(62 days)
No
The 510(k) summary describes a mechanical spinal implant system and does not mention any software, algorithms, or AI/ML capabilities.
Yes
The device is described as providing "immobilization of spinal segments as an adjunct to fusion" and restoring "the integrity of the spinal column," which are therapeutic actions.
No
The device is an implantable system intended for spinal immobilization and fusion, not for diagnosing medical conditions. Its purpose is therapeutic/surgical.
No
The device description explicitly states it is a "multiple component system comprised of a variety of non-sterile, single use implantable components" made of titanium alloy, including plates, screws, hooks, connectors, rods, and locking caps. This clearly indicates a hardware-based medical device.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly describes a surgical implant system for stabilizing spinal segments as an adjunct to fusion or for temporary stabilization in patients with tumors. This is a therapeutic and structural function, not a diagnostic one.
- Device Description: The device is described as a system of implantable components (plates, screws, hooks, rods, etc.) made of titanium alloy. These are physical devices used in surgery, not reagents, instruments, or software used to examine specimens from the human body.
- Lack of Diagnostic Activity: There is no mention of the device being used to analyze biological samples (blood, tissue, urine, etc.) or to provide information about a patient's health status based on such analysis.
- Performance Studies: The performance studies described are mechanical tests (compression, tension, bending, etc.) to demonstrate the structural integrity and equivalence of the device, which is typical for surgical implants. There are no studies related to diagnostic accuracy or performance with biological samples.
IVD devices are specifically designed to be used in vitro (outside the body) to examine specimens from the human body to provide information for diagnostic, monitoring, or compatibility purposes. The SOLSTICE OCT System is an in vivo (inside the body) surgical implant.
N/A
Intended Use / Indications for Use
The SOLSTICE OCT System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7) and the thoracic spine (T1 to T3): Traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g. pseudoarthrosis); tumors involving the cervical/thoracic spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The SOLSTICE OCT System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
In order to achieve additional levels of fixation, The SOLSTICE OCT System may be connected to the titanium Life Spine NAUTILUS Spine System using the 3.5mm/5.5mm titanium parallel connectors.
Product codes (comma separated list FDA assigned to the subject device)
NKG, KWP
Device Description
The SOLSTICE OCT System is a temporary, titanium alloy (6AL-4V-ELI per ASTM F 136), multiple component system comprised of a variety of non-sterile, single use implantable components. The system consists of an assortment of occipital plates, occipital bone screws, polyaxial screws, hooks, connectors, rods, and locking caps.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
craniocervical junction, the cervical spine (C1 to C7) and the thoracic spine (T1 to T3)
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Static compression, dynamic compression, static dissociation and static tension testing according to ASTM F1717 along with static axial grip, static torsional grip, and static moment bending testing according to ASTM F1798 was presented to demonstrate the substantial equivalency of The SOLSTICE OCT System.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 888.3075 Posterior cervical screw system.
(a)
Identification. Posterior cervical screw systems are comprised of multiple, interconnecting components, made from a variety of materials that allow an implant system to be built from the occiput to the upper thoracic spine to fit the patient's anatomical and physiological requirements, as determined by preoperative cross-sectional imaging. Such a spinal assembly consists of a combination of bone anchors via screws (i.e., occipital screws, cervical lateral mass screws, cervical pedicle screws, C2 pars screws, C2 translaminar screws, C2 transarticular screws), longitudinal members (e.g., plates, rods, including dual diameter rods, plate/rod combinations), transverse or cross connectors, interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors), and closure mechanisms (e.g., set screws, nuts). Posterior cervical screw systems are rigidly fixed devices that do not contain dynamic features, including but not limited to: non-uniform longitudinal elements or features that allow more motion or flexibility compared to rigid systems.Posterior cervical screw systems are intended to provide immobilization and stabilization of spinal segments in patients as an adjunct to fusion for acute and chronic instabilities of the cervical spine and/or craniocervical junction and/or cervicothoracic junction such as: (1) Traumatic spinal fractures and/or traumatic dislocations; (2) deformities; (3) instabilities; (4) failed previous fusions (
e.g., pseudarthrosis); (5) tumors; (6) inflammatory disorders; (7) spinal degeneration, including neck and/or arm pain of discogenic origin as confirmed by imaging studies (radiographs, CT, MRI); (8) degeneration of the facets with instability; and (9) reconstruction following decompression to treat radiculopathy and/or myelopathy. These systems are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.(b)
Classification. Class II (special controls). The special controls for posterior cervical screw systems are:(1) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(2) Nonclinical performance testing must demonstrate the mechanical function and durability of the implant.
(3) Device components must be demonstrated to be biocompatible.
(4) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(5) Labeling must include the following:
(i) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(ii) Intended use and indications for use including levels of fixation;
(iii) Device specific warnings, precautions, and contraindications that include the following statements:
(A) “Precaution: Preoperative planning prior to implantation of posterior cervical screw systems should include review of cross-sectional imaging studies (
e.g., CT and/or MRI) to evaluate the patient's cervical anatomy including the transverse foramen, neurologic structures, and the course of the vertebral arteries. If any findings would compromise the placement of these screws, other surgical methods should be considered. In addition, use of intraoperative imaging should be considered to guide and/or verify device placement, as necessary.”(B) “Precaution: Use of posterior cervical pedicle screw fixation at the C3 through C6 spinal levels requires careful consideration and planning beyond that required for lateral mass screws placed at these spinal levels, given the proximity of the vertebral arteries and neurologic structures in relation to the cervical pedicles at these levels.”
(iv) Identification of magnetic resonance (MR) compatibility status;
(v) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user, and;
(vi) Detailed instructions of each surgical step, including device removal.
0
Image /page/0/Picture/2 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features the department's name encircling a stylized caduceus symbol. The caduceus is composed of three human profiles facing right, with a triple helix design representing health and medicine. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 18, 2017
Life Spine, Inc. Mr. Randy Lewis General Manger 13951 South Ouality Drive Huntley, Illinois 60142
Re: K170804
Trade/Device Name: The Solstice OCT System Regulatory Class: Unclassified Product Code: NKG, KWP Dated: March 17, 2017 Received: March 17, 2017
Dear Mr. Lewis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
1
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Lori A. Wiggins -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
2
Indications for Use
510(k) Number (if known) K170804
Device Name The Solstice OCT System
Indications for Use (Describe)
The SOLSTICE OCT System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical junction, the cervical spine (C1 to C7) and the thoracic spine (T1 to T3): Traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g. pseudoarthrosis); tumors involving the cervical/thoracic spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The SOLSTICE OCT System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
In order to achieve additional levels of fixation, The SOLSTICE OCT System may be connected to the titanium Life Spine NAUTILUS Spine System using the 3.5mm/5.5mm titanium parallel connectors.
Type of Use (Select one or both, as applicable) |
---|
Prescription Use (Part 21 CFR 801 Subpart D) |
Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
3
510(k) Summary The SOLSTICE OCT System
| Submitted By: | Life Spine, Inc.
13951 S. Quality Drive
Huntley, IL 60142
Telephone: 847-884-6117
Fax: 847-884-6118 |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Contact: | Randy Lewis
Life Spine, Inc.
13951 S. Quality Drive
Huntley, IL 60142
Telephone: 847-884-6117
Fax: 847-884-6118 |
| Date Prepared: | May 11th, 2017 |
| Trade Name: | The Solstice OCT System |
| Common Name: | OCT System |
| Classification: | NKG, Unclassified
KWP, 21 CFR 888.3050, Class II |
| Primary Predicate : | Synapse Occipital-Cervical-Thoracic (OCT) System (K142838) |
| Reference Device: | Solstice OCT System (K142253)
Solstice OCT System (K143249) |
Device Description:
The SOLSTICE OCT System is a temporary, titanium alloy (6AL-4V-ELI per ASTM F 136), multiple component system comprised of a variety of non-sterile, single use implantable components. The system consists of an assortment of occipital plates, occipital bone screws, polyaxial screws, hooks, connectors, rods, and locking caps.
Intended Use of the Device:
The SOLSTICE OCT System is intended to provide immobilization and stabilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the
4
craniocervical junction, the cervical spine (C1 to C7) and the thoracic spine (T1 to T3): Traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g. pseudoarthrosis); tumors involving the cervical/thoracic spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The SOLSTICE Spinal System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
In order to achieve additional levels of fixation, the SOLSTICE Spinal System may be connected to the titanium Life Spine NAUTILUS Spine System using the 3.5mm/5.5mm titanium parallel connectors.
Technological Characteristics:
The SOLSTICE OCT System is substantially equivalent to the predicate systems in terms of design, materials, indications for use and sizing.
Performance Data:
Static compression, dynamic compression, static dissociation and static tension testing according to ASTM F1717 along with static axial grip, static torsional grip, and static moment bending testing according to ASTM F1798 was presented to demonstrate the substantial equivalency of The SOLSTICE OCT System.
Substantial Equivalence:
The SOLSTICE OCT System was shown to be substantially equivalent to the predicate devices in indications for use, design, function, materials used and mechanical performance.