K Number
K173955

Validate with FDA (Live)

Date Cleared
2018-01-26

(29 days)

Product Code
Regulation Number
876.5980
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

MIC-KEY* SF Over-the-Wire Stoma Measuring Device is indicated for measuring the length of a stoma prior to placement of a low profile feeding tube.

Device Description

MIC-KEY* SF Over-the-Wire Stoma Measuring Device is designed for measuring the length of a stoma prior to placement of a low profile feeding tube. The Stoma Measuring Device comprises a tubular shaft 10 Fr with graduated scale, an inflation valve and retaining balloon. The MIC-KEY* SF Over the Wire Stoma Measuring Device is made of polyurethane tubing, EtO sterilized and for single use. Like the predicate device, it is intended to be use in hospital environment.

AI/ML Overview

This document describes the non-clinical tests performed on the MIC-KEY* SF Over-the-Wire Stoma Measuring Device to demonstrate its performance and substantial equivalence to a predicate device.

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance CriterionResult
Tensile Strength, Head/Tube≥ 15 N per EN 1615:2000Pass
Balloon Concentricity< 2:1 per ASTM F 2528-06Pass
Deflated Balloon ODPass through 4 Fr size larger ring gauge per ASTM F 2528-06 (without lubricant)Pass
Deflated Balloon ODPass through 2 Fr size larger ring gauge per ASTM F 2528-06 (with lubricant)Pass
Tube Outer DiameterPass through 1 Fr size larger ring gauge per ASTM F 2528-06Pass
Balloon Valve Luer Tip Comp.Able to inflate balloon with a Luer tip syringePass
Printed Ink DurabilityPrinting remains legible following cleaningPass
Device Removal Force≤ 0.48 lbfPass
Balloon System Leak TestNo leakage when the balloon is inflated for five minutesPass
Collar MovementThe collar can slide along the tubing for the entire length of the printed rulerPass
System Leak Test, Cap/HeadNo leakage when pressurized to 0.5 psiPass
Guidewire InsertionDevice can slide along a 0.038" diameter guidewirePass
Device RadiopacityASTM F640-7Pass
Sterile BarrierNo bubbles per ASTM F 2096-11Pass

Biocompatibility tests were also conducted and all met acceptance criteria:

  • ISO 10993-1
  • AAMI ANSI ISO 10993-5
  • ISO 10993-10
  • ISO 10993-11

2. Sample size used for the test set and the data provenance:

The document does not explicitly state the specific sample size for each test. However, it indicates "Performance testing on the device under submission was conducted to demonstrate the modified device continued to meet performance specification." The data provenance is not specified, but these are non-clinical verification tests, likely conducted in a laboratory setting by the manufacturer, Halyard Health, Inc.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

Not applicable. This study focuses on physical and material performance of a medical device, not on expert interpretation of data. The "ground truth" for these tests are the defined engineering and material standards (e.g., N, Fr size, lbf, psi).

4. Adjudication method for the test set:

Not applicable. These are objective engineering and material tests with pass/fail criteria based on established standards. There is no mention of human adjudication for these non-clinical tests.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This device is a physical medical instrument for measuring stoma length, not an AI or diagnostic imaging device that involves human readers.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

Not applicable. This is a physical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The ground truth used for the non-clinical tests are established engineering standards, material specifications, and quality control metrics (e.g., EN 1615:2000, ASTM F 2528-06, ISO 14971:2012, ISO 10993 series).

8. The sample size for the training set:

Not applicable. This is a non-clinical performance study of a physical device, not an AI algorithm requiring a training set.

9. How the ground truth for the training set was established:

Not applicable. There is no training set for this type of device performance study.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 26, 2018

Halyard Health, Inc. David M. Lee, JD Associate Director, Regulatory Affairs 5405 Windward Parkway Alpharetta, GA 30004

Re: K173955

Trade/Device Name: MIC-KEY* SF Over-the-Wire Stoma Measuring Device Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal Tube and Accessories Regulatory Class: II Product Code: KNT Dated: December 26, 2017 Received: December 28, 2017

Dear David M. Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K173955

Device Name

MIC-KEY* SF Over-the-Wire Stoma Measuring Device

Indications for Use (Describe)

MIC-KEY* SF Over-the-Wire Stoma Measuring Device is indicated for measuring the length of a stoma prior to placement of a low profile feeding tube.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) SUMMARY- K173955

Date: January 20, 2018

Applicant's Name, Address, Telephone, Contact Person

Halvard Health. Inc. 5405 Windward Parkway Alpharetta, Georgia 30004

Contact Name: David M. Lee, JD Associate Director, Requlatory Affairs Halvard Health 470 448 5182 (Phone) David.lee@hyh.com

1. CLASSIFICATION, COMMON OR USUAL NAME, DEVICE NAME

Classification:Class II, per 21 CFR 876.5980
Classification Name:Gastrointestinal tube and accessories
Common/Usual Name:Stoma Measuring Device
Product Code:KNT
Trade Name:MIC-KEY* SF Over-the-Wire Stoma Measuring Device

2. PREDICATE DEVICE

MIC-KEY* SF Over the Wire Stoma Measuring Device cleared under the MIC-KEY*LOW PROFILE GASTROSTOMY TUBE AND ACCESSORIES-K122653.

3. INDICATIONS FOR USE

MIC-KEY* SF Over-the-Wire Stoma Measuring Device is indicated for measuring the length of a stoma prior to placement of a low profile feeding tube.

4. DESCRIPTION OF DEVICE

MIC-KEY* SF Over-the-Wire Stoma Measuring Device is designed for measuring the length of a stoma prior to placement of a low profile feeding tube. The Stoma Measuring Device comprises a tubular shaft 10 Fr with graduated scale, an inflation valve and retaining balloon. The MIC-KEY* SF Over the Wire Stoma Measuring Device is made of polyurethane tubing, EtO sterilized and for single use. Like the predicate device, it is intended to be use in hospital environment.

5. TECHNOLOGICAL CHARACTERISTICS

The MIC-KEY* SF Over-the-Wire Stoma Measuring Device is substantially equivalent to the predicate device. The following changes from the predicate device were made:

  • . the printed device labeling (printed marking was added to the polyurethane molded head indicating the balloon inflation volume)
  • minor changes to the geometry of the molded polyurethane head ●
  • minor change to the geometry of the molded tethered cap ●
  • the ink used for device marking was changed o

There are no other significant differences in the design, materials, performance, and technological characteristics from the predicate device.

{4}------------------------------------------------

6. SUMMARY OF NONCLINICAL TESTS

Performance testing on the device under submission was conducted to demonstrate the modified device continued to meet performance specification. Results of design verification and validation activities did not raise any new issues of safety or effectiveness. Non-clinical verification was conducted through the risk management process according to ISO 14971:2012. The following verification tests were conducted:

TestAcceptance CriterionResult
Tensile Strength, Head/Tube≥ 15 N per EN 1615:2000Pass
Balloon Concentricity< 2:1 per ASTM F 2528-06Pass
Deflated Balloon Outer DiameterPass through 4 Fr size larger ring gaugeper ASTM F 2528-06 (without lubricant)Pass
Deflated Balloon Outer DiameterPass through 2 Fr size larger ring gaugeper ASTM F 2528-06 (with lubricant)Pass
Tube Outer DiameterPass through 1 Fr size larger ring gaugeper ASTM F 2528-06Pass
Balloon Valve Luer Tip CompatibilityAble to inflate balloon with a Luer tipsyringePass
Printed Ink DurabilityPrinting remains legible following cleaningPass
Device Removal Force≤ 0.48 lbfPass
Balloon System Leak TestNo leakage when the balloon is inflatedfor five minutesPass
Collar MovementThe collar can slide along the tubing forthe entire length of the printed rulerPass
System Leak Test, Cap/HeadNo leakage when pressurized to 0.5 psiPass
Guidewire InsertionDevice can slide along a 0.038" diameterguidewirePass
Device RadiopacityASTM F640-7Pass
Sterile BarrierNo bubbles per ASTM F 2096-11Pass

The following biocompatibility tests were also conducted, and all met acceptance criteria:

ISO 10993-1 Fourth edition 2009-10-15 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process

AAMI ANSI ISO 10993-5, 2009: Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity

ISO 10993-10:2010: Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization

ISO 10993-11:2006/(R2010): Biological evaluation of medical devices - Part 11: Tests for systemic toxicity

7. OVERALL PERFORMANCE CONCLUSIONS

The above nonclinical studies demonstrate MIC-KEY* SF Over-the-Wire Stoma Measuring Device is substantial equivalence to the predicate MIC-KEY* SF Over-the-Wire Stoma Measuring Device (K122653) in intended use, design, materials, performance, and biocompatibility attributes. There are no new questions of safety and effectiveness as compared to the predicate device.

§ 876.5980 Gastrointestinal tube and accessories.

(a)
Identification. A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).(b)
Classification. (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.