(78 days)
The Thunderbolt™ Minimally Invasive and Lancer™ Open Pedicle Screw Systems are intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease (DDD; defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.
When used in a posterior percutaneous approach with MIS instrumentation, the Thunderbolt™ Minimally Invasive System is intended for non-cervical pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.
When used for posterior, non-cervical pedicle, and non-pedicle fixation, the Lancer™ Open Pedicle Screw System is indicated for the following: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. Overall levels of fixation are T1 to the Sacrum/Ilium. When used for fixation to the ilium, the lateral offset connectors on the Lancer™ Open Pedicle Screw System must be used in conjunction with pedicle screws placed at the S1 or S2 spinal level.
The Thunderbolt™ Minimally Invasive and Lancer™ Open Pedicle Screw Systems include implant components made of implant grade titanium alloy (Ti-6Al-4V ELI; ASTM F136) and cobalt chrome alloy (Co-28Cr-6Mo; ASTM F1537). The systems also include instruments made of PEEK (ASTM F2826), Tantalum (ASTM F560), stainless steel (ASTM F899/A564) and aluminum (ASTM B221). These components are available in various designs and sizes that allow the surgeon to build an implant construct suited to a patient's anatomical and physiological requirements.
The components include: polyaxial pedicle screws, set screws, rods, instruments and sterilizer trays. The Lancer™ Open Pedicle Screw System also includes connector and hook components.
The provided text describes a medical device, the Thunderbolt™ Minimally Invasive and Lancer™ Open Pedicle Screw Systems, and its substantial equivalence submission to the FDA. It does not describe an AI/ML device or a study comparing its performance against acceptance criteria in the context of AI/ML.
Therefore, I cannot extract the information required to answer your request about acceptance criteria and a study proving a device meets acceptance criteria in an AI/ML context. The document focuses on the mechanical and material equivalence of a surgical implant system to previously cleared predicate devices, as assessed through non-clinical (mechanical) testing.
Here's why the requested information cannot be found in the provided text:
- AI/ML Device: The "Thunderbolt™ Minimally Invasive and Lancer™ Open Pedicle Screw Systems" are physical surgical implants (pedicle screws, rods, connectors) used for spinal stabilization. They are not AI/ML devices that perform diagnostic, predictive, or assistive functions involving data analysis.
- Acceptance Criteria (for AI/ML): The document discusses "Substantial Equivalence" based on "intended use/indications for use, materials, design features, operational principles, and levels of attachment" and mentions non-clinical (mechanical) testing. These are not acceptance criteria for an AI/ML model's performance (e.g., sensitivity, specificity, AUC).
- Study Proving Device Meets Acceptance Criteria (for AI/ML): The "Non-Clinical Evaluation" describes mechanical testing (Static Compression-Bending, Dynamic Compression-Bending, and Static Torsion tests per ASTM F1717). This is a purely physical performance evaluation, not a study involving test sets, ground truth, expert readers, or sample sizes relevant to AI/ML performance.
- Other AI/ML related questions: All other points in your request (sample size for test set, data provenance, number of experts, adjudication, MRMC study, standalone performance, type of ground truth, training set size, training ground truth) are specifically applicable to the evaluation of AI/ML algorithms and are therefore not present or relevant in this document.
In summary, the provided text describes a traditional medical device (spinal implants) and its regulatory clearance process, which relies on demonstrating substantial equivalence to already-cleared predicate devices through mechanical testing, not through AI/ML performance evaluation.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a symbol of three stylized human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 3, 2017
Choice Spine Ms. Kim Finch Manager of Regulatory Affairs 400 Erin Drive Knoxville, Tennessee 37919
Re: K171456
Trade/Device Name: Thunderbolt™ Minimally Invasive and Lancer™ Open Pedicle Screw Systems Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB, KWP Dated: June 16, 2017 Received: June 19, 2017
Dear Ms. Finch:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K171456
Device Name
Thunderbolt™ Minimally Invasive and Lancer™ Open Pedicle Screw Systems
Indications for Use (Describe)
The Thunderbolt™ Minimally Invasive and Lancer™ Open Pedicle Screw Systems are intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: desc disease (DDD; defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.
When used in a posterior percutaneous approach with MIS instrumentation, the Thunderbolt™ System is intended for non-cervical pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.
When used for posterior, non-cervical pedicle, and non-pedicle fixation The Lancer™ Open Pedicle Screw System is indicated for the following: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); turnor; pseudoarthrosis; and failed previous fusion. Overall levels of fixation are T1 to the Sacrum/Ilium. When used for fixation to the ilium, the lateral offset connectors of the Lancer™ Open Pedicle Screw System must be used in conjunction with pedicle screws placed at the S1 or S2 spinal level.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(K) Summary
| Date Prepared: | June 16, 2017 |
|---|---|
| Sponsor: | Choice Spine, LP400 Erin DriveKnoxville, TN 37919 |
| Phone: | 865-243-3969 |
| Fax: | 865-246-3334 |
| Contact Person: | Kim Finch, Manager of Regulatory Affairs |
| Proposed ProprietaryTrade Name: | The Thunderbolt™ Minimally Invasive and Lancer™ Open Pedicle Screw Systems |
| Product Class: | Class II |
| Classification Name: | 888.3070 – Thoracolumbosacral Pedicle Screw System |
| Device Product Code: | NKB, KWP |
| Purpose ofSubmission: | The purpose of this submission is to gain clearance for additional connectors tothe already cleared Lancer™ Open Pedicle Screw System. |
| Device Description: | The Thunderbolt™ Minimally Invasive and Lancer™ Open Pedicle Screw Systemsinclude implant components made of implant grade titanium alloy (Ti-6Al-4V ELI;ASTM F136) and cobalt chrome alloy (Co-28Cr-6Mo; ASTM F1537). The systemsalso include instruments made of PEEK (ASTM F2826), Tantalum (ASTM F560),stainless steel (ASTM F899/A564) and aluminum (ASTM B221). Thesecomponents are available in various designs and sizes that allow the surgeon tobuild an implant construct suited to a patient's anatomical and physiologicalrequirements.The components include: polyaxial pedicle screws, set screws, rods, instrumentsand sterilizer trays. The Lancer™ Open Pedicle Screw System also includesconnector and hook components. |
| Indications: | The Thunderbolt™ Minimally Invasive and Lancer™ Open Pedicle Screw Systemsare intended to provide immobilization and stabilization of spinal segments inskeletally mature patients as an adjunct to fusion in the treatment of the |
{4}------------------------------------------------
following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease (DDD; defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.
When used in a posterior percutaneous approach with MIS instrumentation, the Thunderbolt™ Minimally Invasive System is intended for non-cervical pedicle fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor, pseudoarthrosis; and failed previous fusion in skeletally mature patients.
When used for posterior, non-cervical pedicle, and non-pedicle fixation, the Lancer™ Open Pedicle Screw System is indicated for the following: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. Overall levels of fixation are T1 to the Sacrum/Ilium. When used for fixation to the ilium, the lateral offset connectors on the Lancer™ Open Pedicle Screw System must be used in conjunction with pedicle screws placed at the S1 or S2 spinal level.
Thunderbolt™ Minimally Invasive and Lancer™ Open Pedicle Screw Systems Predicate Devices: K132049 and K160775 (Primary Predicate). K151488 Orthofix FIREBIRD (Additional Predicate)
Summary of Technological
Characteristics:
Substantial equivalence is based on consideration of the following characteristics as well as a non-clinical evaluation, detailed below:
o Intended Use/Indications for Use:
The proposed Lancer™ Open Pedicle Screw System with additional connectors has similar indications as the predicate devices.
-
Materials:
The proposed and predicate systems utilize the same materials in their implants. Those materials are Ti-6Al-4V ELI per ASTM F136 band Co-28Cr-6Mo per ASTM F1537. -
. Design Features:
The proposed connectors and those of the predicate systems have similar design features in regards to style, function and size offerings.
{5}------------------------------------------------
Operational Principles:
The proposed connectors and those of the predicate systems have similar operational principles in regards to intended purpose, levels of use, method of attachment and locking mechanism.
● Levels of Attachment:
The proposed and predicate systems can be used in the Thoracic, Lumbar and Sacral/Iliac spine segments.
Non-Clinical Evaluation:
The proposed devices were subjected to Static Compression-Bending, Dynamic Compression-Bending and Static Torsion tests per ASTM F1717. The results of these tests were compared to the same results obtained for the primary predicate system. Results show that Lancer™ Open Pedicle Screw System with the proposed connectors does not create a new worst-case construct and is sufficient for its intended use.
Substantial Equivalence
Conclusion:
Choice Spine concludes that the Thunderbolt™ Minimally Invasive and Lancer™ Open Pedicle Screw Systems with the additional proposed connectors are equivalent to Thunderbolt™ Minimally Invasive and Lancer™ Open Pedicle Screw Systems (K132049 and K160775; Primary Predicate), as well as the Orthofix FIREBIRD System (K151488; Secondary Predicate). Equivalence is based on similarities in the indications for/intended use, materials, design, product offering, operational principles and levels of attachment. The proposed system is substantially equivalent to the Thunderbolt™ Minimally Invasive and Lancer™ Open Pedicle Screw Systems (K132049) based also on mechanical performance. The technological differences between the subject and predicate devices do not raise new questions regarding the safety and effectiveness of the device.
N/A