K Number
K171974
Manufacturer
Date Cleared
2017-10-16

(108 days)

Product Code
Regulation Number
866.3980
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Solana® RSV+hMPV Assay is a qualitative in vitro diagnostic test for the detection and differentiation of RSV and hMPV viral RNA in nasal and nasopharyngeal swabs from patients with signs and symptoms of respiratory infection. This test is intended for use as an aid in the differential diagnosis of RSV and hMPV viral infections in humans in conjunction with clinical and epidemiological risk factors. This test is not intended to differentiate the two subtypes of RSV or the four genetic sub-lineages of hMPV.

Negative results do not preclude RSV infection and/or hMPV infection and should not be used as the sole basis for diagnosis, treatment or other patient management decisions.

Conversely, positive results do not rule-out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease. The use of additional laboratory testing and clinical presentation must be considered in order to obtain the final diagnosis of respiratory viral infection.

Device Description

The Solana RSV+hMPV Assay amplifies and detects viral RNA present in viral transport media containing nasopharyngeal or nasal swab specimens obtained from symptomatic patients.

The assay consists of two major steps: 1) specimen preparation, and 2) amplification and detection of target sequences specific to RSV and/or hMPV using isothermal Reverse Transcriptase - Helicase-Dependent Amplification (RT-HDA) in the presence of targetspecific fluorescence probes.

A patient nasal or nasopharyngeal swab specimen in viral transport media is transferred to a Process Buffer Tube, subjected to heat treatment at 95°C for 5 minutes and mixed. The processed sample is transferred to a Reaction Tube. The Reaction Tube contains lyophilized RT-HDA reagents, dNTPs, primers and probes. Once rehydrated with the processed sample, the Reaction Tube is placed in Solana for amplification and detection of RSV and hMPV specific target sequences. In Solana, the target sequences are amplified by RSV and hMPV specific primers and detected by RSV and hMPV specific fluorescence probes, respectively. A process control (PRC) is included in the Process Buffer Tube to monitor sample processing, inhibitory substances in clinical samples, reagent failure or device failure. The PRC target is amplified by RSV and hMPV specific primers and detected by a PRC specific fluorescence probe.

The two target probes and PRC probe are labeled with a quencher on one end and a fluorophore on the other end. In addition, the two target probes and PRC probe incorporate one or more RNA bases. Upon annealing to RSV, hMPV or PRC amplicons, the fluorescence probes are cleaved by RNaseH2 and the fluorescence signal increases due to physical separation of fluorophore from quencher. Solana measures and interprets the fluorescent signal, using on-board method-specific algorithms. Solana then reports the test results to the user on its display screen, and it can print out the results via an attached printer.

AI/ML Overview

Here's an analysis of the provided text, focusing on the acceptance criteria and study that proves the device meets them:

The document describes the Solana RSV+hMPV Assay, a qualitative in vitro diagnostic test for the detection and differentiation of RSV and hMPV viral RNA. The evaluation of this device primarily focuses on its analytical performance (reproducibility, detection limit, analytical specificity) and clinical performance (comparison to an FDA-cleared molecular assay).

Acceptance Criteria and Reported Device Performance

The acceptance criteria are implicitly defined by the performance benchmarks achieved in the analytical and clinical studies. While explicit numerical acceptance criteria (e.g., "PPA must be >90%") are not stated as a formalized table, the performance results represent the presented evidence of meeting these criteria.

Here's a table summarizing the reported device performance, which serves as evidence of meeting the implicit acceptance criteria:

Acceptance Criterion (Implicit)Reported Device Performance
Analytical Reproducibility: Consistent results across different operators, sites, and days for various concentrations (high negative, low positive, moderate positive samples).RSV A strain A2 (VR-1540):- High Negative (0.2x LOD): 71.1% Overall Agreement (95% CI: 56.6 to 82.3)- Low Positive (1x LOD): 100% Overall Agreement (95% CI: 92.1 to 100)- Moderate Positive (2x LOD): 100% Overall Agreement (95% CI: 92.1 to 100)RSV B strain Wash/18537/62 (VR-1580):- High Negative (0.2x LOD): 48.9% Overall Agreement (95% CI: 30.9 to 58.8)- Low Positive (1x LOD): 100% Overall Agreement (95% CI: 92.1 to 100)- Moderate Positive (2x LOD): 100% Overall Agreement (95% CI: 92.1 to 100)hMPV 20 Type A2 (IA14-2003 G gene):- High Negative (0.2x LOD): 77.8% Overall Agreement (95% CI: 63.7 to 87.5)- Low Positive (1x LOD): 100% Overall Agreement (95% CI: 92.1 to 100)- Moderate Positive (2x LOD): 100% Overall Agreement (95% CI: 92.1 to 100)hMPV 4 Type B2 (Peru1-2002, B2):- High Negative (0.2x LOD): 60.0% Overall Agreement (95% CI: 45.5 to 73.0)- Low Positive (1x LOD): 100% Overall Agreement (95% CI: 92.1 to 100)- Moderate Positive (2x LOD): 100% Overall Agreement (95% CI: 92.1 to 100)Negative control and positive control samples showed 100% agreement. The results suggest no significant differences between users, sites, or days.
Analytical Sensitivity (Limit of Detection - LOD): Ability to detect low concentrations of target viruses.RSV:- RSV A, A2 (VR-1540): 7.9x10^3 TCID50/mL- RSV B, Wash/18537/62 (VR-1580): 3.9x10^2 TCID50/mLhMPV:- hMPV 16 Type A1, IA10-2003: 3.7x10^2 TCID50/mL- hMPV 20 Type A2 IA14-2003 G gene: 1.2x10^4 TCID50/mL- hMPV 5 Type B1, Peru3-2003: 3.8x10^3 TCID50/mL- hMPV 4 Type B2, Peru1-2002: 2.3x10^3 TCID50/mL
Analytical Specificity (Cross-Reactivity): No false positives due to other common respiratory microorganisms or interfering substances.Cross-Reactivity: No cross-reactivity observed with 46 microorganisms (25 bacteria, 1 yeast, 20 viruses) at tested concentrations (10^6 CFU/mL or higher for bacteria/yeast; 10^5 pfu/mL or TCID50/mL or higher for viruses).Interference: No evidence of interference from 20 potentially interfering substances (e.g., mucin, blood, nasal sprays, common medications) when tested with RSV and hMPV at 3x LOD concentrations.
Analytical Reactivity (Inclusivity): Ability to detect different strains of the target viruses.All four additional tested RSV strains (2 RSV A, 2 RSV B) and four additional hMPV strains (1 each of A1, A2, B1, B2) were inclusively detected at concentrations near their LOD.
Clinical Sensitivity/Specificity (Agreement with Comparator): High agreement with an FDA-cleared molecular comparator assay.RSV (Compared to FDA-cleared molecular assay):- Positive Percent Agreement (PPA): 95.5% (95% CI: 91.0 to 97.8) for all specimens (Fresh & Frozen)- Negative Percent Agreement (NPA): 99.9% (95% CI: 91.0 to 97.8) for all specimens (Fresh & Frozen)Note: Discrepant results adjudicated using an alternative FDA-cleared molecular device.hMPV (Compared to FDA-cleared molecular assay):- Positive Percent Agreement (PPA): 95.6% (95% CI: 89.1 to 98.3) for all specimens (Fresh & Frozen)- Negative Percent Agreement (NPA): 99.8% (95% CI: 99.6 to 99.9) for all specimens (Fresh & Frozen)Note: Discrepant results adjudicated using an alternative FDA-cleared molecular device.

Study Details:

The study described is a premarket submission for an in vitro diagnostic device, primarily focused on analytical and clinical performance.

  1. Sample sizes used for the test set and the data provenance:

    • Analytical Reproducibility Test Set:
      • Four-sample panel (3 levels of combined RSV and hMPV, and a negative sample).
      • Panel run by 2 operators at each of 3 testing sites for 5 non-consecutive days.
      • Each sample tested in 3 replicates.
      • Total of 45 results per level for each virus strain (2 operators * 5 days * 3 sites * 3 replicates).
      • Provenance: In-house laboratory (Athens, Quidel Corp), Lab Alliance, and Medical Center of Wisconsin (US based). The study appears to be prospective for this analytical validation.
    • Analytical Sensitivity (LOD) Test Set:
      • Quantified cultures of 6 virus strains (1 RSV A, 1 RSV B, 1 hMPV A1, 1 hMPV A2, 1 hMPV B1, 1 hMPV B2) serially diluted in negative nasopharyngeal matrix.
      • Each dilution run as 20 replicates.
      • Provenance: This is an analytical study using spiked samples, likely conducted internally or by a contracted lab.
    • Analytical Specificity (Cross-Reactivity/Interference) Test Set:
      • 46 microorganisms (25 bacteria, 1 yeast, 20 viruses) diluted in negative nasal matrix.
      • 20 potentially interfering substances evaluated.
      • Provenance: Analytical study using spiked samples.
    • Analytical Reactivity (Inclusivity) Test Set:
      • 4 additional RSV strains and 4 additional hMPV strains.
      • Provenance: Analytical study using spiked samples.
    • Clinical Studies Test Set:
      • Sample Size: 2064 prospectively collected specimens (nasal or nasopharyngeal swabs) initially included.
      • After exclusions (4 protocol deviations, 14 invalid Solana tests), 2046 specimens were included in the final analysis.
      • Provenance: Collected between January and May 2017 at six sites across the United States. A mix of fresh (769) and frozen (1291) specimens. This is a prospective clinical study.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • For this in vitro diagnostic device, "ground truth" is established through laboratory methods, not human expert interpretation of images.
    • For the analytical studies, the "ground truth" is defined by the known concentrations and identities of the spiked viral material or microorganisms.
    • For the clinical study, the "ground truth" is established by "an FDA-cleared RSV+hMPV molecular assay" (the comparator method). This is a laboratory-based gold standard, not a consensus of human expert readers.
    • There is no mention of experts (e.g., radiologists) in the context of establishing ground truth for this type of diagnostic assay.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • For the clinical study, there was an adjudication for discordant results between the Solana assay and the FDA-cleared molecular comparator.
    • Method: For discordant specimens (Solana Positive/Comparator Negative or Solana Negative/Comparator Positive), an "alternative FDA-cleared molecular device" was used for re-testing. This acts as a tie-breaker or confirmatory test.
    • For RSV: 1 Solana+/Comparator- specimen was positive by an alternative device. 6 of 7 Solana-/Comparator+ specimens were positive by an alternative device.
    • For hMPV: All 3 Solana+/Comparator- specimens were negative by an alternative device. All 4 Solana-/Comparator+ specimens were positive by an alternative device.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is not an imaging-based AI device. It is a molecular diagnostic assay for detecting viral RNA. Therefore, an MRMC study and evaluation of human reader improvement with AI assistance are not applicable. The device's performance is standalone or compared to another laboratory diagnostic method.
  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Yes, the performance characteristics (analytical and clinical) presented are for the Solana RSV+hMPV Assay system (device + instrument) operating qualitatively.
    • The "Solana instrument measures and interprets the fluorescent signal, using on-board method-specific algorithms" and "reports the test results to the user on its display screen." This indicates that the results are generated by the algorithm/instrument system itself without real-time human interpretation affecting the qualitative positive/negative result. Humans perform the sample preparation and loading, and read the final result from the instrument.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Analytical Studies: "Ground truth" was established by spiking known quantities of specific viral strains/microorganisms into negative matrix, representing a controlled laboratory standard.
    • Clinical Studies: "Ground truth" was established by an FDA-cleared RSV+hMPV molecular assay (the comparator device), and for discordant results, a different "alternative FDA-cleared molecular device" was used as a confirmatory ground truth. This is a reference standard based on a previously validated laboratory test.
  7. The sample size for the training set:

    • The document describes performance studies (analytical and clinical validation) for a finished device. It does not specify a separate "training set" in the context of machine learning model development. This is typical for traditional in vitro diagnostic device submissions, where the focus is on validation studies rather than the development lifecycle of an "AI" algorithm. The "Solana instrument" has "on-board method-specific algorithms," but how these algorithms were developed or "trained" (if indeed they involve learning from data in a modern ML sense) is not described in this document. Often, such "algorithms" in traditional IVDs are rule-based or statistical models rather than deep learning models.
  8. How the ground truth for the training set was established:

    • As no explicit "training set" (in the machine learning sense) is mentioned, there is no information on how its ground truth would have been established. The document focuses on the validation of the device's performance against established laboratory standards and a clinical comparator.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing to the right, stacked on top of each other. The profiles are connected and appear to be emerging from a single form.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 16, 2017

Quidel Corporation Ronald H. Lollar Sr. Director, Clinical, Regulatory and Scientific Affairs 2005 East State Street, Suite 100 Athens, OH 45701

Re: K171974

Trade/Device Name: Solana RSV+hMPV Assay Regulation Number: 21 CFR 866.3980 Regulation Name: Respiratory viral panel multiplex nucleic acid assay Regulatory Class: II Product Code: OCC Dated: September 15, 2017 Received: September 18, 2017

Dear Mr. Lollar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tamara V. Feldblyum -S for

Uwe Scherf, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K171974

Device Name Solana RSV+hMPV Assay

Indications for Use (Describe)

The Solana RSV+hMPV Assay is a qualitative in vitro diagnostic test for the detection and differentiation of RSV and hMPV viral RNA in nasal and nasopharyngeal swabs from patients with signs and symptoms of respiratory infection. This test is intended for use as an aid in the differential diagnosis of RSV and hMPV viral infections in conjunction with clinical and epidemiological risk factors. This test is not intended to differentiate the two subtypes of RSV or the four genetic sub-lineages of hMPV.

Negative results do not preclude RSV infection and/or hMPV infection and should not be used as the sole basis for diagnosis, treatment or other patient management decisions.

Conversely, positive results do not rule-out bacterial infection with other viruses. The agent detected may not be the definite cause of disease. The use of additional laboratory testing and clinical presentation must be considered in order to obtain the final diagnosis of respiratory viral infection.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Applicant:

Quidel Corporation 12544 High Bluff Drive, Suite 200 San Diego, California 92130 Telephone: 858-552-7910 Fax: 858-646-8045

Contact Information:

Ronald H. Lollar, Senior Director Clinical and Regulatory Affairs 2005 East State Street, Suite 100 Athens, Ohio 45701 740-589-3300 - Corporate number 740-589-3373 – Desk phone 858-552-6451—Fax Ron.Lollar@quidel.com

Date of preparation of 510(k) summary:

June 29, 2017

A. 510(k) Number:

K171974

B. Purpose for Submission:

To obtain substantial equivalence for the Solana "RSV+hMPV Assay when performed on the Solana

C. Measurand:

RSV: Matrix Gene; hMPV: Fusion Protein Gene

D. Type of Test:

Reverse Transcriptase - Helicase-Dependent Amplification (RT-HDA)

{4}------------------------------------------------

E. Applicant:

Quidel Corporation

F. Proprietary and Established Names:

Solana RSV+hMPV Assay

G. Regulatory Information:

Table 1.Regulatory Information
Product CodeClassificationRegulation SectionPanel
OCCClass II21 CFR 866.3980 Respiratoryviral panel multiplex nucleic acidassayMicrobiology(83)
OEM21 CFR 866.3980 HumanMetapneumovirus (Hmpv) RnaAssay System

H. Intended Use:

1. Intended Use(s):

The Solana® RSV+hMPV Assay is a qualitative in vitro diagnostic test for the detection and differentiation of RSV and hMPV viral RNA in nasal and nasopharyngeal swabs from patients with signs and symptoms of respiratory infection. This test is intended for use as an aid in the differential diagnosis of RSV and hMPV viral infections in humans in conjunction with clinical and epidemiological risk factors. This test is not intended to differentiate the two subtypes of RSV or the four genetic sub-lineages of hMPV.

Negative results do not preclude RSV infection and/or hMPV infection and should not be used as the sole basis for diagnosis, treatment or other patient management decisions.

Conversely, positive results do not rule-out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease. The use of additional laboratory testing and clinical presentation must be considered in order to obtain the final diagnosis of respiratory viral infection.

{5}------------------------------------------------

2. Indication(s) for Use:

Same as Intended Use

3. Special conditions for use statement(s):

  • For in vitro diagnostic use only
  • For prescription use only

4. Special instrument requirements:

Solana® Instrument

l. Device Description:

The Solana RSV+hMPV Assay amplifies and detects viral RNA present in viral transport media containing nasopharyngeal or nasal swab specimens obtained from symptomatic patients.

The assay consists of two major steps: 1) specimen preparation, and 2) amplification and detection of target sequences specific to RSV and/or hMPV using isothermal Reverse Transcriptase - Helicase-Dependent Amplification (RT-HDA) in the presence of targetspecific fluorescence probes.

A patient nasal or nasopharyngeal swab specimen in viral transport media is transferred to a Process Buffer Tube, subjected to heat treatment at 95°C for 5 minutes and mixed. The processed sample is transferred to a Reaction Tube. The Reaction Tube contains lyophilized RT-HDA reagents, dNTPs, primers and probes. Once rehydrated with the processed sample, the Reaction Tube is placed in Solana for amplification and detection of RSV and hMPV specific target sequences. In Solana, the target sequences are amplified by RSV and hMPV specific primers and detected by RSV and hMPV specific fluorescence probes, respectively. A process control (PRC) is included in the Process Buffer Tube to monitor sample processing, inhibitory substances in clinical samples, reagent failure or device failure. The PRC target is amplified by RSV and hMPV specific primers and detected by a PRC specific fluorescence probe.

The two target probes and PRC probe are labeled with a quencher on one end and a fluorophore on the other end. In addition, the two target probes and PRC probe incorporate one or more RNA bases. Upon annealing to RSV, hMPV or PRC amplicons, the fluorescence probes are cleaved by RNaseH2 and the fluorescence signal increases due to physical separation of fluorophore from quencher. Solana measures and interprets the

{6}------------------------------------------------

fluorescent signal, using on-board method-specific algorithms. Solana then reports the test results to the user on its display screen, and it can print out the results via an attached printer.

Materials Provided:

Solana® RSV+hMPV Assay Kit: M306 48 Tests per kit

Table 2.Kit Components
ComponentQuantityStorage
Process Buffer48 tubes/kit 1.55 mL2°C to 8°C
Reaction Tubes48 tubes/kit2°C to 8°C

Materials required but not provided:

  • . External controls for RSV and hMPV (e.g. Solana RSV+hMPV Control Set, which contains positive and negative controls, serves as an external processing control)
  • Sterile DNase-free filter-blocked positive displacement micropipettor tips ●
  • Micropipettor
  • Stopwatch or timer
  • Scissors or a blade
  • Workflow tray
  • Transfer Rack
  • Heat block capable of 95 ± 2°C temperature ●
  • Thermometer
  • Solana instrument ●
  • Transport Media (BD/Copan UTM, Remel M4, Remel M4RT, Remel M5, Remel M6, ● or Copan eSwab)

J. Substantial Equivalence Information:

    1. Predicate device name(s):
      Lyra® RSV+hMPV Assay
    1. Predicate 510(k) number(s): K131813

{7}------------------------------------------------

3. Comparison with predicate:

ItemSolana® RSV+hMPV AssayLyra® RSV+hMPV Assay (K131813)
Intended UseThe Solana® RSV+hMPV Assay is aqualitative in vitro diagnostic test forthe detection and differentiation of RSVand hMPV viral RNA in nasal andnasopharyngeal swabs from patientswith signs and symptoms of respiratoryinfection. This test is intended for useas an aid in the differential diagnosis ofRSV and hMPV viral infections inhumans in conjunction with clinical andepidemiological risk factors. This test isnot intended to differentiate the twosubtypes of RSV or the four geneticsub-lineages of hMPV.Negative results do not preclude RSVinfection and/or hMPV infection andshould not be used as the sole basis fordiagnosis, treatment or other patientmanagement decisions.Conversely, positive results do not rule-out bacterial infection or co-infectionwith other viruses. The agent detectedmay not be the definite cause ofdisease. The use of additionallaboratory testing and clinicalpresentation must be considered inorder to obtain the final diagnosis ofrespiratory viral infection.The Lyra RSV + HMPV Assay is amultiplex Real-Time PCR (RT-PCR)assay for the qualitative detectionand identification of respiratorysyncytial virus (RSV) and humanmetapneumovirus (hMPV)ribonucleic acid (RNA) extractedfrom nasal and nasopharyngealswab specimens from patientswith signs and symptoms ofrespiratory infection. This in vitrodiagnostic test is intended to aidin the differential diagnosis of RSVand hMPV infections in humans inconjunction with clinical andepidemiological risk factors. Thistest is not intended todifferentiate the two subtypes ofRSV or the four genetic sub-lineages of hMPV.Negative results do not precludeRSV infection and/or hMPVinfection and should not be usedas the sole basis for diagnosis,treatment or other patientmanagement decisions.Conversely, positive results do notrule-out bacterial infection or co-infection with other viruses. Theagent detected may not be thedefinite cause of disease. The use
Table 3.Similarities
ItemSolana® RSV+hMPV AssayLyra® RSV+hMPV Assay (K131813)
of additional laboratory testingand clinical presentation must beconsidered in order to obtain thefinal diagnosis of respiratory viralinfection.The Quidel Molecular RSV + hMPVAssay can be performed usingeither the Life TechnologiesQuantStudio™Dx RT-PCRInstrument, the AppliedBiosystems® 7500 Fast Dx RT-PCRInstrument, or the CepheidSmartCycler® II System.
Sample Typesnasal swab and nasopharyngeal swabSame
DetectionTechniquesAutomated multiplex assay usingdifferent reporter dyes for each targetSame

{8}------------------------------------------------

Table 4. Differences
ItemSolana® RSV+hMPV AssayLyra® RSV+hMPV Assay (K131813)
Viral TargetRSV: Matrix Gene;hMPV: Fusion ProteinRSV: L viral polymerase and NS2 geneshMPV: RNA polymerase gene
AmplificationTechnologyReverse Transcriptase - Helicase-Dependent Amplification (RT-HDA)Real Time PCR-based system fordetecting the presence or absenceof viral RNA in clinical specimens
ExtractionMethodsNonebioMérieux easyMAG® AutomatedMagnetic Extraction Reagents
InstrumentSolana®Life Technologies QuantStudio®Dx, the Applied Biosystems® 7500Fast Dx, or the CepheidSmartCycler® II

{9}------------------------------------------------

K. Standard/Guidance Document Referenced (if applicable):

Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ ucm180307.htm

Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays – https://www.fda.gov/RegulatoryInformation/Guidances/ucm180308.htm

Guidance for Industry and FDA Staff: Statistical Guidance on Reporting Results from Studies Evaluating Diagnostic Tests (Final, 3/13/2007) http://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidance/GuidanceD ocuments/ucm071287.pdf

Guidance on Informed Consent for In Vitro Diagnostic Device Studies Leftover Human Specimens that are Not Individually Identifiable (April 2006) – http://www.fda.gov/cdrh/oivd/guidance/1588.pdf.

Guidance for Industry and Food and Drug Administration Staff - eCopy Program for Medical Device (December 2012)

http://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidance/GuidanceD ocuments/UCM313794.pdf

L. Test Principle:

The Solana RSV+hMPV Assay amplifies and detects viral RNA present in viral transport media containing nasopharyngeal or nasal swab specimens obtained from symptomatic patients.

The assay consists of two major steps: 1) specimen preparation, and 2) amplification and detection of target sequences specific to RSV and/or hMPV using isothermal Reverse

{10}------------------------------------------------

Transcriptase - Helicase-Dependent Amplification (RT-HDA) in the presence of targetspecific fluorescence probes.

A patient nasal or nasopharyngeal swab specimen in viral transport media is transferred to a Process Buffer Tube, subjected to heat treatment at 95°C for 5 minutes and mixed. The processed sample is transferred to a Reaction Tube. The Reaction Tube contains lyophilized RT-HDA reagents, dNTPs, primers and probes. Once rehydrated with the processed sample, the Reaction Tube is placed in Solana for amplification and detection of RSV and hMPV specific target sequences. In Solana, the target sequences are amplified by RSV and hMPV specific primers and detected by RSV and hMPV specific fluorescence probes, respectively. A competitive process control (PRC) is included in the Process Buffer Tube to monitor sample processing, inhibitory substances in clinical samples, reagent failure or device failure. The PRC target is amplified by RSV and hMPV specific primers and detected by a PRC specific fluorescence probe.

The two target probes and PRC probe are labeled with a quencher on one end and a fluorophore on the other end. In addition, the two target probes and PRC probe incorporate one or more RNA bases. Upon annealing to RSV, hMPV or PRC amplicons, the fluorescence probes are cleaved by RNaseH2 and the fluorescence signal increases due to physical separation of fluorophore from quencher. Solana measures and interprets the fluorescent signal, using on-board method-specific algorithms. Solana then reports the test results to the user on its display screen, and it can print out the results via an attached printer.

M. Performance Characteristics:

    1. Analytical performance:
    • a. Precision/Reproducibility:

Reproducibility

A four-sample panel consisting of three levels of a combined RSV and hMPV (two (2) strains of each virus) contrived samples and a negative contrived sample were tested in this study. RSV A strain A2 (VR-1540) and hMPV 20 Type A2 (IA14-2003 G gene) (Set 1), or RSV B strain Wash/18537/62 (VR-1580) and hMPV 4 Type B2 (Peru1-2002, B2) (Set 2) were diluted in negative nasal matrix to 2x LOD for moderate positive, 1x LOD for low positive and diluted to C20 to C80 for high negative / low positive. Negative nasal matrix without spiked virus was used for the negative sample. Positive and negative controls were run in triplicate along with the panels. The panels were run by two operators at each testing site for five (5) nonconsecutive days. The study was completed by the in-house laboratory (Athens facility of Quidel Corp), Lab Alliance, and Medical Center of Wisconsin. The Solana RSV+hMPV assay was used per the instructions for use.

{11}------------------------------------------------

Panels and controls were tested at each site by two operators per instrument for five days, each sample tested in three (3) replicates, for a total of 45 results per level for each virus strain (2 operators x 5 days x 3 sites x 3 replicates).

Table 5.Reproducibility Summary
SITEOverall Percent Agreement95% Confidence Interval
Site #1Site #2Site #3
# Expected Result/# tested% Agreement with Expected Result# Expected Result/# tested% Agreement with Expected Result# Expected Result/# tested% Agreement with Expected Result
RSV A strain A2(VR-1540)High Negative (0.2x LOD(1.6x103 TCID50/mL)*8/1553.314/1593.310/1566.732/4571.156.6 to 82.3
RSV A strain A2(VR-1540)Low Positive (1x LOD)(7.9x103 TCID50/mL)15/1510015/1510015/1510045/4510092.1 to 100
RSV A strain A2(VR-1540)Moderate Positive(2x LOD)(1.6x104 TCID50/mL)15/1510015/1510015/1510045/4510092.1 to 100
RSV B strainWash/18537/62(VR-1580)High Negative (0.2x LOD(1.4 x102 TCID50/mL)*6/1540.010/1566.76/154022/4548.930.9 to 58.8
RSV B strainWash/18537/62(VR-1580)Low Positive (1x LOD)(4.7x102 TCID50/mL)15/1510015/1510015/1510045/4510092.1 to 100
RSV B strainWash/18537/62(VR-1580)Moderate Positive(2x LOD)(9.4x102 TCID50/mL)15/1510015/1510015/1510045/4510092.1 to 100
Negative0/301000/301000/301000/9010096.0 to 100
RSV Positive Control30/3010030/3010030/3010090/9010096.0 to 100
RSV Negative Control0/301000/301000/301000/9010096.0 to 100
Table 5: Reproducibility Summary
SITE
Site #1Site #2Site #3
# ExpectedResult/#tested% Agreementwith ExpectedResult# ExpectedResult/#tested% Agreementwith ExpectedResult# ExpectedResult/#tested% Agreementwith ExpectedResultOverall PercentAgreementWith Expected Results95%ConfidenceInterval
hMPV 20 Type A2(IA14-2003 G gene)High Negative (0.2x LOD(2.4x102 TCID50/mL)*11/1573.314/1593.310/1566.735/4577.863.7 to 87.5
hMPV 20 Type A2(IA14-2003 G gene)Low Positive (1x LOD)(1.2x104 TCID50/mL)15/1510015/1510015/1510045/4510092.1 to 100
hMPV 20 Type A2(IA14-2003 G gene)Moderate Positive(2x LOD)(2.4x104 TCID50/mL)15/1510015/1510015/1510045/4510092.1 to 100
hMPV 4 Type B2(Peru1-2002, B2)High Negative (0.2x LOD(4.6x102 TCID50/mL)*9/15609/15609/156027/4560.045.5 to 73.0
hMPV 4 Type B2(Peru1-2002, B2)Low Positive (1x LOD)(2.3x103 TCID50/mL)15/1510015/1510015/1510045/4510092.1 to 100
hMPV 4 Type B2(Peru1-2002, B2)Moderate Positive(2x LOD)(4.6x103 TCID50/mL)15/1510015/1510015/1510045/4510092.1 to 100
Negative0/301000/301000/301000/9010096.0 to 100
hMPV Positive Control30/3010030/3010030/3010090/9010096.0 to 100
hMPV Negative Control0/301000/301000/301000/9010096.0 to 100
  • An expected result for the high negative sample is a negative result.

{12}------------------------------------------------

Solana® RSV + hMPV Assay 6/29/2017 Page 10 of 20

510(k) Summary

  • An expected result for the high negative sample is a negative result.

The results suggest that there are no significant differences between different users and different sites on different days.

  • b. Linearity/assay reportable range:
    Not applicable – This assay is qualitative.

{13}------------------------------------------------

  • Traceability, Stability, Expected values (controls, calibrators, or methods): C.
    Traceability:

Not applicable. This assay is qualitative.

Specimen Stability:

RSV (RSV A, A2 (VR-1540) and RSV B, Wash/18537/62 (VR-1580)) and hMPV (hMPV 16 Type A1, IA10-2003 and hMPV 5 Type B1, Peru3-2003) were formulated in six (6) transport media pooled negative matrix (Copan UTM, Remel M4, Remel M5, Remel M6, Remel M4RT or Copan ESwab transport media) at a final concentration of 2 to 3x LOD level.

The transport media systems containing the contrived samples were stored at 2° to 8°C up to 9 days or -70°C up to 5 weeks. The samples were processed per the instructions for use. Each transport media was tested in 3 replicates at Day 0, 24 hours, 48 hours, 72 hours, Day 7 and Day 9 for 2 to 8°C storage or Day 0, week 1, week 2, week 3, week 4 and week 5 for -70°C storage.

RSV and hMPV are stable in transport media BD UTM, Remel M4, Remel M4RT, Remel M5, Remel M6 and Copan eSwab at 2° to 8°C for up to 8 days and at -70°C for up to 10 weeks.

Controls:

Controls (Solana RSV+hMPV Control Set, which contains positive and negative controls and serves as an external processing control) were run on the Solana RSV+hMPV Assay each day of testing. These controls are described as follows:

  • a. The process control is used to monitor sample processing, to detect HDA inhibitory specimens, to confirm the integrity of assay reagents and the operation of the Solana instrument. The process control is included in the Reaction Mix tube.
  • b. The external positive control may be treated as a patient specimen. The control should be sampled and tested as if it were a patient specimen and processed as described above in the Assay Procedure. The external positive control is intended to monitor substantial reagent and instrument failure.

{14}------------------------------------------------

  • The external negative control may be treated as a patient specimen. The C. control should be sampled and tested as if it were a patient specimen and processed as described above in the Assay Procedure. The external negative control is used to detect reagent or environmental contamination (or carryover) by RSV and hMPV RNA or amplicon.
    It is recommended that the reactivity of each new lot and each new shipment of the Solana RSV+hMPV Assay be verified on receipt and before use. External control tests should be performed thereafter in accordance with appropriate federal, state and local guidelines. The Solana RSV+hMPV Assay should not be used in patient testing if the external controls do not produce the correct results.

  • d. Detection limit:
    The analytical sensitivity (limit of detection or LOD) of the Solana RSV+hMPV Assay was determined using quantified (TCID50/mL) cultures of one RSV A, one RSV B, one hMPV A1, one hMPV A2, one hMPV B1 and one hMPV B2 strain, serially diluted in negative nasopharyngeal matrix. Each dilution was run as 20 replicates in the Solana RSV+hMPV assay. Analytical sensitivity (LOD) is defined as the lowest concentration at which at least 95% of all replicates tested positive. The demonstrated LOD for each strain tested is shown below:

VirusTCID50/mL
RSV
RSV A, A2 (VR-1540)7.9x103
RSV B, Wash/18537/62 (VR-1580)3.9x102
hMPV
hMPV 16 Type A1, IA10-20033.7x102
hMPV 20 Type A2 IA14-2003 G gene1.2x104
hMPV 5 Type B1, Peru3-20033.8x103
hMPV 4 Type B2, Peru1-20022.3x103
  • e. Analytical specificity:

Cross Reactivity:

A study was performed to evaluate the cross-reactivity of the Solana RSV+hMPV Assay with forty-six (46) microorganisms (25 bacteria, 1 yeast, 20 viruses) potentially

{15}------------------------------------------------

found in specimens that are collected from patients symptomatic for RSV and/or hMPV. Each microorganism was diluted in negative nasal matrix to the desired concentration (10 or higher CFU/mL for bacteria, yeast and 105 or higher pfu/mL or TCID50/mL for viruses) and tested with the Solana RSV+hMPV Assay. No cross reactivity was observed with the organisms at concentrations shown in the table below.

Table 7. Potential Cross-reactive Organisms
OrganismConcentration TestedUnits
Adenovirus 11.00E+05TCID50/mL
Adenovirus 111.00E+05TCID50/mL
Bordetella bronchiseptica1.00E+06CFU/mL
Bordetella pertussis1.00E+06CFU/mL
Candida albicans1.00E+06CFU/mL
Chlamydophila pneumoniae1.00E+06IFU/mL
Chlamydia trachomatis1.00E+06IFU/mL
Coronavirus 229E1.00E+05TCID50/mL
Corynebacterium diptheriae1.00E+06CFU/mL
Coxsackievirus B5/10/20061.00E+05TCID50/mL
Cytomegalovirus (VR-977)1.00E+05TCID50/mL
Echovirus 111.00E+05TCID50/mL
Echovirus 61.00E+05TCID50/mL
Enterovirus, Type 711.00E+05TCID50/mL
Epstein Barr virus1.00E+05TCID50/mL
Escherichia coli1.00E+06CFU/mL
Haemophilus influenzae1.00E+06CFU/mL
HSV 2 G strain1.00E+05TCID50/mL
hMPV Peru1-2002, B211.00E+05TCID50/mL
Influenza A/Texas/50/20121.00E+05TCID50/mL
Influenza B/Panama/45/901.00E+05TCID50/mL
Klebsiella pneumoniae1.00E+06CFU/mL
Lactobacillus plantarum1.00E+06CFU/mL
Legionella pneumophila1.00E+06CFU/mL
Measles1.00E+05TCID50/mL
Moraxella catarrhalis1.00E+06CFU/mL
Mumps1.00E+05TCID50/mL
Table 7. Potential Cross-reactive Organisms
OrganismConcentration TestedUnits
Mycobacterium avium1.00E+06CFU/mL
Mycobacterium tuberculosis1.00E+06CFU/mL
Mycoplasma pneumoniae1.00E+06CFU/mL
Neisseria gonorrhoeae1.00E+06CFU/mL
Neisseria meningitidis1.00E+06CFU/mL
Parainfluenza Type 11.00E+05TCID50/mL
Parainfluenza Type 21.00E+05TCID50/mL
Parainfluenza Type 31.00E+05TCID50/mL
Proteus mirabilis1.00E+06CFU/mL
Proteus vulgaris1.00E+06CFU/mL
Pseudomonas aeruginosa1.00E+06CFU/mL
Rhinovirus Type 71.00E+05TCID50/mL
RSV A2 (VR-1540)²1.00E+05TCID50/mL
Staphylococcus aureus1.00E+06CFU/mL
Staphylococcus epidermidis1.00E+06CFU/mL
Streptococcus mutans1.00E+06CFU/mL
Streptococcus pneumoniae1.00E+06CFU/mL
Streptococcus pyogenes1.00E+06CFU/mL
Streptococcus salivarius1.00E+06CFU/mL

{16}------------------------------------------------

1All three replicates tested positive for hMPV and negative for RSV in the Solana RSV+hMPV assay. ²All three replicates tested negative for hMPV and positive for RSV in the Solana RSV+hMPV assay No cross-reactivity was observed with the forty-six (46) microorganisms (25 bacteria, 1 yeast, 20 viruses) tested with the Solana® RSV+hMPV Assay.

Interference:

The performance of Solana RSV+hMPV Assay was evaluated with twenty (20) potentially interfering substances that may be present in nasal and nasopharyngeal specimens. The potentially interfering substances were evaluated with RSV (RSV A, A2 (VR-1540) and RSV B, Wash/18537/62 (VR-1580)) and hMPV (hMPV 16 Type A1, IA10-2003 and hMPV 5 Type B1, Peru3-2003) at concentrations of 3x LOD. There was no evidence of interference caused by the substances tested at the concentrations shown below.

Table 8.Interfering Substances
SubstanceActive IngredientConcentration Tested
Table 8. Interfering Substances
SubstanceActive IngredientConcentration Tested
Purified mucin proteinMucin protein2.5 mg/mL
Blood (human)Blood5.0%
Afrin - nasal sprayOxymetazoline5.0%
Saline nasal spraySodium Chloride15.0%
Neo-SynephrinePhenylephrine hydrochloride15.0%
FlonaseFluticasone5.0%
Zicam Gentle Allergy ReliefNasalGelGalphimia glauca, Histaminumhydrochloricum, Luffa operculata,Sulfur5.0%
BactrobanMupirocin12.0 mg/mL
TamiFluOseltamivir$2.2 \mu g/mL$
RelenzaZanamivir282.0 ng/mL
TobramycinTobramycin Sulfate2.5 mg/mL
Chloraseptic sprayBenzocaine, Menthol0.68 g/mL
SYMMETREL®Amantadine hydrochloride282.0 ng/mL
Nasocort Allergy 24 hourTriamcinolone5.0%
Sinus Buster Nasal SprayCapsicum annuum (Capsaicin)5.0%
NasalCrom Nasal AllergySprayCromolyn Sodium5.0%
RhinocortBudesonide (Glucocorticoid)5.0%
Air-Vita Allergy Multi-Symptom ReliefAllium cepa, Ambrosiaartemisiaefolia, Apis mellifica,Chamomilla, Eucalyptol, Eucalyptusglobulus, Euphrasia officinalis,Galphimia glauca, Histaminumhydrochloricum, Natrummuriaticum, Nux vomica, Quercusrobur, Silicea, Wyethia helenioides5.0%
Atrovent® Nasal SprayIpratropium bromide10.0 mg/mL
Table 8.Interfering Substances
SubstanceActive IngredientConcentration Tested
Patanase Nasal SprayOlopatadine hydrochloride10.0 mg/mL

{17}------------------------------------------------

{18}------------------------------------------------

Analytical Reactivity (Inclusivity):

The reactivity of the Solana® RSV+hMPV Assay was evaluated against four (4) additional strains of RSV, which include two (2) RSV A and two (2) RSV B and four (4) additional strains of hMPV, which include one (1) each of hMPV A2, hMPV B1 and hMPV B2 at concentrations near the level of detection (LOD) of the assay.

Table 9.Inclusivity Strains
StrainTCID50/mLInclusive (Yes or No)
RSV
RSV A, strain Long(VR-26)$1.6x10^4$Yes
RSV A, strain 4/2015 Isolate #1$1.6x10^4$Yes
RSV B, strain 9320(VR-955)$7.9x10^2$Yes
RSV B, strain WV/14617/85 (VR-1400)$7.9x10^2$Yes
hMPV
hMPV 9 Type A1, strain IA3-2002$7.4x10^2$Yes
hMPV 27 Type A2, strain IA27-2004$2.4x10^4$Yes
hMPV 3 Type B1, strain Peru2-2002$7.6x10^3$Yes
hMPV 18 Type B2, strain IA18-2003$4.5x10^3$Yes
  • f. Assay cut-off:
    Not applicable.

    1. Comparison studies:
    • a. Method comparison with predicate device:

Not applicable

  • b. Matrix comparison:
    Not applicable

{19}------------------------------------------------

3. Clinical studies:

  • a. Clinical Sensitivity:
    Performance characteristics of the Solana RSV+hMPV Assay were established during a prospective study with specimens collected between January and May 2017. Two thousand sixty-four (2064) specimens prospectively collected specimens have been included in this study at six (6) sites across the United States. Specimens were tested fresh (773) or after freezing (1291) at -70°C. A single nasal or nasopharyngeal swab specimen (300 and 1760, respectively) was collected per patient in viral transport media (BD/Copan UTM, Remel M5, Remel M6). Four (4) of the fresh specimens were removed from the study due to protocol deviations (inappropriate specimen type). All specimens were transported to a central location for extraction with the NucliSENS® easyMAG® and testing with a FDA-cleared RSV+hMPV molecular assay. The specimens were processed and tested with Solana RSV+hMPV Assay (frozen (1291) and fresh (769)) on the Solana instrument at one of the six (6) sites.

COMPARISON WITH A FDA-CLEARED RSV+HMPV MOLECULAR ASSAY

Two thousand sixty (2060) specimens were processed using the NucliSENS® easyMAG® and tested with a FDA-cleared RSV+hMPV molecular assay per the assay's package insert.

Of 2060 specimens evaluated in the study, 769 specimens were tested fresh (nonfrozen) using the Solana RSV+hMPV Assay and the Lyra RSV+hMPV Assay for the presence of RSV and hMPV. One thousand two hundred ninety-one (1291) specimens were frozen after collection and stored at -70°C prior to testing with the Solana® RSV+hMPV Assay and the Lyra RSV+hMPV Assay for the presence of RSV and hMPV. Fourteen (14) specimens (9 fresh and 5 frozen) were invalid in the Solana® Assay when initially tested and upon repeat testing (invalid rate of 0.7%, with 95% Cl 0.4% to 1.1%). These fourteen (14) specimens have been excluded from further analysis.

Table 10 details the positive percent agreement (PPA) and the negative percent agreement (NPA) of the Solana RSV+hMPV Assay results for RSV, as compared with an FDA cleared molecular comparator, for the remaining two thousand forty-six (2046) specimens.

{20}------------------------------------------------

Table 10. Percent Agreement of the Solana® RSV+hMPV Assay for RSV Comparedto an FDA cleared RSV+hMPV Molecular Assay (Across all Sites Combined)
SourceCategoryNTPFPTNFNPPA(95% CI)NPA(95% CI)
Fresh760120747192.3(66.7 to 98.6)100(99.5 to 100)
Frozen128613611143695.8(91.1 to 98.0)99.9(99.5 to 100)
All204614811890795.5(91.0 to 97.8)99.9(91.0 to 97.8)

There were a total of eight (8) discordant specimens among the two thousand forty-six (2046) specimens evaluated. The one (1) discordant specimens (Solana Positive/Comparator Negative) reported in Table 10 was positive by an alternative FDA-cleared molecular device. Of the seven (7) discordant specimens (Solana Negative/ Comparator Positive) reported in Table 10, six (6) of these specimens were positive by an alternative FDA-cleared molecular device.

Table 11 details the positive percent agreement (PPA) and the negative percent agreement (NPA) of the Solana RSV+hMPV Assay results for hMPV, as compared with an FDA cleared molecular comparator, for the remaining two thousand forty-six (2046) specimens.

Table 11.Percent Agreement of the Solana® RSV+hMPV Assay for hMPV Comparedto an FDA-cleared RSV+hMPV Molecular Assay (Across all Sites Combined)
SourceCategoryNTPFPTNFNPPA(95% CI)NPA(95% CI)
Fresh760242733196.0(80.5 to 99.3)99.7(99.0 to 99.9)
Frozen12866211220395.4(87.3 to 98.4)99.9(99.5 to 100)
All20468631953495.6(89.1 to 98.3)99.8(99.6 to 99.9)

There were a total of seven (7) discordant specimens among the two thousand forty-six (2046) specimens evaluated. Of the three (3) discordant specimens (Solana Positive/ Comparator Negative) reported in Table 11, all of these specimens were negative by an alternative FDA-cleared molecular device. Of the four (4) discordant specimens (Solana Negative/ Comparator Positive) reported in Table 11, all of these specimens were positive by an alternative FDA-cleared molecular device.

  • b. Clinical specificity:
    See Section 3a.

  • Other clinical supportive data (when a. and b. are not applicable): C.
    Not applicable

{21}------------------------------------------------

4. Clinical cut-off:

Not applicable

    1. Expected values:
      The expected values of the Solana RSV+hMPV Assay were established during a prospective study conducted between January and May 2017. Two thousand sixty-four (2064) specimens (fresh (773) and frozen (1291)) have been included in this study at six (6) sites across the United States. Four (4) of the fresh specimens were removed from the study due to protocol deviations (inappropriate specimen type). A single specimen was collected per patient. The specimens were processed and tested with Solana RSV+hMPV Assay on the Solana instrument at the sites.

The expected value of RSV and hMPV with the Solana RSV+hMPV Assay has been calculated for the combined sites based on the age of the patient.

Eighteen (18) of the two thousand sixty-four (2064) specimens were removed from analysis: (four (4) specimens were removed from the study due to protocol deviations (inappropriate specimen type); fourteen (14) specimens were invalid). Table 12 provides the percentage of RSV and hMPV positive cases per specified age group, as determined by the Solana RSV+hMPV Assay, for the remaining two thousand forty-six (2046) specimens.

Table 12. Expected Values (N=2046) (January to May 2017)
RSVhMPV
Age GroupNumber of PatientsNumber of PositivesPositive Detection rateNumber of PatientsNumber of PositivesPositive Detection rate
<1 year2354619.6%235156.4%
1 to 5 years3904110.5%390307.7%
6 to 10 years18663.2%186115.9%
11 to 15 years12554.0%12543.2%
16 to 21 years10921.8%10900.0%
22 to 50 years358113.1%358113.1%
51 to 65 years259155.8%25972.7%
> 65 years384236.0%384112.9%
Combined Age Groups20461497.3%2046894.3%

N. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:

Instrument: Solana Instrument

{22}------------------------------------------------

O. System Descriptions:

1. Modes of Operation:

The Solana instrument heats each reaction tube to 58°C. If present, the target RNA sequence is reverse transcribed into cDNA by the Reverse Transcriptase and RSV or hMPV specific primers that are present in the reaction mix. After the completion of the Reverse Transcriptase step, the Solana instrument heats each reaction tube to 65°C where the isothermal DNA polymerase amplifies the cDNA strands using the RSV or hMPV specific primers. RSV or hMPV specific fluorescence probes are also included in the Reaction Tube. The target probes are labeled with a quencher on one end and a fluorophore specific for the recognition sequence on the other end. In addition, the target probes carry a ribonucleic acid. Upon annealing to amplicons, the fluorescence probes are cleaved by RNaseH2 and the fluorescence signal increases due to physical separation of fluorophore from quencher. The Solana instrument measures and interprets the fluorescent signal, using on-board method-specific algorithms. Solana instrument will then report the test results to the user on its display screen, and it can print out the results via a printer.

2. Software:

FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:

Yes Yes × No

P. Proposed Labeling:

The labeling satisfies the requirements of 21 CFR Part 809.10, 21 CFR 801.109, and the special controls.

§ 866.3980 Respiratory viral panel multiplex nucleic acid assay.

(a)
Identification. A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B;
(2) Influenza A subtype H1 and Influenza A subtype H3;
(3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B;
(4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus;
(5) Human Metapneumovirus;
(6) Rhinovirus; and
(7) Adenovirus.
(b)
Classification. Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.