(106 days)
The indications for use of this internal fixation device include:
- Open and closed tibial fractures
- Pseudoarthrosis and correction osteotomy
- Pathologic fractures, impending pathologic fractures, and tumor resections
- Fractures involving osteopenic and osteoporotic bone
- Nonunion and malunion
The End Cap Lower Extremity and the Nail Holding Screw Tibia / Femur PF may also be used in conjunction with the T2 Alpha Femur Antegrade GT/PF Nailing System.
The indications for use of these internal fixation devices include:
- Fixation of subtrochanteric, ipsilateral neck/shaft, comminuted proximal femoral shaff fractures
- Femoral fixation required as a result of pathological disease
- Temporary stabilization of fractures of the femoral shaft ranging from the femoral neck to the supracondylar regions of the femur
- Open and closed femoral fractures
- Pseudoarthrosis and correction osteotomy
- Pathologic fractures, impending pathologic fractures and tumor resections
- Ipsilateral femur fractures
- Fractures proximal to a total knee arthroplasty
- Nonunions and malunions
- Fractures involving osteopenic and osteoporotic bone
The IMN Screws System is intended to stabilize the intramedullary nail-bone construct for temporary stabilization.
The IMN Instruments System is intended to enable the implantation of intramedullary nail and screw.
The T2 Alpha Tibia Nailing System is a fracture fixation system and includes sterile implants (intramedullary nails in various diameter, compression screw and end caps) as well as non-sterile instruments (targeting devices). The sterile implants are made of titanium alloy (Ti6A14V ELI) as per ASTM F136. The T2 Alpha Tibia Nailing System will be used with the locking screws cleared in K003018 (Titan Tibial Nail) and K172774 (IMN Screws System). Further, End Cap Lower Extremity and the Nail Holding Screw Tibia / Femur PF of T2 Alpha Tibia Nailing System can be used with T2 Alpha Femur Antegrade GT/PF Nailing System (K172774). The T2 Alpha Tibia Nailing System will be used with the existing instruments previously cleared in K131365 (T2 Tibial Nailing System) and 510(k) exempt devices. The T2 Alpha Tibia Nailing System is intended for use with IMN Screws System and IMN Instruments System.
The IMN Screws System includes bone screws (locking screws previously cleared in K172774 and advanced locking screws) that are inserted through the intramedullary nail to stabilize the nail-bone construct. The screws are made of titanium alloy (Ti6A14V ELI) as per ASTM F136. The IMN Screws System is intended for use with T2 Alpha Tibia Nailing System, T2 Alpha Femur Antegrade GT/PF Nailing System and IMN Instruments System.
The IMN Instruments System includes distal targeting device femur antegrade previously cleared in K172774 as well as surgical instruments that support the implantation and extraction of intramedullary nails and screws. The IMN Instruments System is intended for use with T2 Alpha Tibia Nailing System, T2 Alpha Femur Antegrade GT/PF Nailing System and IMN Screws System.
The T2 Alpha Femur Antegrade GT/PF Nailing System previously cleared in K172774 is a fracture fixation system and includes sterile implants (femoral nails in various diameter and sizes, compression screw, set screws and end caps) as well as non-sterile instruments.
The T2 Alpha Femur Antegrade GT/PF Nailing System will be used with the locking screws originally cleared in K003018 (Titan Tibial Nail) that have subsequently also received clearance for use in locking femoral nailing systems (K010801, T2 Femoral Nail System) as well as locking screws previously cleared in K172774 (IMN Screws System) and the T2 Lag Screw Recon previously cleared in K032898 (T2 Recon Nail System). Further, it will be used with the existing instruments previously cleared in K172774 (IMN Instruments System) and 510(k) exempt devices. The T2 Alpha Femur Antegrade GT/PF Nailing System is intended for use with IMN Screws System and IMN Instruments System.
The provided document is a 510(k) Premarket Notification for a medical device, specifically orthopaedic implants and instruments (T2 Alpha Tibia Nailing System, T2 Alpha Femur Antegrade GT/PF Nailing System, IMN Screws System, and IMN Instruments System).
This document does not describe a study involving an AI or software-as-a-medical-device (SaMD) product that requires human-in-the-loop or standalone performance evaluation with clinical data and expert ground truth. Instead, it focuses on demonstrating substantial equivalence of traditional medical devices to previously cleared predicate devices through bench testing and biocompatibility assessments. Therefore, most of the requested information regarding acceptance criteria, sample sizes for test/training sets, expert qualifications, ground truth establishment methods, MRMC studies, and effect sizes of AI assistance are not applicable to this type of submission.
The document describes non-clinical performance data primarily from bench testing to demonstrate the substantial equivalence of the new devices or line extensions to previously cleared devices.
Here's a breakdown of the applicable and non-applicable information based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not provide a quantitative table of acceptance criteria and reported device performance in the format typically seen for AI/SaMD studies. Instead, it states that "Testing demonstrated that T2 Alpha Tibia Nailing System is equivalent in mechanical performance to the predicate devices (K003018, K082770, and K040762)." Similar statements are made for the other systems.
The types of tests performed indicate the implicit acceptance criteria:
- T2 Alpha Tibia Nailing System:
- Testing of mechanical properties per ASTM F1264
- Fatigue strength testing (proximal and distal)
- Cut-out testing
- Targeting accuracy testing
- Biocompatibility testing (ISO 10993-1, -5, -18)
- Bacterial Endotoxin Testing
- IMN Screws System (advanced locking screws):
- Testing of mechanical properties as per ASTM F543
- Biocompatibility assessment (ISO 10993-1)
- Bacterial Endotoxin Testing
- IMN Instruments System:
- Performance testing in cadaver lab
- Compatibility evaluation
- Biocompatibility testing (ISO 10993-1, -5, -18)
- T2 Alpha Femur Antegrade GT/PF Nailing System:
- Compatibility assessment
The reported "performance" is primarily a statement of equivalence or suitability to the predicate devices based on these tests. No specific numerical thresholds or results are presented.
2. Sample sizes used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
This information is not provided in typical AI/SaMD study terms. The "test set" here refers to samples of the physical medical devices used in bench testing. The document does not specify the number of physical samples tested for each mechanical or biocompatibility test. Data provenance in terms of 'country of origin of data' or 'retrospective/prospective' is not relevant as it's not a clinical study on human subjects with data collection over time.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
Not applicable. Ground truth in the context of this submission refers to the established mechanical properties, biocompatibility standards, and functional performance of the devices as measured by standardized tests. It does not involve expert readers reviewing clinical cases.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not applicable. There is no adjudication method described as it is not a clinical study requiring consensus among multiple readers.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This submission is for traditional medical devices (implants and instruments), not an AI/SaMD product. Therefore, no MRMC study or AI assistance evaluation was performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This document describes physical medical devices, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for this submission is established through objective, standardized bench testing for mechanical properties, biocompatibility standards (e.g., ISO standards), and functional performance (e.g., targeting accuracy, cadaver lab performance). It is not derived from expert consensus on medical images, pathology, or clinical outcomes data.
8. The sample size for the training set
Not applicable. This is not an AI/ML product that requires a training set.
9. How the ground truth for the training set was established
Not applicable. This is not an AI/ML product that requires a training set.
In summary, this 510(k) submission for conventional orthopaedic devices focuses on demonstrating substantial equivalence through non-clinical bench testing and adherence to recognized standards, rather than clinical studies or evaluations of AI performance. Therefore, most of the detailed questions about acceptance criteria, sample sizes, expert involvement, and ground truth establishment methods as typically understood for AI/SaMD products are not relevant to this document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 6, 2018
Stryker Trauma GmbH Heike Gustke Staff Regulatory Affairs Specialist Prof-Kuentscher-Str. 1-5 Schoenkirchen, 24232 De
Re: K180436
Trade/Device Name: T2 Alpha Tibia Nailing System. T2 Alpha Femur Antegrade GT/PF Nailing System, IMN Screws System, IMN Instruments System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB, HWC Dated: April 30, 2018 Received: May 4, 2018
Dear Dr. Gustke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Vincent J. Devlin -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | Form Approved: OMB No. 0910-0120 |
|---|---|
| Food and Drug Administration | Expiration Date: 06/30/2020 |
| Indications for Use | See PRA Statement below. |
| 510(k) Number (if known) |
Device Name
T2 Alpha Tibia Nailing System
Indications for Use (Describe)
The indications for use of this internal fixation device include:
-
Open and closed tibial fractures
-
Pseudoarthrosis and correction osteotomy
-
Pathologic fractures, impending pathologic fractures, and tumor resections
-
Fractures involving osteopenic and osteoporotic bone
-
Nonunion and malunion
The End Cap Lower Extremity and the Nail Holding Screw Tibia / Femur PF may also be used in conjunction with the T2 Alpha Femur Antegrade GT/PF Nailing System.
Type of Use (Select one or both, as applicable)
[ Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
{3}------------------------------------------------
Traditional 510(k) Premarket Notification
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
K180436
Device Name
T2 Alpha Femur Antegrade GT/PF Nailing System
Indications for Use (Describe)
The indications for use of these internal fixation devices include:
-
Fixation of subtrochanteric, ipsilateral neck/shaft, comminuted proximal femoral shaff fractures
-
Femoral fixation required as a result of pathological disease
-
Temporary stabilization of fractures of the femoral shaft ranging from the femoral neck to the supracondylar regions of the femur
-
Open and closed femoral fractures
-
Pseudoarthrosis and correction osteotomy
-
Pathologic fractures, impending pathologic fractures and tumor resections
-
Ipsilateral femur fractures
-
Fractures proximal to a total knee arthroplasty
-
Nonunions and malunions
-
Fractures involving osteopenic and osteoporotic bone
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}------------------------------------------------
Traditional 510(k) Premarket Notification
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | Form Approved: OMB No. 0910-0120 |
|---|---|
| Food and Drug Administration | Expiration Date: 06/30/2020 |
| Indications for Use | See PRA Statement below. |
| 510(k) Number (if known) | K180436 |
|---|---|
| Device Name | IMN Screws System |
Indications for Use (Describe)
The IMN Screws System is intended to stabilize the intramedullary nail-bone construct for temporary stabilization.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
{5}------------------------------------------------
Traditional 510(k) Premarket Notification
| DEPARTMENT OF HEALTH AND HUMAN SERVICESFood and Drug AdministrationIndications for Use | Form Approved: OMB No. 0910-0120Expiration Date: 06/30/2020See PRA Statement below. |
|---|---|
| ------------------------------------------------------------------------------------------------ | --------------------------------------------------------------------------------------------- |
| 510(k) Number (if known) | K180436 |
|---|---|
| -------------------------- | --------- |
Device Name
IMN Instruments System
Indications for Use (Describe)
The IMN Instruments System is intended to enable the implantation of intramedullary nail and screw.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{6}------------------------------------------------
Section 9: 510(k) Summary I. SUBMITTER
| I. SUBMITTER | Regulation Number / Name: | T2 Alpha Tibia Nailing System21CFR 888.3020 (Intramedullary fixation rod)T2 Alpha Femur Antegrade GT/PF Nailing System21CFR 888.3020 (Intramedullary fixation rod)IMN Screws System21CFR 888.3040 (Smooth or threaded metallic bone fixationfastener)IMN Instruments System21CFR 888.3020 (Intramedullary fixation rod) | |
|---|---|---|---|
| Sponsor: | Stryker Trauma GmbHProf.-Kuentscher-Str. 1-524232 Schoenkirchen / Germany | Product Code: | T2 Alpha Tibia Nailing SystemHSB (Rod, fixation, intramedullary and accessories)T2 Alpha Femur Antegrade GT/PF Nailing SystemHSB (Rod, fixation, intramedullary and accessories)IMN Screws SystemHWC (Screw, fixation, bone)IMN Instruments SystemHSB (Rod, fixation, intramedullary and accessories) |
| Contact Person: | Dr. Heike GustkeStaff Regulatory Affairs SpecialistProf.-Kuentscher-Str. 1-524232 Schoenkirchen / Germanyheike.gustke@stryker.comPhone: +49 4348 702 637Fax: +49 4348 702 8637 | Regulatory Class: | Class II |
| Date Prepared: | February 5, 2018 | ||
| II. DEVICE | |||
| Name of Device: | T2 Alpha Tibia Nailing SystemT2 Alpha Femur Antegrade GT/PF Nailing SystemIMN Screws SystemIMN Instruments System | ||
| Common Name: | T2 Alpha Tibia Nailing SystemRod, fixation, intramedullary and accessoriesT2 Alpha Femur Antegrade GT/PF Nailing SystemRod, fixation, intramedullary and accessoriesIMN Screws SystemScrew, fixation, boneIMN Instruments SystemRod, fixation, intramedullary and accessories |
{7}------------------------------------------------
III. PREDICATE DEVICE
IMN Screws System
| T2 Alpha Tibia Nailing System | |
|---|---|
| Primary predicate device: | Titan Tibial Nail (K003018) |
| Additional predicate devices: | Zimmer Natural Nail System Tibial Nail (K082770) |
| Synthes (USA) Tibial Nail System EX (K040762) |
| Predicate device: | IMN Screws System (K172774) |
|---|---|
| Reference device: | Titan Tibial Nail (K003018) |
{8}------------------------------------------------
Traditional 510(k) Premarket Notification
IMN Instruments System
| Predicate device: | IMN Instruments System (K172774) |
|---|---|
| ------------------- | ---------------------------------- |
T2 Alpha Femur Antegrade GT/PF Nailing System
Predicate device: T2 Alpha Femur Antegrade GT/PF Nailing System (K172774)
IV. DEVICE DESCRIPTION
A Traditional 510(k) submission is being supplied to the U.S. FDA to gain clearance to market:
- The T2 Alpha Tibia Nailing System, ●
- . Line extension of IMN Screws System (addition of advanced locking screws),
- . Line extension of IMN Instruments System (addition of surgical instruments),
- . Compatibility of T2 Alpha Femur Antegrade GT/PF Nailing System with IMN Screws System as well as IMN Instruments System, and devices of T2 Alpha Tibia Nailing System
The intended use and indications for use of existing T2 Alpha Femur Antegrade GT/PF Nailing System, IMN Screws System and IMN Instruments System previously cleared in K172774 remain unchanged.
This submission encompasses multiple systems (T2 Alpha Tibia Nailing System, T2 Alpha Femur Antegrade GT/PF Nailing System, IMN Screws System and IMN Instruments System) that have similar intended use and will be used together during the surgical procedure.
T2 Alpha Tibia Nailing System
The T2 Alpha Tibia Nailing System is a fracture fixation system and includes sterile implants (intramedullary nails in various diameter, compression screw and end caps) as well as non-sterile instruments (targeting devices). The sterile implants are made of titanium alloy (Ti6A14V ELI) as per ASTM F136. The T2 Alpha Tibia Nailing System will be used with the locking screws cleared in K003018 (Titan Tibial Nail) and K172774 (IMN Screws System). Further, End Cap Lower Extremity and the Nail Holding Screw Tibia / Femur PF of T2 Alpha Tibia Nailing System can be used with T2 Alpha Femur Antegrade GT/PF Nailing System (K172774). The T2 Alpha Tibia Nailing System will be used with the existing instruments previously cleared in
{9}------------------------------------------------
K131365 (T2 Tibial Nailing System) and 510(k) exempt devices. The T2 Alpha Tibia Nailing System is intended for use with IMN Screws System and IMN Instruments System.
IMN Screws System
The IMN Screws System includes bone screws (locking screws previously cleared in K172774 and advanced locking screws) that are inserted through the intramedullary nail to stabilize the nail-bone construct. The screws are made of titanium alloy (Ti6A14V ELI) as per ASTM F136. The IMN Screws System is intended for use with T2 Alpha Tibia Nailing System, T2 Alpha Femur Antegrade GT/PF Nailing System and IMN Instruments System.
IMN Instruments System
The IMN Instruments System includes distal targeting device femur antegrade previously cleared in K172774 as well as surgical instruments that support the implantation and extraction of intramedullary nails and screws. The IMN Instruments System is intended for use with T2 Alpha Tibia Nailing System, T2 Alpha Femur Antegrade GT/PF Nailing System and IMN Screws System.
T2 Alpha Femur Antegrade GT/PF Nailing System
The T2 Alpha Femur Antegrade GT/PF Nailing System previously cleared in K172774 is a fracture fixation system and includes sterile implants (femoral nails in various diameter and sizes, compression screw, set screws and end caps) as well as non-sterile instruments.
The T2 Alpha Femur Antegrade GT/PF Nailing System will be used with the locking screws originally cleared in K003018 (Titan Tibial Nail) that have subsequently also received clearance for use in locking femoral nailing systems (K010801, T2 Femoral Nail System) as well as locking screws previously cleared in K172774 (IMN Screws System) and the T2 Lag Screw Recon previously cleared in K032898 (T2 Recon Nail System). Further, it will be used with the existing instruments previously cleared in K172774 (IMN Instruments System) and 510(k) exempt devices. The T2 Alpha Femur Antegrade GT/PF Nailing System is intended for use with IMN Screws System and IMN Instruments System.
{10}------------------------------------------------
Traditional 510(k) Premarket Notification
V. INTENDED USE
T2 Alpha Tibia Nailing System
The T2 Alpha Tibia Nailing System is intended for temporary stabilization of bone segments or fragments until bone consolidation has been achieved.
IMN Screws System
The IMN Screws System is intended to stabilize the intramedullary nail-bone construct for temporary stabilization.
IMN Instruments System
The IMN Instruments System is intended to enable the implantation and extraction of intramedullary nail and screw.
T2 Alpha Femur Antegrade GT/PF Nailing System
The T2 Alpha Femur Antegrade GT/PF Nailing System is intended for temporary stabilization of bone segments or fragments until bone consolidation has been achieved.
VI. INDICATION FOR USE
T2 Alpha Tibia Nailing System
The indications for use of this internal fixation device include:
- . Open and closed tibial fractures
- . Pseudoarthrosis and correction osteotomy
- . Pathologic fractures, impending pathologic fractures, and tumor resections
- . Fractures involving osteopenic and osteoporotic bone
- . Nonunion and malunion
The End Cap Lower Extremity and the Nail Holding Screw Tibia / Femur PF may also be used in conjunction with the T2 Alpha Femur Antegrade GT/PF Nailing System.
IMN Screws System
The IMN Screws System is intended to stabilize the intramedullary nail-bone construct for temporary stabilization.
{11}------------------------------------------------
IMN Instruments System
The IMN Instruments System is intended to enable the implantation and extraction of intramedullary nail and screw.
T2 Alpha Femur Antegrade GT/PF Nailing System
The indications for use of these internal fixation devices include:
- . Fixation of subtrochanteric, intertrochanteric, ipsilateral neck/shaft, comminuted proximal femoral shaft fractures
- . Femoral fixation required as a result of pathological disease
- . Temporary stabilization of fractures of the femoral shaft ranging from the femoral neck to the supracondylar regions of the femur
- . Open and closed femoral fractures
- . Pseudoarthrosis and correction osteotomy
- . Pathologic fractures, impending pathologic fractures and tumor resections
- . Ipsilateral femur fractures
- . Fractures proximal to a total knee arthroplasty
- . Nonunions and malunions
- Fractures involving osteopenic and osteoporotic bone ●
VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
T2 Alpha Tibia Nailing System
Device comparison demonstrated that the T2 Alpha Tibia Nailing System is substantially equivalent to the previously cleared Titan Tibial Nail (K003018), Zimmer Natural Nail System Tibial Nail (K082770) and Synthes (USA) Tibial Nail System EX (K040762) in regards to intended use, indications for use, technological characteristics (design features, material and performance) as well as operating principle.
At a high level, the subject device and predicate devices are based on the following same technological elements:
- . Intramedullary nailing systems to provide a fracture fixation of the tibia,
- Nail and screw design (length, diameter, and shape), and .
{12}------------------------------------------------
-
Nails, compression screw and end caps manufactured from titanium alloy (Ti6Al4V ELI) . as per ASTM F136.
The following technological difference exists between the subject and predicate devices: -
Hole configurations in proximal and distal part of nail. .
The difference has been addressed through performance testing which demonstrates that the subject device is as safe and effective as the predicate devices.
IMN Screws System
Device comparison demonstrated that the IMN Screws System is substantially equivalent to the previously cleared IMN Screws System (K172774) in regards to intended use, indications for use, technological characteristics (material and performance) as well as operating principle. The reference device (Titan Tibial Nail System, K003018) is suitable to provide technical information to help address the safety and effectiveness of a new technological characteristic. At a high level, the subject device and predicate device are based on the following same technological elements:
- Stabilization of intramedullary nail-bone construct, .
- . Used for proximally and distally locking of nail-bone construct,
- . Design (length and diameter), and
- . Manufactured from titanium alloy (Ti6A14V ELI) as per ASTM F136.
The following technological difference exists between the subject and predicate devices:
- . Thread design
The difference has been addressed through performance testing which demonstrates that the subject device is as safe and effective as the predicate devices.
IMN Instruments System
Device comparison demonstrated that the IMN Instruments System is substantially equivalent to the previously cleared IMN Instruments System (K172774) in regards to intended use, indications for use, technological characteristics as well as operating principle.
At a high level, the subject device and predicate devices are based on the following same technological elements:
- Used for the implantation and extraction of intramedullary nail and screw, and .
- Compatibility with T2 Alpha nailing systems and IMN Screws System.
{13}------------------------------------------------
T2 Alpha Femur Antegrade GT/PF Nailing System
Device comparison demonstrated that the T2 Alpha Femur Antegrade GT/PF Nailing System is substantially equivalent to the previously cleared T2 Alpha Femur Antegrade GT/PF Nailing System (K172774) in regards to intended use, technological characteristics (design features, material and performance) as well as operating principle.
At a high level, the subject device and predicate devices are based on the following same technological elements:
- . Intramedullary nailing systems to provide a fracture fixation of the long bone,
- Nail and screw design (length, diameter, and shape),
- . Nails, compression screw and end caps manufactured from titanium alloy (Ti6Al4V ELI) as per ASTM F136,
- Set screws are manufactured from titanium alloy (Ti6A14V ELI) as per ASTM F136 and . PEEK,
- . Hole configurations in proximal and distal part of nail, and
- Locking configurations. ●
VIII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
T2 Alpha Tibia Nailing System
Non-clinical testing was performed for T2 Alpha Tibia Nailing System under worst case scenario to determine substantial equivalence. The following bench testing was conducted:
- . Testing of mechanical properties per ASTM F1264
- . Fatigue strength testing (proximal and distal)
- . Cut-out testing
- . Targeting accuracy testing
Testing demonstrated that T2 Alpha Tibia Nailing System is equivalent in mechanical performance to the predicate devices (K003018, K082770, and K040762).
Non-clinical testing has demonstrated that T2 Alpha Tibia Nailing System is MR Conditional.
Biocompatibility testing (ISO 10993-1, ISO 10993-5 and ISO 10993-18) was performed to evaluate the biological safety of T2 Alpha Tibia Nailing System according to FDA Guidance
{14}------------------------------------------------
'Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process'. Testing demonstrated that T2 Alpha Tibia Nailing System is biocompatible.
The Bacterial Endotoxin Testing demonstrated that the sterile implants of T2 Alpha Tibia Nailing System meet the specified endotoxin limit.
IMN Screws System
Non-clinical testing was performed for the advanced locking screws of IMN Screws System. The following bench testing was conducted:
- Testing of mechanical properties as per ASTM F543 ●
Testing demonstrated that the advanced locking screws are equivalent in mechanical performance to the predicate device (K172774) and reference device (K003018).
Non-clinical testing has demonstrated that the advanced locking screws are MR Conditional.
Biocompatibility assessment (ISO 10993-1) was performed according to FDA Guidance 'Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process'. It has been demonstrated that the advanced locking screws are biocompatible. The Bacterial Endotoxin Testing demonstrated that the advanced locking screws meet the specified endotoxin limit.
IMN Instruments System
Non-clinical testing was performed for the surgical instruments of IMN Instruments System. The following bench testing was conducted:
- . Performance testing in cadaver lab, and
- . Compatibility evaluation
Testing demonstrated that the surgical instruments are equivalent in performance to the predicate device (K172774).
Biocompatibility testing (ISO 10993-1, ISO 10993-5, and ISO 10993-18) was performed according to FDA Guidance 'Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process'. It has been demonstrated that the surgical instruments are biocompatible.
{15}------------------------------------------------
T2 Alpha Femur Antegrade GT/PF Nailing System
Non-clinical testing was performed for T2 Alpha Femur Antegrade GT/PF Nailing System to determine substantial equivalence. Compatibility assessment demonstrated that T2 Alpha Femur Antegrade GT/PF Nailing System is suitable to be used with the advanced locking screws of IMN Screws System and the surgical instruments of IMN Instruments System as well as the End Cap Lower Extremity and the Nail Holding Screw Tibia / Femur PF of T2 Alpha Tibia Nailing System.
IX. CLINICAL TESTING
T2 Alpha Tibia Nailing System
No clinical testing of the T2 Alpha Tibia Nailing System has been conducted.
IMN Screws System
No clinical testing of the IMN Screws System has been conducted.
IMN Instruments System
No clinical testing of the IMN Instruments System has been conducted.
T2 Alpha Femur Antegrade GT/PF Nailing System
No clinical testing of the T2 Alpha Femur Antegrade GT/PF Nailing System has been conducted.
X. CONCLUSION
T2 Alpha Tibia Nailing System
The T2 Alpha Tibia Nailing System is substantially equivalent to the predicate devices (Titan Tibial Nail (K003018), Zimmer Natural Nail System Tibial Nail (K082770), and Synthes (USA) Tibial Nail System EX (K040762)) identified in this premarket notification.
IMN Screws System
The IMN Screws System is substantially equivalent to the predicate device (IMN Screws System (K172774) identified in this premarket notification.
{16}------------------------------------------------
IMN Instruments System
The IMN Instruments System is substantially equivalent to the predicate device (IMN Instruments System (K172774)) identified in this premarket notification.
T2 Alpha Femur Antegrade GT/PF Nailing System
The T2 Alpha Femur Antegrade GT/PF Nailing System is substantially equivalent to the predicate device (T2 Alpha Femur Antegrade GT/PF Nailing System (K172774)) identified in this premarket notification.
§ 888.3020 Intramedullary fixation rod.
(a)
Identification. An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.(b)
Classification. Class II.