K Number
K170781

Validate with FDA (Live)

Device Name
MX50N(MX50YQS)
Manufacturer
Date Cleared
2017-04-04

(20 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

MX50N(MX50YQS) LCD Monitor System is intended to be used in displaying and viewing digital medical images for review and analysis by trained medical practitioners. It is specifically designed for digital mammography applications and digital breast tomosynthesis applications.

Device Description

MX50N(MX50YQS) is a flat panel hi-resolution LCD monitor system for displaying digital medical images. The system consists of a state-of-the-art LCD monitor and a high-resolution graphic control board that connects to a PACS workstation for grayscale image display. The WIDE controller board is installed into the PACS workstation computer or other computer system to display PACS medical images.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the MX50N(MX50YQS) LCD Monitor System, which is intended for displaying and viewing digital medical images, specifically for digital mammography and digital breast tomosynthesis applications. The submission aims to demonstrate substantial equivalence to a predicate device (K160326).

Here's an analysis of the acceptance criteria and study information provided:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implied by the comparison to the predicate device and the compliance with international standards. The performance metrics are a direct comparison between the subject device and the predicate device.

ParameterAcceptance Criteria (Predicate Device Performance)Reported Device Performance (MX50N(MX50YQS))
Intended UseSame as predicateSame as predicate
Technological CharacteristicsSame as predicateSame as predicate
Response Time (Typical)25ms (On/Off)25ms (On/Off)
LCD Panel Size21.3"21.3"
Resolution2560 x 20482560 x 2048
Pixel Pitch0.165mm x 0.165mm0.165mm x 0.165mm
Brightness1200cd/m²1200cd/m²
Contrast Ratio1200 : 11000 : 1
Input SignalDVI-IDVI-I, DisplayPort
Power Supply100~240 VAC, 50/60Hz100~240 VAC, 50/60Hz
Color/MonochromeMonochromeMonochrome

Note regarding "Acceptance Criteria": For the parameters where the subject device’s performance is directly aligned with the predicate (e.g., Response Time, LCD Panel Size), the predicate's performance acts as the de facto acceptance criterion for demonstrating substantial equivalence. For parameters with slight differences (like Contrast Ratio), the acceptance is based on the argument that the difference is not critical.

2. Sample Size Used for the Test Set and the Data Provenance

The document does not describe a test set in the traditional sense of a dataset for evaluating an AI algorithm. This device is a display monitor, not an AI diagnostic tool that processes clinical data.

Therefore, this information is Not Applicable in the context of this device and submission. There is no mention of clinical data or patient samples being used for testing. The testing performed is non-clinical, related to the display's technical specifications and safety.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

As there is no "test set" of clinical data to establish ground truth for this display monitor, this information is Not Applicable.

4. Adjudication Method for the Test Set

As there is no "test set" of clinical data, this information is Not Applicable.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC study was conducted. This device is a display monitor, not an AI algorithm intended to assist human readers directly or to be evaluated for its impact on reader performance.

Therefore, this information is Not Applicable.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This device is not an algorithm, but a hardware display monitor. Thus, a standalone algorithm performance study is Not Applicable.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

Ground truth in the clinical sense (expert consensus, pathology) is Not Applicable, as this is a display monitor and not an AI diagnostic algorithm. The "ground truth" for the device's technical performance is established by the specifications and measurements according to recognized engineering and medical device standards.

8. The Sample Size for the Training Set

No training set is mentioned or applicable. This device is a display monitor, not a machine learning model.

Therefore, this information is Not Applicable.

9. How the Ground Truth for the Training Set was Established

As no training set is applicable, this information is Not Applicable.


Summary of the Study that Proves the Device Meets Acceptance Criteria:

The study that proves the device meets the acceptance criteria is a non-clinical performance evaluation and compliance testing against recognized international standards and a comparison to a legally marketed predicate device.

  • Type of Study: Non-clinical (bench testing, technical specification validation, standard compliance). No clinical studies were considered necessary or performed.
  • Methodology:
    • Substantial Equivalence Comparison: The MX50N(MX50YQS) was directly compared, parameter by parameter, with the predicate device (5MP Monochrome LCD Monitor MS55i2, K160326). This comparison covered intended use, technological characteristics, and various technical specifications (response time, LCD panel size, resolution, pixel pitch, brightness, contrast ratio, input signal, power supply, color/monochrome). The conclusion was that the devices were substantially equivalent, with any differences (e.g., contrast ratio) not adversely affecting the device's use.
    • Compliance with Consensus Standards: The device underwent testing to comply with the following international and FDA-recognized consensus standards:
      • IEC 60601-1: Medical Electrical Equipment -- Part 1: General Requirements For Basic Safety And Essential Performance.
      • IEC 60601-1-2: Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests.
  • Performance Metrics Evaluated: Tests were conducted for resolution, luminance, contrast, and noise.
  • Conclusion: The results of these tests demonstrated that the MX50N(MX50YQS) met the acceptance criteria specified in the standards and was adequate for its intended use. The device was found to be as safe and effective as the predicate device.

In essence, the "study" is a comprehensive engineering and electrical safety validation against established industry benchmarks and a direct side-by-side technical comparison with a cleared device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 4, 2017

WIDE Corporation % YeoJin Yun RA Manager 12 Wongomae-Ro, Giheung-Gu Yongin-Si, Gyeonggi-Do 17086 REPUBLIC OF KOREA

Re: K170781

Trade/Device Name: MX50N(MX50YOS) Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: PGY Dated: March 10, 2017 Received: March 17, 2017

Dear YeoJin Yun:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D'Hara

For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K170781

Device Name MX50N(MX50YQS)

Indications for Use (Describe)

MX50N(MX50YQS) LCD Monitor System is intended to be used in displaying and viewing digital medical images for review and analysis by trained medical practitioners. It is specifically designed for digital mammography applications and digital breast tomosynthesis applications.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

[As required by 21 CFR 807.92]

This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR 807.92

1. Date Prepared [21 CFR 807.92(a) (1)]

03/10//2017

2. Submitter's Information [21 CFR 807.92(a) (1)]

Name of Sponsor:WIDE Corporation.
Address:12 Wongomae-Ro, Giheung-Gu, Yongin-Si, Gyeonggi-Do17086, Republic of Korea
Contact Name:YeoJin Yun
Telephone #: +82-31-218-1675
Fax #: +82-31-218-7400
Email: yyjin@widecorp.com
Registration Number:3004082357
Name of Manufacturer:Same as Sponsor

3. Trade Name, Common Name, Classification [21 CFR 807.92(a) (2)]

Model Name:MX50N(MX50YQS)
Common Name:TFT LCD Medical Monitor System
Classification Name:Display, Diagnostic Radiology
Regulation Number:21 CFR 892.2050
Product Code:PGY
Device Class:2
Review Panel:Radiology

4. Identification of Predicate Device(s) [21 CFR 807.92(a) (3)]

510(k) Number: K160326
Applicant: JVC KENWOOD Corporation
Model Name: 5MP Monochrome LCD Monitor MS55i2(ML21055, MD211G5)
Common Name: TFT LCD Medical Monitor System
Classification Name: Display, Diagnostic Radiology
Regulation Number: 21 CFR 892.2050
Product Code: PGY
Device Class: 2

{4}------------------------------------------------

5. Description of the Device [21 CFR 807.92(a) (4)]

MX50N(MX50YQS) is a flat panel hi-resolution LCD monitor system for displaying digital medical images. The system consists of a state-of-the-art LCD monitor and a high-resolution graphic control board that connects to a PACS workstation for grayscale image display. The WIDE controller board is installed into the PACS workstation computer or other computer system to display PACS medical images.

Subject of this new 510(k) premarket notification is an addition to the indications for use for the MX50N(MX50YQS). The device was cleared via premarket notification K160353. There is no any change of configuration/components and functionality of the subject device as compared with cleared device (K160353).

6. Intended Use [21 CFR 807.92(a) (5)]

The MX50N(MX50YQS) LCD Monitor System is intended to be used in displaying and viewing digital medical images for review and analysis by trained medical practitioners. It is specifically designed for digital mammography applications and digital breast tomosynthesis applications.

7. Technological Characteristics [21 CFR 807.92(a) (6)]

The device is an image display system which consists of computer software and components. The device does not contact the patient, nor does it control any life sustaining devices. A physician or trained medical practitioner provides ample opportunity for competent human intervention to interpret images and information being displayed.

ParameterSubject DevicePredicate Device
510(k) NumberUnknownK160326
Model NameMX50N(MX50YQS)5MP Monochrome LCDMonitor MS55i2(ML21055, MD211G5)
ManufacturerWIDE Corporation.JVC KENWOOD Corporation
Common NameTFT LCD Medical Monitor System
Classification NameDisplay, Diagnostic Radiology
Classification PanelRadiology
Classification Regulation21 CFR 892.2050
Product CodePGY
Device ClassClass II
Intended UseThe MX50N(MX50YQS) LCDMonitor System is intended to beused in displaying and viewing5MP Monochrome LCDMonitor MS55i2 (ML21055,MD211G5) are intended to be

8. Substantial Equivalence [21 CFR 807.92(b) (1) and 807.92]

{5}------------------------------------------------

ParameterSubject DevicePredicate Device
digital medical images for reviewand analysis by trained medicalpractitioners. It is specificallydesigned for digital mammographyapplications and digital breasttomosynthesis applications.used in displaying and viewingmedical images for diagnosis bytrained medical practitioners.They are to be used in digitalmammography PACSmodalities including FFDM, andbreast tomosynthensis.
Response Time(Typical)25ms (On/Off)25ms (On/Off)
LCD Panel Size21.3"21.3"
Resolution2560 x 20482560 x 2048
Pixel pitch0.165mm x 0.165mm0.165mm x 0.165mm
Brightness1200cd/m²1200cd/m²
Contrast Ratio1000 : 11200 : 1
Input SignalDVI-I, DisplayPortDVI-I
Power Supply100~240 VAC, 50/60Hz100~240 VAC, 50/60Hz
Color/MonochromeMonochromeMonochrome

When compared to the predicate devices (K160326), the MX50N(MX50YQS) presented in this submission has the same of the followings:

  • Intended Use
  • Technological characteristics
  • Response Time
  • LCD Panel Size
  • Resolution
  • Pixel pitch
  • · Brightness

The two devices share the similar performance as the following:

  • Contrast Ratio
    There is no significant difference between the MX50N(MX50YQS) and the predicate device that would adversely affect the use of the product. The subject device is substantially equivalent to the predicate device in intended use, technological characteristics, panel size, resolution, pixel pitch, Brightness and contrast ratio.

{6}------------------------------------------------

9. Summary of Non-Clinical Data

MX50N(MX50YQS) comply with the following international and FDA-recognized consensus standards:

IEC 60601-1:Medical Electrical Equipment -- Part 1: GeneralRequirements For Basic Safety And Essential Performance
IEC 60601-1-2:Medical Electrical Equipment - Part 1-2: GeneralRequirements For Basic Safety And Essential Performance -Collateral Standard: Electromagnetic Compatibility -Requirements And Tests

The tests conducted for resolution, luminance, contrast and noise all met the acceptance criteria specified in the standards.

10. Summary of Clinical Data

No clinical studies were considered necessary and performed.

11. Conclusion [21 CFR 807.92(b) (3)]

Subject Device is substantially equivalent to the currently marketed and predicate devices in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness. Even though the predicate device and the subject device differ in the brightness, the difference is not a critical in the effectiveness because subject device covers brightness range of predicate device.

Additionally, the safety of the subject device was validated through tests including IEC60601-1 and IEC 60601-1-2. The effectiveness of the device was validated through tests for resolution, luminance, contrast and noise.

The results of these tests demonstrate that MX50N(MX50YQS) meets the acceptance criteria and is adequate for this intended use. The comparison of technological characteristics, nonclinical performance data, safety testing demonstrates that the device is as safe and effective as the predicate device and performs as well as the predicate device.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).