K Number
K180341
Date Cleared
2018-08-03

(177 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.

Device Description

FUJIFILM 600 Series Endoscope EG-600WR v2 is an upper gastrointestinal endoscope that captures images when used in combination with a video processor and light source. Light travels from the light source, through the glass fiber bundles in the endoscope, and out the tip of the insertion portion to illuminate the body cavity. This provides enough light to the CMOS image sensor to capture an image and display it on the monitor.

AI/ML Overview

The provided text describes the Fujifilm 600 Series Endoscope EG-600WR v2, a medical device intended for the visualization and treatment of the upper digestive tract. It details a 510(k) premarket notification for this device, claiming substantial equivalence to a predicate device (Fujifilm 600 Series Endoscopes EC-600WL and EG-600WR, K132210).

However, the document does not contain information about an AI/ML-driven medical device, nor does it describe a study involving algorithms, human-in-the-loop performance, or reader studies with experts for ground truth establishment.

The provided text focuses on the endoscope device itself, and the modifications made to it. The performance testing outlined is for the physical device's characteristics (e.g., field of view, bending capability, resolution, safety, and reprocessing).

Therefore, I cannot provide the requested information regarding acceptance criteria and studies for an AI/ML device because the document describes a physical medical endoscope and not an AI/ML system.

If the request refers to the acceptance criteria for the physical endoscope, the available information is as follows:

1. Table of Acceptance Criteria and Reported Device Performance:

The document states: "In all cases, the device met the pre-defined acceptance criteria for the test." However, the specific numerical acceptance criteria for each performance parameter are not explicitly listed in the provided text. Only the parameters tested are mentioned.

Acceptance Criteria (Not Explicitly Stated)Reported Device Performance
Field of viewMet pre-defined criteria
Bending capabilityMet pre-defined criteria
Air supply rateMet pre-defined criteria
Water supply rateMet pre-defined criteria
Suction rateMet pre-defined criteria
Working lengthMet pre-defined criteria
Forceps channel diameterMet pre-defined criteria
Viewing directionMet pre-defined criteria
ResolutionMet pre-defined criteria
LG output (Light Guide Output)Met pre-defined criteria
Electrical safety (ANSI/AAMI ES60601-1, etc.)Compliant
Biocompatibility (ISO 10993)Compliant
Endoscopic performance (ISO 8600-1)Compliant
Cleaning, high-level disinfection, sterilization validationCompliant
Storage and transportation validationCompliant

2. Sample sized used for the test set and the data provenance: Not applicable as this is a physical device and not a data-driven AI/ML system. The testing involved samples of the physical endoscope.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth as typically understood for AI/ML models (e.g., expert consensus on image findings) is not relevant for the performance testing of a physical endoscope.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document is not about an AI-assisted device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithm.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc): For the physical device, performance was evaluated against technical specifications and consensus standards (e.g., ISO 8600-1 for endoscopic performance, IEC and ISO for safety and biocompatibility).

8. The sample size for the training set: Not applicable. This is a physical device, not an AI/ML model.

9. How the ground truth for the training set was established: Not applicable. This is a physical device, not an AI/ML model.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 3, 2018

FUJIFILM Corporation % Jeffrey Wan Specialist, Regulatory Affairs FUJIFILM Medical Systems U.S.A., Inc. 10 High Point Drive Wayne, NJ 07470

Re: K180341

Trade/Device Name: FUJIFILM 600 Series Endoscope EG-600WR v2 Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FDS Dated: July 5, 2018 Received: July 6, 2018

Dear Jeffrey Wan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Glenn B. Bell -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K180341

Device Name

FUJIFILM 600 Series Endoscope EG-600WR v2

Indications for Use (Describe)

The device is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K180341 Page 1 of 5

510(k) SUMMARY

FUJIFILM 600 Series Endoscope EG-600WR v2

Submitter's Information:

FUJIFILM Corporation 798 Miyanodai Kaisei-Machi Ashigarakami-Gun, Kanagawa, 258-8538, Japan FDA Establishment Registration Number: 3001722928

Contact Person:

Jeffrey Wan Specialist, Regulatory Affairs Fujifilm Medical Systems U.S.A., Inc. Telephone: (973) 709-2219 Facsimile: (201) 995-2452 E-Mail: jeffrey.wan@fujifilm.com

Date Prepared: July 5th, 2018

Identification of the Proposed Device:

Proprietary/Trade Name:FUJIFILM 600 Series Endoscope EG-600WR v2
Common Name:Video Endoscope
Device Class:Class II
Review Panel:Gastroenterology/Urology
Classification:Endoscope and accessories, 21 C.F.R. § 876.150
Product Code:FDS

Predicate Device:

Fujifilm 600 Series Endoscopes EC-600WL and EG-600WR, Fujifilm Medical Systems U.S.A, K132210

Reference Devices:

    1. Fujifilm Video Colonoscope Model EC-600WLv2, FUJIFILM Medical Systems U.S.A., K160196
    1. Fujifilm Endoscope Models EC-600HL and EC-600LS, FUJIFILM Medical Systems U.S.A., K162622

{4}------------------------------------------------

Intended Use / Indications for Use:

The device is intended for the visualization of the upper digestive tract, specifically for the observation, diagnosis, and endoscopic treatment of the esophagus, stomach, and duodenum.

Device Description:

FUJIFILM 600 Series Endoscope EG-600WR v2 is an upper gastrointestinal endoscope that captures images when used in combination with a video processor and light source. Light travels from the light source, through the glass fiber bundles in the endoscope, and out the tip of the insertion portion to illuminate the body cavity. This provides enough light to the CMOS image sensor to capture an image and display it on the monitor.

Technoloqical Characteristics:

The proposed device EG-600WR v2 differs from the predicate device in the following minor modifications:

  • o Use of a new epoxy resin in the manufacturing process instead of the original epoxy resin. Both the new and original epoxy resins can be used for repair.
  • Minor design changes to the optical fiber, distal end cap, and air/water nozzle .
  • Addition of VP/BL-7000 (K163675) and VP-4440FN/XL-4450FN (K162836) as compatible . video processors and light sources.
  • Expansion of transport and storage conditions
  • Use of an improved Air/Water button AW-500 and Suction button SB-500 ●
  • Continued compatibility with an improved Water Tank WT-4 ●
  • o J Tube JT-500 is changed from a mandatory accessory to an optional accessory.

A comparison of the technological characteristics between the modified and predicate devices is provided in the table below.

Proposed DevicePredicate Device
Device nameEG-600WR v2EG-600WR
Common nameSame as K132210Endoscope and accessories
ManufacturerSame as K132210FUJIFILM Corporation
510(k) numberSame as K132210K132210
IntendedUse/Indicationsfor UseSame as K132210The device is intended for the visualization of theupper digestive tract, specifically for the observation,diagnosis, and endoscopic treatment of theesophagus, stomach, and duodenum.
AppearanceSame as K132210Image: Endoscope
Viewing directionSame as K132210Forward / 0 degrees
ObservationrangeSame as K1322102-100mm
Field of viewSame as K132210140 degrees
F# of the objectivelensSame as K1322109.1
ResolutionSame as K132210At 5mm of working distance: 0.07mm of line pair onthe square wave chart is readable.At 100mm of working distance: 0.9mm of line pair onthe square wave chart is readable.
DistortioncharacteristicsSame as K132210Orthogonal Projection
Magnification oflens(es)Same as K1322100.48-0.01
Focal lengthSame as K1322101.2mm
Image sensorsSame as K132210CMOS
# of Light GuideFiber BundlesSame as K1322102800
Distal enddiameterSame as K1322109.2mm
Flexible portiondiameterSame as K1322109.3mm
Maximuminsertion diameterSame as K13221010.7mm
BendingUpSame as K132210210 degrees
capabilityDownSame as K13221090 degrees
LeftSame as K132210100 degrees
RightSame as K132210100 degrees
Forceps channeldiameterSame as K1322102.8mm
Working lengthSame as K1322101100mm
Total lengthSame as K1322101400mm
WJ FunctionSame as K132210?
Location of WJinletSame as K132210On the light guide connector
Video ProcessorLight source: BL-7000Processor: VP-7000EPX-4440HDLight source: XL-4450Processor: VP-4440HD
EPX-4440HDLight source: XL-4450Processor: VP-4440HDEPX-4440FNLight source: XL-4450FN
Processor: VP-4440FN500 Series
Video/Light guide connectorSame as K132210
PeripheralsSame as K132210Water Tank WT-2
Water Tank WT-4
Monitor
Printer
Electrosurgical Instruments
Foot Switch
Cart
AccessoriesCleaning Brush (WB11002FW2)Cleaning Brush (WB11002FW2)
Cleaning Brush (WB4321FW2)Cleaning Brush (WB4321FW2)
Cleaning Adapter Kit (CA-510/A)Cleaning Adapter Kit (CA-510/A)
Forceps Valve (FOV-DV7)Forceps Valve (FOV-DV7)
Ventilation Adapter (AD-7)Ventilation Adapter (AD-7)
Air/Water button (AW-500)J Tube (JT-500)
Suction button (SB-500)Air/Water button (AW-500)
Water Jet Inlet capSuction button (SB-500)
Water Jet Inlet cap
Optional ItemsAir leak tester LT-7FAir leak tester LT-7F
J Tube (JT-500)
Electrical Safety ComplianceANSI AAMI ES60601-1 Edition 3.1ANSI AAMI ES60601-1 Edition 3.0

{5}------------------------------------------------

{6}------------------------------------------------

Substantial equivalence was determined based on the performance testing described below.

Performance Data:

Electrical safety testing was conducted on the proposed device EG-600WR v2 using the following consensus standards: ANSI/AAMI ES60601-1:2005+AMD1:2012 CSV, IEC 60601-1-2:2007, IEC 60601-1-6:2013, and IEC 60601-2-18:2009.

The proposed device EG-600WR v2 was adopted into biocompatibility testing conducted on the reference device EC-600WL v2 using the following consensus standards: ISO 10993-1:2009, ISO 10993-5:2009, and ISO 10993-10:2010.

Endoscopic performance testing was conducted on the proposed device EG-600WR v2 according to the consensus standard ISO 8600-1:2015.

Cleaning, high-level disinfection, and sterilization testing was conducted on the proposed device EG-600WR v2 in accordance with the FDA guidance "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling" (March 17, 2015).

The proposed device EG-600WR v2 was adopted into storage and transportation validation testing conducted on the reference devices EC-600HL/LS.

{7}------------------------------------------------

Fujifilm conducted the following performance testing on the proposed device EG-600WR v2 to ensure that the modified device performs equivalently to the predicate device:

  • . Field of view
  • Bending capability
  • Air supply rate
  • Water supply rate
  • Suction rate
  • o Working length
  • Forceps channel diameter
  • Viewing direction ●
  • o Resolution
  • LG output

In all cases, the device met the pre-defined acceptance criteria for the test.

Substantial Equivalence:

The company's EG-600WR v2 has the same intended use as the previously cleared predicate EG-600WR (K132210). In addition, the proposed device EG-600WR v2 has similar indications for use, technological characteristics, and principles of operation as its predicate. The minor differences between the proposed device and its predicate device do not raise new or additional questions of safety or effectiveness of the proposed device. Thus, the proposed device EG-600WR v2 is substantially equivalent to its predicate device.

Conclusions:

The modified EG-600WR v2 is substantially equivalent to the predicate EG-600WR and conforms to applicable medical device safety and performance standards.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.