(206 days)
Axiostat Chitosan Hemostatic Dressing is indicated to control bleeding of lacerations, minor cuts and abrasions.
Axiostat is a single use, hemostatic dressing made of 100% chitosan is a wellknown natural polysaccharide generally derived from shellfish which has widely recognized hemostatic properties. When applied on the bleeding site with firm pressure, the Axiostat Chitosan Hemostatic Dressing acts as a mechanical barrier against bleeding. The Axiostat is intended for a maximum duration of use of 24 hours, including bandage changes that may be needed. Axiostat is individually packaged in foil bags and sterilized using gamma irradiation. Axiostat is provided in six different sizes for over the counter use as listed below.
This looks like a 510(k) premarket notification for a medical device. Based on the provided text, the device is the Axiostat Chitosan Hemostatic Dressing. Here's an analysis of the acceptance criteria and supporting studies:
1. Table of Acceptance Criteria and Reported Device Performance
The submission is a 510(k) for a medical device, which means the primary "acceptance criterion" is demonstrating substantial equivalence to a legally marketed predicate device. The document explicitly states that the device is substantially equivalent to the predicate device Chito-SAM™ Active (K133121).
The acceptance criteria are not explicitly numerical thresholds for performance metrics (like sensitivity/specificity for AI, for example), but rather a demonstration that the Axiostat Chitosan Hemostatic Dressing is as safe and effective as the predicate device.
Here's a table summarizing the comparison:
| Characteristic | Acceptance Criterion (Predicate: Chito-SAM™ Active K133121) | Reported Device Performance (Axiostat Chitosan Hemostatic Dressing) | Notes |
|---|---|---|---|
| Regulatory Status | Unclassified; Product Code FRO | Unclassified; Product Code FRO (initially); QSY (subsequent administrative change) | Both products are unclassified and initially shared the same product code. The change to QSY for Axiostat was an administrative categorization by the FDA and did not involve a re-review of the 510(k) submission. |
| Material | Chitosan | Chitosan | Same. "The chemical composition of the chitosan material in both products is the same." |
| Intended Use | Hemostasis | Hemostasis | Same. |
| Indications for Use | Control bleeding of lacerations, minor cuts and abrasions. | Control bleeding of lacerations, minor cuts and abrasions. | Identical. |
| Biocompatibility | Non-cytotoxic, Negative (Skin Irritation, Sensitization, Acute Intravenous Systemic Toxicity, Acute Intraperitoneal Systemic Toxicity, Hemolysis) | Non-cytotoxic, Negative (Skin Irritation, Sensitization, Acute Intravenous Systemic Toxicity, Acute Intraperitoneal Systemic Toxicity, Hemolysis, Pyrogenicity Test) | The subject device includes a Pyrogenicity Test, which the predicate device's data does not explicitly list in the provided table comparison, but "identical biocompatibility characteristics" is claimed. All other listed tests show Same results (e.g., Non-cytotoxic for cytotoxicity). |
| Sterilization | Gamma irradiation | Gamma irradiation | Same. |
| OTC/Prescription | OTC | OTC | Same. |
| Anatomical Site | External wounds | External wounds | Same. |
| Package Material | Foil bag | Foil Bag | Same. |
| Physical Properties | Not explicitly listed for predicate, but included in bench performance. | Evaluated (Physical, Absorbency, Moisture, pH, Tensile strength) | The "bench performance testing" for the subject device indicates these were tested, and the conclusion is that "the Axiostat Chitosan Hemostatic Dressing functions to its specifications, performs as intended, and exhibits the appropriate characteristics of a hemostatic dressing," implying they are equivalent to what would be expected from the predicate. |
| Hemostatic Activity | Not explicitly listed for predicate, but included in bench performance. | Evaluated (In-Vitro Hemostatic Activity, In Vivo rabbit hemostasis study) | Same as above. The "In Vivo rabbit hemostasis study" directly supports the hemostatic function. |
2. Sample Size Used for the Test Set and Data Provenance
Since this is a hemostatic dressing (a physical device, not an AI/software device), the "test set" concept as used in AI/software evaluation (e.g., for sensitivity/specificity) doesn't directly apply here. Instead, it refers to the samples used in non-clinical testing.
- Biocompatibility testing: "performed per the requirements of ISO 10993-1:2009, under the section 'surface devices used in breached and compromised surfaces with limited exposure (<24 hrs)'." The specific number of samples for each test (Cytotoxicity, Skin Irritation, etc.) is not provided in this summary.
- Heavy metal testing: "The subject device was tested for cadmium, lead, arsenic and mercury." The number of samples is not provided.
- Bench performance testing: "Physical Properties, Absorbency Testing, Moisture Content, pH Testing, Tensile strength, In-Vitro Hemostatic Activity." The number of samples for these tests is not provided.
- In Vivo rabbit hemostasis study: This is an animal study, likely involving a controlled number of rabbits. The exact sample size is not provided.
- Sterilization and Packaging Validation: Includes real-time stability studies and accelerated aging tests per ASTM F-1980, Package seal strength per ISO11607-1:2006, Dye penetration test as per ASTM F1929:15, and Sterility test as per US Pharmacopeia <71>. The sample sizes for these tests are not provided.
Data Provenance: The studies are non-clinical (laboratory and animal studies) conducted by Advamedica for their device. The country of origin for the data is not specified but would likely be the location of the testing laboratories. These are prospective tests conducted specifically for this device submission.
3. Number of Experts and Qualifications for Ground Truth
This question is typically relevant for studies involving subjective human assessment, particularly in fields like radiology where experts establish "ground truth" for AI models. For a hemostatic dressing, the "ground truth" is established by direct measurement and observation in laboratory and animal studies, rather than expert consensus on interpretive tasks. Therefore, this section is not applicable in the same way it would be for an AI device. The ground truth is objective (e.g., cytotoxicity present/absent, metal levels, pH value, successful hemostasis).
4. Adjudication Method for the Test Set
As explained above, the concept of an adjudication method (like 2+1 or 3+1 for human readers) is not applicable here because the "test set" involves non-clinical studies with objective measurements rather than subjective expert interpretation for ground truth.
5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study
This is not applicable for this type of device. An MRMC study is typically used for diagnostic or interpretive devices (often AI) to assess how human readers' performance changes with or without AI assistance. The Axiostat Chitosan Hemostatic Dressing is a physical therapeutic device, not an interpretive one.
6. Standalone Performance Study (Algorithm Only)
This is not applicable as the Axiostat Chitosan Hemostatic Dressing is a physical product, not an algorithm or AI.
7. Type of Ground Truth Used
The ground truth for the non-clinical studies relies on:
- Standardized laboratory assays: For biocompatibility tests (e.g., cytotoxicity, irritation, hemolysis, pyrogenicity), heavy metal analysis, and bench performance tests (e.g., pH, tensile strength, absorbency).
- In-vivo observation and measurement: For the rabbit hemostasis study, where the ability of the dressing to control bleeding is directly observed and measured within a living system.
- Validated microbiological and physical testing: For sterility assurance level (SAL), package integrity, and shelf-life studies.
Essentially, the ground truth is based on objective scientific measurements and observations according to established standards (ISO, USP, ASTM).
8. Sample Size for the Training Set
This is not applicable. The concept of a "training set" is specific to machine learning algorithms. This device is a physical product, and its development and testing do not involve machine learning.
9. How the Ground Truth for the Training Set Was Established
As section 8 is not applicable, this section is also not applicable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left, and the FDA logo is on the right. The FDA logo includes the agency's name, "U.S. Food & Drug Administration," in blue text.
April 21, 2023
Advamedica Inc. % Alan Donald President Matrix Medical Consulting, Inc. 8880 Rio San Diego Drive, Suite 800 San Diego, California 92108
Re: K172324
Trade/Device Name: Axiostat Chitosan Hemostatic Dressing Regulatory Class: Unclassified Product Code: QSY
Dear Alan Donald:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated February 23, 2018. Specifically, FDA is updating this SE Letter because FDA has better categorized your device technology under product code QSY.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Julie Morabito, OHT4: Office of Surgical and Infection Control Devices, 240-402-3839, Julie.Morabito@fda.hhs.gov.
Sincerely,
Julie A. Morabito -S
Julie Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box. To the right of the blue box is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 23, 2018
Advamedica Inc. % Alan Donald President Matrix Medical Consulting, Inc. 8880 Rio San Diego Drive, Suite 800 San Diego, California 92108
Re: K172324
Trade/Device Name: Axiostat Chitosan Hemostatic Dressing Regulatory Class: Unclassified Product Code: FRO Dated: January 22, 2018 Received: January 23, 2018
Dear Alan Donald:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may; therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
David Krause -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications for Use
510(k) Number (if known) K172324
Device Name
Axiostat Chitosan Hemostatic Dressing
Indications for Use (Describe)
Axiostat Chitosan Hemostatic Dressing is indicated to control bleeding of lacerations, minor cuts and abrasions.
Type of Use (Select one or both, as applicable)
__ Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for Advamedica. The logo consists of a stylized teal-colored mountain range to the left of the company name. The company name "ADVAMEDICA" is written in a gray sans-serif font, with all letters capitalized.
510(k) Summary
{5}------------------------------------------------
K172324 p 2/6
510(k) SUMMARY - K172324
Advamedica, Inc.
AXIOSTAT CHITOSAN HEMOSTATIC DRESSING
ADMINISTRATIVE INFORMATION
- Date of Preparation: February 23, 2018 a.
| b. | Submitter: | Advamedica Inc.485 Massachusetts Avenue,Suite 300, Cambridge,Massachusetts 02139, USATel: +1 973-718-7575Fax: +1 888-584-8237 |
|---|---|---|
| C. | Official Contact: | Mr. Leo Mavely, PresidentEmail: office@advamedica.comWeb: www.advamedica.com |
| d. | Prepared By: | Mr. Leo Mavely, PresidentAdvamedica, Inc. |
2. DEVICE NAME AND CLASSIFICATION
| a. | Trade/Proprietary Names: | Axiostat Chitosan Hemostatic Dressing |
|---|---|---|
| b. | Common Name: | Hemostatic Dressing |
| c. | Classification Name: | Dressing, Wound, Drug |
| d. | Device Class: | Unclassified |
| e. | Product Codes: | FRO |
| f. | Classification Panel: | General and Plastic Surgery |
3. PREDICATE DEVICE
| Proprietary/Trade name | Chito-SAM™ Active |
|---|---|
| Manufacturer | SAM Medical Products |
| Classification Name | Dressing, Wound, Drug |
| Device Class | Unclassified |
| Panel | General & Plastic Surgery |
| Product Code | FRO |
| 510(k) Number | K133121 |
DEVICE DESCRIPTION 4.
Axiostat is a single use, hemostatic dressing made of 100% chitosan is a wellknown natural polysaccharide generally derived from shellfish which has widely recognized hemostatic properties. When applied on the bleeding site with firm pressure, the Axiostat
{6}------------------------------------------------

Chitosan Hemostatic Dressing acts as a mechanical barrier against bleeding. The Axiostat is intended for a maximum duration of use of 24 hours, including bandage changes that may be needed.
Axiostat is individually packaged in foil bags and sterilized using gamma irradiation. Axiostat is provided in six different sizes for over the counter use as listed below.
| Axiostat Sizes |
|---|
| 10 x 10 cm |
| 8 x 8 cm |
| 8 x 5 cm |
| 5 x 5 cm |
| 3.5 x 3.5 cm |
| 2 x 2 cm |
INDICATIONS FOR USE ડ.
Axiostat Chitosan Hemostatic Dressing is indicated to control bleeding of lacerations, minor cuts and abrasions.
NON-CLINICAL TESTING 6.
The subject device has been evaluated through a series of nonclinical studies.
A. Biocompatibility testing:
Biocompatibility tests have been performed per the requirements of ISO 10993-1:2009, under the section "surface devices used in breached and compromised surfaces with limited exposure (<24 hrs)". The following tests have been performed as per these requirements,
-
- Cytotoxicity
-
- Skin Irritation
-
- Skin Sensitization
-
- Acute Intravenous Systemic Toxicity
- Acute Intraperitoneal Systemic Toxicity న్.
-
- Hemolysis Test
-
- Pyrogenicity Test
B. Heavy metal testing
The subject device was tested for cadmium, lead, arsenic and mercury. The levels for all elements were found to be fewer than 1 ppm. which meets the USP-232 limits.
C. Bench performance testing:
-
- Physical Properties
-
- Absorbency Testing
-
- Moisture Content
{7}------------------------------------------------
K172324 p 4/6
-
- pH Testing
- న్. Tensile strength
- In-Vitro Hemostatic Activity 6.
-
- In Vivo rabbit hemostasis study
STERILIZATION & PACKAGING 7.
Axiostat Chitosan Haemostatic Dressing is supplied sterile in a foil bag. It is sterilized using gamma radiation to a sterility assurance level (SAL) of 106. Gamma sterilization validation was performed in compliance with the ISO 11137 Part-1:2006 and ISO 11137 Part-2:2013 standards.
Following the gamma sterilization, the packaging was subjected to sterile barrier testing to validate a shelf life of three (3) years as per ISO and ASTM standards.
- The stability and effectiveness of packaging of the sterilized product during the shelf-life a. period was confirmed by real time stability studies and accelerated aging test, per ASTM F-1980.
- Package seal strength (Wet/Dry) per ISO11607-1:2006 b.
- Dye penetration test as per ASTM F1929:15 C.
- d. Sterility test as per US Pharmacopeia <71>.
SUBSTANTIAL EQUIVALENCE DETERMINATION 8.
The Axiostat Chitosan Hemostatic Dressing is substantially equivalent in intended use, indications for use, technology, material, design, packaging, sterilization, and performance claims to the predicate device, Chito-SAM™ Active (K133121). The following tables show the key characteristics of the subject and predicate devices.
| Subject device | Predicate device | Notes | |
|---|---|---|---|
| Manufacturer | Advamedica Inc. | SAM Medical Products | -- |
| Model/TradeName | Axiostat Chitosan HemostaticDressing | Chito-SAM™ Active | -- |
| 510K Identifier | New device | K133121 | -- |
| Common Name | Hemostatic dressing | Hemostatic Dressing | Same |
| Classification | Unclassified | Unclassified | Same |
| Product Code | FRO | FRO | Same |
| Material | Chitosan | Chitosan | Same |
| Intended Use | Hemostasis | Hemostasis | Same |
| Indications for use | Axiostat Chitosan HemostaticDressing is indicated to controlbleeding of lacerations, minor cutsand abrasions. | Chito-SAM™ Active is indicatedto control bleeding of lacerations,minor cuts and abrasions. | Same |
Table 1: Comparison of technological characteristics of subject and predicate devices
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the words "TRADITIONAL 510K" in bold, uppercase letters. Below that, the words "Axiostat Chitosan Hemostatic Dressing" are also in bold. The text appears to be advertising a traditional product called Axiostat Chitosan Hemostatic Dressing.
MEDICA
| OTC/ Prescription | OTC | OTC | Same |
|---|---|---|---|
| Anatomical Site | External wounds | External wounds | Same |
| Package material | Foil Bag | Foil bag | Same |
| Sterilization | Gamma irradiation | Gamma irradiation | Same |
| Sizes | Axiostat:10x10cm8x8cm8x5cm5x5cm3.5x3.5cm2x2cm | 4 x 4-inch3-inch x 6-foot Z-fold | -- |
Table 2: Comparison of biocompatibility tests on subject and predicate devices
| Biocompatibility Tests | Subject deviceAxiostat Chitosan HemostaticDressing | Predicate deviceChito-SAM™ Active(K133121) |
|---|---|---|
| Cytotoxicity Test | Non-cytotoxic | Non-cytotoxic |
| Skin Irritation Study | Negative | Negative |
| Skin Sensitization Study | Negative | Negative |
| Acute Intravenous SystemicToxicity Study | Negative | Negative |
| Acute Intraperitoneal SystemicToxicity Study | Negative | Negative |
| Hemolysis | Negative | Negative |
| Pyrogenicity Test | Negative | -- |
The intended use and "Indications for Use" for the predicate device are identical to those of the Axiostat Chitosan Hemostatic Dressing. The chemical composition of the chitosan material in both products is the same. The technological and performance characteristics of the Axiostat are equivalent to the predicate device. No new issues regarding safety or efficacy are introduced by the Axiostat Chitosan Hemostatic Dressing.
9. SIMILARITY AND DIFFERENCES
The subject device has the following similarities to the predicate devices:
-
the same intended use
- V the same material (chitosan)
-
same indications of use
-
identical biocompatibility characteristics
-
same sterilization method.
The minor differences in dimensions and form between the proposed device and predicate device do not affect the performance of proposed device or raise any safety concerns.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for ADVAMEDICA. The logo consists of a stylized "M" shape in teal, followed by the word "ADVAMEDICA" in gray. Above the logo, the text "K172324 p 6/6" is displayed in black.
10. CONCLUSION
The Axiostat Chitosan Hemostatic Dressing shares its chemical composition and performance characteristics with the predicate device. The results of the performance testing confirm that the Axiostat Chitosan Hemostatic Dressing functions to its specifications, performs as intended, and exhibits the appropriate characteristics of a hemostatic dressing. The Axiostat is substantially equivalent to the predicate device in terms of materials, technological characteristics, intended use, indications for use, and performance. No new issues of safety or effectiveness are raised by the Axiostat Chitosan Hemostatic Dressing.
N/A