K Number
K171605
Manufacturer
Date Cleared
2018-02-20

(264 days)

Product Code
Regulation Number
866.6010
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Immunological in vitro assay for quantitative determination of CA 15-3 in human serum and Li-heparin plasma to aid in the management of breast cancer patients. In conjunction with other clinical and diagnostic procedures, serial testing with this assay is an aid

  • · in the early detection of recurrence in previously treated stage II and III breast cancer patients
  • · for monitoring response to therapy in metastatic breast cancer patients
    The electrochemiluminescence immunoassay "ECLIA" is intended for use on cobas e immunoassay analyzers.
Device Description

The Elecsys CA 15-3 II Test System is a two-step sandwich immunoassay with streptavidin microparticles and electrochemiluminescence detection for the quantitative determination of CA 15-3 in human serum and plasma. It is intended for use on the cobas e 411 immunoassay analyzer. The cobas e family of analyzers employs the electrochemiluminescence "ECLIA" technology.
Results are determined using a calibration curve that is generated specifically on each instrument by a 2-point calibration and a master curve provided with the reagent barcode.

AI/ML Overview

The provided text describes the 510(k) Premarket Notification for the Elecsys CA 15-3 II diagnostic test system. This is an in vitro immunoassay designed to quantitatively determine CA 15-3 in human serum and Li-heparin plasma to aid in the management of breast cancer patients. It is not an AI/ML powered device, therefore many of the requested elements pertaining to AI/ML specific studies (e.g., MRMC studies, training set details, expert ground truth establishment for AI) are not applicable.

However, I can extract information related to the performance studies conducted to prove the device meets acceptance criteria, focusing on the analytical performance characteristics relevant to an immunoassay.

Acceptance Criteria and Reported Device Performance

The document doesn't explicitly present a single "acceptance criteria" table with corresponding "reported device performance" values in a concise format. Instead, it describes various non-clinical performance evaluations and states that "specification was met" or "deviations were within predetermined acceptance criteria" for each. I will summarize the key performance parameters and the general findings related to meeting criteria.

Note: For each study, the acceptance criteria are implied by the statement that the "specification was met" or "deviations were within predetermined acceptance criteria." The document does not provide the numerical thresholds for these criteria.

Performance ParameterAcceptance Criteria (Implied)Reported Device Performance
1. PrecisionPerformance meets CLSI EP5-A3 guidelines.Evaluated on one cobas e 801 Immunoassay Analyzer with one reagent lot. Performed 2 replicates of controls and 5 pooled human sera samples per run, 2 runs/day for 21 days. Implied: Met.
2. Analytical Sensitivity
2a. Limit of Blank (LoB)Highest observed measurement values for samples free of analyte determined as 95th percentile.Determined according to CLSI EP17-A2. Based on 1 run/day from 3 reagent lots on 1 cobas e 801 for 6 days (60 total blank replicates/lot). Reported: 1.0 U/mL (from comparison table). Implied: Met.
2b. Limit of Detection (LoD)Lowest amount of analyte in a sample detectable with 95% probability.Determined according to CLSI EP17-A2. Based on 1 run/day from 3 reagent lots on 1 cobas e 801 for 6 days (60 total replicates/sample/lot). Reported: 1.5 U/mL (from comparison table). Implied: Met.
2c. Limit of Quantitation (LoQ)Performance meets CLSI EP17-A2 guidelines.Based on 10 human serum samples (single donors) for 3 reagent lots. 1 run/day for 5 days (25 replicates/sample/lot). Reported: 3 U/mL (from comparison table). Implied: Met.
3. High-Dose Hook EffectHook concentration reported corresponds to the highest analyte concentration that generates a signal >10% above the upper limit of the measuring range.Assessed on cobas e 801 in 3-fold determination with 2 spiked samples. Implied: Met for substances tested.
4. Linearity/Assay Reportable RangeAll deviations within predetermined acceptance criteria. Performance meets CLSI EP6-A guidelines.Evaluated using cobas e 801. Three high analyte serum and three high analyte plasma samples diluted into 15 concentrations (3-fold determination in single run). Implied: Met.
5. Endogenous InterferencesNon-interference at spiked concentrations for substances tested.Tested effect of Bilirubin, Lipemia, Hemoglobin, Biotin, HAMA, and Rheumatoid Factors on cobas e 801 using low, mid, and high CA 15-3 concentrations. Implied: Met for all substances.
6. Exogenous Interferences – AnticoagulantsPerformance consistent across different anticoagulants.Determined by comparing values from serum and Li-Heparin plasma. 49 native samples tested on cobas e 801. Evaluation by Passing/Bablok regression. Implied: Met.
7. Method ComparisonAgreement with predicate device (cobas e 601 immunoassay analyzer). Performance meets CLSI EP09-A3 guidelines.Performed with 190 human serum samples (single donors, native) covering the measuring range (5.14 - 290 U/mL for e 601; 4.24 – 273 U/mL for e 801). Evaluated by Passing/Bablok and linear regression. Implied: Met.
8. Reagent StabilityOnboard reagent stability and shelf-life stability claims supported.Onboard: 16 weeks (new calibration every 28 days). Tested on cobas e 801. Real-time: Ongoing, tested at 0, 12/13/15, 17/19/25/30 months for 3 lots. Implied: Met.
9. Sample StabilitySample stability maintained at various storage conditions.2-8°C: 9 days (10 samples/type). Room Temp (20-25°C): 8 days (10 samples/type). -20°C: 53 weeks (10 samples/type). All tested on cobas e 601. Implied: Met.
10. Calibration StabilityCalibration lot stability (12 weeks) and on-board calibration stability (28 days) are maintained.Lot: 12 weeks recommended. Tested for 13 weeks on cobas e 801 (5 HS samples). On-board: 28 days. Tested on cobas e 801 (6 HS samples). Implied: Met.

Study Details:

  1. Sample Size Used for the Test Set and Data Provenance:

    • Precision: 2 replicates of controls and 5 pooled human sera samples per run, 2 runs per day for 21 days.
    • LoB/LoD: 60 total blank replicates per reagent lot (LoB) and 60 total replicates per sample per reagent lot (LoD).
    • LoQ: 25 replicates per sample per reagent lot.
    • High-Dose Hook Effect: 2 samples.
    • Linearity: 3 high analyte serum samples and 3 high analyte plasma samples, diluted to 15 concentrations.
    • Endogenous Interferences: For each interfering substance, three human serum samples (low, mid, high CA 15-3 concentrations).
    • Exogenous Interferences (Anticoagulants): 49 native samples (serum and Li-Heparin plasma).
    • Method Comparison: 190 human serum samples (single donors, native samples).
    • Reagent Stability:
      • Onboard: Six human serum samples.
      • Real-time: PreciControl Tumor Marker Level 1 and 2 (3 lot numbers over various time points).
    • Sample Stability: 10 samples for each sample type (Serum, Li-Heparin plasma) for each storage condition.
    • Calibration Stability:
      • Lot: Five human serum samples.
      • On-board: Six human serum samples.
    • Data Provenance: The document states "human serum" and "human plasma" samples were used, and refers to Roche Diagnostics GmbH in Mannheim, Germany, and Penzberg, Germany for establishment registration. It appears to be primarily laboratory-generated data for analytical validation, not specific patient cohorts with country of origin. The term "native samples" implies real patient samples, but no specifics on their origin are given beyond being "human." The studies are retrospective analytical investigations to characterize the device's technical performance.
  2. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

    • This is an in vitro diagnostic (IVD) immunoassay, not an AI/ML device relying on interpretation of medical images or clinical data by experts. The "ground truth" for these analytical studies is the quantitative concentration of CA 15-3 in the sample, measured using reference methods or established spiked concentrations. No human experts were involved in establishing the "ground truth" in the way it would be for an AI device. The ground truth for analytical performance is based on the inherent properties of the samples and reference materials.
  3. Adjudication Method for the Test Set:

    • Not applicable. This is an IVD immunoassay, not an AI/ML interpretation study requiring human adjudication.
  4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    • No, an MRMC study was not done. This is an IVD immunoassay. MRMC studies are typically performed for imaging interpretation/AI devices to assess the impact of AI on human reader performance.
  5. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done:

    • Yes, the entire study is essentially a "standalone" performance evaluation of the immunoassay system (Elecsys CA 15-3 II on cobas e analyzers). Its output is a quantitative value (CA 15-3 concentration), not an interpretation that a human would then use as "assistance." The performance metrics evaluated (precision, linearity, limits, interference, method comparison) are intrinsic to the device's analytical function without human interaction for result generation.
  6. The Type of Ground Truth Used:

    • The ground truth for this device's performance studies is based on:
      • Reference materials/controls: For precision, stability studies (e.g., PreciControl Tumor Marker).
      • Known (spiked) concentrations: For linearity, hook effect, interference studies, where samples are prepared with precisely known amounts of analyte or interfering substances.
      • Native samples: For method comparison and sample type evaluations, where the "ground truth" is established by the predicate device or a trusted reference measurement.
      • Blank samples: For Limit of Blank determination.
  7. The Sample Size for the Training Set:

    • Not applicable in the context of an AI/ML algorithm. This device is a biochemical immunoassay. Its "training" or "calibration" involves internal calibration procedures (e.g., 2-point calibration and master curve provided with reagent barcode), which are distinct from the data-driven training of a machine learning model. The reference to "calibration curve" and "master curve" relates to the instrument's operational setup, not a machine learning training dataset.
  8. How the Ground Truth for the Training Set Was Established:

    • Not applicable in the AI/ML sense. Calibration is performed using "CalSet" (calibration materials) and is standardized against previous Elecsys CA 15-3 assays, which were in turn standardized against Enzymun Test CA 15-3 and CA 15-3 RIA by Fujirebio Diagnostics. This establishes traceability to recognized reference methods for the quantitative measurement.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.

February 27, 2018

Roche Diagnostics Adennis Cora Regulatory Affairs Consultant 9115 Hague Road Indianapolis, Indiana 46250

Re: K171605

Trade/Device Name: Elecsys CA 15-3 II Regulation Number: 21 CFR 866.6010 Regulation Name: Tumor-associated antigen immunological test system Regulatory Class: Class II Product Code: MOI Dated: May 26, 2017 Received: June 1, 2017

Dear Adennis Cora:

This letter corrects our substantially equivalent letter of February 20, 2018.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

{1}------------------------------------------------

801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Kelly O

For

Lea Carrington Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K171605

Device Name Elecsys CA 15-3 II

Indications for Use (Describe)

Immunological in vitro assay for quantitative determination of CA 15-3 in human serum and Li-heparin plasma to aid in the management of breast cancer patients. In conjunction with other clinical and diagnostic procedures, serial testing with this assay is an aid

  • · in the early detection of recurrence in previously treated stage II and III breast cancer patients
  • · for monitoring response to therapy in metastatic breast cancer patients

The electrochemiluminescence immunoassay "ECLIA" is intended for use on cobas e immunoassay analyzers.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Elecsys CA 15-3 II 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

In accordance with 21 CFR 807.87, Roche Diagnostics hereby submits official notification as required by Section 510(k) of the Federal Food, Drug and Cosmetics Act of our intention to market the device described in this Premarket Notification 510(k).

The purpose of this Traditional 510(k) Premarket Notification is to obtain FDA review and clearance for the Elecsys CA 15-3 II.

Submitter NameRoche Diagnostics
Address9115 Hague RoadP.O. Box 50416Indianapolis, IN 46250-0457
ContactAdennis CoraPhone: (317) 521-3915FAX: (317) 521-2324Email: adennis.cora_gress@roche.com
Date PreparedApril 14, 2017
Proprietary NameElecsys CA 15-3 II
Common NameElecsys Cancer Antigen 15-3 II
Classification NameSystem, Test, Immunological, Antigen, Tumor
Product Codes,Regulation NumbersMOI, 866.6010
Predicate DevicesCA 15-3 II Assay (K001468)
Establishment RegistrationFor the Elecsys CA 15-3 II, the establishment registration number for RocheDiagnostics GmbH in Mannheim, Germany is 9610126, and for Penzberg,Germany, 9610529. The establishment registration number for RocheDiagnostics in the United States is 1823260.

{4}------------------------------------------------

1. DEVICE DESCRIPTION

The Elecsys CA 15-3 II Test System is a two-step sandwich immunoassay with streptavidin microparticles and electrochemiluminescence detection for the quantitative determination of CA 15-3 in human serum and plasma. It is intended for use on the cobas e 411 immunoassay analyzer. The cobas e family of analyzers employs the electrochemiluminescence "ECLIA" technology.

Results are determined using a calibration curve that is generated specifically on each instrument by a 2-point calibration and a master curve provided with the reagent barcode.

1.1. Reagents

The cobas e pack includes:

  • Streptavidin-coated microparticles .
  • . Reagent 1 (Biotinylated monoclonal antibody (115D8; mouse) 1.75 mg/L)
  • . Reagent 2 (Monoclonal anti-CA 15-3 antibody (DF3; mouse) labeled with ruthenium complex 10 mg/L)

1.2. Calibration

Traceability: This method has been standardized against the Elecsys CA 15 3 assay. This in turn has been standardized against the Enzymun Test CA 15 3 method and CA 15 3 RIA by Fujirebio Diagnostics. The predefined master curve is adapted to the analyzer using the relevant CalSet. Calibration frequency: Calibration must be performed once per reagent lot using fresh reagent (i.e. not more than 24 hours since the cobas e pack was registered on the analyzer). Calibration interval may be extended based on acceptable verification of calibration by the laboratory.

1.3. -Control

PreciControl Tumor Marker is used for the quality control verification of the performance of the assay on the cobas e analyzers.

{5}------------------------------------------------

2. INDICATIONS FOR USE

Immunological in vitro assay for quantitative determination of CA 15-3 in human serum and Liheparin plasma to aid in the management of breast cancer patients. In conjunction with other clinical and diagnostic procedures, serial testing with this assay is an aid

  • in the early detection of recurrence in previously treated stage II and III breast cancer . patients
  • for monitoring response to therapy in metastatic breast cancer patients. .

The electrochemiluminescence immunoassay "ECLIA" is intended for use on cobas e immunoassay analyzers.

TECHNOLOGICAL CHARACTERISTICS 3.

The following table compares the Elecsys CA 15-3 II with its predicate device, CA 15-3 (K001468). In summary, the Elecsys CA15-3 II described in this submission is substantially equivalent to the predicate device.

FeatureCA 15-3 II(Cleared K001468)Elecsys CA 15-3 II
Intended useFor the quantitative determination ofCA 15-3 in human serum andplasma.for quantitative determination of CA15 3 in human serum andLi-heparin plasma
Indication for useTo aid in the management of breastcancer patients. In conjunction withother clinical and diagnosticprocedures, serial testing with CA 15-3 assay is an aid in the earlydetection of recurrence in previouslytreated Stage II and III breast cancerpatients. For monitoring response totherapy in metastatic breast cancerpatients.Same
Sample TypeHuman serum and plasma treatedwith heparin (Li-, Na-) or K3-EDTAplasmahuman serum andLi-heparin plasma
Analytical SpecificityBased on monoclonal 115D8 andDF3 antibodies available fromCentocorSame

{6}------------------------------------------------

FeatureCA 15-3 II(Cleared K001468)Elecsys CA 15-3 II
Instrument PlatformElecsys 2010
MODULAR ANALYTICS E170
cobas e 411cobas e 801
cobas e 601
cobas e 602
Measuring Range1.0 - 300 U/mL3 - 300 U/mL
Determined by LDL (2SD)Determined by LOQ (Per EP17-A2)
BiotinAdded to R2Same
PrewashNo PrewashPrewash Added
Sample Dilution1:101:20
LoD< 1.0 U/mL1.5 U/mL
LoBN/A1.0 U/mL
LoQN/A3 U/mL
Lower Detection Limit< 1.00 U/mLN/A

4. NON-CLINICAL PERFORMANCE EVALUATION

Non-clinical performance evaluation for the Elecsys CA 15-3 II are briefly summarized below:

4.1. Precision

Precision of the Elecsys CA 15-3 II assay was evaluated on one cobas e 801 Immunoassay Analyzer according to CLSI EP5-A3 guideline. One reagent lot was evaluated.

The protocol consisted of testing 2 replicates of each control (PC Tumor Marker = PreciControl Tumor Marker) and five pooled human sera (HS) samples per run, 2 runs per day for 21 days. The samples were run in randomized order on the cobas e 801 analyzer. To obtain samples with higher concentration, low sample pools were spiked with a high concentrated native sample (pool)..

Analytical Sensitivity 4.2.

4.2.1. Limit of Blank (LoB)

LoB of the Elecsys CA 15-3 II assay on the cobas e 801 Immunoassay Analyzer has been determined according to CLSI EP17-A2.

{7}------------------------------------------------

Limit of Blank determines the highest observed measurement values for samples free of analyte. The Limit of Blank was determined as the 95th percentile of measurements of blank samples. The Limit of Blank was determined by performing one run per day from three reagent lots on one cobas e 801 instrument, over six days. The testing included one blank sample with ten replicates per run for a total of sixty total blank replicates per reagent lot.

4.2.2. Limit of Detection (LoD)

LoD of the Elecsys CA 15-3 assay on the cobas e 801 Immunoassay Analyzer has been determined according to CLSI EP17-A2.

The LoD was determined as the lowest amount of analyte in a sample that can be detected with 95% probability. The Limit of Detection was determined by performing one run per day from three reagent lots on one cobas e 801 instrument, over six days. The testing included two replicates per sample (5 samples) per run for a total of sixty total blank replicates per sample per reagent lot.

Limit of Quantitation (LoQ) 4.2.3.

The LoQ of the Elecsys CA 15-3 II assay was determined on the cobas e 801 Immunoassay Analyzer according to CLSI Guideline EP17-A2.

A set of ten human serum samples (single donors, native as well as diluted) with concentrations in the specified LoQ-area was evaluated for each of 3 reagent lots.

The mean value and the intermediate precision as coefficient of variation (CV) and standard deviation (SD) for each LoQ sample were calculated. The Limit of Quantitation was determined by performing one run per day from three reagent lots on one cobas e 801 instrument, over five days. The testing included ten samples, each with five replicates per daily run over 5 days for a total of twenty five replicates per sample per reagent lot.

4.3. High-Dose Hook Effect

The high-dose hook effect of the Elecsys CA 15-3 II assay was assessed on the cobas e 801 analyzer in three-fold determination. Two samples were spiked with analyte to achieve high CA 15-3 concentrations. For each sample, a dilution series was performed using native human

{8}------------------------------------------------

serum. The hook concentration reported corresponds to the highest analyte concentration that generates a signal >10% above the upper limit of the measuring range.

Linearity/Assay Reportable Range 4.4.

Linearity of the Elecsys CA 15-3 II assay was evaluated using the cobas e 801 Immunoassay Analyzer according to CLSI EP6-A.

Three high analyte serum samples and three high analyte plasma samples (serum/plasma pool, spiked) were diluted with Elecsys Universal Diluent. 15 concentrations (dilutions) throughout the measuring range were prepared. Samples were assayed in 3-fold determination within a single run.

The linearity data were analyzed with regards to linear, quadratic and cubic polynomials according to CLSI EP6-A. In a first step, a linearity check was performed with a first order (linear) regression and then with higher order models (quadratic and cubic).

All deviations were within predetermined acceptance criteria.

Endogenous Interferences 4.5.

The effect on quantitation of analyte in the presence of endogenous interfering substances using the Elecsys CA 15-3 II was determined on the cobas e 801 Immunoassay analyzer using human serum samples.

For each interfering substance three human serum samples containing low (pooled sera), mid (pooled spiked sera) and high (pooled spiked sera) concentrations of CA 15-3 were tested.

{9}------------------------------------------------

4.6. Hemolysis/Bilirubin/Lipemia/Biotin

Biotin:

One aliguot of each serum sample was spiked with the interfering substance, another aliquot was spiked with the same volume of isotonic NaCl solution (dilution pool). The interfering pool was then diluted into the dilution pool in 10% increments.

The recovery for each sample was calculated by comparison to the reference (unspiked) sample.

In conclusion, the specification was met for all substances tested, as each substance was found to be non-interfering at the spiked concentration.

Lipemia:

One aliquot of each serum sample was spiked with the interfering substance, and another aliguot was spiked with the same volume of isotonic NaCl solution (dilution pool). The interfering pool was then diluted into the dilution pool in 10% increments.

The recovery for each sample was calculated by comparison to the reference (unspiked) sample.

In conclusion, the specification was met for all substances tested, as each substance was found to be non-interfering at the spiked concentration.

Hemoglobin:

Fresh hemolysate was prepared from fresh EDTA or heparin blood. One aliquot of each serum sample was spiked with the hemolysate (interfering pool), another aliquot was spiked with the same volume of isotonic NaCl solution (dilution pool). The interfering pool was then diluted into the dilution pool in 10% increments. The recovery for each sample was calculated by comparison to the reference (unspiked) sample.

In conclusion, the specification was met for all substances tested, as each substance was found to be non-interfering at the spiked concentration.

Bilirubin:

One aliquot of each serum sample was spiked with the interfering substance, and another aliquot with was spiked with the same volume of the solvent (0.1 mol/L NaOH) of the interfering

{10}------------------------------------------------

substance (dilution pool). The interfering pool was then dilution pool in 10% increments.

HAMA:

The effect of the presence of human anti-mouse antibodies on the Elecsys CA 15-3 II assay was assessed on the cobas e 801 Immunoassay Analyzer.

The specified HAMA-serum with a concentration of 805 ng/mL and the related reference serum (without interferent) was measured in two-fold determination over 3 levels. Recovery of the HAMA-serum compared to the reference serum sample was calculated.

In conclusion, the specification was met for all substances tested, as each substance was found to be non-interfering at the spiked concentration.

Rheumatoid Factors Interference:

One aliquot of each serum sample was spiked with the interfering substance. and another aliquot with the same volume of the solvent (buffer matrix) of the interfering substance (dilution pool). The interfering pool was then diluted into the dilution pool in 10% increments.

The recovery for each sample was calculated by comparison to the reference sample.

In conclusion, the specification was met for all substances tested, as each substance was found to be non-interfering at the spiked concentration.

4.7. Exogenous Interferences – Anticoaqulants

The effect on quantitation of analyte in the presence of anticoagulants with the Elecsys CA 15-3 II Immunoassay was determined by comparing values obtained from samples (native) drawn into Serum, and Li-Heparin plasma.

Forty nine native samples were tested with one reagent lot on one cobas e 801 Immunoassay Analyzer. Potential effects are assessed by Passing/Bablok regression analysis.

{11}------------------------------------------------

4.8. Method Comparison to cobas e 601 Immunoassay Analyzer

A method comparison was performed using the cobas e 601 immunoassay analyzer as predicate device. The study was performed on the cobas e 601 immunoassay analyzer (X) and on the cobas e 801 analyzer (Y).

A total of 190 human serum samples (all single donors, native samples) were measured with Elecsys CA 15-3 II immunoassay on both analyzers in singleton covering the entire measuring range. Mean values ranges between 5.14 - 290 U/mL for the cobas e 601 analyzer and between 4.24 – 273 U/mL for the cobas e 801 analyzer.

Evaluation was performed for all samples within the measuring range of both assays and in compliance with Standard Operation Procedures of Roche Diagnostics GmbH. Passing/Bablok and linear regression analysis was performed according to CLSI EP09-A3.

4.9. Reagent Stability

To test reagent stability, two studies were executed with one study completed including:

Study 1. On board reagent stability (16 weeks)

Study 2. A real-time stability study is ongoing to support shelf-life stability claim.

Study 1- Onboard reagent stability (for 16 weeks) 4.9.1. -

Elecsys CA 15-3 II reagent kits can be stored on board of the analyzers for up to 16 weeks.

A new calibration of the kit kept on-board is recommended every 28 days.

Onboard Reagent Stability and Calibration Stability for the Elecsys CA 15-3 II assay was tested on one cobas e 801 Immunoassay Analyzer.

A fresh cobas e pack was placed on the analyzer and calibrated. Reference values for the samples tested were determined (day 0). After 5, 9, 13 and 17 weeks the same samples were measured with the same reagent kit kept under onboard condition. Re-calibration was performed at every measuring time point.

{12}------------------------------------------------

Samples tested include six human serum (HS) sample was tested in two-fold determination. The human serum samples used were pooled sera for sample ranges low, Decision range 1 and 2, and pooled spiked sera for sample ranges medium and high.

Study 2 - Reagent Real-time Stability 4.9.2.

In the real-time stability study, the Elecsys CA 15-3 II reagent was stored at 2-8°C. The stored assay reagents were tested at time point T=0 (after manufacturing) and at specified intervals over the shelf life of the device. Testing was performed using PreciControl Tumor Marker Level 1 and 2 (1yophilized, stored at +2 to +8°C) for lot-numbers 156423, 158287 and 174942.

Data for lot no. (in months):

156423 for the time-points at 0, 15, 19, 30

158287 for the time-points at 0, 13, 17, 25

174942 for the time-points at 0, 12, 15, 19

The average on-test recovery value will be calculated as percent recovery compared to the reference value (assigned value for PreciControl Tumor Marker Level 1 and 2).

4.10. Sample Stability

To test sample stability, three studies were completed, including:

Study 1. Sample stability at 2-8°C

Study 2. Sample stability at Room Temperature (20-25°C)

  • Study 3. Sample stability -20°C (± 5 °C)
  • 4.10.1. Study 1-Sample Stability at 2-8°C

Ten samples for each sample type (Serum, and Li-Heparin plasma) were aliquoted and measured after storage at 2-8°C for 9 days. The aliguot for the reference value was stored at -80°C and measured at the same time point. Measurements were performed with three-fold determination on a cobas e 601 Immunoassay Analyzer and recovery was calculated as percent of the reference value. The samples used were all single donors (native, spiked as well as diluted).

{13}------------------------------------------------

4.10.2. Study 2-Sample Stability at Room Temperature (20-25°C)

Ten samples for each sample type (Serum, and Li-Heparin plasma) were aliquoted and measured after storage at 20-25℃ for 8 days. The aliquot for the reference value was stored at -80℃ and measured at the same time point.

Measurements were performed with three-fold determination on a cobas e 601 Immunoassay Analyzer and recovery was calculated as percent of the reference value.

The samples used were all single donors (native, spiked as well as diluted).

4.10.3.   Study 3-Sample Stability at -20°C (+5°C)

Ten samples for each sample type (Serum, and Li-Heparin plasma) were aliquoted and measured after storage at - 20°C for 53 weeks. The aliquot for the reference value was stored at -80°C and measured at the same time point.

Measurements were performed with three-fold determination on a cobas e 601 Immunoassav Analyzer and recovery was calculated as percent of the reference value.

The samples used were all single donors (native, spiked as well as diluted).

4.11. Calibration Stability

To test calibration stability, two studies were completed, including:

Study 1. Lot calibration stability

Study 2. On-board calibration stability

4.11.1. Study 1-Lot Calibration Stability

Calibration of an Elecsys CA 15-3 II reagent lot is recommended every 12 weeks. During that time period fresh reagent kits of the same lot can be used without calibration using the calibration curve of the Day 0 reagent kit.

Elecsys CA 15-3 II was calibrated with a fresh reagent kit on Day 0 using a cobas e 801 Immunoassay Analyzer. After 13weeks a new reagent kit of the same lot was used and recovery of samples was determined using the calibration curve of day 0.

{14}------------------------------------------------

Five human serum (HS) samples were tested; each sample was tested with two-fold determination. The human serum samples used were pooled diluted sera for sample range low. pooled sera for Decision range, and pooled spiked sera for sample ranges medium and high.

4.11.2. Study 2-On-board Calibration Stability

Elecsys CA 15-3 II reagent kits can be stored on board of the analyzers for up to 28 days without a new calibration.

Reagent On-board Calibration stability for the Elecsys CA 15-3 II assay was tested on one cobas e 801 Immunoassay Analyzer.

A fresh Reagent cobas e pack was placed on the analyzer and calibrated. All samples were measured on day 0. On day 29 the same samples were measured with the same reagent kit kept under on-board conditions using the calibration curve established on day 0.

Samples tested include six human serum (HS) samples. Each sample was tested in two-fold determination. The human serum samples used were pooled diluted sera for sample range low. pooled sera for Decision range, and pooled spiked sera for sample ranges medium and high,

5. CONCLUSIONS

The information provided in this 510(k) Premarket Notification will support a determination of substantial equivalence for the Elecsys CA 15-3 II test system. The data supports a safe, effective device which performs as well as or better than the predicate device.

§ 866.6010 Tumor-associated antigen immunological test system.

(a)
Identification. A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.(b)
Classification. Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.