(294 days)
TheraBreath Dry Mouth Oral Rinse is intended for relief of dry mouth symptoms and is indicated for use to refresh, moisturize, lubricate and soothe the oral cavity, and alleviate discomfort due to dry mouth.
TheraBreath Dry Mouth Lozenges are intended for relief of dry mouth symptoms, and are indicated for use to refresh, moisturize, lubricated, and soothe the oral cavity, and alleviate discomfort due to dry mouth.
TheraBreath Dry Mouth Oral Rinse and TheraBreath Dry Mouth Lozenges are specifically formulated as an artificial saliva substitute with moisturizers, humectants and salivary enzymes with lubricating, moisturizing and soothing properties to relieve the symptoms of dry mouth. In addition. TheraBreath Dry Mouth Lozenges continuously moisten the mouth to relieve the symptoms of Dry Mouth and to refresh breath. TheraBreath Dry Mouth Oral Rinse and TheraBreath Dry Mouth Lozenges are sugar-free, alcohol-free, and gluten-free.
TheraBreath Dry Mouth Oral Rinse is packaged in 16 ounce size bottles and is Tingling Mint flavored. TheraBreath Dry Mouth Lozenges are Mandarin Mint flavored and are sold in boxes of 24, 72, or 100 lozenges.
The provided text is a 510(k) summary for medical devices (TheraBreath Dry Mouth Oral Rinse and TheraBreath Dry Mouth Lozenges), seeking to establish substantial equivalence to predicate devices. This document does not contain information about an AI/ML medical device, nor does it describe acceptance criteria and a study proving a device meets those criteria for such a device.
Instead, it's a submission for an over-the-counter dry mouth relief product. The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to non-clinical tests (biocompatibility, shelf-life, physical properties) to demonstrate the safety and performance of the new products are comparable to existing, legally marketed products.
Therefore, I cannot fulfill your request for detailed information about an AI/ML device's acceptance criteria and studies based on the provided text. The document does not describe:
- A table of acceptance criteria and reported device performance for an AI/ML device.
- Sample sizes, data provenance, expert numbers/qualifications, or adjudication methods for an AI/ML test set.
- MRMC comparative effectiveness studies or standalone AI algorithm performance.
- Methods for establishing ground truth for training or test sets in an AI/ML context.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 16, 2018
Dr. Harold Katz, LLC (dba TheraBreath) % Barbara Fant President Clinical Research Consultants, Inc. 3308 Jefferson Avenue, Upper Level Cincinnati. Ohio 45220
Re: K171542
Trade/Device Name: TheraBreath Dry Mouth Oral Rinse, TheraBreath Dry Mouth Lozenges Regulatory Class: Unclassified Product Code: LFD Dated: February 7, 2018 Received: February 9, 2018
Dear Barbara Fant:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Andrew I. Steen -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
4. Indications for Use Statement
510(k) Number (if known): To Be Assigned K171542
Device Name:
TheraBreath Dry Mouth Oral Rinse TheraBreath Dry Mouth Lozenges
Indications for Use:
TheraBreath Dry Mouth Oral Rinse is intended for relief of dry mouth symptoms and is indicated for use to refresh, moisturize, lubricate and soothe the oral cavity, and alleviate discomfort due to dry mouth.
TheraBreath Dry Mouth Lozenges are intended for relief of dry mouth symptoms, and are indicated for use to refresh, moisturize, lubricated, and soothe the oral cavity, and alleviate discomfort due to dry mouth.
Type of Use (Select one or both, as applicable)
□ Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
5. 510(k) Summary
510(k) Submission, Traditional; as required by (21CFR§807.92(c))
| 510(k) Owner: | Dr. Harold Katz, LLC (dba TheraBreath)5802 Willoughby AvenueLos Angeles, California 90038Telephone: 323-762-8518Facsimile: 323-993-8327E-Mail: harold@drkatz.com |
|---|---|
| Contact Person: | Barbara S. Fant, Pharm.D.Clinical Research Consultants, Inc.3308 Jefferson Avenue, Upper LevelCincinnati, Ohio 45220Telephone: (513) 961-8200Facsimile: (513) 961-2858E-Mail: bsfant@crc-regulatory.com |
| Date: | March 15, 2018 |
| Trade Name: | TheraBreath Dry Mouth Oral RinseTheraBreath Dry Mouth Lozenges |
| Common name: | Dry mouth oral rinseDry mouth mouthwashDry mouth lozenges |
| Classification Name: | Saliva, Artificial; Unclassified |
| Product Code: | LFD |
Identification of a Legally Marketed Predicate Device
TheraBreath Dry Mouth Oral Rinse and TheraBreath Dry Mouth Lozenges are substantially equivalent to the legally marketed predicate device: MedActive Oral Relief Spray and MedActive Oral Relief Gel, marketed by Medactive Oral Pharmaceuticals, LLC, Premarket Notification Number: K152201, FDA Product Code LFD (Saliva, Artificial; Unclassified); and
{4}------------------------------------------------
to the reference devices: Dr. Fresh Orazyme Dry Mouth Mouthwash marketed by Dr. Fresh, LLC, 510(k) Premarket Notification Number: K111250, FDA Product Code LFD (Saliva, Artificial; Unclassified); and, SST 100 Lozenges marketed by Hexim Pharmaceuticals, 510(k) Premarket Notification Number: K023046, FDA Product Code LFD (Saliva, Artificial; Unclassified).
General Description
TheraBreath Dry Mouth Oral Rinse and TheraBreath Dry Mouth Lozenges are specifically formulated as an artificial saliva substitute with moisturizers, humectants and salivary enzymes with lubricating, moisturizing and soothing properties to relieve the symptoms of dry mouth. In addition. TheraBreath Dry Mouth Lozenges continuously moisten the mouth to relieve the symptoms of Dry Mouth and to refresh breath. TheraBreath Dry Mouth Oral Rinse and TheraBreath Dry Mouth Lozenges are sugar-free, alcohol-free, and gluten-free.
TheraBreath Dry Mouth Oral Rinse is packaged in 16 ounce size bottles and is Tingling Mint flavored. TheraBreath Dry Mouth Lozenges are Mandarin Mint flavored and are sold in boxes of 24, 72, or 100 lozenges.
Intended Use/Indications for Use
TheraBreath Dry Mouth Oral Rinse is intended for relief of dry mouth symptoms and is indicated for use to refresh, moisturize, lubricate and soothe the oral cavity, and alleviate discomfort due to dry mouth.
TheraBreath Dry Mouth Lozenges are intended for relief of dry mouth symptoms, and are indicated for use to refresh, moisturize, lubricated, and soothe the oral cavity, and alleviate discomfort due to dry mouth.
TheraBreath Dry Mouth Program
When the TheraBreath dry mouth products are used together along with routine oral cleansing (i.e., toothbrushing), it provides a simple three-step solution to clean and moisturize the mouth:
- Step 1. Cleansing: Brush with TheraBreath Dry Mouth Toothpaste at least twice daily.
- Step 2, Rinsing: Rinse with TheraBreath Dry Mouth Oral Rinse after brushing, for up to ● 12 hours of dry mouth relief.
- Step 3, Moisturizing: Use TheraBreath Dry Mouth Lozenges for convenience and ● moisturizing during the day, as often as required, to effectively relieve oral dryness by temporarily lubricating and moisturizing the mouth.
{5}------------------------------------------------
Comparison of Technological Characteristics
TheraBreath Dry Mouth Oral Rinse and TheraBreath Dry Mouth Lozenges are substantially equivalent to Dr. Fresh Orazyme Dry Mouth Mouthwash (K111250) and to MedActive Oral Relief Spray and Gel (K152201) in intended use, design, materials, chemical composition, and performance. The products are all artificial saliva agents designed for relief from dry mouth symptoms. TheraBreath Dry Mouth Oral Rinse and Dr. Fresh Orazyme Dry Mouth Mouthwash both contain salivary enzymes. The TheraBreath and MedActive products contain Spilanthes Extract as a flavor. TheraBreath Dry Mouth Oral Rinse and TheraBreath Dry Mouth Lozenges are both manufactured for and distributed in the U.S.A. by Dr. Harold Katz. LLC (dba TheraBreath; Los Angeles, California). Dr. Fresh Orazyme Dry Mouth Mouthwash is manufactured and distributed in the U.S.A. by Dr. Fresh, LLC (Buena Park, California). MedActive Oral Relief Spray and Gel are manufactured and distributed in the U.S.A. by MedActive Oral Pharmaceuticals, LLC (Odessa, Florida).
The primary technological differences between the oral rinses are that: (1) the proprietary composition of the salivary enzymes in TheraBreath and Orazyme products are very similar, but not identical; (2) the product raw materials and chemical composition are similar, but not identical; and, (3) the TheraBreath products are directed for use as needed, whereas, Orazyme is directed for use 2 to 3 times per day.
Brief Summary of Non-Clinical Tests and Results
Testing of the TheraBreath dry mouth products was performed to determine product composition and demonstrate conformity with product specifications for color, appearance, pH, specific gravity, and preservative efficacy. Biocompatibility testing of the TheraBreath oral rinse and lozenge formulations was performed according to ISO 10993-5 for cytotoxicity and according to ISO 10993-10 for mucosal sensitization and irritation. The products passed all tests in conformity with the standards and testing determined that the TheraBreath dry mouth products were non-cytotoxic, non-sensitizing, and non-irritating. Accelerated aging testing was performed to establish a 3 year shelf-life for TheraBreath Dry Mouth Oral Rinse and real-time aging was performed to establish a 2 year shelf-life for TheraBreath Dry Mouth Lozenges.
Comparative laboratory testing was performed for TheraBreath Dry Mouth Oral Rinse and Dr. Fresh Orazyme Dry Mouth Mouthwash and confirmed that the physical properties were similar for the two products for pH, specific gravity, odor, color and appearance.
{6}------------------------------------------------
| TheraBreathDry MouthOral Rinse | TheraBreathDry MouthLozenges | Dr. FreshOrazyme DryMouthMouthwash(ReferenceDevice) | MedActiveOral ReliefSpray, Gel(Predicate) | Same orDifferent | |
|---|---|---|---|---|---|
| Biocompatible | Yes | Yes | Yes | Yes | Same.TheraBreathproducts passedtesting criteriain conformitywith ISO10993-5 and10993-10 forcytotoxicity,irritation, andsensitization. |
| PhysiologicPhysico-ChemicalCharacteristics | Yes | Yes | Yes | Yes | Similar pH,specific gravity,color, odor andappearance fororal rinseformulations |
| TheraBreathDry MouthOral Rinse | TheraBreathDry MouthLozenges | Dr. FreshOrazyme DryMouthMouthwash(ReferenceDevice) | MedActiveOral ReliefSpray, Gel(Predicate) | Same orDifferent | |
| Intended Use | Relief of drymouthsymptoms | Relief of drymouthsymptoms | Relief of drymouthsymptoms | Relief of drymouthsymptoms | Same |
| Indications for Use | TheraBreathDry MouthOral Rinse isintended forrelief of drymouthsymptoms andis indicated foruse to refresh,moisturize,lubricate andsoothe the oralcavity, andalleviatediscomfort dueto dry mouth. | TheraBreathDry MouthLozenges areintended forrelief of drymouthsymptoms, andare indicatedfor use torefresh,moisturize,lubricated, andsoothe the oralcavity, andalleviatediscomfort dueto dry mouth. | Relieves thesymptoms ofdry mouth,whilemoisturizingandlubricating oraldryness. | Providessymptom relieffrom DryMouth and lowsaliva including:Oral discomfort,Mucosal softtissue dryness,Oral side effectsof illness,therapies, andmedications.Soothesmoistens andlubricates.Hydrates softtissue. | Indications forartificialsalivas differslightly basedon productoffering |
| Disease State | Xerostomia | Xerostomia | Xerostomia | Xerostomia | Same |
| Area of Use | Oral cavity | Oral cavity | Oral cavity | Oral cavity | Same |
| Product Offering | Liquid oralrinse | Oral Lozenge | Liquid OralRinse | Spray, Gel | Different,unique to eachProduct |
| Enzymes | Yes | No | Yes | No | Similar forproductscontainingenzymes |
| Flavor/Moisturizer | Yes | Yes | Yes | Yes | Similar,formulationunique to eachproduct |
| Sterility | Non-sterile | Non-sterile | Non-sterile | Non-sterile | Same |
| Xylitol | Yes | Yes | Yes | Yes | Same |
| Color Additive | No | No | No | No | Same |
| Sugar-Free/Gluten-Free | Yes | Yes | Yes | Yes | Same |
| Intended Population | Unsupervised | Unsupervised | Unsupervised | Unsupervised | Same |
| Applications Per Day | As needed | As needed | 2-3 times/day | As needed | Similar, instructions for use are unique to each product offering |
Basis of Substantial Equivalence
TheraBreath Dry Mouth Oral Rinse and TheraBreath Dry Mouth Lozenges are substantially equivalent to Dr. Fresh Orazyme Dry Mouth Mouthwash (K111250) and to MedActive Oral Relief Spray and Gel (K152201) in intended use, design, materials, chemical composition, and performance as artificial saliva products designed for relief from dry mouth symptoms. The ingredients used to formulate TheraBreath Dry Mouth Oral Rinse and TheraBreath Dry Mouth Lozenges are used in the predicate and reference devices, which are legally marketed for the same intended use and indications. Technological characteristics of the TheraBreath products and predicate and reference devices are summarized in the table below and support the substantial equivalence of the products.
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
| TheraBreathDry MouthOral Rinse | TheraBreathDry MouthLozenges | Dr. FreshOrazyme DryMouthMouthwash(ReferenceDevice) | MedActiveOral ReliefSpray, Gel(Predicate) | Same orDifferent | |
|---|---|---|---|---|---|
| Biocompatible | Yes | Yes | Yes | Yes | Same.TheraBreathproductspassed testingcriteria inconformitywith ISO10993-5 and10993-10 forcytotoxicity,irritation, andsensitization. |
| Physiologic Physico-ChemicalCharacteristics | Yes | Yes | Yes | Yes | Similar pH,specificgravity, color,odor andappearance fororal rinseformulations |
N/A