(249 days)
The Universal Mesh is indicated for reconstruction, stabilization subsequent to craniotomy, craniectomy, cranioplasty, and cranial fractures in adults and adolescents (age 12 and higher).
The 3D Preformed meshes are added to the Universal Mesh system in order to offer a more comprehensive selection of mesh solutions for cranial reconstruction. Simply put, the 3D Preformed meshes are the 1.5/1.7mm Dynamic Mesh predicate design provided in a preformed shape. The 3D Preformed meshes are provided non-sterile in 2 different variations; these meshes are offered in a Symmetrical Preformed mesh and an Anatomically preformed lateral mesh. The Anatomically preformed lateral 3D meshes add a 1.0mm thick mesh option to the portfolio, which was not available with the predicate device.
The provided text describes the Stryker Universal Mesh, a device indicated for reconstruction, stabilization, and/or rigid fixation subsequent to craniotomy, craniectomy, cranioplasty, and cranial fractures in adults and adolescents (age 12 and higher). The information needed to answer the request is scattered throughout sections V, VI, and VII of the document.
Here's an analysis of the acceptance criteria and the studies that prove the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
| Test Category | Test / Criterion | Reported Device Performance |
|---|---|---|
| Biocompatibility | Cytotoxicity determined using DIN EN 10993-5:2009. | Tests supported the biocompatibility of the device. |
| Performance Bench Testing | Sterilization Validation: Validate sterilization instructions according to internal standard and referenced standards (ISO 14937, ISO 17664, ISO 17665-1&2, AAMI TIR 12, AAMI ST 81, AAMI ST 79). | Method validated. |
| Static Load Max. Defect: Determination of defined load and deformation over maximum defect according to labeling of fixated meshes with bone simulation material. | Acceptance criterion were met, tests were passed. | |
| Static Load (70mm diameter defect): Determination of defined load and deformation over 70mm diameter defect according to labeling of fixated meshes with bone simulation material. Comparison to predicate device tested over same defect. | Acceptance criterion were met, tests were passed. | |
| End User Validation Testing: Surgeon evaluation. | Acceptance criterion was met, test was passed. | |
| Internal Handling Test: Non-trained user evaluation. | Acceptance criterion was met, test was passed. | |
| MR Compatibility: Magnetic Resonance (MR) testing for radiofrequency-induced heating and establishing safety and compatibility in the MR environment, according to FDA guidance (March 22, 2016, and December 11, 2014, respectively). | Stryker Universal mesh will be labeled as MR conditional. | |
| Overall Performance | The Subject device met all pre-defined acceptance criteria and standards and was found to not represent a new worst case. | All acceptance criteria and standards were met. |
2. Sample size used for the test set and the data provenance:
The document does not specify sample sizes for any of the performance bench tests. It also does not provide information on data provenance (e.g., country of origin, retrospective or prospective) for these tests, as these are typically laboratory bench tests rather than clinical studies with patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
For the "End User Validation Testing" (surgeon evaluation), the document simply states "Surgeon evaluation" but does not specify the number of surgeons or their qualifications.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
The document does not provide information on adjudication methods for any of the tests.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No MRMC comparative effectiveness study was done. This device is a physical implant (mesh), not an AI-assisted diagnostic tool.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
This question is not applicable as the device is a physical implant, not an algorithm or AI.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
For the physical performance tests (static load, sterilization, MR compatibility), the ground truth is established by meeting defined engineering standards and regulations (e.g., ASTM F67, ISO standards, FDA guidance). For the user validation tests, the 'ground truth' is subjective user (surgeon/non-trained user) evaluation against implicit criteria of usability and fit.
8. The sample size for the training set:
This question is not applicable as the device is a physical implant and not an AI/algorithm-based product that requires a training set.
9. How the ground truth for the training set was established:
This question is not applicable for the same reason as above.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 13, 2017
Stryker Jonathan Schell Staff Regulatory Affairs Specialist 750 Trade Centre Way-Suite 200 Portage, Michigan 49002
Re: K170725
Trade/Device Name: Stryker Universal Mesh Regulation Number: 21 CFR 882.5320 Regulation Name: Preformed Alterable Cranioplasty Plate Regulatory Class: Class II Product Code: GWO Dated: October 10, 2017 Received: October 11, 2017
Dear Mr. Schell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael J. Hoffmann -S
Carlos L. Peña. PhD. MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K170725
Device Name Stryker Universal Mesh
Indications for Use (Describe)
The Universal Mesh is indicated for reconstruction, stabilization subsequent to craniotomy, craniectomy, cranioplasty, and cranial fractures in adults and adolescents (age 12 and higher).
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Stryker logo in black. The logo is a stylized wordmark with the word "stryker" in a bold, sans-serif font. A small circle with the letter 'c' inside is located to the upper right of the wordmark.
510(k) Summary
This section provides a summary of 510(k) information in accordance with the requirements of 21 CFR 807.92.
- SUBMITTER [§807.92(a)(1)] I.
| 510(k) Owner: | Stryker Leibinger GmbH & Co. KGBoetzinger Strasse 4179111 Freiburg, Germany |
|---|---|
| Submitter/ ContactPerson: | Jonathan SchellStaff Regulatory Affairs SpecialistStryker Craniomaxillofacial750 Trade Centre WayPortage, MI 49002Phone: 269-389-5596Fax: 877-648-7114 |
Date prepared: November 6, 2017
- DEVICE [§807.92(a)(2)] II.
Trade Name: Stryker Universal Mesh
Bone Plate Common or Usual name:
Classification Preformed Alterable Cranioplasty Plate name:
Regulatory Class: Class II
Product Code: GWO
III. PREDICATE DEVICE [§807.92(a)(3)]
Universal Mesh System - K161821
IV. DEVICE DESCRIPTION [§807.92(a)(4)]
The 3D Preformed meshes are added to the Universal Mesh system in order to offer a more comprehensive selection of mesh solutions for cranial reconstruction. Simply put, the 3D Preformed meshes are the 1.5/1.7mm Dynamic Mesh predicate design provided in a
{4}------------------------------------------------
preformed shape. The 3D Preformed meshes are provided non-sterile in 2 different variations; these meshes are offered in a Symmetrical Preformed mesh and an Anatomically preformed lateral mesh. The Anatomically preformed lateral 3D meshes add a 1.0mm thick mesh option to the portfolio, which was not available with the predicate device.
INDICATIONS FOR USE [§807.92(a)(5 )] V.
| Subject Device | Predicate – K161821 | |
|---|---|---|
| Indications forUse | The Universal Mesh is indicatedfor reconstruction, stabilizationand/or rigid fixation subsequentto craniotomy, craniectomy,cranioplasty, and cranial fracturesin adults and adolescents (age 12and higher). | The Universal Mesh is indicatedfor reconstruction, stabilizationand/or rigid fixation subsequent tocraniotomy, craniectomy,cranioplasty, and cranial fracturesin adults and adolescents (age 12and higher). |
TABLE 1: COMPARISON OF INDICATIONS FOR USE
The predicate device is indicated for use in the cranial skeleton. The Indications for Use statement of the subject Universal 3D Preformed Mesh is unchanged and remains identical to the predicate device.
COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE [§807.92(a)(6)]
The Stryker Universal Mesh is compared to its predicate devices for substantial equivalence based on the following criteria:
- A. Principle of Operation
- B. Technological Characteristics
A. Principle of Operation
The Stryker Universal 3D Preformed Mesh implants are designed for the reconstruction, stabilization and rigid fixation of the bony areas of the cranial skeleton. The necessary instrumentation are included with the system.
The Principle of Operation of the subject Universal 3D Preformed Mesh is unchanged and remains identical to the predicate device.
B. Technological and Operational Characteristics
At a high level, the Subject devices and Predicate devices are based on the following technological elements:
{5}------------------------------------------------
- Same materials: Both the subject device and predicate device are made of titanium according to ASTM F67
- -Same mode of fixation: implant fixation with screws
- -Same mesh pattern: The Subject device has the same Dynamic Mesh pattern as the predicate device, however is provided in 1.0mm thickness.
- -Similar Design/Sizes: The Subject device 3D Preformed mesh is designed to offer the surgeon a pre-bent version of an existing mesh, to minimize the time the surgeon needs for bending the implant into the desired shape. The 3D Preformed mesh are designed to be modified by the surgeon as needed to fit the patient needs, the same as the predicate mesh.
VII. PERFORMANCE DATA [§807.92(b)(1)]
The following performance data were provided in support of the substantial equivalence determination.
Biocompatibility Testing
The biocompatibility evaluation for the subject device was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"" May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The tests supported the biocompatibility of the device.
The implant is made from titanium which conforms to ASTM F67 for chemical composition, which is the same as the predicate. Cytotoxicity testing was performed using DIN EN 10993-5:2009, Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity. The tests supported the biocompatibility of the device.
Performance Bench Testing
| Test | Test Method Summary | Results |
|---|---|---|
| SterilizationValidation | Validate the sterilization instructions for the Subject device accordingto internal standard. Referenced standards: ISO 14937, ISO 17664,ISO 17665 - 1&2, AAMI TIR 12, AAMI ST 81, AAMI ST 79. | Methodvalidated. |
| Static Load Max.Defect | Determination of defined load and deformation over maximum defectaccording to labeling of fixated meshes with bone simulation material | Acceptancecriterionwere met, testswere passed. |
The following performance bench tests were completed.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "stryker" in a bold, sans-serif font. The word is all lowercase, and the letters are closely spaced together. A small registered trademark symbol is located to the right of the letter "r".
| Static Load | Determination of defined load and deformation over 70mm diameter defect according to labeling of fixated meshes with bone simulation material. Comparison to predicate device tested over same defect. | Acceptance criterion were met, tests were passed. |
|---|---|---|
| End User Validation Testing | Surgeon evaluation | Acceptance criterion was met, test was passed |
| Internal Handling Test | Non-trained user evaluation | Acceptance criterion was met, test was passed |
| MR Compatibility | Magnetic Resonance (MR) testing was performed on the subject device mesh implants in accordance with FDA guidance for industry "Assessment of Radiofrequency - Induced Heating in the Magnetic Resonance (MR) Environment for Multi-Configuration Passive Medical Devices,"March 22, 2016. MR heat testing has been performed on the Subject device according to FDA final guidance "Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment," December 11, 2014. | Based on the results of these evaluations, the Stryker Universal mesh will be labeled as MR conditional. |
The Subject device met all pre-defined acceptance criteria and standards and was found to not represent a new worst case. Overall, the results of the performance bench tests support the substantial equivalence of the Subject device to the predicate.
Animal Testing
Animal testing was not required as a basis for substantial equivalence.
Clinical Testing [§807.92(b)(2)]
Clinical testing was not required as a basis for substantial equivalence.
VIII. CONCLUSIONS [§807.92(b)(3)]
The results of the non-clinical data demonstrate the Stryker Universal 3D Preformed Mesh will perform as intended in the specified use conditions. According to the comparison based on the requirements of 21 CFR 807.92 and the information provided herein, it is concluded that the information included in this submission supports substantial equivalence.
§ 882.5320 Preformed alterable cranioplasty plate.
(a)
Identification. A preformed alterable cranioplasty plate is a device that is implanted into a patient to repair a skull defect. It is constructed of a material, e.g., tantalum, that can be altered or reshaped at the time of surgery without changing the chemical behavior of the material.(b)
Classification. Class II (performance standards).