use (Describe)

The CoaguChek XS Plus system is intended for use by professional healthcare providers for quantitative prothrombin time testing for the monitoring of warfarin therapy. The system uses fresh capillary or non-anticoagulated venous whole blood.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

CoaguChek XS Plus Super Sani Cloth claims extension

Special 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

In accordance with 21 CFR 807.87. Roche Diagnostics hereby submits official notification as required by Section 510(k) of the Federal Food, Drug and Cosmetics Act of our intention to market the device described in this Premarket Notification 510(k).

The purpose of this Special 510(k) Premarket Notification is to obtain FDA review and clearance for the CoaguChek XS Plus Super Sani Cloth claims extension. The CoaguChek XS Plus System labeling will be modified to add a new cleaning and disinfecting product for use with our device, Super Sani-Cloth® (EPA #9480-4).

The CoaguChek XS Plus System is a prothrombin time test system, a Class II medical device according to 21 CFR 864.7750. Super Sani-Cloth is a General Purpose Disinfectant, a Class I exempt product, according to 21 CFR 880.6890.

The CoaguChek XS Plus System has not changed since its clearance on K071041. No technological, material, performance, or design changes to the CoaguChek XS Plus System have been implemented since its clearance on K071041. This submission pertains only to the performance of Super Sani- Cloth wipes for the effective cleaning and disinfecting of the CoaguChek XS Plus System housing and components.

Please note that we intend to add the Super Sani-Cloth to the CoaguChek XS Plus System product labeling for cleaning and disinfection purposes. Previously, 70% ethanol or isopropyl alcohol and 10% sodium hypochlorite solution were cleared as recommended cleaning/disinfecting solutions. The Super Sani-Cloth will replace the sodium hypochlorite solutions in the product labeling.

{4}------------------------------------------------

Submitter NameRoche Diagnostics
Address9115 Hague RoadP.O. Box 50416Indianapolis, IN 46250-0457
ContactBin SunPhone: (317) 521-3107FAX: (317) 521-2324Email: bin.sun.bs2@roche.comAngie ClementsPhone: (317) 521-7338FAX: (317) 521-2324Email: angie.clements@roche.com
Date PreparedMarch 9, 2018
Proprietary NameCoaguChek XS Plus System
Common NameTest, Time, Prothrombin
Classification NameProthrombin time test
Product Codes,Regulation NumbersGJS, Class II, 21 CFR 864.7750
Predicate DevicesThe 510(k) history for the CoaguChek XS Plus System is K071041, cleared on May 11, 2007.Please note that in this submission we did not make any changes or enhancements to the CoaguChek XS Plus system.
Establishment RegistrationFor the CoaguChek XS Plus System, the establishment registration number for RocheDiagnostics GmbH in Mannheim, Germany is 9610126.

{5}------------------------------------------------

1. DEVICE DESCRIPTION

No technological, material, performance, or design changes to the CoaguChek XS Plus System have been implemented since its clearance on K071041.

Thus, the device description for the CoaguChek XS Plus System remains the same as that presented and cleared in K071041.

This submission deals only with the performance of Super Sani-Cloth wipes for the effective cleaning and disinfection of the CoaguChek XS Plus System housing and components.

Please note that we intend to modify the CoaguChek XS Plus System labeling by adding the Super Sani-Cloth for cleaning and disinfection of the system. Previously, 70% ethanol or isopropyl alcohol and 10% sodium hypochlorite solution were cleared as recommended cleaning/disinfecting solutions. The Super Sani-Cloth will replace the sodium hypochlorite solutions in the product labeling.

2. INDICATIONS FOR USE

The use of Super-Sani Cloth wipes for the cleaning and disinfection of the CoaguChek XS Plus System is the focus of this Special 510(k) and does not change the Intended Use or Indications for Use of the system.

Therefore, the Intended Use and Indications for Use for the CoaguChek XS Plus System remains the same as the cleared CoaguChek XS Plus System (K071041).

The CoaguChek XS Plus system is intended for use by professional healthcare providers for quantitative prothrombin time testing for the monitoring of warfarin therapy. The system uses fresh capillary or non-anticoagulated venous whole blood.

TECHNOLOGICAL CHARACTERISTICS 3.

The fundamental scientific technology of the CoaguChek XS Plus System (K071041) has not changed.

ACCESSORY NAME (SUPER SANI-CLOTH) 4.

Proprietary name: Super Sani-Cloth (manufactured by PDI)

{6}------------------------------------------------

Classification names: Disinfectant, Medical Devices

Subsequent Product Code: LRJ, General Purpose Disinfectant

510(k) History: n/a, Class I exempt

INSTRUMENT CLEANING AND DISINFECTION ROBUSTNESS STUDY 5.

No technological, material, performance, or design changes to the CoaguChek XS Plus System have been implemented since its clearance on K071041.

The corresponding test results on the device components demonstrate that the Super Sani-Cloth wipes are effective for the cleaning and disinfection of the device.

Robustness and effectiveness cleaning and disinfecting testing on the CoaguChek XS Plus System for Super Sani-Cloths demonstrated that the device meets the performance requirements for its intended use. Acceptance criteria were met when testing with venous blood the CoaguChek XS Plus meters that had been subjected to 10.950 C&D cycles representing three-years lifetime of the meter based on a testing frequency of 10 tests per day. Neither meter accuracy nor meter functionality were impacted by the cleaning and disinfection procedure.

Overall, the data demonstrate that the CoaguChek XS Plus System operates in the same manner when cleaning and disinfecting occurs with the Super Sani-Cloth wipes.

Please note that we intend to modify the CoaguChek XS Plus System labeling the Super Sani-Cloth for cleaning and disinfection of the system. Previously, 70% ethanol or isopropyl alcohol and 10% sodium hypochlorite solution were cleared as recommended cleaning/disinfecting solutions. The Super Sani-Cloth will replace the sodium hypochlorite solutions in the product labeling.

UPDATED LABELING 6.

The CoaguChek XS Plus Operator's Manual has been updated to reflect Super Sani-Cloth. This is included in Section 10. The Maintenance & Care topic starts on page 121.

§ 864.7750 Prothrombin time test.

(a)
Identification. A prothrombin time test is a device used as a general screening procedure for the detection of possible clotting factor deficiencies in the extrinsic coagulation pathway, which involves the reaction between coagulation factors III and VII, and to monitor patients receiving coumarin therapy (the administration of one of the coumarin anticoagulants in the treatment of venous thrombosis or pulmonary embolism).(b)
Classification. Class II (performance standards).