(89 days)
The Celerity 20 Steam Challenge Pack is used for qualification, routine microbial monitoring, and load monitoring of steam sterilizers.
The validated steam sterilization cycles include:
- 250°F (121°C) 30-minute gravity .
- 270°F (132°C) 15-minute gravity .
Celerity 20 Steam Process Challenge Device for Gravity Cycles (PCD), contains a Celerity 20 Steam Biological Indicator (SCBI) and a steam chemical integrator, sealed within a plastic tray with foil cover. The tray has a molded channel which serves as the tortuous pathway for air removal/steam penetration during steam sterilization. The PCD is designed and validated to be equivalent to the standard 16-towel test pack described in ANSI/AAMI ST79. The proposed and predicate are identical except for the BI in the PCD.
The provided text describes the CELERITY 20 Steam Process Challenge Device for Gravity Cycles, a medical device for monitoring steam sterilizers. The document is a 510(k) premarket notification, indicating a claim of substantial equivalence to a predicate device.
Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:
1. A table of acceptance criteria and the reported device performance:
| Test | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Simulated Use - BI Performance | Performance of the BI (Biological Indicator) in the PCD (Process Challenge Device) is equivalent to the performance of the BI in the AAMI (Association for the Advancement of Medical Instrumentation) reference pack in their respective sterilization processes. | PASS |
| Simulated Use - Chemical Integrator Performance | Performance of the chemical integrator in the PCD is equivalent to the performance of the chemical integrator in AAMI reference pack in their respective processes. | PASS |
| Simulated Use - Challenge Level | PCD provides an equivalent or greater challenge than the AAMI standardized test pack. | PASS |
| BI in pack vs BI outside of pack | PCD provides a greater challenge to the process than the BI itself. | PASS |
| CI in pack vs CI outside of pack | PCD provides a greater challenge to the process than the integrator by itself. | PASS |
| Chemical Integrator Timing | Chemical integrator does not reach endpoint before BI is inactivated. | PASS |
2. Sample size used for the test set and the data provenance:
The document mentions "Performance testing" and "nonclinical tests" but does not specify the sample size used for the test set of the device.
The data provenance is not explicitly stated in terms of country of origin or whether it was retrospective or prospective. However, given that this is a 510(k) submission for a new device claiming substantial equivalence, the testing would typically be prospective, and conducted by the manufacturer (STERIS Corporation, based in Mentor, Ohio, USA).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
This information is not provided in the document. The tests seem to rely on established physical/chemical principles and AAMI standards for sterilization challenges, rather than on expert human interpretation of results in the way image-based diagnostics might.
4. Adjudication method for the test set:
This information is not applicable to the type of device and testing described. The "adjudication method" usually refers to how discrepancies in expert ratings are resolved for ground truth in diagnostic studies. Here, the "ground truth" is determined by the failure or success of sterilization (based on BI growth or chemical integrator change) in defined challenge conditions.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This information is not applicable to this device. An MRMC study is relevant for diagnostic devices (especially AI-powered ones) where human interpretation is involved. The CELERITY 20 is a process challenge device, and its performance is assessed by biological and chemical indicators reacting to sterilization conditions, not by human readers interpreting images or data for diagnostic purposes.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
This information is not applicable to this device. There is no AI algorithm involved in the function or interpretation of this sterilization process challenge device. Its performance is inherent to its physical and biological components.
7. The type of ground truth used:
The ground truth used for this device's performance evaluation is based on biological inactivation (no growth of the Geobacillus stearothermophilus spores in the biological indicator) and chemical reaction (the appropriate change in the chemical integrator) under specified sterilization conditions, relative to established standards like the AAMI reference pack.
Essentially, the ground truth is:
- AAMI standardized test pack performance: The CELERITY 20 PCD is compared against the performance of the standard 16-towel test pack described in ANSI/AAMI ST79.
- Biological indicator viability: Absence of microbial growth after incubation signifies a successful sterilization cycle and inactivation of the biological indicator. Presence of growth signifies failure.
- Chemical integrator endpoint: The chemical integrator changes visibly to indicate exposure to sterilization conditions.
8. The sample size for the training set:
This information is not applicable as this device does not involve machine learning or a "training set."
9. How the ground truth for the training set was established:
This information is not applicable as there is no "training set" for this device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
February 21, 2018
STERIS Corporation Anthony Piotrkowski 5976 Heislev Rd Mentor, Ohio 44060
Re: K173629
Trade/Device Name: CELERITY 20 Steam Process Challenge Device for Gravity Cycles Regulation Number: 21 CFR 880,2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: FRC Dated: November 22, 2017 Received: November 24, 2017
Dear Anthony Piotrkowski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known)
K173629
Device Name
CELERITY 20 Steam Process Challenge Device for Gravity Cycles
IThe Celerity 20 Steam Challenge Pack is used for qualification, routine microbial monitoring, and load monitoring of steam sterilizers.
The validated steam sterilization cycles include:
- 250°F (121°C) 30-minute gravity .
- 270°F (132°C) 15-minute gravity .
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
[X] Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for STERIS. The logo consists of the word "STERIS" in a bold, sans-serif font, with the registered trademark symbol to the right of the word. Below the word is a graphic of several horizontal, wavy lines in blue, resembling water waves.
510(k) Summary For CELERITY 20 Steam Process Challenge Device for Gravity Cycles
Sponsor Facility
STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 357-9198
Manufacturing Facility
STERIS Corporation 9325 Pinecone Drive Mentor, OH 44060 Phone: (440) 392-7800 Fax No: (440) 392-7896
Tony Piotrkowski Contact:
Telephone: (440) 392-7437 Fax No: e-mail: tony_piotrkowski.@steris.com
January 23, 3018 Submission Date:
Premarket Notification Number: K173629
STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600
{4}------------------------------------------------
Device Name 1.
| Trade Name: | CELERITY 20 Steam Process Challenge Device forGravity Cycles |
|---|---|
| Common/usual Name: | Biological Indicator Pack (PCD) |
| Device Classification: | Class II |
| Classification Name: | Sterilization Process Indicator(21 CFR 880.2800, FRC) |
2. Predicate Device
VERIFY Assert STEAM Process Challenge Device for Gravity Cycles, K170070
3. Description of Device
Celerity 20 Steam Process Challenge Device for Gravity Cycles (PCD), contains a Celerity 20 Steam Biological Indicator (SCBI) and a steam chemical integrator, sealed within a plastic tray with foil cover. The tray has a molded channel which serves as the tortuous pathway for air removal/steam penetration during steam sterilization. The PCD is designed and validated to be equivalent to the standard 16-towel test pack described in ANSI/AAMI ST79. The proposed and predicate are identical except for the BI in the PCD.
4. Intended Use/ Indications for Use
The Celerity 20 Steam Challenge Pack is used for qualification, routine microbial monitoring, and load monitoring of steam sterilizers.
The validated steam sterilization cycles include:
- 250°F (121°C) 30-minute gravity
- . 270°F (132°C) 15-minute gravity
ട. Summary of Technical Characteristics
A comparison of technical characteristics are summarized in Table 5-1.
{5}------------------------------------------------
K173629 STERIS TRADITIONAL 510(k) PREMARKET NOTIFICATION Celerity 20 Steam Process Challenge Device for Gravity Cycles
| Feature | Celerity Gravity PCD(K173629) Proposed | Assert Gravity PCD(K170070) Predicate | Comparison |
|---|---|---|---|
| IntendedUse | The Celerity 20 SteamChallenge Pack is usedfor qualification, routinemicrobial monitoring,and load monitoring ofsteam sterilizers.The validated steamsterilization cyclesinclude:· 250°F (121°C) 30-minute gravity· 270°F (132°C) 15-minute gravity | The VERIFY™ASSERT™ STEAMProcess Challenge Devicefor Gravity Cycles is usedfor qualification, routinemicrobial monitoring, andcycle monitoring of steamsterilizers.The validated steamsterilization cyclesinclude:· 250°F (121°C) 30-minute gravity· 270°F (132°C) 15-minute gravityThe VERIFY Assert Self-Contained Indicator withinthe Process challengedevice must be used withthe VERIFY Incubator forAssert Self ContainedBiological Indicators.When used in conjunctionwith the VERIFYIncubator for Assert SelfContained BiologicalIndicators, the VERIFYAssert Self-ContainedIndicator within theProcess challenge deviceprovides a fluorescentresult within 40 minutes. | Both are intended formonitoring steamsterilization cycles. |
| GeneralDesign | Indicators are sealed inplastic tray with channelto limit steampenetration/air removal. | Indicators are sealed inplastic tray with channelto limit steampenetration/air removal. | The PCD housing isidentical for the twodevices |
| BiologicalIndicator | Celerity 20 SteamBiological Indicator(20-minute fluorescentresult) | Verify Assert SelfContained BiologicalIndicator (40-minutefluorescent result) | Both contain 510(k)Biological Indicators withrapid fluorescent results. |
| ChemicalIntegrator | Chemical Integrator | Chemical Integrator | Both contain the samechemical integrator |
| Means todistinguishprocessedPCD fromunprocessed | Proposed device'sinternal integrator isvisible through thepack. | Proposed device'sinternal integrator isvisible through thepack. | Same |
Table 5-1. Summary of PCD Physical Description and Technological Properties
{6}------------------------------------------------
Any differences between the predicate and proposed device are addressed through the testing described in Section 18 of this submission.
6. Summary of Nonclinical Tests
Performance testing to demonstrate substantial equivalence to the predicate has been completed and is summarized in Table 5-2 below.
| Test | Acceptance Criteria | Conclusion |
|---|---|---|
| Simulated Use | • Performance of the BI in the PCD is equivalent tothe performance of the BI in the AAMI referencepack in their respective sterilization processes | PASS |
| • Performance of the chemical integrator in the PCDis equivalent to the performance of the chemicalintegrator in AAMI reference pack in their | PASS | |
| respective processesPCD provides an equivalent or greater challengethan the AAMI standardized test pack | PASS | |
| BI in pack vs BIoutside of pack | PCD provides a greater challenge to the process thanthe BI itself. | PASS |
| CI in pack vs CIoutside of pack | PCD provides a greater challenge to the process thanthe integrator by itself. | PASS |
| Chemical Integrator | Chemical integrator does not reach endpoint before BIis inactivated. | PASS |
Table 5-2. Summary of Non-clinical Testing
7. Conclusion
The Celerity 20 Steam Process Challenge Device for Gravity Cycles has met the established performance criteria. The results of the studies demonstrate that the biological indicator performs as intended, and based on the nonclinical tests performed, the subject device is as safe, as effective, and performs as the legally marketed predicate device, VERIFY Assert STEAM Process Challenge Device for Gravity Cycles, cleared in K170070 (21 CFR 880.2800, Product code FRC).
§ 880.2800 Sterilization process indicator.
(a)
Biological sterilization process indicator —(1)Identification. A biological sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization.(2)
Classification. Class II (performance standards).(b)
Physical/chemical sterilization process indicator —(1)Identification. A physical/chemical sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.(2)
Classification. Class II (performance standards).