K Number
K172835

Validate with FDA (Live)

Date Cleared
2018-06-11

(265 days)

Product Code
Regulation Number
884.1730
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GRI-Alleset Veress Needle is a disposable, single-use, sterile device intended to be used in minimally-invasive abdominal procedures for the establishment of pneumoperitoneum.

Device Description

The GRI-Alleset Veress Needle is a disposable, single-use, sterile surgical instrument used during minimally invasive surgery for the establishment of peritoneum of the abdominal cavity prior to abdominal surgery. The device is comprised of a 14 gauge stainless steel needle (120mm and 150mm in length) which is attached to a plastic handle that contains a stopcock with a luer lock connector. The spring-loaded, blunt stylet extends beyond the needle tip and retracts as the needle is pushed through the abdominal tissue; it automatically advances forward once the peritoneum is penetrated. A red safety indicator, attached to the proximal end of the stylet, is exposed above the handle as an indication of an exposed needle tip. The device is provided sterile via ethylene oxide.

AI/ML Overview

This document describes the premarket notification (510(k)) for the GRI-Alleset Veress Needle, a disposable, single-use, sterile device intended for establishing pneumoperitoneum in minimally invasive abdominal procedures. The submission aims to demonstrate substantial equivalence to a predicate device, the Ethicon Endopath Ultra Veress Needle (K983925).

Here's an analysis of the acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly listed in a separate table as "acceptance criteria." Instead, the document outlines performance tests that were conducted and states that the "results show acceptable results and demonstrate that the GRI-Alleset Veress Needle meets its specifications." The direct comparisons are made against the predicate device.

Test / Performance MetricAcceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance
Material BiocompatibilityISO 10993 standards complianceBiocompatible
CytotoxicityISO 10993-5:2009 complianceAcceptable
IrritationISO 10993-10:2010 complianceAcceptable
SensitizationISO 10993-10:2010 complianceAcceptable
SterilizationISO 11135-1:2014 complianceValidated
ETO ResidualsISO 10993-7:2008 complianceAcceptable
Mechanical Bench TestingPerformance equivalent or superior to predicateAcceptable, met specifications
Gas flowComparable to predicateAcceptable
LeakageComparable to predicateAcceptable
Max Puncture ForceComparable to predicateAcceptable
Rotational Valve OperationComparable to predicateAcceptable
Stylet AlignmentComparable to predicateAcceptable
Stylet StrengthComparable to predicateAcceptable
Connector FittingComparable to predicateAcceptable
Audible RateComparable to predicateAcceptable
Shelf LifeMaintain functional performance and sterilityMaintains over proposed shelf life
Functional PerformanceMaintained throughout shelf lifeAcceptable
Packaging SterilityMaintained throughout shelf lifeAcceptable

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample sizes for the test sets used in the biocompatibility, sterilization, mechanical bench testing, or shelf-life studies. It only mentions that these tests were "conducted in comparison with the predicate." The data provenance is not specified, but typically such testing would be performed in a controlled laboratory environment. The studies are prospective in the sense that the new device was evaluated against established standards and a predicate.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the document. For medical devices like a Veress needle, "ground truth" is typically established through adherence to engineering standards, validated test methods, and comparison data from a legally marketed predicate, rather than expert consensus on a test set in the way AI algorithms are evaluated (e.g., image interpretation). The "experts" would be the engineers, quality specialists, and toxicologists conducting and interpreting the tests according to relevant ISO standards.

4. Adjudication Method for the Test Set

Not applicable in the context of this device and the tests performed. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies, particularly for diagnostic accuracy assessments where there might be disagreement among human readers or interpreters. For device performance testing against engineering standards, results are typically objective and measurable.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for evaluating the performance of diagnostic tools or AI algorithms where human interpretation is involved. The GRI-Alleset Veress Needle is a surgical instrument, and its effectiveness is assessed through physical and mechanical performance tests, not through reader interpretation.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

This question is not applicable. The GRI-Alleset Veress Needle is a physical surgical instrument, not an algorithm or AI system. Therefore, standalone algorithm performance is not relevant.

7. The Type of Ground Truth Used

The "ground truth" for this device's evaluation is primarily based on:

  • Established ISO Standards: For biocompatibility (ISO 10993 series) and sterilization (ISO 11135-1).
  • Performance Specifications: The device must meet its own design specifications and perform comparably to the legally marketed predicate device (Ethicon Endopath Ultra Veress Needle).
  • Predicate Device Performance: The predicate device itself serves as a benchmark for acceptable performance.

8. The Sample Size for the Training Set

This question is not applicable. Training sets are used for machine learning models. The GRI-Alleset Veress Needle is a hardware medical device, and its development does not involve machine learning training sets.

9. How the Ground Truth for the Training Set Was Established

This question is not applicable, as there is no training set for a hardware medical device.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around a symbol. To the right of the seal, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG" are written in blue, with the word "ADMINISTRATION" written in a smaller font size below.

June 11, 2018

GRI Medical and Electronic Technology Co., Ltd. % Sharon Marrow Consultant Sharon Marrow 2907 Cherry Branch Drive Knoxville, TN 37948

Re: K172835

Trade/Device Name: GRI-Alleset Veress Needle Regulation Number: 21 CFR§ 884.1730 Regulation Name: Laparoscopic Insufflator Regulatory Class: II Product Code: HIF Dated: April 13, 2018 Received: May 10, 2018

Dear Sharon Marrow:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K172835

Device Name GRI-Alleset Veress Needle

Indications for Use (Describe)

The GRI-Alleset Veress Needle is a disposable, single-use, sterile device intended to be used in minimally-invasive abdominal procedures for the establishment of pneumoperitoneum.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Global Resources International (GRI). The logo features the letters "gri" in a stylized, lowercase font, with a blue curved line above the "i". Below the letters, the words "Global Resources International" are written in a smaller, sans-serif font.

510(K) SUMMARY

1. 510(k) Summary

Date Prepared: June 8, 2018

1.1 Applicant: Owner and Official Correspondent

GRI Medical & Electronic Technology Co., Ltd. 1805 Honggao Rd. Xiuzhou Industry Zone Jiaxing, CN-33 CHINA 314031 Phone: 86-135-11359722

Contact: Martin D. Paugh GLOBAL RESOURCES INTERNATIONAL, INC. 4142 Industry Way Flowery Branch, GA 30549 Phone: 770 9253251 Fax: 770 9259469 Email: mpaugh@gri-usa.com

1.1. Device Trade Name: Trade/Proprietary Name

GRI-Alleset Veress Needle

Common or Usual Name:Veress Needle
Regulation Name:Laparoscopic insufflator
Regulation Number:21 CFR 884.1730
Class:II
Product Code:HIF (insufflator, laparoscopic)

1.2. Predicate Device:

Device Name:Endopath Ultra Veress Needle
510(k) Number:K983925
Manufacturer:Ethicon Endosurgery, Inc.

The predicate device has not been subject to a design related recall.

1.3. Description of the Device

The GRI-Alleset Veress Needle is a disposable, single-use, sterile surgical instrument used during minimally invasive surgery for the establishment of peritoneum of the abdominal cavity prior to abdominal surgery. The device is comprised of a 14 gauge stainless steel needle (120mm and 150mm in length) which is attached to a plastic handle that contains a stopcock with a luer lock

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for GRI, which likely stands for Global Resources International. The logo features the letters 'gri' in a bold, sans-serif font, with the 'i' having a curved line above it, resembling a check mark or a stylized accent. Below the letters, there is some text that is not legible.

connector. The spring-loaded, blunt stylet extends beyond the needle tip and retracts as the needle is pushed through the abdominal tissue; it automatically advances forward once the peritoneum is penetrated. A red safety indicator, attached to the proximal end of the stylet, is exposed above the handle as an indication of an exposed needle tip. The device is provided sterile via ethylene oxide.

1.4. Indications for Use

The GRI-Alleset Veress Needle is a disposable, single-use, sterile device intended to be used in minimally-invasive abdominal procedures for the establishment of pneumoperitoneum.

1.5. Comparison to the Predicate Device

The table below provides an overview of the comparison of the new device to the predicate device. As noted in the table below, the subject and predicate device have the same intended use. There are also differences in technological characteristics between the subject and predicate device; however, these differences do not raise different questions of safety or effectiveness.

Element of ComparisonSubject DevicePredicate Device
GeneralManufacturerGRIEthicon(K983925)
Trade NameVeress NeedleUltra Veress Needle
ClassIIII
Regulation NameLaparoscopic insufflatorLaparoscopicinsufflator
Product CodeHIFHIF, FHP
Intended UseIntended UserSurgeons in operating roomSurgeons inoperating room
Intended UseThe Veress Needle has application in minimally invasive abdominal procedures for the establishment of pneumoperitoneum.The ENDOPATHUltra VeressInsufflation Needlehas application ingynecologiclaparoscopy andother minimallyinvasive abdominalprocedures forestablishment ofpneumoperitoneum.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for GRI, which stands for Global Reporting Initiative. The logo is in black and features the letters 'gri' in a stylized, italicized font. Above the 'i' is a blue curved line, adding a modern touch to the design. The logo is simple, clean, and easily recognizable.

Element of ComparisonSubject DevicePredicate Device
TechnologicalCharacteristicsDevice Description14 gauge, 120mm and 150mm,two way valve with luer lockconnector, red safety indicator14 gauge, 120mmand 150mm,optional stop cockwith luer lockconnector, redsafety indicator
Materials/CompositionStainless steel, ABSStainless Steel, ABS
SterilityProvided Sterile via EtO; Single-Use DisposableProvided Sterile viaGamma Radiation;Single-UseDisposable
Length120mm, 150mm120mm
Diameter2.1mm (14 gauge)14 gauge
Luer Lock ConnectorYesYes
Spring LoadedYesYes
Indicator for when inperitoneal cavityYes (red indicator)Yes (red ball)
Sterile PackagingTyvek to PET peel pouchTyvek to poly peelpouch
Design DifferencesHandle design (shape)Handle Design(shape); indicator forUltra Veress Needleis red ball, hasoptional luerconnector

1.6. Summary of non-clinical tests submitted

Cytotoxicity, irritation and sensitization testing were conducted according to ISO 10993-1:2009, ISO 10993-5:2009, ISO 10993-10:2010 and ISO 10993-12:2012. The results demonstrate that the subject device is biocompatible.

Sterilization validation was performed according to ISO 11135-1:2014. Ethylene Oxide residuals were evaluated according to ISO 10993-7:2008 and are acceptable.

Mechanical bench testing, including assessments of gas flow, leakage, max puncture force, rotational valve operation, stylet alignment, stylet strength, connector fitting, and audible rate, were conducted in comparison with the predicate, Ethicon Endopath Ultra Veress Needle. All

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for GRI, which is a company that provides sustainability reporting standards. The logo is black text with a blue arc over the "i". The text is in a sans-serif font and is slightly italicized.

performance tests show acceptable results and demonstrate that the GRI-Alleset Veress Needle meets its specifications.

Shelf life testing demonstrates that the subject device maintains its functional performance and its packaging maintains device sterility over the duration of the proposed shelf life.

1.7. Conclusion of the tests

The results of performance testing demonstrate that the GRI-Alleset Veress Needle is substantially equivalent to the Ethicon Endopath Ultra Veress Needle.

§ 884.1730 Laparoscopic insufflator.

(a)
Identification. A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.(b)
Classification. (1) Class II (performance standards).(2) Class I for tubing and tubing/filter kits which include accessory instruments that are not used to effect intra-abdominal insufflation (pneumoperitoneum). The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.