K Number
K171699
Date Cleared
2017-07-28

(50 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Intuitive Surgical Endoscopic Instrument Control System, Model 184000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparosopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Surgical System Model IS3000) is intended to r no multive accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp assist in the dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation approximation against, orceral procedures, general laparoscopic surgical procedures, gynecologic probos and acoussent anning actrage tal otolaryngology surgical procedures restricted to benign and malignant itumors classified as T1 and T2, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy rumors clasmed as 1 x and 12, gateda adjunctive mediastinotomy to perform coronary anastomosis during procedurer The system is indicated for adult and pediatric use. It is intended to be used by trained earchies in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Surgical System, Model 184200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparosopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

Device Description

The da Vinci Surgical Systems (Models IS3000, IS4000 and IS4200) are software-controlled, electromechanical systems designed for surgeons to perform minimally invasive surgery. The da Vinci Surgical Systems consist of a Surgeon Console, a Patient Side Cart (PSC), and a Vision Side Cart (VSC) and are used with an Endoscope, EndoWrist Instruments, and Accessories.

AI/ML Overview

The provided text is a 510(k) summary for the da Vinci Surgical System, focusing on a labeling change rather than a study proving performance against acceptance criteria for a new device or significant modification.

Therefore, most of the requested information regarding acceptance criteria, study design, sample sizes, expert involvement, and ground truth establishment is not available in this document. This 510(k) is specifically for a minor administrative change (a labeling update to contraindications), not a clinical performance study.

Here's an explanation of what can and cannot be answered based on the provided text:

1. A table of acceptance criteria and the reported device performance

  • Not applicable / Not provided. The document describes a change to labeling (contraindications) for existing devices already on the market, not a new device or a performance enhancement that would require a study with acceptance criteria.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not applicable / Not provided. No test set or clinical study data is presented as this submission is for a labeling change.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable / Not provided. No ground truth establishment is described as there is no clinical study.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable / Not provided. No test set or adjudication process is mentioned.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable / Not provided. This device is a surgical system, not an AI-assisted diagnostic tool or an "AI" in the typical sense that would involve human "readers" or an MRMC study comparing AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable / Not provided. The da Vinci system is a human-controlled surgical system. There is no standalone "algorithm-only" performance to evaluate in this context.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

  • Not applicable / Not provided. No ground truth is established as there is no performance study.

8. The sample size for the training set

  • Not applicable / Not provided. No new algorithm or trainable component is described in this 510(k) submission that would require a training set.

9. How the ground truth for the training set was established

  • Not applicable / Not provided. No training set is involved.

Summary of the document:

This 510(k) submission (K171699) by Intuitive Surgical, Inc. is a Special 510(k) for their da Vinci Surgical Systems (Models IS3000, IS4000, IS4200). A Special 510(k) is typically used for well-defined modifications to a manufacturer's own legally marketed device that do not affect the indications for use or alter the fundamental scientific technology of the device, and for which the risk of the modification can be adequately assessed by the manufacturer's design controls.

In this specific case, the modification is a labeling change to the "General Warnings and Cautions" section in the User Manuals, specifically concerning the endoscopic contraindication statement. The submitter argues that this change:

  • Does not change device usage or the patient population.
  • Does not change the risk profile of the devices.
  • Adequately instructs the user.

Therefore, the submission concludes that the modified devices (with the updated labeling) are substantially equivalent to the predicate devices because the labeling modification has no impact on the safety or effectiveness of the da Vinci Surgical Systems themselves. This type of submission does not involve clinical studies or performance data against specific acceptance criteria for a new or significantly modified device function.

{0}------------------------------------------------

July 28, 2017

Intuitive Surgical, Inc. % Ms. Cindy Domecus Principal, Domecus Consulting Services, LLC & Chief Regulatory Advisor, Intuitive Surgical Domecus Consulting Services LLC 1171 Barroilhet Drive Hillsborough, California 94010

Re: K171699

Trade/Device Name: da Vinci Xi Surgical System, da Vinci Si Surgical System, da Vinci X Surgical System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: NAY Dated: June 7, 2017 Received: June 8, 2017

Dear Cindy Domecus:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{1}------------------------------------------------

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K171699

Device Name

da Vinci Surgical System, Model IS4000, and EndoWrist Instruments and Accessories

Indications for Use (Describe)

The Intuitive Surgical Endoscopic Instrument Control System, Model 184000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparosopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

Device Name

da Vinci Surgical System, Model IS3000, and EndoWrist Instruments and Accessories

Indications for Use (Describe)

The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Surgical System Model IS3000) is intended to r no multive accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp assist in the dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation approximation against, orceral procedures, general laparoscopic surgical procedures, gynecologic probos and acoussent anning actrage tal otolaryngology surgical procedures restricted to benign and malignant itumors classified as T1 and T2, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy rumors clasmed as 1 x and 12, gateda adjunctive mediastinotomy to perform coronary anastomosis during procedurer The system is indicated for adult and pediatric use. It is intended to be used by trained earchies in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the The burden time for the Selloons, search existing data sources, gather and maintain the data needed and complete time to rowly the collection. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Indications for Use

510(k) Number (if known)

Device Name

da Vinci Surgical System, Model IS4200, and EndoWrist Instruments and Accessories

Indications for Use (Describe)

The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Surgical System, Model 184200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparosopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

K171699

Special 510(k) Summary (21 CFR § 807.92(c))

I: SUBMITTER INFORMATION

Submitter:Intuitive Surgical, Inc.1266 Kifer RoadSunnyvale, CA 94086
Contact:Cindy Domecus, R.A.C. (US & EU)Principal, Domecus Consulting Services LLCDomecusConsulting@comcast.net(650) 343-4813
Date Summary Prepared:July 11, 2017
II: SUBJECT DEVICE INFORMATION
Device Trade Name:Intuitive Surgical® da Vinci® Surgical Systems (Models: IS3000, IS4000,IS4200)
Common Name:Endoscopic Instrument Control System
Classification Name:System, Surgical, Computer Controlled Instrument (21 CFR §876.1500)
Regulatory Class:II
Product Code:NAY

III: PREDICATE DEVICE INFORMATION

da Vinci Surgical Systems (K090993, K131861, K171294) Predicate Devices:

IV: DEVICE DESCRIPTION

The da Vinci Surgical Systems (Models IS3000, IS4000 and IS4200) are software-controlled, electromechanical systems designed for surgeons to perform minimally invasive surgery. The da Vinci Surgical Systems consist of a Surgeon Console, a Patient Side Cart (PSC), and a Vision Side Cart (VSC) and are used with an Endoscope, EndoWrist Instruments, and Accessories.

V: INDICATIONS FOR USE

Model IS3000:

The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Surgical System Model IS3000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, transoral otolaryngology surgical procedures restricted to benign and malignant tumors classified as T1 and T2, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric

{6}------------------------------------------------

use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

Models IS4000 and IS4200:

The Intuitive Surgical Endoscopic Instrument Control System (da Vinci Surgical System Models: IS4000 and IS4200) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments including rigid endoscopes, blunt and sharp endoscopic dissectors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories, during urologic surgical procedures, general laparoscopic surgical procedures, gynecologic laparoscopic surgical procedures, general thoracoscopic surgical procedures and thoracoscopically-assisted cardiotomy procedures. The system can also be employed with adjunctive mediastinotomy to perform coronary anastomosis during cardiac revascularization. The system is indicated for adult and pediatric use. It is intended to be used by trained physicians in an operating room environment in accordance with the representative, specific procedures set forth in the Professional Instructions for Use.

VI: MODIFICATIONS OF SUBJECT DEVICE

No modifications are being made to the technological characteristics of the subject devices or the cleared indications for use. This 510(k) is being submitted for a labeling change to the Instructions for Use only. Specifically, the subject modification is for a change to the endoscopic contraindication statement in the "General Warnings and Cautions" section in the User Manuals. The original and modified language are noted below:

Original:

Any and all relative and absolute contraindications to endoscopic surgical technique applicable to the use of conventional endoscopic surgical instruments apply to the da Vinci' System. General, non-procedure specific, contraindications to endoscopic surgery include bleeding diathesis, morbid obesity and pregnancy.

Modified:

Any and all relative and absolute contraindications to endoscopic, thoracoscopic, laparoscopic, and transoral otolaryngology surgical techniques applicable to the use of conventional endoscopic surgical instruments apply to the use of the da Vinci System. The device is not intended for use when endoscopic techniques are contraindicated.

VII: DESIGN CONTROL ACTIVITIES

The proposed change to the contraindication statement does not change device usage or the patient population in which the devices are used. The same general contraindications regarding endoscopic surgery are still applicable and are clearly stated in the revised statement. As such, the proposed change does not change the risk profile of the devices. The proposed statement adequately instructs the user to

- The "Si", "Xi" or "X" System is specified in each of the pertinent PIFUs.

{7}------------------------------------------------

refrain from device usage when endoscopic surgical techniques are contraindicated.

VIII: CONCLUSION

The labeling modification has no impact on the safety or effectiveness of the da Vinci Surgical Systems. As such, the modified devices are substantially equivalent to the predicate devices.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.