(179 days)
Not Found
No
The description focuses on standard software tools for image overlay, measurement, and implant positioning, without mentioning any AI/ML capabilities. The "Mentions AI, DNN, or ML" section explicitly states "Not Found".
No
The device is indicated for preoperative planning of total hip arthroplasty, and it assists healthcare professionals in selecting and positioning hip implant models. It does not directly treat or alleviate a medical condition.
No
Explanation: The device is described as assisting in "preoperative planning of total hip arthroplasty (THA)" by allowing "overlaying of 3D/2D hip implant models on radiological images" and including "tools for performing measurements." It plans interventions and exports planning reports. This indicates a pre-surgical planning tool, not a device used to diagnose a medical condition.
Yes
The device description explicitly states "hipEOS 3.0 allows surgeons to perform preoperative surgical planning... The software provides surgical tools... This software can be used to plan interventions... hipEOS is accessible on any computer via ONEFIT Management System". The summary of performance studies also focuses on "Software verification and validation testing". There is no mention of any hardware component being part of the device itself.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In Vitro Diagnostic devices are used to examine specimens taken from the human body (like blood, urine, tissue) to provide information for diagnosis, monitoring, or screening.
- Device Function: The hipEOS software is used for preoperative planning of total hip arthroplasty (THA). It works with radiological images and 3D bone reconstructions, and helps healthcare professionals select and position hip implants.
- No Specimen Analysis: The description clearly states the software uses radiological images and 3D models of bone. There is no mention of analyzing any specimens taken from the body.
Therefore, the device's function and the type of data it processes fall outside the scope of an In Vitro Diagnostic device. It is a medical device used for surgical planning based on imaging data.
N/A
Intended Use / Indications for Use
hipEOS software is indicated to be used for assisting healthcare professionals in preoperative planning of total hip arthroplasty (THA). The software allows for overlaying of 3D/2D hip implant models on radiological images or 3D reconstruction of bone, and includes tools for performing measurements on the image or 3D model of bones and for selecting and positioning the hip implant models. Clinical judgments and experience are required to properly use the software.
Product codes
LLZ
Device Description
hipEOS 3.0 allows surgeons to perform preoperative surgical planning of hip surgeries in case of Total Hip Arthroplasty (THA). The software provides surgical tools for the implantation of a Total Hip Prosthesis. This software can be used to plan interventions using implants from different manufacturers. The images displayed are x-rays from EOS System (EOS Imaging) and 3D model of the hip from sterEOS Workstation (EOS Imaging), hipEOS also displays preoperative parameters and updated values of parameters after planning. It exports a planning report containing the planning parameters: model, size and position of implants. hipEOS is accessible on any computer via ONEFIT Management System (Class I device -Product code LMD - 510(k) Exempt) that provides a secure internet interface and storage through authentication mechanisms.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
2D images (x-rays), 3D model of bone.
Anatomical Site
hip
Indicated Patient Age Range
Not Found
Intended User / Care Setting
healthcare professionals, surgeons
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". The software was considered as a "moderate" level of concern, since a failure or latent design flaw could directly result in minor injury to the patient or operator, or could indirectly result in minor injury to the patient or operator through incorrect or delayed information or through the action of a care provider.
Key Metrics
Not Found
Predicate Device(s)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 892.2050 Medical image management and processing system.
(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
0
April 27, 2018
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
oneFIT medical % Ms. Camille Vernay QA / RA Manager 18 Rue Alain Savary 25000 Besancon FRANCE
Re: K173390
Trade/Device Name: hipEOS Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: March 16, 2018 Received: March 19, 2018
Dear Ms. Vernay:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
1
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara For
Robert Ochs. Ph.D Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
2
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
1 K173390
Device Name
hipEOS
Indications for Use (Describe)
hipEOS software is indicated to be used for assisting healthcare professionals in preoperative planning of total hip arthroplasty (THA). The software allows for overlaying of 3D/2D hip implant models on radiological images or 3D reconstruction of bone, and includes tools for performing measurements on the image or 3D model of bones and for selecting and positioning the hip implant models. Clinical judgments and experience are required to properly use the software.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services [301) 443-6740 EF
32
3
510(k) SUMMARY
oneFIT medical's hipEOS
Submitter
oneFIT medical 18, rue Alain Savary 25000 Besançon France
Phone: +33(0) 3 81 25 24 27 Facsimile: +33(0) 3 81 25 53 51
Contact Person: Ms Camille VERNAY, QA/RA Manager
Date Prepared: March 16th, 2018
Device:
Name of Device: hipEOS Classification Name: Picture Archiving and Communication System (21 CFR 892.2050) Regulatory Class: II Product Code: LLZ
Predicate Devices
Primary predicate: hipEOS 2.5 (K161479) by oneFIT medical Second predicate: DYONICS PLAN Hip Impingement Planning System (K132636) by Smith & Nephew, Inc.
Device Description
hipEOS 3.0 allows surgeons to perform preoperative surgical planning of hip surgeries in case of Total Hip Arthroplasty (THA). The software provides surgical tools for the implantation of a Total Hip Prosthesis. This software can be used to plan interventions using implants from different manufacturers. The images displayed are x-rays from EOS System (EOS Imaging) and 3D model of the hip from sterEOS Workstation (EOS Imaging), hipEOS also displays preoperative parameters and updated values of parameters after planning. It exports a planning report containing the planning parameters: model, size and position of implants. hipEOS is accessible on any computer via ONEFIT Management System (Class I device -Product code LMD - 510(k) Exempt) that provides a secure internet interface and storage through authentication mechanisms.
Intended Use / Indications for Use
hipEOS software is indicated to be used for assisting healthcare professionals in preoperative planning of total hip arthroplasty (THA). The software allows for overlaying of 3D/2D hip implant models on radiological images or 3D reconstruction of bone, and includes tools for performing
4
measurements on the image or 3D model of bones and for selecting and positioning the hip implant models. Clinical judgments and experience are required to properly use the software.
Comparison of Technological Characteristics with the predicate devices
Preoperative planning for hip surgery is the technological principle for both hipEOS and its predicate devices. The aim for these software is to provide to the surgeon realistic tools for performing the preoperative planning of hip surgeries.
At a high level, the subject and its predicate devices are based on the following same technological elements:
- Access to the software is limited only to surgeons because surgical experience is required. -
- Use of 2D images and 3D model of the hip. -
- -Use of the workflow to guide the surgeon through the steps of planning: observation, analysis and action.
- An interface that allows for the maximum amount of information (including images) and tools, visible simultaneously on the same page.
- -Display values of preoperative parameters and updated values after planning.
The following technological differences exist between the subject and its predicate devices:
-
(1) Use of seated patient images, if available
They are optional in hipEOS 3.0 and they can be used to provide more information to the surgeon (ROM simulation in sitting and theoretical sitting). This difference does not raise new questions of safety or performance for hipEOS 3.0 because this type of imagery was already used by the primary predicate. -
(2) Workflow available
The appearance of steps in hipEOS 3.0 is related to the decomposition of the planning to ensure that the user will not confuse the preoperative information with those from the planning and also to simplify the interface by minimizing the information present at each step. This difference does not raise new questions of safety or performance for hipEOS 3.0 because the user still has the same information at his disposal. -
(3) The range of motion simulation
This type of analysis tool has already been integrated with planning software for total hip arthroplasty (predicate 2). Although the range of motion of the predicate 2 and the one of the submitted device are different, it does not raise new questions of safety or performance for hipEOS 3.0: the ROM of hipEOS 3.0 allows to evaluate for each planning the risk of collision between the components of the implant. This feature is only based on objects of which the 3D geometry is accurate. -
(4) The generation of a second planning report, in DICOM format
This difference does not raise new questions of safety or performance for hipEOS 3.0 because the information in the planning report was already provided to the user and the chosen format is universal.
5
Performance Data
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". The software was considered as a "moderate" level of concern, since a failure or latent design flaw could directly result in minor injury to the patient or operator, or could indirectly result in minor injury to the patient or operator through incorrect or delayed information or through the action of a care provider.
Conclusions
The device has the same intended use, and similar indications for use, technological characteristics, and principles of operation as its predicate device. The minor differences between the device and its predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that the hipEOS 3.0 is as safe and effective as its predicates hipEOS 2.5 (K161479) and DYONICS PLAN Hip Impingement Planning System (K132636) and, thus, is substantially equivalent.