(18 days)
The TBL-ABI is indicated for use on adult subjects at risk of having or developing peripheral arterial disease (PAD). TBL-ABI is intended for rapid measurement of ankle-brachial pressure index (ABPI) or ankle-brachial index (ABI) in adults and pulse volume recording (PVR) / volume plethysmography in adults.
It is suitable for use in wound care assessing symptomatic PAD, and as a screening device for PAD. It may also be used on patients with venous or arterial ulcers prior to the application of compression therapy.
TBL-ABI can be used on patients with unilateral lower limb amputation.
The TBL-ABI System is intended to be used to spot-check patients. The TBL-ABI provides information regarding patient risk. The physician has the responsibility of making proper judgments based on this information.
TBL-ABI System comprises:
- 3 Bluetooth blood pressure devices with attached cuffs.
- Software installed on a computer
It is intended to measure a patient's Ankle Brachial Index (ABI) and provide Pulse Volume Recording (PVR) / volume plethysmography.
This is done through an automated process.
The operator places the three devices with different color-coded electronic boxes on the right or left arm, and on each leg as described in the instructions for use, and the devices connect to the computer via Bluetooth.
Once the devices are connected, the operator clicks start on the software to start the measurement. The devices will then automatically control the inflation of the cuffs and monitor the variations in individual pressures to determine values to be used for the calculation of the ABI values for both the left and right of the patient.
TBL-ABI uses pneumo-plethysmography in order to obtain physiologic measurements from the patient's limbs.
Measurements are conducted as a single occurrence on the three limbs, thus eliminating any requirement to rest the patient between measurements.
The test period takes approximately 3 minutes.
ABI values, as well as the Pulse Volume recording (PVR), are displayed on the software installed in a computer. The results are saved in a backup file and can also be printed.
The TBL-ABI System, a device for measuring Ankle-Brachial Index (ABI) and Pulse Volume Recording (PVR), was reviewed for substantial equivalence to a legally marketed predicate device (TM-ABI, K143152). The provided document is a 510(k) summary, which typically presents a summary of verification and validation testing, rather than detailed study protocols and results for clinical performance. Therefore, information regarding human reader studies, ground truth establishment for clinical data, and specific sample sizes for clinical accuracy testing (if any were performed on human subjects) is not extensively detailed.
However, based on the provided text, the device primarily relies on non-clinical performance and engineering validation to demonstrate substantial equivalence, particularly concerning the accuracy of its physiological measurements and the safety of its technological changes (e.g., Bluetooth communication).
Here's an analysis of the acceptance criteria and the non-clinical study that proves the device meets these criteria, as described in the document:
1. Table of Acceptance Criteria (Performance Specifications) and Reported Device Performance (as claimed by the manufacturer and deemed sufficient by FDA):
| Acceptance Criteria (Performance Specification) | TBL-ABI Reported Device Performance | Notes on Acceptance Basis |
|---|---|---|
| Measurement Ranges: | ||
| Pressure: 0 to 270 mmHg | Pressure: 0 to 270 mmHg | Meets its specified range. The predicate device had a slightly higher maximum pressure (299 mmHg), but this difference was deemed acceptable and not to affect safety or effectiveness. |
| Limit values of Measurement Errors: | ||
| Pressure: ± 3mmHg | Pressure: ± 3mmHg | Matches the predicate device's accuracy specification, indicating equivalent physiological measurement accuracy. |
| Ankle Brachial Pressure Index: ± 0.1 | Ankle Brachial Pressure Index: ± 0.1 | Matches the predicate device's accuracy specification for ABI, indicating equivalent physiological measurement accuracy. |
| Operating Environment: | ||
| Working environment: 5 to 40°C, 15 to 80% relative air humidity | Working environment: 5 to 40°C, 15 to 80% relative air humidity | Meets its specified range. Differences from the predicate device (TM-ABI: 10 to 40°C, 30 to 85% relative air humidity) were assessed and found not to affect safety or effectiveness. |
| Transport and storage: -20 to 55°C, up to 85% relative air humidity | Transport and storage: -20 to 55°C, up to 85% relative air humidity | Meets its specified range. Differences from the predicate device (TM-ABI: 0 to 60°C, up to 85% relative air humidity) were assessed and found not to affect safety or effectiveness. |
| Electrical Safety: | ||
| Compliant with standards: 60601-1, 60601-1-2, 80601-2-30 | Compliant with standards: 60601-1, 60601-1-2, 80601-2-30 | Confirms adherence to essential medical electrical equipment and particular sphygmomanometer standards, ensuring safety. |
| Type of protection against electric shock: Class II, BF | Type of protection against electric shock: Class II, BF | Indicates compliance with electrical safety classifications. |
| Biocompatibility: | ||
| Cuffs made of lycra fabric, raised fabric, compliant with ISO 10993-10: 2010, ISO 10993-10: 2012, and ISO 10993-5: 2009 | Cuffs made of lycra fabric, raised fabric, compliant with ISO 10993-10: 2010, ISO 10993-10: 2012, and ISO 10993-5: 2009 | Ensures patient safety regarding material contact through adherence to biocompatibility standards. |
| Software Performance: | ||
| Accurately capture, store, and analyze the data measured by the hardware. | CRC (Cyclic redundancy check) Testing performed. | Software verification (SRS/SDS/STD/STR) was performed to ensure proper data handling and functionality. Bluetooth communication was tested and proven safe. |
| Cybersecurity: | ||
| Assessed against cybersecurity risks. | Cybersecurity risk assessment performed. | Ensures the device's system is secure, especially with the new Bluetooth communication. |
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size: The document does not specify a "test set" in terms of human subjects or patient data for clinical performance evaluation. The study primarily relies on non-clinical testing (bench testing, engineering validation, software verification). The performance claims related to pressure accuracy and ABI accuracy (± 3mmHg and ± 0.1 respectively) are presented as specifications that are met, likely through calibration and validation against established standards or measurement principles rather than a patient-based test set.
- Data Provenance: Not applicable as the "study" described is verification and validation, predominantly on the device itself and its components. The document implies comparison to the predicate device's established performance, rather than new clinical data generation.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
- Not applicable. As the evaluation focused on non-clinical performance and substantial equivalence to a predicate device with established performance specifications, "ground truth" was likely established through engineering standards, calibration references, and predicate device specifications. No mention of expert radiologists or similar clinical experts is found for establishing ground truth from patient data.
4. Adjudication Method for the Test Set:
- Not applicable, as the evaluation focused on non-clinical performance testing.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
- No, a MRMC comparative effectiveness study was not done. The device is a diagnostic tool that provides objective measurements (ABI, PVR) and does not involve human interpretation of images or signals that would typically necessitate an MRMC study to show human reader improvement with AI assistance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- The device inherently operates as a "standalone" algorithm in that it performs automated measurements and calculations (Ankle-Brachial Index and Pulse Volume Recordings). The performance specifications for pressure and ABI accuracy (± 3mmHg, ± 0.1 ABI respectively) represent the algorithm's performance. The "study" (non-clinical testing) demonstrated that the device (including its algorithm for measurement calculation) meets these specified accuracy limits.
7. The Type of Ground Truth Used:
- For the core physiological measurements (pressure, ABI), the ground truth is implicitly based on metrological standards and established physiological measurement accuracy limits (e.g., ± 3mmHg for blood pressure, which is a widely accepted standard for non-invasive blood pressure devices). The device's performance is claimed to meet these standard accuracy limits, similar to its predicate device.
- For software and hardware functionality, the ground truth is derived from engineering specifications, test protocols (SRS/SDS/STD/STR), and standard compliance (e.g., IEC 60601 series).
- For biocompatibility, the ground truth is established by ISO 10993 standards.
8. The Sample Size for the Training Set:
- Not applicable. The document describes a 510(k) submission for substantial equivalence. This type of submission relies on demonstrating that a new device is as safe and effective as a legally marketed predicate device, often through a combination of non-clinical testing and comparison of technological characteristics. It does not typically involve training a machine learning model, which would require a training set. The device performs calculations based on established physical principles (pneumo-plethysmography), not learned patterns from data.
9. How the Ground Truth for the Training Set Was Established:
- Not applicable, as there was no training set for a machine learning model.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below that.
December 19, 2017
LD Technology LLC Albert Maarek Quality Manager 100 N. Biscayne Blvd Suite 502 Miami. Florida 33132
Re: K173696
Trade/Device Name: TBL-ABI System Regulation Number: 21 CFR 870.2780 Regulation Name: Hydraulic, Pneumatic, Or Photoelectric Plethysmographs Regulatory Class: Class II Product Code: JOM Dated: November 3, 2017 Received: December 1, 2017
Dear Albert Maarek:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
M.A. Hilleman
for Bram D. Zuckerman, M.D.
Director Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number K173696
Device Name: TBL-ABI System
Indications for Use (Describe)
The TBL-ABI is indicated for use on adult subjects at risk of having or developing peripheral arterial disease (PAD). TBL-ABI is intended for rapid measurement of ankle-brachial pressure index (ABPI) or ankle-brachial index (ABI) in adults and pulse volume recording (PVR) / volume plethysmography in adults.
It is suitable for use in wound care assessing symptomatic PAD, and as a screening device for PAD. It may also be used on patients with venous or arterial ulcers prior to the application of compression therapy.
TBL-ABI can be used on patients with unilateral lower limb amputation.
The TBL-ABI System is intended to be used to spot-check patients. The TBL-ABI provides information regarding patient risk. The physician has the responsibility of making proper judgments based on this information.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
K173696
Image /page/3/Picture/1 description: The image shows a logo for LD Technology. The logo consists of a circular emblem on the left and the text "LD Technology" on the right. The emblem features a green sphere with plant-like structures extending from it, all enclosed within a circle. The text "LD Technology" is written in a light gray color, with the "LD" appearing larger and slightly above the word "Technology".
510(k) Premarket Notification Number: 173696 Date of preparation: November 04 .2017
510(k) Summary TBL-ABI device
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92
1. Submitter's Identification: Manufacturer: L.D TECHNOLOGY Address:
100 N. Biscayne Blvd. Suite 502 Miami, FL, 33132, USA Tel: 305-379-9900 E mail: albert.ldteck@gmail.com
2. Device Name / Classification
Trade name: Patient monitor Device Name and Model: TBL-ABI Regulation number: 21CFR 870.2780 Product Code: JOM Device Class: Class II Classification Name: Hydraulic, pneumatic, or photoelectric plethysmograph Classification Panel: Cardiology
3. Predicate legally marketed devices
TM-ABI K143152 Applicant: LD TECHNOLOGY LLC. Product Code: JOM.
4. Device Description
TBL-ABI System comprises:
- 2 3 Bluetooth blood pressure devices with attached cuffs.
- ∑ Software installed on a computer
It is intended to measure a patient's Ankle Brachial Index (ABI) and provide Pulse Volume Recording (PVR) / volume plethysmography.
This is done through an automated process.
The operator places the three devices with different color-coded electronic boxes on the right or left arm, and on each leg as described in the instructions for use, and the devices connect to the computer via Bluetooth.
Once the devices are connected, the operator clicks start on the software to start the measurement. The devices will then automatically control the inflation of the cuffs and monitor the variations in individual pressures to determine values to be used for the calculation of the ABI
{4}------------------------------------------------
values for both the left and right of the patient.
TBL-ABI uses pneumo-plethysmography in order to obtain physiologic measurements from the patient's limbs.
Measurements are conducted as a single occurrence on the three limbs, thus eliminating any requirement to rest the patient between measurements.
The test period takes approximately 3 minutes.
ABI values, as well as the Pulse Volume recording (PVR), are displayed on the software installed in a computer. The results are saved in a backup file and can also be printed.
5. Intended use and indications for use
The TBL-ABI is indicated for use on adult subjects at risk of having or developing peripheral arterial disease (PAD).
TBL-ABI is intended for the rapid measurement of ankle-brachial pressure index (ABPI), or ankle-brachial index (ABI), and pulse volume recording (PVR) / volume plethysmography in adults. It is suitable for use in wound care assessment, for assessing symptomatic PAD, and as a screening device for PAD. It may also be used on patients with venous or arterial ulcers prior to the application of compression therapy.
TBL-ABI can be used on patients with unilateral lower limb amputation.
The TBL-ABI System is intended to be used to spot-check patients. The TBL-ABI provides information regarding patient risk. The physician has the responsibility of making proper iudgments based on this information.
*Prescription Use: Federal law restricts the use of this device to sale by or on the order of a physician
6. Performances, technical specifications and materials Performances
TBL-ABI comprises 3 portable medical devices, powered by internal Li-Po battery. Each device has a rechargeable battery, and a Bluetooth connection with the software installed in a computer. Each device has a different box color.
The front panel of each device has an on/off button to connect the Bluetooth.,
The main purpose of the TBL-ABI is to measure ABI, SYS and DIA of all limbs. It also allows for checking measurement history.
Technical specifications i.e. table of comparison with the predicate device Patient contact materials:
The materials in contact with the patient are the cuffs. The Cuff material is lycra fabric, raised fabric.
7. Contra-indications
- Σ Patients undergoing external defibrillation.
-
Patients connected to electronic life support devices, or any implanted electronic device.
- 网 Patients moving or undergoing long term monitoring.
-
Do not use this device in the presence of:
-
Magnetic resonance imaging (MR or MRI) equipment. MRI equipment may deliver induced current to the device.
-
{5}------------------------------------------------
-
Strong electromagnetic sources, such as electro surgery equipment.
-
Computed Tomography (CT) equipment.
-
WIFI router at distance less than 10 feet and in the same room.
-
Cellphones at distance less than 10 feet.
-
Microwaves ovens, cables and connectors associated with satellite dishes, nearby power
-
lines or power stations, cordless telephones, wireless speakers, some monitors and
-
displays, cameras, baby monitors.
The distance between hardware and the computer must be less than 10 feet and in the same room.
- ∑ Arterial catheters, or with an AV fistula or pressure dressing.
- 2 Venous pulsations may cause erroneous reading in blood pressure (e.g. tricuspid valve regurgitation).
- C Take caution with patients that have low perfusion. Using the blood pressure device may cause skin erosion and/or pressure necrosis.
8. Undesirable side effects:
No side effects or adverse reactions are known to date.
9. Substantial equivalence
Predicate legally marketed device: TM-ABI, K143152. Applicant: LD TECHNOLOGY LLC. Product Code: JOM.
Substantial Equivalence Discussion
-
Step 1: Does the new device have the same indication statement? Yes
Device Modifications are not affected the intended use of the device. -
V Step 3: The new device has the same intended use and may be "substantially equivalent." Yes
-
Step 5: Does the new device have the same technological characteristics, e.g. design, materials, etc.?
Similarities:
- ✔ Same intended use
- Scientific Background (oscillometric measurement of the blood pressure and PVR)
- ✔ Placement of the cuff
- V Power supply using a rechargeable battery
- Classification Degree of protection against electric shocks
- V Operating mode
Device Modifications:
- √ The TBL-ABI and predicate device have different dimension and weight
- The TBL-ABI and predicate device use a battery with different voltage and intensity rate.
{6}------------------------------------------------
- √ The TBL-ABI cuffs are in different material and are directly attached to each device versus cuffs connected with tubes for TM-ABI.
- ✔ The TBL-ABI has a maximal pressure of 270 mmHg versus 299 mmHg for TM-ABI
- V The TM-ABI has small difference in temperature and humidity conditions for storage and working environment
- √ The TBL-ABI uses 3 Bluetooth devices versus one USB connection device for TM-ABI
Rationale for the Device modifications:
-
Bluetooth communication
- √ Cuffs attached to 3 electronic boxes with pump Allow to remove the tubes and the patient feel more comfortable and it is more convenient for the technician.
Step 6: If it has new technological characteristics, could they affect safety or effectiveness? Discussion why any differences between the subject and predicate(s) do not impact safetv and effectiveness [see section 513(i) (1) (A) of the FD&C Act and 21 CFR 807.87(f)] See "
Does the change in communication to the software affect the safety or effectiveness?
The Bluetooth communication is safe, and we use the same CRC coding test as the previous USB communication. CRC (Cyclic redundancy check) Coding was performing to demonstrate the software performance to accurately capture, store, and analyze the data measured by the hardware. Testing of the Bluetooth has been performed.
Do the other changes affect the safety or effectiveness?
The fact that the TBL-ABI has:
- √ A different dimension and weight
- V Battery with lower voltage and intensity rate
-
Different cuff material and cuffs directly attached to the device
-
A small lower maximal blood pressure value
- √ A small difference in temperature and humidity environment
do not affect the performances and the effectiveness of the TBL-ABI system compared to the predicate device as shown by the laboratory testing (60601-1-1-1, 60601-1-2, IEC 80601-2-30. ISO 10993 and IEC 62133) and testing performed comparing the results of 2 devices as well as the cybersecurity risk assessment.
{7}------------------------------------------------
Table of comparison
| Information | TBL-ABI | TM-ABI K143152 |
|---|---|---|
| Intended use | The TBL-ABI is indicated for use on adultsubjects at risk of having or developingperipheral arterial disease (PAD).TBL-ABI is intended for the rapid measurementof ankle-brachial pressure index (ABPI), orankle-brachial index (ABI), and pulse volumerecording (PVR) / volume plethysmography inadults. It is suitable for use in wound careassessment, for assessing symptomatic PAD, andas a screening device for PAD. It may also beused on patients with venous or arterial ulcersprior to the application of compression therapy.TBL-ABI can be used on patients with unilaterallower limb amputation. | The TM-ABI is indicated for use on adultsubjects at risk of having or developingperipheral arterial disease (PAD).TM-ABI is intended for the rapid measurementof ankle-brachial pressure index (ABPI), orankle-brachial index (ABI), and pulse volumerecording (PVR) / volume plethysmography inadults. It is suitable for use in wound careassessment, for assessing symptomatic PAD,and as a screening device for PAD. It may alsobe used on patients with venous or arterialulcers prior to the application of compressiontherapy.TM-ABI can be used on patients with unilaterallower limb amputation. |
| Dimensions/weight | For each device: Width: 60mm, height: 30mm,depth:132mm Weight: 260g | width: 250 mm, height: 730mm, depth: 200 mm,weight: 0.60kg |
| Power Supply | Output: 3.7 VDC. Battery type: rechargeablelithium polymerCapacity: 530 mAh,number of measurements per charge: 60 | Output: 5VDC. Battery type: rechargeable lithiumpolymerCapacity: 2,300mAh,number of measurements per charge: 30 |
| Applied parts in contactwith the patient | 3 cuffs lycra fabric raised fabric ISO 10993- 102010. ISO 10993-10: 2012 andISO 10993-5: 2009 | 3 cuffs, nylon and bladders ISO 10993-10: 2010 |
| Testing Bench | Type of protection against electric shock: ClassII. BFCompliant with standards:60601-1,60601-1-280601-2-30. | Type of protection against electric shock: ClassII. BFCompliant with standards:60601-1,60601-1-280601-2-30. |
| Measurement types | Using the pneumo-plethysmographic method:Right and left Ankle brachial pressure indexSystolic bloodDiastolic blood pressure | Using the pneumo-plethysmographic method:Right and left Ankle brachial pressure indexSystolic bloodDiastolic blood pressure |
| Measurement ranges | Pressure: 0 to 270 mmHg | Pressure: 0 to 299 mmHg |
| Limit values ofmeasurement errors | Pressure: ± 3mmhgAnkle brachial pressure index: ± 0.1 | Pressure: ± 3mmhgAnkle brachial pressure index: ± 0.1 |
| Cuffs inflation and deflation | Automatic inflation using an air pump and deflation using an electromagnetic valve. | Automatic inflation using an air pump and deflation using an electromagnetic valve. |
| Temperature and humidity range | working environment: 5 to 40°C, 15 to 80%relative air humidity, IPX21 protection,transport and storage: -20 to 55°C, up to 85%relative air humidity. | working environment: 10 to 40°C, 30 to 85%relative air humidity, IPX0 protection,transport and storage: 0 to 60°C, up to 85%relative air humidity. |
| Target population | Adult | Adult |
| Where used | Clinical environment | Clinical environment |
| PC Data transmission | Bluetooth version 4.0 | USB |
{8}------------------------------------------------
10. Non-clinical, Performances and Effectiveness
CRC (Cyclic redundancy check) Coding was performing to demonstrate the software performance to accurately capture, store, and analyze the data measured by the hardware. Testing comprises:
-
- Calibration tests
-
- Bluetooth testing and Cybersecurity risk design
-
- Laboratory tests
-
- Software verification (SRS/SDS/STD/STR)
11. General Safety Concerns
The laboratory test reports of the TBL-ABI System and Bluetooth communication technology testing have demonstrated the general safety of the system compared to the legally marketed predicate device.
12. Standards
6 | 7
- V IEC 60601-1: Medical electrical equipment Part 1: General requirements for basic safety and essential performance. Third Edition December 2006
-
IEC60601-1-2: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests. Third Edition 05/17/2007
-
IEC 80601-2-30: Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers. 10/31/2010 Second Edition
-
ISO 10993-10: Tests for irritation and skin sensitization.2010 and 2012
-
ISO 10993-5: Biological evaluation of medical devices. 2009
-
IEC 62133: Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications. 2nd Edition.2012.
-
Guidance for: Industry; FDA staff; and Content of Premarket Submissions for Software Contained in Medical Devices Document issued on: May 11, 2005
-
Wireless Medical device FDA Guidance. 2014.
{9}------------------------------------------------
- ✔ Cybersecurity FDA Guidance. 2013.
- V ISO 14971: Medical devices Application of risk management to medical devices. March 01, 2007
Conclusion
TBL-ABI is equivalent in performance, technology, safety and efficacy to the legally marketed predicate device. Signature:
Signature:
Albert MAAREK
scribble
Premarket notification [510K] Number: 173696
§ 870.2780 Hydraulic, pneumatic, or photoelectric plethysmographs.
(a)
Identification. A hydraulic, pneumatic, or photoelectric plethysmograph is a device used to estimate blood flow in a region of the body using hydraulic, pneumatic, or photoelectric measurement techniques.(b)
Classification. Class II (performance standards).