K Number
K180500

Validate with FDA (Live)

Device Name
VariAx 2 System
Manufacturer
Date Cleared
2018-06-04

(98 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VariAx 2 System is indicated for adult and pediatric patients, where the implant would not cross open growth plates, for the treatment of normal or osteopenic bone for the following conditions or procedures:

  • · Fracture fixation, including single, segmental, and comminuted fractures
  • · Revision, including nonunion and malunion
  • · Intra- and extra-articular fractures
  • · Compression fracture
  • · Displaced fracture
  • · Reconstruction
  • · Replantation
  • · Arthrodesis
  • · Osteotomy
Device Description

VariAx 2 is a system used for internal fixation applications and is composed of sterile and nonsterile screws, washers, and instruments. In addition to independent use, screws of this system are also used with compatible, internal fixation systems. These devices are made of titanium alloy, with Type III anodization, and are available in a variety of sizes and diameters, as well as locking and non-locking types.

AI/ML Overview

I am sorry, but based on the provided text, there is no information about acceptance criteria or a study that proves a device meets those criteria, as well as no mention of sample sizes, data provenance, expert panels, adjudication methods, multi-reader multi-case studies, standalone performance, or ground truth establishment.

The document is a 510(k) premarket notification summary for the VariAx 2 System, which is a medical device for internal fixation. It discusses the device's substantial equivalence to predicate devices based on non-clinical mechanical testing.

Therefore, I cannot provide the requested information.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Stryker GmbH Paul Nelson Senior Regulatory Affairs Specialist 325 Corporate Drive Mahwah, New Jersey 07430

Re: K180500

Trade/Device Name: VariAx 2 System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: April 26, 2018 Received: April 27, 2018

Dear Paul Nelson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

June 4, 2018

{1}------------------------------------------------

K180500

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Katherine D. Kavlock -S

for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K180500

Device Name VariAx 2 System

Indications for Use (Describe)

The VariAx 2 System is indicated for adult and pediatric patients, where the implant would not cross open growth plates, for the treatment of normal or osteopenic bone for the following conditions or procedures:

  • · Fracture fixation, including single, segmental, and comminuted fractures
  • · Revision, including nonunion and malunion
  • · Intra- and extra-articular fractures
  • · Compression fracture
  • · Displaced fracture
  • · Reconstruction
  • · Replantation
  • · Arthrodesis
  • · Osteotomy

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

K180500 Page 1 of 1

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

Submitter:Stryker GmbHBohnackerweg lCH-2545 SelzachSwitzerland
Contact Person:Paul NelsonSenior Regulatory Affairs SpecialistPhone: (201) 831-5691Fax: (201) 831-6500
Date Prepared:February 2, 2018
Name of Device:VariAx 2 System
Common or Usual Name:Bone Screws
Classification Name:Smooth or threaded metallic bone fixation fastener(21 CFR § 888.3040)
Regulatory Class:Class II
Product Codes:HWC
Subject:VariAx 2 System, K180500
Primary Predicate:VariAx 2 System, K173135
Additional Predicate:Profyle System, K062498

Description:

VariAx 2 is a system used for internal fixation applications and is composed of sterile and nonsterile screws, washers, and instruments. In addition to independent use, screws of this system are also used with compatible, internal fixation systems. These devices are made of titanium alloy, with Type III anodization, and are available in a variety of sizes and diameters, as well as locking and non-locking types.

{4}------------------------------------------------

Indications for Use:

The VariAx 2 System is indicated for adult and pediatric patients, where the implant would not cross open growth plates, for the treatment of normal or osteopenic bone for the following conditions or procedures:

  • · Fracture fixation, including single, segmental, and comminuted fractures
  • · Revision, including nonunion and malunion
  • Intra- and extra-articular fractures
  • Compression fracture
  • Displaced fracture
  • Reconstruction
  • Replantation
  • Arthrodesis
  • Osteotomy

Summary of Technologies:

A comparison of the systems demonstrated that the subject VariAx 2 System is substantially equivalent to the previously cleared VariAx 2 system, K173135, and to the Profyle System, K062498, in regards to intended use, material, design, and operating principles.

Performance Data:

Non-Clinical Testing

Comparative mechanical testing to the predicate system demonstrated substantial equivalence.

The following tests were performed:

  • Torque Strength ●
  • . Insertion Torque
  • Pull-Out

Testing demonstrated that the VariAx 2 System is equivalent in mechanical performance to the predicate device of the Profyle System.

LAL testing was performed to establish that the subject devices meet the less than 20 EU/device limit.

Clinical Testing

Clinical testing was not a requirement for this submission.

Conclusion:

The VariAx 2 System described in this submission has the same indications, intended use, target patient population, technological characteristics, and materials as the predicate devices. The mechanical testing demonstrates the performance of the proposed devices is equivalent to the predicate devices.

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.