K Number
K180389
Device Name
Venera 508 Deep Vein Thrombosis (DVT) Prevention System
Date Cleared
2018-06-29

(136 days)

Product Code
Regulation Number
870.5800
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The Venera 508 Deep Vein Thrombosis (DVT) Prevention Therapy System is a portable treatment to help prevent the onset of DVT in patients by stimulating blood flow in the extremities (simulating muscle contractions). This system can be applied in either home or clinical settings to: - . Aid in the prevention of DVT - . Enhance blood circulation - . Diminish post-operative pain and swelling - . Reduce wound healing time - . Aid in the treatment of: stasis dermatitis, venous stasis ulcers, arterial and diabetic leg ulcers, chronic venous insufficiency and reduction of edema in the lower limbs - . As a prophylaxis for DVT by persons expecting to be stationary for long periods of time
Device Description
The Venera™ 508 Deep Vein Thrombosis Prevention System is an easy to use portable pneumatic compression system that noninvasively helps prevent the onset of DVT in patients by simulating blood flow in the extremities (simulating muscle contractions). The Venera 508 system consists of a pair of pump and cuff assemblies. The device will alternatively inflate and deflate the garment (cuff) to stimulate blood flow in the extremities (muscle contraction). The device provides a 50mmHg pressure and followed by 50 seconds of deflation period once it reaches the desired pressure.
More Information

Not Found

No
The description focuses on pneumatic compression and mechanical function, with no mention of AI or ML terms or capabilities.

Yes
The device is clearly indicated for "Deep Vein Thrombosis (DVT) Prevention Therapy" and aims to "aid in the prevention of DVT", "enhance blood circulation", and "diminish post-operative pain and swelling", all of which are therapeutic purposes.

No

The device is a DVT prevention therapy system that stimulates blood flow; it does not diagnose medical conditions.

No

The device description explicitly states the system consists of "a pair of pump and cuff assemblies" which are hardware components. The performance studies also include tests related to hardware function (pressure accuracy, battery life, etc.).

Based on the provided information, the Venera 508 Deep Vein Thrombosis (DVT) Prevention Therapy System is not an IVD (In Vitro Diagnostic) device.

Here's why:

  • IVD devices are used to examine specimens derived from the human body. The Venera 508 system works by applying external pneumatic compression to the extremities to stimulate blood flow. It does not analyze any biological samples (like blood, urine, tissue, etc.).
  • The intended use and device description clearly outline a physical therapy/prevention method. The system is designed to mechanically aid in blood circulation and prevent DVT through external pressure, not through laboratory analysis of patient samples.

Therefore, the Venera 508 system falls under the category of a therapeutic or preventative medical device, not an in vitro diagnostic device.

N/A

Intended Use / Indications for Use

The Venera 508 Deep Vein Thrombosis (DVT) Prevention Therapy System is a portable treatment to help prevent the onset of DVT in patients by stimulating blood flow in the extremities (simulating muscle contractions).

This system can be applied in either home or clinical settings to:

  • . Aid in the prevention of DVT
  • . Enhance blood circulation
  • . Diminish post-operative pain and swelling
  • . Reduce wound healing time
  • . Aid in the treatment of: stasis dermatitis, venous stasis ulcers, arterial and diabetic leg ulcers, chronic venous insufficiency and reduction of edema in the lower limbs
  • . As a prophylaxis for DVT by persons expecting to be stationary for long periods of time

Product codes

JOW

Device Description

The Venera™ 508 Deep Vein Thrombosis Prevention System is an easy to use portable pneumatic compression system that noninvasively helps prevent the onset of DVT in patients by simulating blood flow in the extremities (simulating muscle contractions). The Venera 508 system consists of a pair of pump and cuff assemblies. The device will alternatively inflate and deflate the garment (cuff) to stimulate blood flow in the extremities (muscle contraction). The device provides a 50mmHg pressure and followed by 50 seconds of deflation period once it reaches the desired pressure.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Extremities (limbs)

Indicated Patient Age Range

Not Found

Intended User / Care Setting

home or clinical settings

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies

To verify that the device design met its function and performance requirements, samples of the device underwent function and mechanical testing.

The following tests were conducted:

  • Pressure Accuracy Test Report
  • Pressure Switch Test Report
  • Venera 508 Cycle Time Test Report
  • Alarm Function Test Report
  • Battery Life Test Report

The conclusions drawn from the performance tests demonstrate that the device is performing as intended, and is safe and effective.

Key Metrics

Not Found

Predicate Device(s)

K151189

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 870.5800 Compressible limb sleeve.

(a)
Identification. A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.(b)
Classification. Class II (performance standards).

0

June 29, 2018

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Suzhou Minhua Medical Apparatus Supplies Co. Ltd. Alex Wang Regulatory Affairs Specialist No.789, Wu Fang Road, Friendship Industrial Zone Song Ling Town, Wu Jiang Suzhou, 215222 China

Re: K180389

Trade/Device Name: Venera 508 Deep Vein Thrombosis (DVT) Prevention System Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II Product Code: JOW Dated: May 15, 2018 Received: June 4, 2018

Dear Alex Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820);

1

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Nicole M. Gillette -S 2018.06.29 08:28:57 -04'00'

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K180389

Device Name

Venera 508 Deep Vein Thrombosis Prevention Therapy System

Indications for Use (Describe)

The Venera 508 Deep Vein Thrombosis (DVT) Prevention Therapy System is a portable treatment to help prevent the onset of DVT in patients by stimulating blood flow in the extremities (simulating muscle contractions).

This system can be applied in either home or clinical settings to:

  • . Aid in the prevention of DVT
  • . Enhance blood circulation
  • . Diminish post-operative pain and swelling
  • . Reduce wound healing time
  • . Aid in the treatment of: stasis dermatitis, venous stasis ulcers, arterial and diabetic leg ulcers, chronic venous insufficiency and reduction of edema in the lower limbs
  • . As a prophylaxis for DVT by persons expecting to be stationary for long periods of time

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
-------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

510K Summary

Submitter:

Suzhou Minhua Medical Apparatus Supplies Co. Ltd.

No.789, Wu Fang Road, Friendship Industrial Zone, Song Ling Town,

Wu Jiang District, Su Zhou City, Jiangsu Province, China

Postal Code: 215222

Tel: +86-0512-63091066

Contact Person: Alex Wang

Date Prepared: Jan 15, 2018

Device:

Common Names:Intermittent Pneumatic Compression Device
Proprietary Name:Venera™ 508 Deep Vein Thrombosis Prevention System
Regulatory Class:II
Product Code:JOW

Predicate Devices:

The Venera 508 Deep Vein Thrombosis Prevention System is equivalent to the following:

Predicate DeviceManufacturer510(k)#
Cirona 6300 Deep Vein
Thrombosis(DVT) Prevention SystemDevon Medical ProductsK151189

Device Description

The Venera™ 508 Deep Vein Thrombosis Prevention System is an easy to use portable pneumatic compression system that noninvasively helps prevent the onset of DVT in patients by

4

simulating blood flow in the extremities (simulating muscle contractions). The Venera 508 system consists of a pair of pump and cuff assemblies. The device will alternatively inflate and deflate the garment (cuff) to stimulate blood flow in the extremities (muscle contraction). The device provides a 50mmHg pressure and followed by 50 seconds of deflation period once it reaches the desired pressure.

Indications for Use:

The Venera™ 508 Deep Vein Thrombosis Prevention System is a portable treatment system intended to help prevent the onset of DVT in patients by stimulating blood flow in the extremities (simulating muscle contractions).

This system can be applied in either home or clinical settings to:

  • Aid in the prevention of DVT .
  • . Enhance blood circulation
  • . Diminish post-operative pain and swelling
  • . Reduce wound healing time
  • . Aid in the treatment of: stasis dermatitis, venous stasis ulcers, arterial and diabetic leg ulcers, chronic venous insufficiency and reduction of edema in the lower limbs
  • As a prophylaxis for DVT by persons expecting to be stationary for long periods of time

Technological Characteristics:

Below is a table of comparison for the technological characteristics against the predicate device:

PredicateCirona 6300 DVT System
Indication for UseS
ComponentsS
MaterialSE
BiocompatibilitySE
Principle of OperationSE
Treatment parametersS
User InterfaceS

*SE – Substantial Equivalent *S - Same

5

The manufacturer believes that the technological characteristics of the Venera™ 508 Deep Vein Thrombosis Prevention System are substantially similar to those of the predicate device. Venera 508 DVT Prevention System has similar components to its predicate device and identical principles of operation.

Performance Tests

To verify that the device design met its function and performance requirements, samples of the device underwent function and mechanical testing.

The following tests were conducted:

Function Performance Tests
Pressure Accuracy Test Report
Pressure Switch Test Report
Venera 508 Cycle Time Test Report
Alarm Function Test Report
Battery Life Test Report

The conclusions drawn from the performance tests demonstrate that the device is performing as intended, and is safe and effective.

Biocompatibility

Biocompatibility test evaluation for Venera 508 DVT Prevention System is done in accordance to the FDA Good Laboratory Practice quidelines. The following tests were done:

  • Cytotoxicity い
  • Sensitization ﻬﺎ ﻓﻬﻮ ﺍﻟﻤﺪﻳﻨﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪ
  • Irritation い

Sterilization and Shelf Life

Sterilization and shelf life are not applicable to Venera 508 DVT Prevention System.

Electrical Safety and Electromagnetic Compatibility (EMC)

Electrical safety and EMC tests were conducted according to the following standards:

  • IEC 60601-1: 2005 + A1:2012 Medical Electrical Equipment Part 1: General requirements , for basic safety and essential performance

6

  • IEC 60601-1-2: 2014 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests
  • IEC 60601-1-6:2010 + A1:2013 Medical electrical equipment Part 1-6: General requirements ﻬﺎ ﻓﻬﻮ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪ for safety - Collateral Standard: Usability including IEC 62366: Application of usability engineering to medical devices
  • IEC 60601-1-11: 2015 Medical electrical equipment Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

Software Verification and Validation

Software verification and validation were conducted and documentation is provided. The software was considered as a Moderate level of concern, since a failure or latent flaw in the software could directly or indirectly result in minor injury to the patient or operator.

Animal Study and Clinical Study

No animal study or clinical study was conducted.

Statement of Substantial Equivalence

The Venera 508 Deep Vein Thrombosis Prevention System is substantially equivalent in its intended use, technology, function, operating parameters to predicate devices that are currently commercially available and in distribution, and does not raise any new concerns regarding safety or effectiveness.