K Number
K180381
Date Cleared
2018-11-07

(268 days)

Product Code
Regulation Number
880.6250
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is disposable non-sterile device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Powder Free Vinyl Patient Examination Gloves are Patient Examination Gloves, Disposable, single use only and non-sterile. The gloves are made of vinyl materials and are powder free. The physical and performance characteristics of the devices meet all requirements of ASTM D5250-06 (Reapproved 2015) Standard Specification For Poly(Vinyl Chloride) Gloves For Medical Application.

AI/ML Overview

This document is a 510(k) summary for "Powder Free Vinyl Patient Examination Gloves," a Class I medical device. The information provided pertains to the substantial equivalence of the subject device to a predicate device, focusing on non-clinical performance characteristics to demonstrate safety and effectiveness.

Here's the breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

1. A table of acceptance criteria and the reported device performance

The acceptance criteria are generally established by the ASTM D 5250-06 (Reapproved 2015) Standard Specification for Poly(Vinyl Chloride) Gloves For Medical Application. The predicate device and the subject device both aim to meet these requirements.

CharacteristicsAcceptance Criteria (ASTM D 5250-06 (2015) requirements)Reported Device Performance (Subject Device K180381)Result of Comparison (Predicate vs Subject)Conclusion (Subject Device performance)
Physical Properties:
Tensile Strength (Before aging)11 Mpa minimum14.4-17.1 MpaSame (Met/Exceeded)Pass (Meets/Exceeds criteria)
Tensile Strength (After aging)11 Mpa minimum14.1-16.9 MpaSame (Met/Exceeded)Pass (Meets/Exceeds criteria)
Elongation (Before aging)300% minimum390-480%Same (Met/Exceeded)Pass (Meets/Exceeds criteria)
Elongation (After aging)300% minimum380-470%Same (Met/Exceeded)Pass (Meets/Exceeds criteria)
Freedom from PinholeG-I, AQL 2.5 (Meets ASTM D 5250-06 (2015) requirements)Meets AQL 2.5 requirementsSamePass (Meets criteria)
Dimensions:Meets ASTM D 5250-06 (2015) requirementsSamePass (Meets criteria)
- Overall Length (mm)230 for all sizes minimumS: 233-241mm, M:231-242mm, L:230-244mm, XL: 234-245mmN/A (Directly compared to standard)Pass (Meets/Exceeds criteria)
- Width (mm)S: 85 ± 5, M: 95 ± 5, L: 105±5, XL: 115±5S: 87-88 mm, M: 97-98 mm, L: 107-108 mm, XL: 118-119 mmN/A (Directly compared to standard)Pass (Meets criteria)
- Palm Thickness (mm)0.08mm minimum0.08mmN/A (Directly compared to standard)Pass (Meets criteria)
- Finger Thickness (mm)0.05mm minimum0.09-0.11mmN/A (Directly compared to standard)Pass (Meets/Exceeds criteria)
Residual PowderNot more than 2 mg per glove (Meets Applicable Definition for Powder Free)0.20 mg/gloveSame (Met/Exceeded)Pass (Meets/Exceeds criteria)
Biocompatibility:
- Primary Skin IrritationUnder conditions of the study, not an irritantUnder conditions of the study, not an irritantSamePass (Not an irritant)
- Dermal SensitizationUnder conditions of the study, not an irritantUnder conditions of the study, not an irritantSamePass (Not an irritant)
- Cytotoxicity TestN/A (Predicate had no data)Under the conditions of this study, not a cytotoxic potentialDifferent (Subject had data)Pass (Not a cytotoxic potential)

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not explicitly state the specific numerical sample sizes for each test. Instead, it refers to "Inspection Level and AQL" (Acceptable Quality Level) from the ASTM standards for some tests (e.g., S-2, AQL 4.0 for dimensions and tensile strength; G-I, AQL 2.5 for pinhole). These AQLs imply a sampling plan rather than a fixed sample size.

  • Pinhole: Inspection Level G-I, AQL 2.5.
  • Dimensions, Tensile Strength, Ultimate Elongation: Inspection Level S-2, AQL 4.0.
  • Residual Powder: N=5 (5 gloves were tested).
  • Biocompatibility Tests (Primary Skin Irritation, Dermal Sensitization, Cytotoxicity Test): No sample sizes are specified for these in the table; the results are qualitative ("not an irritant," "not a cytotoxic potential").

The data provenance is from Shanxi Hongjin Plastic Technology Co., Ltd in China, as indicated by their address on the document. The studies are non-clinical lab tests conducted to support the 510(k) submission, implying a prospective data generation process for the purpose of regulatory approval.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This section is not applicable as the device is a medical glove, and the "ground truth" is based on objective physical, chemical, and biological performance characteristics measured by laboratory tests and adherence to industry standards (ASTM D5250-06), not expert medical interpretation of images or clinical outcomes.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in studies involving subjective human interpretation (e.g., radiologists reading medical images) to establish consensus ground truth. The tests for these gloves are objective measurements.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. This type of study relates to AI-assisted diagnostic devices and human reader performance, not physical medical devices like examination gloves.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable. This concept applies to AI/software as a medical device, where the algorithm performs a task independently. The tests performed here are direct measurements of a physical product's attributes.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for this device's performance is established by objective measurements against pre-defined specifications and compliance with recognized industry standards, specifically ASTM D 5250-06 (Reapproved 2015). For biocompatibility, it's based on standard biological evaluation tests determining if the material is an irritant or cytotoxic.

8. The sample size for the training set

This is not applicable. The concept of a "training set" refers to data used to train machine learning models. This document describes the testing and performance of a physical product according to established standards, not an AI/ML device.

9. How the ground truth for the training set was established

This is not applicable for the same reasons as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

November 7, 2018

Shanxi Hongjin Plastic Technology Co., Ltd Kathy Liu Project Manager Hongray USA Medical Products Inc 3973 Schaefer Ave. Chino, California 91710

Re: K180381

Trade/Device Name: Powder Free Vinyl Patient Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LYZ Dated: September 26, 2018 Received: October 2, 2018

Dear Kathy Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Clarence W. Murray lii III -S

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180381

Device Name

Powder Free Vinyl Patient Examination Gloves

Indications for Use (Describe)

A patient examination glove is disposable non-sterile device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Coal Bed Gas Industrial Zone, Qu'e Town, Daning County, Linfen City, Shanxi Province, China 042399

Trade Name: Powder Free Vinyl Patient Examination Gloves

510(K) SUMMARY

This summary of 510(K) is being submitted in accordance with 21 CFR 807.92.

The assigned 510(K) number is: K180381

1. Owner's Identification:

Mr. Wu Zhigang Shanxi Hongjin Plastic Technology Co., LTD Coal Bed Gas Industrial Zone, Qu'e Town, Daning County, Linfen City, Shanxi Province, China 042399

Tel: 86-311-66179653 Fax: 86-311-83616934

Contact: Ms. Kathy Liu, Project Manager or Ms Monica Yu Address: 3973 Schaefer Ave., Chino, CA 91710 Tel: 909-590-1611 Fax: 909-590-1511 Date Summary Prepared: November 5, 2018

2. Name of the Device:

Trade Name: Powder Free Vinyl Patient Examination Gloves Common Name: Exam Gloves Classification Name: Patient Examination Glove Classification Regulation: 880.6250 Classification Panel: 880 General Hospital and Personal Use Product Code: LYZ Device Class: Class I

3. Predicate Device Information:

Ever Light Plastic Products Co., Ltd. Powder Free Vinyl Patient Examination Gloves (K142571)

4. Device Description:

Powder Free Vinyl Patient Examination Gloves are Patient Examination Gloves, Disposable, single use only and non-sterile. The gloves are made of vinyl materials and are powder free. The physical and performance characteristics of the devices meet all requirements of ASTM D5250-06 (Reapproved 2015) Standard Specification For Poly(Vinyl Chloride) Gloves For Medical Application.

{4}------------------------------------------------

Coal Bed Gas Industrial Zone, Qu'e Town, Daning County, Linfen City, Shanxi Province, China 042399

Trade Name: Powder Free Vinyl Patient Examination Gloves

5. Intended Use of the Device:

A patient examination glove is disposable non-sterile device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

6. Technological Characteristic Comparison Table:

Shown below is the comparison of technological characteristic between the subject and predicate device.

CharacteristicsDevice PerformanceResult ofcomparison
Predicate deviceK142571Subject Device K180381
Product CodeLYZLYZSame
Intended UsePredicate device isdisposable non-steriledevice intended formedical purpose that isworn on the examiner'shand or finger to preventcontamination betweenpatient and examiner.Subject device isdisposable non-steriledevice intended formedical purpose that isworn on the examiner'shand or finger to preventcontamination betweenpatient and examiner.Same
LabelingThere are no speciallabeling claimsThere are no speciallabeling claimsSame
Device MaterialsVinylVinylSame
ColorClearClearSimilar
Specifications & Performance Data:
physical Properties11Mpa minimumElongation 300%Meets ASTM D 5250-06(2015)requirements11Mpa minimumElongation:300%Meets ASTM D 5250-06(2015)requirementsSame
Freedom frompinholesG -1,AQL 2.5Meets ASTM D 5250-06(2015)requirementsG -1,AQL 2.5Meets ASTM D 5250-06(2015)requirementsSame
Dimensions:Meets ASTM D 5250-06(2015)requirementsMeets ASTM D 5250-06(2015)requirementsSame
Residual powderMeets ApplicableDefinition for PowderFree; ≤ 2 mg per gloveMeets ApplicableDefinition for PowderFree; ≤ 2 mg per gloveSame
BiocompatibilityPrimary skinUnder conditions of theUnder conditions of theSame

{5}------------------------------------------------

Coal Bed Gas Industrial Zone, Qu'e Town, Daning County, Linfen City, Shanxi Province, China 042399

Trade Name: Powder Free Vinyl Patient Examination Gloves

Device PerformanceResult of
CharacteristicsPredicate deviceK142571Subject Device K180381comparison
irritation teststudy, not an irritantstudy, not an irritant
DermalsensitizationassayUnder conditions of thestudy, not an irritantUnder conditions of thestudy, not an irritantSame
Cytotoxicity Test/Under the conditions ofthis study, not a cytotoxicpotentialDifferent

Shanxi Hongjin Plastic Technology Co., Ltd's Powder Free Vinyl Patient Examination Gloves shares the same or similar technology characteristics compared to the predicate device.

7. Discussion of Non-Clinical Tests

Testing ItemsFDA-recognizedStandardRequirementsInspectionLevel and AQLActual TestingResultsConclusion
OverallLength(mm)230 for all sizesminimumS-2, AQL4.0S: 233-241mmM:231-242mmL:230-244mmXL: 234-245mmPass
Width(mm)S: 85 ± 5S-2, AQL4.087-88 mmPass
M: 95 ± 597-98 mm
L: 105±5107-108 mm
XL: 115±5118-119 mm
PalmThickness(mm)0.08mmminimumS-2, AQL4.00.08mmPass
FingerThickness(mm)0.05mmminimumS-2, AQL4.00.09-0.11mmPass
Tensile Strength (Mpa)
Before aging11Mpa minimumS-2, AQL4.014.4-17.1MpaPass
After aging11Mpa minimum14.1-16.9MpaPass
Ultimate Elongation (%)
Before aging300% minimumS-2, AQL4.0390-480%Pass
After aging300% minimum380-470%Pass

{6}------------------------------------------------

Coal Bed Gas Industrial Zone, Qu'e Town, Daning County, Linfen City, Shanxi Province, China 042399

Pinhole2.5G-Imeet AQL2.5 requirementsPass
ResidualPowderNot more than2mg per gloveN=50.20 mg/glovePass
(a) PrimarySkinIrritation TestUnder conditions of the study, notan irritantUnder conditions of the study, notan irritant
(b) DermalSensitizationStudyUnder conditions of the study, notan irritantUnder conditions of the study, notan irritant
(c)CytotoxicityTestUnder the conditions of this study,not a cytotoxic potentialUnder the conditions of this study,not a cytotoxic potential

Trade Name: Powder Free Vinyl Patient Examination Gloves

8. Discussion of Clinical Tests Performed:

Not applicable

9. Conclusion:

Based on the performed nonclinical tests, demonstrate that the device is as safe, as effective, and performs as well as the legally marketed predicate device.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.