K Number
K170978
Date Cleared
2017-07-19

(107 days)

Product Code
Regulation Number
892.1000
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Embrace Neonatal MRI System is indicated for use as a magnetic resonance imaging device for producing axial, sagittal, coronal and oblique images that displays the internal structure of neonatal head with circumference of up to 38 cm and weight between 1Kg and 4.5 Kg. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.

Device Description

The Embrace Neonatal MRI System is a 1Tesla Permanent MRI system producing MR images of Neonatal head. During MRI scan, body parts to be imaged are held within a uniform static magnetic field, and are subject to sequences of RF pulses and gradient magnetic fields. The signal from the precession of the magnetization created by these fields is sampled and processed to produce image data.

AI/ML Overview

The provided document is a 510(k) Premarket Notification for the Embrace Neonatal MRI System. It details the device's indications for use, comparison to a predicate device, and the non-clinical testing performed. However, it explicitly states that no clinical studies were conducted to establish substantial equivalence for this device. Therefore, it does not contain the acceptance criteria or study details regarding device performance from a clinical perspective.

Here's the information that can be extracted or inferred:

1. A table of acceptance criteria and the reported device performance:

Based on the document, direct clinical acceptance criteria and reported device performance from a clinical study are not available. The document focuses on non-clinical testing and comparison to a predicate device for substantial equivalence.

Acceptance Criteria (Non-Clinical)Reported Device Performance (Non-Clinical)
Compliance with ES/IEC 60601-1Demonstrated (implied by submission and clearance)
Compliance with IEC 60601-1-2Demonstrated (implied by submission and clearance)
Compliance with IEC 60601-2-33Demonstrated (implied by submission and clearance)
Compliance with IEC 60601-2-20Demonstrated (implied by submission and clearance)
Compliance with IEC 60601-1-6Demonstrated (implied by submission and clearance)
Compliance with IEC 62304Demonstrated (implied by submission and clearance)
Compliance with NEMA MS-1, MS-2, MS-3, MS-5, MS-12Demonstrated (implied by submission and clearance)
Demonstrated performance via System and Software V&VDemonstrated (implied by submission and clearance)
Demonstrated diagnostic quality images from phantom studiesDemonstrated (stated in "Summary of Clinical Tests" section)

2. Sample size used for the test set and the data provenance:

  • Test set for clinical performance: Not applicable, as no clinical studies were performed.
  • Test set for non-clinical performance (phantom images, V&V): Not specified in the document. The provenance of this data is from Aspect Imaging Ltd., likely internally generated during device development.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

Not applicable, as no clinical studies with expert-established ground truth were performed. The document mentions "When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis," but this refers to the intended use of the device, not a performance study.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

Not applicable, as no clinical studies with adjudication were performed.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. The device is an MRI system, not an AI diagnostic tool designed to assist human readers. Moreover, no clinical studies were performed.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

Not applicable. This is an MRI system, not a standalone algorithm. Its performance is evaluated on its ability to produce diagnostic quality images, which are then interpreted by a human physician.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

Not applicable for a clinical study. For the non-clinical phantom images, the "ground truth" would be the known properties and geometry of the phantoms themselves, used to assess image quality, resolution, and artifact levels.

8. The sample size for the training set:

Not applicable. This is an MRI system, not an AI model that requires a specific training set in the conventional sense. The "training" of the system would involve engineering and calibration using test data and phantoms.

9. How the ground truth for the training set was established:

Not applicable. As described in point 8, this is an MRI system, not an AI model with a training set. If considering the calibration and engineering process, the "ground truth" would be established through physical measurements, known phantom compositions, and established MRI physics principles.

In summary, the provided FDA 510(k) letter and summary explicitly state that no clinical studies were conducted to support substantial equivalence for the Embrace Neonatal MRI System. The device was cleared based on non-clinical testing, including phantom images, to demonstrate its ability to provide diagnostic quality images, and comparison to a legally marketed predicate device (Wrist 3 MRI System). The rationale given for not performing clinical studies was the "vulnerable patient population" (neonates) and that the "benefits of sample clinical images did not outweigh the risks associated with acquisition of those images."

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 20, 2017

Aspect Imaging Ltd. % Israel Citron Vice President Quality Assurance and Regulatory Affairs 27 Shaked Street Industrial Area, Hevel Modi'in P.O. Box 926 Shoham, 6085001 Israel

Re: K170978

Trade/Device Name: Embrace Neonatal MRI System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH Dated: June 4, 2017 Received: June 7, 2017

Dear Israel Citron:

This letter corrects our substantially equivalent letter of July 19, 2017.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Michael D'Hara For

Robert A. Ochs. Ph.I Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K170978

Device Name Embrace Neonatal MRI System

Indications for Use (Describe)

The Embrace Neonatal MRI System is indicated for use as a magnetic resonance imaging device for producing axial, sagittal, coronal and oblique images that displays the internal structure of neonatal head with circumference of up to 38 cm and weight between 1Kg and 4.5 Kg. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image contains the logo for Aspect Imaging. The logo consists of a stylized cube made up of three layers, with each layer a different color: teal, green, and yellow. To the right of the cube is the company name, "aspectimaging," written in a sans-serif font, with the word "aspect" in purple and "imaging" in gray.

510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:
Date:March 28, 2017
Submitter:Aspect Imaging Ltd.
27 Shaked St.
Industrial Area Hevel Modi'in
P.O. Box 926
Shoham, 6085001
ISRAEL
Primary Contact Person:Israel A. Citron
VP Quality Assurance and Regulatory Affairs
Aspect Imaging LTD
Tel: +972-73-223-9000 Ext 90
Fax: +972-3-716-5606
E mail: icitron@aspectimaging.com
Device Trade Name:Embrace Neonatal MRI System
Common/Usual Name:MRI System
Product ClassificationII
Classification Name:Magnetic Resonance Diagnostic Device (21 CFR892.1000)
Product Code:LNH, MOS
Predicate Device(s):K160185 Wrist 3 MRI System
Reference Device(s)K161918 SREE MRI Transport Incubator**referenced for the Patient Bed main specifications
Device Description:The Embrace Neonatal MRI System is a 1Tesla PermanentMRI system producing MR images of Neonatal head. DuringMRI scan, body parts to be imaged are held within a uniformstatic magnetic field, and are subject to sequences of RF pulsesand gradient magnetic fields. The signal from the precession ofthe magnetization created by these fields is sampled andprocessed to produce image data.
Intended Use:The Embrace Neonatal MRI System is indicated for use as amagnetic resonance imaging device for producing axial, sagittal,coronal and oblique images that displays the internal structure of

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for Aspect Imaging. The logo consists of a geometric shape on the left and the company name on the right. The geometric shape is made up of three stacked squares, each in a different color: blue, green, and yellow. The company name is written in a sans-serif font, with the word "aspect" in purple and the word "imaging" in gray.

510(k) Premarket Notification Submission

Comparison ofSpecificationsPredicate DeviceWrist 3 MRI System (K160185)Proposed DeviceEmbrace Neonatal MRISystem
Intended Use/Indication forUseThe Wrist 3 MRI System isintended for use as a magneticresonance imaging device forproducing axial, sagittal,coronal, and oblique images ofthe internal structure of theWrist and Hand. Wheninterpreted by a trainedphysician, these images provideinformation that can be usefulin determining a diagnosis.The Embrace Neonatal MRISystem is indicated for use as amagnetic resonance imaging devicefor producing axial, sagittal,coronal and oblique images thatdisplays the internal structure ofneonatal head with circumferenceof up to 38 cm and weight between1Kg and 4.5 Kg. When interpretedby a trained physician, theseimages provide information thatcan be useful in determining adiagnosis.
Patient PopulationPatients requiring MR images ofthe wrist and HandPatients requiring MR images ofthe Neonatal Head
Anatomical SitesWrist and HandNeonatal Head
Environment of UseHospital or clinical settingHospital setting
Energy Used and/or deliveredMagnetic ResonanceMagnetic Resonance
Human FactorsThe Wrist 3 MRI System isdesigned similar to othercommercially available MRISystems and therefore is familiarand easy for use for the user.Furthermore, the device contains auser-friendly software interfacethrough which the user may easilyaccess all device functions.The Embrace Neonatal MRISystem is designed similar toother commercially available MRISystems and therefore is familiarand easy for use for the user.Furthermore, the device contains auser-friendly software interfacethrough which the user may easilyaccess all device functions.
Hardware Changes
Magnet:
- Physical Dimensions- Bore opening- WeightField StrengthGradient:138x87x82 cm85x220 mm1300 (1700 with Trolley) Kg1 Tesla Permanent Magnet171x145x229 cm184x260 mm5500 (5680 with PB) Kg1 Tesla Permanent Magnet
-Strength215 mT/m150 mT/m
-Rise Time0.2 mSec0.3 mSec
-Slew Rate1074 T/m/Sec500 T/m/Sec
Computer:
-Display:15" Touch Display24" LED Display
RF CoilsCoil Type2 Coils: Small/LargeTX/RX1 Head CoilsTX/RX
Coil GeometryOvalCylindrical
Inner dimensions (mm)60X100X169 / 75X135X192143 Diameter
Coil DesignLinear VolumeLinear Volume
Patient Bed:Reference deviceProposed Device
Comparison ofSREE MRI TransportEmbrace Neonatal MRI
SpecificationsIncubator (K161918)System
Target populationpre- and term newborns and infants upto 1 month of age and 4.5Kg wholebody weight or 55cm whole bodylength (21.7").neonates with head circumferenceof up to 38 cm and weightbetween 1Kg and 4.5 Kg
-Dimensions121.9cm L x 40.6cm W x 30.5cn H60.6cm W x 120cm H x 140cm L
-Patient weight capacity4.5Kg Max4.5 Kg Max
Compartment length55 cm58 cm
Set Temperature28 °C - 39°C20.5 °C - 36.5°C
Warm Up time45 minutes50 minutes
Temperature controlAirAir
Humidity controlnono
MRI compatibility1.5 and 3 TAspect Embrace 1T

{5}------------------------------------------------

510(k) Premarket Notification Submission

Determination of Substantial Equivalence:

Summary of Nonclinical Testing:

The Embrace Neonatal MRI System and its applications were tested to comply with the below voluntary consensus standards.

  • ES/IEC 60601-1
  • IEC 60601-1-2
  • IEC 60601-2-33 l
  • IEC 60601-2-20
  • IEC 60601-1-6
  • ' IEC 62304
  • 트 NEMA MS-1
  • NEMA MS-2
  • NEMA MS-3
  • NEMA MS-5
  • 트 NEMA MS-12

In addition, System and Software verification and Validation testing were performed to demonstrate performance of the device as part of design controls activity.

The following design control activities were applied to the development of the system:

  • Risk Management
  • Requirements Management
  • Design Reviews
  • Unit level module verification
  • System Integration verification
  • Performance verification
  • Safety and EMC testing
  • 트 Simulated use validation testing
  • Sample phantom images

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Aspect Imaging. The logo consists of a stack of three squares in different colors, with the top square being blue, the middle square being green, and the bottom square being yellow. To the right of the squares is the word "aspectimaging" in purple.

Summary of Clinical Tests:

The subject of this premarket submission, Embrace Neonatal MRI System, did not require clinical studies to support substantial equivalence. Since neonates are a vulnerable patient population, for this particular device, the benefits of sample clinical images did not outweigh the risks associated with acquisition of those images. Therefore, substantial equivalence of the Embrace Neonatal MRI System was determined based on phantom images that demonstrate the ability of the system to provide diagnostic quality images, and sample clinical images of the neonatal head were not necessary to support the premarket submission for this device.

Based on bench testing and phantom image studies, Aspect Conclusion: Imaging LTD believes the Embrace Neonatal MRI System is as safe, as effective, as its predicate device and did not raise new safety or effectiveness issues.

Aspect Imaging LTD considers the Embrace Neonatal MRI system to be substantially equivalent to its predicate device.

§ 892.1000 Magnetic resonance diagnostic device.

(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.