(133 days)
A patient examination glove is disposable non-sterile device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Powder Free Vinyl Patient Examination Gloves (Yellow, White) are Patient Examination Gloves, Disposable, single use only and non-sterile. The gloves are made of vinyl materials and are powder free. The physical and performance characteristics of the devices meet all requirements of ASTM D5250-06 (Reapproved 2015) Standard Specification For Poly (Vinyl Chloride) Gloves For Medical Application.
This document is a 510(k) Pre-Market Notification for Powder Free Vinyl Patient Examination Gloves. It focuses on demonstrating substantial equivalence to a predicate device, rather than proving the device meets acceptance criteria through clinical studies in the typical sense of AI/software devices. The "acceptance criteria" and "study that proves the device meets the acceptance criteria" here refer to compliance with recognized standards and performance thresholds for examination gloves.
Here's a breakdown of the requested information based on the provided text, interpreted in the context of a medical glove submission:
1. A table of acceptance criteria and the reported device performance
| Testing Items | Acceptance Criteria (FDA-recognized Standard Requirements) | Reported Device Performance (Actual Testing Results) | Conclusion |
|---|---|---|---|
| Overall Length (mm) | 230 mm for all sizes minimum | S: 233-244mm, M: 231-241mm, L: 230-243mm, XL: 234-242mm | Pass |
| Width (mm) | S: 85±5, M: 95±5, L: 105±5, XL: 115±5 | S: 87-88 mm, M: 97-98 mm, L: 107-108 mm, XL: 118-119 mm | Pass |
| Palm Thickness (mm) | 0.08mm minimum | 0.08mm | Pass |
| Finger Thickness (mm) | 0.05mm minimum | 0.09-0.12mm | Pass |
| Tensile Strength (Mpa) - Before aging | 11Mpa minimum | 15.1-18.1Mpa | Pass |
| Tensile Strength (Mpa) - After aging | 11Mpa minimum | 14.9-17.7Mpa | Pass |
| Ultimate Elongation (%) - Before aging | 300% minimum | 390-470% | Pass |
| Ultimate Elongation (%) - After aging | 300% minimum | 380-490% | Pass |
| Pinhole | ASTM D 5250-06 (2015) requirements / Meet AQL 2.5 requirements | Meets AQL 2.5 requirements | Pass |
| Residual Powder | Not more than 2mg per glove | 0.18 mg/glove | Pass |
| Primary Skin Irritation Test | Under conditions of the study, not an irritant | Under conditions of the study, not an irritant | Pass |
| Dermal Sensitization Study | Under conditions of the study, not a sensitizer | Under conditions of the study, not a sensitizer | Pass |
| Cytotoxicity Test | Under the conditions of this study, not a cytotoxic potential | Under the conditions of this study, not a cytotoxic potential | Pass |
| Overall Performance | Meets ASTM D5250-06 (Reapproved 2015) Standard Specification For Poly (Vinyl Chloride) Gloves For Medical Application | Conforms fully to ASTM D 5250-06 (2015) standard, applicable 21 CFR references, and FDA requirements | Overall Compliance Demonstrated |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: The sample sizes are indicated by the "Inspection Level and AQL" (Acceptable Quality Limit) for most physical and performance tests, such as S-2, AQL4.0, and G-I, AQL2.5. For Residual Powder, the sample size was N=5. These AQL levels refer to sampling plans (e.g., ISO 2859-1 or ANSI/ASQ Z1.4), where a specific number of units are inspected from a lot, and the lot is accepted or rejected based on the number of defects found. The exact number of units tested isn't explicitly stated but would be determined by the AQL tables for the lot size.
- Data Provenance: The manufacturing entity is Shanxi Hongjin Plastic Technology Co., LTD, located in China. The testing would have been conducted by or on behalf of this company. The data is prospective in the sense that it's generated for the purpose of demonstrating compliance for this submission.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This product is a medical device (patient examination gloves) tested against recognized physical, chemical, and biological standards. "Ground truth" in this context is established by the accepted scientific methodologies and criteria outlined in the ASTM and ISO standards, rather than expert human interpretation in the way it applies to diagnostic AI. Therefore, there are no "experts" in the sense of clinicians establishing ground truth for individual cases. The ground truth is the objective measurement against the specified standard requirements.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Not applicable. Adjudication methods like 2+1 or 3+1 are used for resolving discrepancies in expert interpretations (e.g., in medical image reading). For device performance testing against quantitative standards, the measurements are objective.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is a physical product (medical glove), not an AI or software algorithm. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not relevant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an algorithm or AI device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for the performance tests in this submission is the objective measurement against scientifically established and recognized standards (e.g., ASTM D5250-06 for physical properties, specific thresholds for residual powder, and standard biocompatibility tests for irritation and sensitization). It is not pathology, outcomes data, or expert consensus on a diagnostic interpretation.
8. The sample size for the training set
Not applicable. This is not a machine learning/AI device, so there is no training set in that context. Statistical sampling is used for quality control during manufacturing and for conformity testing.
9. How the ground truth for the training set was established
Not applicable. As there is no training set for a machine learning model, this question is not relevant.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 12, 2018
Shanxi Hongjin Plastic Technology Co., Ltd % Kathy Liu Project Manager Hongray USA Medical Products Inc 3973 Schaefer Ave., Chino, CA 91710 US
Re: K180260
Trade/Device Name: Powder Free Vinyl Patient Examination Gloves (Yellow, White) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: Class I Product Code: LYZ Dated: May 16, 2018 Received: May 30, 2018
Dear Kathy Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Elizabeth F. Claverie -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K180260
Device Name
Powder Free Vinyl Patient Examination Gloves (Yellow, White)
Indications for Use (Describe)
A patient examination glove is disposable non-sterile device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Coal Bed Gas Industrial Zone, Qu'e Town, Daning County, Linfen City, Shanxi Province, China 042399
Product: Powder Free Vinyl Patient Examination Gloves (Yellow, White)
510(K) SUMMARY
The assigned 510(K) number is: K180260
1. Owner's Identification :
Mr. Wu Zhigang Shanxi Hongjin Plastic Technology Co., LTD Coal Bed Gas Industrial Zone, Qu'e Town, Daning County, Linfen City, Shanxi Province, China 042399
Tel: 86-311-66179653 Fax: 86-311-83616934
Contact: Ms. Kathy Liu, Project Manager Address: 3973 Schaefer Ave., Chino, CA 91710 Tel: 909-590-1611 Fax: 909-590-1511 Date Summary Prepared: May 16, 2018
2.Name of the Device:
Trade Name: Powder Free Vinyl Patient Examination Gloves (Yellow, White) Common Name: Exam Gloves Classification Name: Patient Examination Glove Classification Regulation: 880.6250 Classification Panel: 880 General Hospital and Personal Use Product Code: LYZ Device Class: Class I
3. Predicate Device Information:
Ever Light Plastic Products Co., Ltd. Powder Free Vinyl Patient Examination Gloves (Yellow, White, Blue, Pink) (K142570)
4. Device Description:
Powder Free Vinyl Patient Examination Gloves (Yellow, White) are Patient Examination Gloves, Disposable, single use only and non-sterile. The gloves are made of vinyl materials and are powder free. The physical and performance characteristics of the devices meet all requirements of ASTM D5250-06 (Reapproved 2015) Standard Specification For Poly (Vinyl Chloride) Gloves For Medical Application.
{4}------------------------------------------------
Coal Bed Gas Industrial Zone, Qu'e Town, Daning County, Linfen City, Shanxi Province, China 042399
Product: Powder Free Vinyl Patient Examination Gloves (Yellow, White)
5. Indications for Use Statement for these Devices:
A patient examination glove is disposable non-sterile device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
6. Technological Characteristics Comparison:
The properties between the subject device and the predicate device are compared in the following table:
| Characteristics | Device Performance | Result ofcomparison | |
|---|---|---|---|
| Predicate device | Subject Device | ||
| Product Code | LYZ | LYZ | Same |
| Intended Use | Predicate device isdisposable non-steriledevice intended formedical purpose that isworn on the examiner'shand or finger to preventcontamination betweenpatient and examiner. | Subject device isdisposable non-steriledevice intended formedical purpose that isworn on the examiner'shand or finger to preventcontamination betweenpatient and examiner. | Same |
| Labeling | There are no speciallabeling claims | There are no speciallabeling claims | Same |
| Device Materials | Vinyl | Vinyl | Same |
| Color | Yellow White Blue Pink | Yellow White | Different |
| Specifications & Performance Data: | |||
| physical Properties | Meets ASTM D 5250-06(2015) requirements | Meets ASTM D 5250-06(2015) requirements | Same |
| Freedom frompinholes | Meets ASTM D 5250-06(2015) requirements | Meets ASTM D 5250-06(2015) requirements | Same |
| Dimensions: | Meets ASTM D 5250-06(2015) requirements | Meets ASTM D 5250-06(2015) requirements | Same |
| Residual powder | Meets ApplicableDefinition for PowderFree; ≤ 2 mg per glove | Meets ApplicableDefinition for PowderFree; ≤ 2 mg per glove | Same |
| Biocompatibility | |||
| Primary skinirritation test | Under conditions of thestudy, not an irritant | Under conditions of thestudy, not an irritant | Same |
| Dermal | Under conditions of the | Under conditions of the | Same |
{5}------------------------------------------------
Coal Bed Gas Industrial Zone, Qu'e Town, Daning County, Linfen City, Shanxi Province, China 042399
Product: Powder Free Vinyl Patient Examination Gloves (Yellow, White)
| Characteristics | Device Performance | Result ofcomparison | |
|---|---|---|---|
| Predicate device | Subject Device | ||
| sensitizationassay | study, not a sensitizer | study, not a sensitizer |
Shanxi Hongjin Plastic Technology Co., Ltd's Powder Free Vinyl Patient Examination Gloves (Yellow, White) shares the same or comparable technology characteristics compared to the predicate device. The subject device performs according to the glove performance standards ASTM D5250-06(2015), biocompatibility requirement and FDA requirements and the labeling claims for the product.
7. Discussion of Non-Clinical Performance Tests
| Testing Items | FDA-recognizedStandardRequirements | InspectionLevel andAQL | Actual TestingResults | Conclusion | |
|---|---|---|---|---|---|
| Overall Length(mm) | 230 for all sizesminimum | S-2, AQL4.0 | S: 233-244mmM:231-241mmL:230-243mmXL: 234-242mm | Pass | |
| Width(mm) | S: 85±5M: 95±5L: 105±5XL: 115±5 | S-2, AQL4.0 | 87-88 mm97-98 mm107-108 mm118-119 mm | Pass | |
| PalmThicknesss (mm) | 0.08mmminimum | S-2, AQL4.0 | 0.08mm | Pass | |
| FingerThicknesss (mm) | 0.05mmminimum | S-2, AQL4.0 | 0.09-0.12mm | Pass | |
| Tensile Strength (Mpa) | Before agingAfter aging | S-2, AQL4.0 | 11Mpa minimum11Mpa minimum | 15.1-18.1Mpa14.9-17.7Mpa | PassPass |
| Ultimate Elongation (%) | Before agingAfter aging | S-2, AQL4.0 | 300% minimum300% minimum | 390-470%380-490% | PassPass |
| Pinhole | 2.5 | G-I | meet AQL2.5requirements | Pass | |
| Residual | Not more than | N=5 | 0.18 mg/glove | Pass |
{6}------------------------------------------------
Coal Bed Gas Industrial Zone, Qu'e Town, Daning County, Linfen City, Shanxi Province, China 042399
Product: Powder Free Vinyl Patient Examination Gloves (Yellow, White)
| Powder | 2mg per glove | ||
|---|---|---|---|
| Primary SkinIrritation Test | Under conditions of the study, notan irritant | Under conditions of the study, notan irritant | |
| DermalSensitizationStudy | Under conditions of the study, nota sensitizer | Under conditions of the study, nota sensitizer | |
| CytotoxicityTest | Under the conditions of thisstudy, not a cytotoxic potential | Under the conditions of this study,not a cytotoxic potential |
8.Clinical Performance Tests:
Clinical testing is not needed for this device.
9._Conclusion:
Shanxi Hongjin Plastic Technology Co., Ltd's Powder Free Vinyl Patient Examination Gloves (Yellow, White) conform fully to ASTM D 5250-06 (2015) standard as well as applicable 21 CFR references, and meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data discussed above. Based on the performed nonclinical tests, the device is as safe, as effective, and performs as well as the legally marketed predicate device.
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.