K Number
K171068
Device Name
OrthoView 7.2
Date Cleared
2017-10-18

(191 days)

Product Code
Regulation Number
892.2050
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
OrthoView is indicated for use when a suitable licensed and qualified healthcare professional requires access to medical images with the intention of using such images to plan or review a surgical procedure. OrthoView provides a set of tools and templates (representing prosthetic and fixation devices) to assist the healthcare professional in planning their surgery. The device is not to be used for mammography.
Device Description
OrthoView 7.2 is dedicated, digital, pre-operative planning and templating software used to create detailed pre-operative plans quickly and easily from digital x-ray images. OrthoView 7.2 is software to be used for medical purposes, performing these purposes without being part of a hardware medical device. The device provides one or more capabilities relating to the acceptance, transfer, display, storage, and digital processing of medical images.
More Information

Not Found

No
The document describes standard image processing and templating software for surgical planning and does not mention AI, ML, or related concepts.

No
The device is described as pre-operative planning and templating software. It assists healthcare professionals in planning surgery by processing medical images and providing tools and templates for prosthetic and fixation devices. It does not directly treat or alleviate a disease, injury, or disability.

No

Explanation: The device is described as pre-operative planning and templating software used to assist in planning surgical procedures. It does not state that it diagnoses conditions or diseases.

Yes

The device description explicitly states "OrthoView 7.2 is software to be used for medical purposes, performing these purposes without being part of a hardware medical device."

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use clearly states that OrthoView is used by healthcare professionals to access and use medical images for planning or reviewing surgical procedures. This is a diagnostic and planning tool used in vivo (on a living patient), not in vitro (on samples taken from the body).
  • Device Description: The description reinforces that it's software for processing and displaying medical images (digital x-ray images) for medical purposes, specifically pre-operative planning and templating.
  • Lack of IVD Characteristics: There is no mention of analyzing biological samples (blood, urine, tissue, etc.), detecting diseases or conditions based on such analysis, or providing information for diagnosis based on laboratory tests.

IVD devices are specifically designed to perform tests on samples taken from the human body to provide information for diagnosis, monitoring, or screening. OrthoView's function is centered around image-based surgical planning, which falls outside the scope of IVD.

N/A

Intended Use / Indications for Use

Ortho View is indicated for use when a suitable licensed and qualified healthcare professional requires access to medical images with the intention of using such images to plan or review a surgical procedure. OrthoView provides a set of tools and templates (representing prosthetic and fixation devices) to assist the healthcare professional in planning their surgery. The device is not to be used for mammography.

Product codes

LLZ

Device Description

OrthoView 7.2 is dedicated, digital, pre-operative planning and templating software used to create detailed pre-operative plans quickly and easily from digital x-ray images. OrthoView 7.2 is software to be used for medical purposes, performing these purposes without being part of a hardware medical device.

The device provides one or more capabilities relating to the acceptance, transfer, display, storage, and digital processing of medical images.

Mentions image processing

Yes

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

X-ray images

Anatomical Site

The anatomical site selected and the corresponding image presented by the device are decided and chosen by the physician. It is therefore beyond the scope of OrthoView 7.2 to specify an anatomical site or target population other than 'any'.

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Suitable licensed and qualified healthcare professional.
Provided for use specifically with third-party PACS equipment or as a standalone device.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

A full test report is provided for each completed test. The test report includes the objective of the test, description of the test methods and procedures, study endpoint(s), predefined pass/fail criteria, results summary and conclusions.

The feature tests of the predicate device now form the regression tests of the subject device verifying that OrthoView 7.2 includes the same functionality as the predicate device.

Non-clinical testing was performed to determine substantial equivalence. This non-clinical testing included the regression tests associated with the predicate device. Testing verified that the accuracy and performance of the system is adequate and performs as intended. This software has been extensively verified via code reviews, automated and manual testing.

All manual testing is performed on a fully configured system installed on hospital representative environments using procedure-specific images to emulate as close as possible intended use. All new features are checked by a surgeon to verify clinical performance.

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Non-clinical tests were performed using regression tests associated with the predicate device. The testing verified the accuracy and performance of the system as adequate and as intended. The software was extensively verified via code reviews, automated and manual testing. Manual testing was performed on fully configured systems installed in hospital representative environments using procedure-specific images to emulate intended use. All new features were checked by a surgeon for clinical performance. Feature additions since OrthoView 4 were verified to be within the scope of the intended use of the predicate, supporting the finding of substantial equivalence.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K063327

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

0

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. Food & Drug Administration".

Meridian Technique Ltd % Mr. Dave Yungvirt CEO Third Party Review Group, LLC The Old Station House 24 Lackawanna Place MILLBURN NJ 07041

Re: K171068

Trade/Device Name: OrthoView 7.2 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: October 9, 2017 Received: October 12, 2017

Dear Mr. Yungvirt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

October 18, 2017

1

Page 2 - Mr. Dave Yungvirt

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

For

Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K171068

Device Name ORTHOVIEW 7.2

Indications for Use (Describe)

Ortho View is indicated for use when a suitable licensed and qualified healthcare professional requires access to medical images with the intention of using such images to plan or review a surgical procedure. OrthoView provides a set of tools and templates (representing prosthetic and fixation devices) to assist the healthcare professional in planning their surgery. The device is not to be used for mammography.

Type of Use (Select one or both, as applicable)
\ Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

3

510(k) Summary

Date Summary Prepared: October 16, 2017

The following section is included as required by the Safe Medical Devices Act (SMDA) of 1990 and 21CFR 807.92.

Submitter information

Table 1

Submitter Company & ManufacturerMeridian Technique Ltd
Establishment registration number3004537781
Street Address2 Venture Road, Southampton Science Park
CitySouthampton
Postal codeSO16 7NP
CountryUK
Phone number+44(0) 2380 119663
Principal Contact personBarbara March
Contact titleQuality Manager
Contact e-mail addressBarbara.march@orthoview.com
Additional contact personAndy Enefer
Contact titleEngineering Manager
Contact e-mail addressAndy.enefer@orthoview.com

Submission information

THE DEVICE

Table 2

The Device Which Is The Subject Of This Submission
Device Trade/Proprietary NameOrthoView 7.2
Device ClassificationClass II
Device PanelRadiology
Common NameImage Processing System Radiological
Classification NamePicture Archiving and Communication Systems
Primary product codeLLZ (21 CFR § 892.2050)
Meridian Technique Ltd
OrthoView 7.2 - 510k Notification
October 16, 2017
Total Pages10
Page Number1

4

THE PREDICATE DEVICE

The Leqally Marketed Predicate Device [807.92(a)(3)]

To demonstrate substantial equivalence of OrthoView 7.2 to a device currently cleared for marketing in the USA, Meridian Technique Ltd will utilize the following device which has not been subject to a design-related recall:

Table 3

The Orthopaedic Predicate Device To Which Substantial Equivalence Is Claimed
ManufacturerMeridian Technique Ltd
Device Trade/Proprietary NameOrthoView 4
Device ClassClass II
Device PanelRadiology
Common NameImage Processing System Radiological
Classification NamePicture Archiving and Communication Systems
Primary product codeLLZ (21 CFR § 892.2050)
510(k) numberK063327
Decision date22nd November 2006

THE DEVICE DESCRIPTION [807.92(a)(4)]

OrthoView 7.2 is dedicated, digital, pre-operative planning and templating software used to create detailed pre-operative plans quickly and easily from digital x-ray images. OrthoView 7.2 is software to be used for medical purposes, performing these purposes without being part of a hardware medical device.

The device provides one or more capabilities relating to the acceptance, transfer, display, storage, and digital processing of medical images.

Conditions of Use

  • OrthoView 7.2 is either a standalone device or is provided for use specifically with third . party PACS equipment.
  • OrthoView 7.2 may be installed on each workstation or deployed onto a workstation from ● a central server.
  • The OrthoView 7.2 software has no direct or indirect patient-contacting components and ● makes no intentional contact with the patient or any other persons.
  • . The OrthoView 7.2 software may communicate with a third party source of digital X Ray images and may therefore be connected to a proprietary network.
  • Some optional features require Internet access but OrthoView 7.2 can be used without such access.

Service and Maintenance Technicians will be required to:

  • Install initial software and software enhancements. ●
  • Maintain access codes and rights. ●
  • Install Templates and Licenses. ●

The scope for contact is the same as for the Operator.

| Meridian Technique Ltd

OrthoView 7.2 - 510k NotificationTotal PagesPage Number
October 16, 2017102

5

Table 4

Minimum Operating System Requirements
ComputerPC or MAC compatible computer
Operating System MAC OSOS X 10.12 or higher
Operating System WindowsWindows 7, 8, 8.1 or 10
Processor1 GHz CPU
Memory2GB RAM
Free Hard Disc Space2GB free disk space
Graphics/Display1280 x 1024 colour graphics display
(Optional) Intranet Connection to PACS.100MBit/Sec
(Optional) Internet Connection (required for
case management, auto template update
and anonymous stat collection features
only).2MBit/Sec.

Note -this is a minimum requirement for OrthoView 7.2 and NOT for the associated PACS system with which it is operating. It is expected that the associated PACS system will exceed this specification.

INTENDED USE & INDICATIONS FOR USE:

The subject of this submission - OrthoView 7.2

Orthoview is indicated for use when a suitable licensed and qualified healthcare professional requires access to medical images with the intention of using such images to plan or review a surgical procedure. OrthoView provides a set of tools and templates (representing prosthetic and fixation devices) to assist the healthcare professional in planning their surgery. The device is not to be used for mammography.

| Meridian Technique Ltd
OrthoView 7.2 - 510k Notification

October 16, 2017Total PagesPage Number
103

6

FeatureOrthoView 7.2Predicate Device - OrthoView 4
Device for premarket notificationFDA marketing clearance K063327
following 510(k) submission November 2006
Indications for UseThe Intended use of OrthoView 7.2 is when a suitably
licensed and qualified healthcare professional requires
access to medical images with the intention of using such
images for pre-operative planning of a potential surgical
procedure.OrthoView 4 has the identical intended use and
uses identical software and algorithms to achieve
this purpose
OrthoView 7.2 is indicated for use when a suitably licensed
and qualified healthcare professional requires access to
medical images with the intention of using such images to
plan or review a surgical procedure. OrthoView 7.2 provides
a set of tools and templates (representing prosthetic and
fixation devices) to assist the healthcare professional in
planning their surgery.
The device is not to be used for mammographyOrthoView 4 is indicated for use when a suitably
licensed and qualified healthcare professional
requires access to medical images with the
intention of using such images, in conjunction with
templates for prosthetic and fixation devices, for the
purposes of choosing the nature and characteristics
of the prosthetic/fixation device to be used when
planning a potential surgical procedure. In addition,
Trauma and Osteotomy modules and Trauma
Templates are provided to extend the range of
functionality available to the healthcare
professional.

Table 5 – Comparison and Explanation of the Instructions For Use Differences [807.92(a)(5)]

Meridian Technique LtdTotal PagesPage Number
OrthoView 7.2 - 510k Notification104
October 16, 2017

7

Table 5 – Comparison and Explanation of the Instructions For Use Differences [807.92(a)(5)]

FeatureOrthoView 7.2Predicate Device - OrthoView 4
Device for premarket notificationFDA marketing clearance K063327
following 510(k) submission November 2006
DiscussionThe indications for use have been changed to reflect accurately the usage of the device and the predicate. The
wording has been slightly changed to reflect that surgeons can plan surgery using either device without actually
using a prosthetic or fixation device. Either device allows the healthcare professional to use a set of wizards to
make measurements and angles between anatomical landmarks. While these measurements and angles are often
used to choose the nature and characteristics of a prosthetic/fixation device they are sometimes sufficient output
in themselves for the surgical plan and so a surgeon may use the device without applying templates/fixation
devices. Several procedures within the predicate and new device do not require templates/fixation devices to be
used. Review has also been added since a healthcare professional may use OrthoView 7.2 or the predicate
device to review the plan or surgery of another surgeon by viewing it within the device. The indications for use
have therefore been modified to provide a more accurate usage statement allowing for situations whereby the full
feature set is not required in order to produce a viable plan for some procedures. In addition, the sentence added
in the predicate device submission has been removed. This is because the additional features mentioned are
already completely covered in the rest of the indications for use (they were originally added to the predicate device
to highlight the extension of functionality in OrthoView 4 added since its previous 510K).
Despite the change of wording the new device is substantially equivalent to the predicate device in that there is no
change in usage but a clarification of possible and permissible current usage.
Meridian Technique LtdTotal PagesPage Number
OrthoView 7.2 - 510k Notification105
October 16, 2017

8

The Application Of The Device And The Intended Patient Population [807,92(a)(5)]

The anatomical site selected and the corresponding image presented by the device are decided and chosen by the physician.

It is evident that the decision regarding whether a patient is included in the target population for either application resides with the patient's physician only and hence, neither application of the device has any direct influence on the target patient population. It is therefore beyond the scope of OrthoView 7.2 to specify an anatomical site or target population other than 'any'.

Technological Characteristics [807.92(a)(6)1

OrthoView 7.2 is medical device software that permits the pre-planning of surgical procedures by permitting image viewing and manipulation within a PACS computer or standalone environment. It is designed to integrate into a third-party provided PACS hardware and software environment. This implies the provision of a computer specification appropriate to a PACS environment. OrthoView 7.2 can be configured to be launched from within a networked environment or as a standalone application. The device is intentionally non-patient contacting patient and does not deliver medication or therapeutic treatment. Energy used or delivered to the user is associated with the related workstation only and not the device.

Performance Data:

Conformance Standards & Guidances

The device complies with the following international and FDA-recognized consensus standards and FDA quidance documents:

  • ISO 14971:2012 & ISO14971:2007 Medical devices Application of risk management ● to medical devices
  • NEMA PS 3.1 - 3.20 (2016), Digital Imaging and Communication in Medicine (DICOM)
  • . IEC 62304:2006 Medical device software - Software life cycle processes
  • IEC 62336:2015. Medical devices Application of usability engineering to medical ● devices
  • . ISO-15223-1:2012-Medical Devices - Symbols To Be Used With Medical Device Labels, Labelling And Information, Part 1 General Requirements
  • . ISO 14155:2011 - Clinical Investigation of Medical Devices for Human Subjects. Good Clinical Practice
  • Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
  • Guidance for the Submission of Premarket Notifications for Medical Image Management Devices
  • . Applying Human Factors and Usability Engineering to Medical Devices

Testing

A full test report is provided for each completed test. The test report includes the objective of the test, description of the test methods and procedures, study endpoint(s), predefined pass/fail criteria, results summary and conclusions.

The feature tests of the predicate device now form the regression tests of the subject device verifying that OrthoView 7.2 includes the same functionality as the predicate device.

Non-clinical tests [807.92(b)(1)]

Each release over time has experienced thorough testing and each new release has had its clinical features evaluated by a surgeon (within a non-clinical environment).

Meridian Technique LtdTotal PagesPage Number
OrthoView 7.2 - 510k Notification
October 16, 2017106

9

Non-clinical testing was performed to determine substantial equivalence. This non-clinical testing included the regression tests associated with the predicate device. Testing verified that the accuracy and performance of the system is adequate and performs as intended. This software has been extensively verified via code reviews, automated and manual testing.

All manual testing is performed on a fully configured system installed on hospital representative environments using procedure-specific images to emulate as close as possible intended use. All new features are checked by a surgeon to verify clinical performance.

Feature additions since the OrthoView 4 have been verified as being within the scope of the intended use of the predicate. In this way, data generated from each test supports the finding of substantial equivalence.

Substantial Equivalence

The characersitics of the device and the predicate device allow the healthcare professional to use a set of wizards to make measurements and angles between anatomical landmarks. These measurements and angles may be used to choose the nature and characteristics of a prosthetic/fixation device and they are also sufficient output for a surgical plan. Ortho/iew 7.2 and the predicate device also allow the healthcare professional to review the plan of another surgeon by viewing it within the device.

Functional Comparison

TECHNOLOGICAL COMPARISON
Physical Properties
OrthoView 7.2 (this submission)OrthoView 4 (Predicate Device)
System Requirements- IDENTICAL
Windows 7, 8.1, 8 or 10, MAC OS X 10.12+
2GB RAM
2GB Free Disk Space
1280 x 1024 Colour Graphics DisplayWindows XP or Windows 2000
512MB RAM (1GB Recommended)
100Mb Free disk Space
1024x768 Colour Graphics Display
Although the System requirements have
changed they are aligned with the improved
hardware available in hospitals since
release of the predicate device. The current
operating environment is identical in that it
reflects the availability of Desktop devices
to the clinicians in the intended use.
Although MAC is a new Operating System
it is in keeping with meeting the needs of
the clinicians using the device.
Installation - IDENTICAL
There is no change to the delivery
installation, licensing and integrations of
OrthoView.To be downloaded from the Internet, installed
and unlocked using a license provided by
Meridian Technique Ltd
Launch - IDENTICAL
The means of launching OrthoView 7.2 are
identical to the predicate device.Can be configured to be launched from within
a PACs workstation environment or as a
standalone application.
Login - IDENTICAL
Access/Login rights are identical but now
include the capability to use Hospital
enforced access rights through Active
Directory.Grant access rights only to authorized users
(via PC password system).

Table 6

| Meridian Technique Ltd
OrthoView 7.2 - 510k Notification
October 16, 2017 | Total Pages
10 | Page Number
7 |

----------------------------------------------------------------------------------------------------------------------

10

| TECHNOLOGICAL COMPARISON

Physical Properties
OrthoView 7.2 (this submission)
Image Loading - IDENTICAL
The Image loading is identical and the
expected modalities are also unchanged.OrthoView 4 (Predicate Device)
Receive X-Ray images in a digital format
from third party X-Ray machines/ X-Ray
digitisers or PACS systems. Orthoview 4
provides the means of recording, storing and
retrieving the templating process steps
performed by the licensed medical
professional when assessing the optimum
prosthetic device for a particular patient.
Image Manipulation - IDENTICAL
Image manipulation functionality such as
Window/level/flip etc is identical. The
function of the existing 90 degree rotation
tools has been extended to horizontal and
vertical alignment.
Templates may now be obtained through
an online server.Orthoview 4 permits to pre-plan surgical
procedures by permitting image viewing and
manipulation and prosthetic template overlay
within a PACS workstation or standalone
environment.
Orthoview 4 processes such images securely
with respect to patient confidentiality, patient
identification and image integrity and allows
the image to be retrieved for processing as
follows:
Scaling of the image.
Selection of appropriate prosthetic and fixing
device manufacturer and size range
templates.
Overlaying the template on the image and
permitting selection of appropriate size of
prosthetic/fixing.
Provide additional functionality in the form of
Trauma Wizards, Pediatrics Wizards
Osteotomy Wizards, Fracture Reduction
Tools and Trauma Templates
Print and archive appropriate reports.
Receive and store templates for prostheses
and fixations supplied by Meridian for
particular manufacturers' range of products.
Provide traceability of operator, date and pre-
operative plans.
Scaling - IDENTICAL
The workflow and practice of scaling
remains unchanged.A marker (not included) is a radio-opaque
object of known size which is placed on the
patient in the same plane as the joint being
imaged. Using markers is a more accurate
method to confirm oversize but a rudimentary
ruler scaling tool is also available to establish
pixel spacing on the x-ray, in the event that
this information is not present on the image.
Analysis Methods - IDENTICAL
Different methods of analysing data for the
pre-operative plan are now available to the
user. The mechanism for selecting the
analysis methods and wizards are the
same, only the presentation of the relevant
wizards has been improved.The analysis or report summary is generated
by the Report Wizard available from the GUI
which offers the options of printing and or
saving the analysis of the pre-operative plan.

| Meridian Technique Ltd
OrthoView 7.2 - 510k Notification
October 16, 2017 | Total Pages
10 | Page Number
8 |

----------------------------------------------------------------------------------------------------------------------

11

TECHNOLOGICAL COMPARISON
Physical Properties
OrthoView 7.2 (this submission)
Landmarks - IDENTICAL
The mechanism for identifying landmarks
on an x-ray is identical. The clinical
relevance of these points is the sameOrthoView 4 (Predicate Device)
Various Wizards are available to identify the
relevant anatomical points or landmarks
according to the clinical procedure being
planned.
Contours - IDENTICAL
Contours are identical to the wizards used
in the predicate device to define cut areas
in Limb Deformity etc.Various wizards are available for
assessment, planning and reduction for
centre of rotation and angulation in planning
limb deforinty correction
Cut Positions - IDENTICAL
Cut positioning functionality is identical to
the wizards used in the predicate device for
osteotomies etc.The Osteotomy wizard defines the position
and angle of cut required when straightening
a shaft.
Reduction - IDENTICAL
No change to this functionality.Various Reduction tools and wizards are
available for manipulating image fragments
and specialized animators for limb deformity
procedures
Measurements - IDENTICAL
The Measurement technology and function
is identical to the predicate device. There is
no difference to the importance or likelihood
of error from these measurements.The many joint specific wizards rely on
measurements in order to correctly pre-plan
surgery
Reporting - IDENTICAL
The reporting function is identical with the
added improvement of an analysis specific
display.Orthoview 4 provides the means of
recording, storing and retrieving the
templating process steps performed by the
licensed medical professional when
assessing the optimum prosthetic device for
a particular patient.
Saving/Commit - IDENTICAL
The Save & Commit functions are identical.An OrthoView 4 planning session can be
'Saved' or 'Committed'. 'Saving' a session
allows it to be re-opened for review.
'Committing' creates a permanent, read-only
record.
Image Storage - IDENTICAL
Images are still stored permanently in
PACS Systems and may be stored
temporarily in files. OrthoView 7.2 allows
those temporary plans to be stored as files
within a Case Management System if
desired.Images are permanently stored within PACS
systems within the Hospital. Interim plans
may be stored and transported as files.

Risk Analysis

A Risk Analysis has been completed for the device in accordance with ISO14971:2007 and risk control measures have been implemented to mitigate identified hazards, including those resulting from intentional or inadvertent misuse of the device. Any risks posed by the design, manufacture and use of OrthoView 7.2 have been evaluated. The risk analysis indicates that the subject device has the same risk profile in the patient environment, using the same materials and construction principles as the predicate device. Accordingly, it is believed that no further clinical work is required.

Meridian Technique LtdTotal PagesPage Number
OrthoView 7.2 - 510k Notification
October 16, 2017109

12

Safety and Effectiveness

OrthoView has been in commercial distribution since 2001, has never been the subject of a recall or medical device report and has proven to be safe and effective in clinical usage. Each release has been thoroughly tested and the clinical features have been evaluated by a surgeon (within a non-clinical environment).

Conclusion

Based on the information submitted in this premarket notification, and based on the indications for use, technological characteristics, performance testing and technological comparison, OrthoView 7.2 raises no new questions of safety and effectiveness and is substantially equivalent to the predicate devices in terms of safety, efficacy and performance.


Meridian Technique LtdTotal PagesPage Number
OrthoView 7.2 - 510k Notification1010
October 16, 2017