(130 days)
The Equinoxe Stemless Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases of the glenohumeral joint where anatomic total shoulder arthroplasty is determined by the surgeon to be the preferred method of treatment.
Clinical indications for anatomic total shoulder arthroplasty are as follows:
- Osteoarthritis, osteonecrosis or post-traumatic degenerative problems
- Congenital abnormalities in the skeletally mature
- Primary and secondary necrosis of the humeral head.
- Pathologies where arthrodesis or resectional arthroplasty of the humeral head are not acceptable
- Revisions of humeral prostheses when other treatments or devices have failed (where adequate fixation can be achieved)
- To restore mobility from previous procedures (e.g. previous fusion)
The Equinoxe Stemless Shoulder humeral components are indicated for press-fit, uncemented use.
The Equinoxe Stemless Shoulder System is intended to be used with the cemented Equinoxe glenoid components.
The Equinoxe Stemless Shoulder is intended to be used with Exactech glenoid components for use in Total Shoulder Arthroplasty. The Exactech glenoid components are indicated for cemented use. The Equinoxe Stemless Shoulder includes humeral components and humeral heads. The proposed Equinoxe Stemless Humeral Components are additively manufactured from Ti-6Al-4V, and have porous regions. The Equinoxe Stemless Humeral Heads are manufactured from Cobalt Chrome. The Equinoxe Stemless Humeral Components are available in two types, Caged and Non-Caged, with each type available in three sizes, with lengths between 17mm and 24mm. The Equinoxe Stemless Humeral Heads are available in 12 sizes, with diameters between 38mm and 53mm and two height options (Extra Short, Short).
This FDA 510(k) premarket notification for the Exactech® Equinoxe® Stemless Shoulder device does not describe an AI/ML-based medical device. Therefore, the questions related to acceptance criteria, AI/ML study design, ground truth, and sample sizes for training/testing sets are not applicable in this context.
This submission is for a medical implant (a shoulder prosthesis) and focuses on demonstrating substantial equivalence to previously approved predicate devices through non-clinical (mechanical) testing, material comparisons, and similar indications for use.
Instead of AI/ML performance metrics, the document presents mechanical performance criteria.
Here's a breakdown of the relevant information provided:
1. A table of acceptance criteria and the reported device performance:
| Acceptance Criteria (Non-Clinical) | Reported Device Performance |
|---|---|
| Fatigue Testing | Performed to demonstrate performance and substantial equivalence. (Specific results not detailed in this summary document, but inferred to meet criteria for substantial equivalence.) |
| Axial Pull Out Testing | Performed to demonstrate performance and substantial equivalence. |
| Torque out Testing | Performed to demonstrate performance and substantial equivalence. |
| Taper Disengagement Testing | Performed to demonstrate performance and substantial equivalence. |
| Pyrogen Testing (USP <161>, USP <85>, ANSVAAMI ST72) | Conducted to ensure components meet recommended limits for sterility. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- Not applicable for AI/ML device. The "test set" here refers to mechanical test samples, not clinical data for an algorithm. The document does not specify the number of mechanical samples used for each test. Data provenance is not relevant as it's mechanical testing of a physical device.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
- Not applicable for AI/ML device. Ground truth, in the context of AI/ML, refers to labels or diagnoses. For a physical implant, ground truth would be established through engineering specifications and standards.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable for AI/ML device. Adjudication methods are used to resolve disagreements in expert labeling for AI/ML ground truth.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is not an AI/ML device, so no MRMC studies or human reader improvement with AI assistance were performed.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is a physical medical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not applicable for AI/ML device. For this physical device, "ground truth" equates to the established engineering specifications, performance standards (e.g., ISO, ASTM for fatigue, pull-out strength), and material properties that the device must meet to function safely and effectively.
8. The sample size for the training set:
- Not applicable for AI/ML device. There is no training set for a physical implant.
9. How the ground truth for the training set was established:
- Not applicable for AI/ML device. There is no ground truth or training set in the AI/ML sense for this device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo, which features a stylized depiction of a human figure. The logo on the right is the FDA U.S. Food & Drug Administration logo, with the FDA acronym in a blue square and the full name of the agency written in blue text to the right of the square.
March 9, 2018
Exactech®, Inc Thomas Mcnamara Senior Regulatory Specialist 2320 NW 66th Court Gainesville, Florida 32653
Re: K173388
Trade/Device Name: Exactech® Equinoxe® Stemless Shoulder Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: PKC Dated: February 6, 2018 Received: February 7, 2018
Dear Thomas McNamara:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
Page 2 - Thomas McNamara
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K173388
Device Name
Exactech® Equinoxe® Stemless Shoulder
Indications for Use (Describe)
The Equinoxe Stemless Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases of the glenohumeral joint where anatomic total shoulder arthroplasty is determined by the preferred method of treatment.
Clinical indications for anatomic total shoulder arthroplasty are as follows:
- · Osteoarthritis, osteonecrosis or post-traumatic degenerative problems
- · Congenital abnormalities in the skeletally mature
- · Primary and secondary necrosis of the humeral head.
- · Pathologies where arthrodesis or resectional arthroplasty of the humeral head are not acceptable
- · Revisions of humeral prostheses when other treatments or devices have failed (where adequate fixation can be achieved)
- · To restore mobility from previous procedures (e.g. previous fusion)
The Equinoxe Stemless Shoulder humeral components are indicated for press-fit, uncemented use.
The Equinoxe Stemless Shoulder System is intended to be used with the cemented Equinoxe glenoid components.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | |
|---|---|
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Exactech® Equinoxe® Stemless Shoulder Traditional 510(k) - 510(k) Summary
| Company: | Exactech®, Inc2320 NW 66th CourtGainesville, FL 32653 |
|---|---|
| Date: | February 6, 2018 |
| Contact Person: | Thomas McNamara, RACSenior Regulatory Specialist |
| Phone: (352) 377-1140Fax: (352) 378-2617 | |
| Proprietary Name: | Exactech® Equinoxe® Stemless Shoulder |
| Common Name: | Shoulder Prosthesis |
| Classification Name: | Prosthesis, Total Anatomic Shoulder, UncementedMetaphyseal Humeral Stem with No Diaphyseal Incursion,Semi-Constrained, 21 CFR 888.3660, Class II, ProductCode PKC |
Legally Marketed Device to Which Substantial Equivalence Is Claimed:
Reference Devices
- Exactech Equinoxe Resurfacing Humeral Head System (K131298) ●
- Exactech Novation® Crown Cup® with InteGrip™ Acetabluar Shell, Cluster-Hole o (K102975)
- Exactech Optetrak Logic® Metaphseal Cones (K153595) ●
Device Description
The Equinoxe Stemless Shoulder is intended to be used with Exactech glenoid components for use in Total Shoulder Arthroplasty. The Exactech glenoid components are indicated for cemented use. The Equinoxe Stemless Shoulder includes humeral components and humeral heads. The proposed Equinoxe Stemless Humeral Components are additively manufactured from Ti-6Al-4V, and have porous regions. The Equinoxe Stemless Humeral Heads are manufactured from Cobalt Chrome. The Equinoxe Stemless Humeral Components are available in two types, Caged and Non-Caged, with each type available in three sizes, with lengths between 17mm and 24mm. The Equinoxe Stemless Humeral Heads are available in 12 sizes, with diameters between 38mm and 53mm and two height options (Extra Short, Short).
{4}------------------------------------------------
Exactech® Equinoxe® Stemless Shoulder Traditional 510(k) - 510(k) Summary
Indications for Use
The Equinoxe Stemless Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases of the glenohumeral joint where anatomic total shoulder arthroplasty is determined by the surgeon to be the preferred method of treatment.
Clinical indications for anatomic total shoulder arthroplasty are as follows:
- Osteoarthritis, osteonecrosis or post-traumatic degenerative problems
- Congenital abnormalities in the skeletally mature
- Primary and secondary necrosis of the humeral head.
- Pathologies where arthrodesis or resectional arthroplasty of the humeral head are not acceptable
- Revisions of humeral prostheses when other treatments or devices have failed (where adequate fixation can be achieved)
- To restore mobility from previous procedures (e.g. previous fusion)
The Equinoxe Stemless Shoulder humeral components are indicated for press-fit, uncemented use.
The Equinoxe Stemless Shoulder System is intended to be used with the cemented Equinoxe glenoid components.
Summary of Technological Characteristics
The rationale for substantial equivalence is based on consideration of the following device use and characteristics:
- . Indications for Use. The proposed Exactech Equinoxe Stemless Shoulder components and the predicate devices have similar indications for use.
- . Materials/Surface Finish/Coatings. The proposed Exactech Equinoxe Stemless Shoulder components and the predicate devices are composed of the same or similar biocompatible substrate materials, and the same or similar surface finish for permanent implants.
- Design Features. The proposed Exactech Equinoxe Stemless Shoulder . components and the predicate devices have similar design features.
- Dimensions. The proposed Exactech Equinoxe Stemless Shoulder components and the predicate devices are dimensionally comparable.
- Sterilization. The proposed Exactech Equinoxe Stemless Shoulder components and the predicate devices are provided sterile for single use only.
- o Performance Requirements. The proposed Exactech Equinoxe Stemless Shoulder components and the predicate devices conform to recognized performance standards for total shoulder replacement devices.
{5}------------------------------------------------
Exactech® Equinoxe® Stemless Shoulder Traditional 510(k) - 510(k) Summary
Non-Clinical Testing
The following engineering analyses were performed to demonstrate that the Exactech Equinoxe Stemless Shoulder perform as intended and are substantially equivalent to the identified predicate devices:
- Fatigue Testing ●
- Axial Pull Out Testing o
- Torque out Testing
- o Taper Disengagement Testing
Pyrogen testing was conducted in accordance with USP <161>, USP <85>, and ANSVAAMI ST72 to ensure the proposed Equinoxe Stemless Shoulder components meet recommended limits per FDA's Guidance Document Submission and Review of Sterility Information in Premarket (510(k)) Submission for Devices Labeled as Sterile.
Substantial Equivalence Conclusion
Based on consideration of indications for use, technological characteristics, and results of non-clinical testing, it was concluded that the Exactech Equinoxe Stemless Shoulder demonstrates substantial equivalence to the referenced predicate devices.
§ 888.3660 Shoulder joint metal/polymer semi-constrained cemented prosthesis.
(a)
Identification. A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).(b)
Classification. Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”