K Number
K170914

Validate with FDA (Live)

Date Cleared
2017-11-21

(238 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TXM Support Catheters are percutaneous, single lumen catheters designed for use in the peripheral vascular system. TXM Support Catheters are intended to guide and support a guide wire during access of the vasculature, allow for wire exchanges and provide a conduit for the delivery of saline solutions or diagnostic contrast agents.

Device Description

The TXM Support Catheter is an over the wire (OTW) single lumen catheter with atraumatic tapered tip. The distal tip catheter shaft has 6 radiopaque markers (including the tip) that works as an aid to estimate positioning within the vasculature. It consists of a lubricious inner liner made from Teflon, and stainless steel braid over the liner and an outer layer that consists of clear Nylon 12 polymer. One gold marker band is positioned on the proximal side of the tip. A polycarbonate hub is attached to the proximal end of the TXM Support Catheter. The TXM Support Catheter system is offered in 8 models, four with a straight tip and four with an angled tip, and the angled tip configuration has an angle of 30°. All models will have an outer diameter of the 4 French size (0.052") and an inner diameter of 0.039" tapered down to 0.0355" towards the distal tip so that is compatible with a 0.035" guidewires. The device is available in lengths of 65cm, 90cm. 135cm, and 150cm. The catheter is coated with a lubricious hydrophilic coating. The device is supplied sterile and is intended for single use only.

AI/ML Overview

This document describes the TXM Support Catheter and its substantial equivalence to a predicate device, focusing on bench and biocompatibility testing. It is a 510(k) submission, used for medical devices that are substantially equivalent to a legally marketed predicate device.

Here's an analysis of the provided text in relation to your request about acceptance criteria and studies:

1. Table of Acceptance Criteria and Reported Device Performance:

The document provides a summary of bench testing performed on the TXM Support Catheter. For each test, it lists the applicable standard or internal test method and indicates "Pass" for both non-aged (T=0) and accelerated-aged (T=2) devices. While it doesn't explicitly state quantitative acceptance criteria for each test in a compact table, the "Pass" result implies that the device met these criteria.

Test NameApplicable Standard or Internal Test MethodAcceptance CriteriaReported Device Performance (T=0 & T=2)
Dimensional And Physical AttributesInternal Test Method(Implicit: Conformance to specifications)Pass
Catheter CompatibilityInternal Test Method(Implicit: Compatibility with specified components)Pass
Air LeakageISO 10555-1(Implicit: No air leakage as per standard)Pass
TrackabilityInternal Test Method(Implicit: Meets specified trackability performance)Pass
CrossabilityInternal Test Method(Implicit: Meets specified crossability performance)Pass
Catheter StiffnessASTM D747-10(Implicit: Meets specified stiffness range)Pass
Gravity Flow RateISO 10555-1(Implicit: Meets specified flow rate)Pass
Power Injection for Flow RateISO 10555-1(Implicit: Meets specified power injection flow rate)Pass
Lubricity of Hydrophilic CoatingInternal Test Method(Implicit: Meets specified lubricity)Pass
Durability of Hydrophilic CoatingInternal Test Method(Implicit: Meets specified coating durability)Pass
Liquid LeakageISO 10555-1(Implicit: No liquid leakage as per standard)Pass
Static Burst PressureISO 10555-1(Implicit: Withstands specified burst pressure)Pass
Tensile StrengthISO 10555-1(Implicit: Withstands specified tensile forces)Pass
RadiopacityASTM-F640-12(Implicit: Meets specified radiopacity levels)Pass
TorqueabilityInternal Test Method(Implicit: Meets specified torqueability performance)Pass
Torque StrengthInternal Test Method(Implicit: Withstands specified torque strength)Pass
Kink ResistanceInternal Test Method(Implicit: Resists kinking under specified conditions)Pass
Corrosion ResistanceISO 10555-1(Implicit: Shows no significant corrosion)Pass
Coating Integrity and ParticulatesInternal Test Method(Implicit: Coating remains intact; minimal particulates)Pass
Packing Integrity (Seal Strength)ASTM F88 / F88M - 15(Implicit: Seal strength meets standard)Pass
Packing Integrity (Packing Dye Penetration)ASTM F1929-15 / F1886-09(Implicit: No dye penetration as per standard)Pass
Shipping and TransitISTA 3A(Implicit: Withstands shipping and transit stresses)Pass
Accelerated AgingASTM F1980-07(Implicit: Demonstrates anticipated shelf-life)Pass
Female Luer VerificationISO 594(Implicit: Conforms to luer connection standard)Pass

2. Sample Size Used for the Test Set and Data Provenance:

The document mentions "Testing was performed on non-aged devices (T=0) as well as on devices subject to 2 years of accelerated aging (T=2)." However, the specific sample sizes for each test are not provided. The data provenance indicates that these were bench tests, meaning they were performed in a laboratory setting. There is no information regarding country of origin of the data as it's not clinical data, and the tests are inherently prospective in nature for a new device.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

This report describes bench testing for a medical device (catheter), not a diagnostic or prognostic algorithm involving human interpretation of medical images or data. Therefore, the concept of "ground truth established by experts" and their qualifications as you might typically find in AI/CADe submissions is not applicable here. The "ground truth" for these tests is defined by the technical specifications, standards (like ISO or ASTM), and internal test methods that define acceptable performance for physical and material properties.

4. Adjudication Method for the Test Set:

Similar to point 3, adjudication methods are not applicable for these types of bench tests. The evaluation of test results against predefined acceptance criteria is typically a technical assessment by engineers or technicians, not a consensus-based adjudication process.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No, a MRMC comparative effectiveness study was not done. This document is for a physical medical device (catheter), not an AI or imaging diagnostic software. Therefore, the concept of human readers improving with or without AI assistance is not relevant.

6. Standalone (Algorithm Only) Performance:

No, a standalone (algorithm only) performance study was not done. This is a physical support catheter, not an algorithm.

7. Type of Ground Truth Used:

The "ground truth" for the performance testing cited is based on engineering specifications, recognized industry standards (ISO, ASTM), and internal test methods. These standards and specifications define the expected physical, chemical, and functional properties of the device. For example, for "Air Leakage," the ground truth is defined by the criteria within ISO 10555-1.

8. Sample Size for the Training Set:

Not applicable. This document describes a physical medical device, not an AI or machine learning model that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

Not applicable. As there is no training set for a physical device, there is no ground truth to be established for it in this context.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

November 21, 2017

Texas Medical Technologies, Inc. % Mr. E.J. Smith Consultant Smith Associates 1468 Harwell Avenue Crofton, Maryland 21114

Re: K170914

Trade/Device Name: TXM Support Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: October 13, 2017 Received: October 16, 2017

Dear Mr. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good

{1}------------------------------------------------

manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Kenneth J. Cavanaugh - S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K170914

Device Name TXM Support Catheter

Indications for Use (Describe)

The TXM Support Catheters are percutaneous, single lumen catheters designed for use in the peripheral vascular system. TXM Support Catheters are intended to guide and support a guide wire during access of the vasculature, allow for wire exchanges and provide a conduit for the delivery of saline solutions or diagnostic contrast agents.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

5.1 - Company Information & Contact Person

Company Name:Texas Medical Technologies Inc.
Company Address:9005 Montana Ave. Ste. A
El Paso, Texas 79925
Telephone:(915) 774-4321
Fax:(915) 774-4323
Contact Person:Cesar Rios, Quality Assurance & Regulatory Manager
Date Prepared:02/28/2017

5.2 – Device Name & Classification

Proprietary Name:TXM Support Catheter
Common Name:Catheter, Support
Classification Name:Catheter, Percutaneous
Regulation Number:21 CFR 870.1250
Product Code:DQY
Device Class:II

5.3 – Predicate Device

Legally Marketed Substantially Equivalent Predicate Device

Proprietary Name:Company Name:Common Name:Classification Name:Regulation Number:Product Code:Device Class:510(k) NumberReference DevicesTrailBlazer™ Support Catheterev3 Inc.Catheter, SupportCatheter, Percutaneous21 CFR 870.1250DQYIIK092299
Proprietary Name:TrailBlazer™ Angled Support Catheter
Company Name:ev3 Inc.
Common Name:Catheter, Support
Classification Name:Catheter, Percutaneous
Regulation Number:21 CFR 870.1250
Product Code:DQY
Device Class:II
510(k) NumberK162384

{4}------------------------------------------------

Proprietary Name:Spectranetics Quick-Cross Extreme Support Catheters
Company Name:Spectranetics Corp.
Common Name:Catheter, Support
Classification Name:Catheter, Percutaneous
Regulation Number:21 CFR 870.1330
Product Code:DQY
Device Class:II
510(k) NumberK082561

Note: Reference devices are only to support the substantial equivalence of the subject device in regards to the angled tip.

5.4 - Device Description

The TXM Support Catheter is an over the wire (OTW) single lumen catheter with atraumatic tapered tip. The distal tip catheter shaft has 6 radiopaque markers (including the tip) that works as an aid to estimate positioning within the vasculature. It consists of a lubricious inner liner made from Teflon, and stainless steel braid over the liner and an outer layer that consists of clear Nylon 12 polymer. One gold marker band is positioned on the proximal side of the tip. A polycarbonate hub is attached to the proximal end of the TXM Support Catheter. The TXM Support Catheter system is offered in 8 models, four with a straight tip and four with an angled tip, and the angled tip configuration has an angle of 30°. All models will have an outer diameter of the 4 French size (0.052") and an inner diameter of 0.039" tapered down to 0.0355" towards the distal tip so that is compatible with a 0.035" guidewires. The device is available in lengths of 65cm, 90cm. 135cm, and 150cm. The catheter is coated with a lubricious hydrophilic coating. The device is supplied sterile and is intended for single use only.

The following table lists the models and sizes available for TXM Support Catheter.

ProductCodeFrenchsizeShaftLength(cm)OuterDiameter(inches)Inner Diameter (inches)MarkerBandMaterialTipShape
SC-35065-S4650.052"0.039" - 0.0355"GoldStraight
SC-35090-S4900.052"0.039" - 0.0355"GoldStraight
SC-35135-S41350.052"0.039" - 0.0355"GoldStraight
SC-35150-S41500.052"0.039" - 0.0355"GoldStraight
SC-35065-A4650.052"0.039" - 0.0355"GoldAngled
SC-35090-A4900.052"0.039" - 0.0355"GoldAngled
SC-35135-A41350.052"0.039" - 0.0355"GoldAngled
SC-35150-A41500.052"0.039" - 0.0355"GoldAngled

Table 5.4 TXM Support Catheter Models and Sizes

{5}------------------------------------------------

5.5 - Indications for Use

The TXM Support Catheters are percutaneous, single lumen catheters designed for use in the peripheral vascular system. TXM Support Catheters are intended to guide and support a guide wire during access of the vasculature, allow for wire exchanges and provide a conduit for the delivery of saline solutions or diagnostic contrast agents.

The Indications for Use are identical, the devices are both intended for peripheral vascular use and the Indications for Use do not change the intended use of the TXM Support Catheter when compared to the predicate device.

5.6 - Summary of Technological Characteristics Comparison

Based on a comparison of the indications for use, fundamental design, technology and principles of operation, materials, performance, sterilization, and packaging, it is determined that the TXM Support Catheter is substantially equivalent to the predicate device. Table 5.6 below provides a comparison of the TXM Support Catheter and the predicate.

Technical Characteristics /Principle of OperationTXM Support CatheterTrailBlazer ™ SupportCatheterSubstantiallyEquivalent?
Length65cm, 90cm, 135 cm,150cm65cm, 90cm, 135 cm,150cmYes
Outer Diameter0.052"0.063"Yes
Inner Diameter0.039"- 0.0355"0.042"-0.036"Yes
ShapeStraight and AngledStraightYes
Inner Liner MaterialPolytetrafluoroethylene(PTFE)*UnknownYes
Braid Reinforcement MaterialStainless Steel*Not ApplicableYes
Radiopaque MarkersA Gold marked band andradiopaque markers ofNylon 12 loaded withTungsten. *Platinum/IridiumYes
Outer Shaft MaterialNylon 12*UnknownYes
Luer MaterialPolycarbonate*UnknownYes
Luer ConnectorFemale Luer Connector*Female Luer ConnectorYes
Anatomical Site UsePeripheral Vascular systemPeripheral Vascular systemYes
Delivery to SiteOver-the-wireOver-the-wireYes
Guidewire CompatibilityMaximum 0.035"Maximum 0.035"Yes
PackagingTyvek PouchTyvek PouchY es
SterilizationEtO GasEtO GasYes

Table 5.6 Comparison of the TXM Support Catheter and the Predicate Device

  • Denotes a patient-contacting material.

{6}------------------------------------------------

5.7 - Performance Testing Summary

The following bench tests were performed to evaluate the design elements and performance characteristics of the TXM Support Catheter and to demonstrate substantial equivalence to the predicate device. The TXM Support Catheter met the predetermined acceptance criteria. Testing was performed on non-aged devices (T=0) as well as on devices subject to 2 years of accelerated aging (T=2). Tests results show that the TXM Support Catheter is substantially equivalent to the predicate device.

5.7.1- Bench Testing Summary

Table 5.7.1 below provides a summary of the bench testing performed on the TXM Support Catheter.

TestNo.Test NameApplicable Standard orInternal Test MethodTest Results
1Dimensional And Physical AttributesInternal Test MethodT=0T=2Pass
2Catheter CompatibilityInternal Test MethodT=0T=2Pass
3Air LeakageISO 10555-1T=0T=2Pass
4TrackabilityInternal Test MethodT=0T=2Pass
5CrossabilityInternal Test MethodT=0T=2Pass
6Catheter StiffnessASTM D747-10T=0T=2Pass
7Gravity Flow RateISO 10555-1T=0T=2Pass
8Power Injection for Flow RateISO 10555-1T=0T=2Pass
9Lubricity of Hydrophilic CoatingInternal Test MethodT=0T=2Pass
10Durability of Hydrophilic CoatingInternal Test MethodT=0T=2Pass
11Liquid LeakageISO 10555-1T=0T=2Pass
12Static Burst PressureISO 10555-1T=0T=2Pass
13Tensile StrengthISO 10555-1T=0T=2Pass
14RadiopacityASTM-F640-12T=0T=2Pass
15TorqueabilityInternal Test MethodT=0T=2Pass
16Torque StrengthInternal Test MethodT=0T=2Pass

Table 5.7.1 Bench Testing TXM Support Catheter

{7}------------------------------------------------

17Kink ResistanceInternal Test MethodT=0T=2Pass
18Corrosion ResistanceISO 10555-1T=0T=2Pass
19Coating Integrity and ParticulatesInternal Test MethodT=0T=2Pass
20Packing Integrity (Seal Strength)ASTM F88 / F88M - 15T=0T=2Pass
21Packing Integrity (Packing DyePenetration)ASTM F1929-15/F1886-09T=0T=2Pass
22Shipping and TransitISTA 3AT=0T=2Pass
23Accelerated AgingASTM F1980-07Pass
24Female Luer VerificationISO 594Pass

5.7.2 - Biocompatibility Testing Summary

The TXM Support Catheter is classified as an Externally Communicating Device, Circulating Blood, Limited Contact (≤24 hours). Biocompatibility testing was performed in accordance with ISO 10993 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process" (2009). Table 5.7.2 below describes the testing performed to determine biocompatibility.

Biological EffectTestCompliance Standard
CytotoxicityL929 MEM ELUTIONISO10993-5
IrritationIntracutaneous InjectionISO10993-10
SensitizationKligman MaximizationMurine Local Lymph AssayISO10993-10
Systemic ToxicitySystemic InjectionISO10993-11
Material MediatedRabbit PyrogenUSP<151>
PyrogenicityLimulus Amebocyte LysateISO10993-11
HemocompatibilityHemolysis Complete (Direct and Indirect)ISO10993-4
Complement ActivationISO10993-4
Thromboresistance Evaluation In vivoISO10993-4

Table 5.7.2. Summary of Biocompatibility Testing for the TXM Support Catheter

5.8 - Sterilization Testing Summary

Table 5.8 Summary of Sterilization Testing

Validation SterilizationProcessComplianceStandardSterility Assurance Level(SAL)ValidationResult
Ethylene Oxide GasISO 11135$10^{-6}$Pass

{8}------------------------------------------------

5.9 – Conclusion

The TXM Support Catheter is substantially equivalent in intended use, fundamental design, technology and principles of operation, materials, performance, sterilization, and packaging to the predicate device. Differences between the devices do not raise any new issues of safety or effectiveness.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).