K Number
K170663
Date Cleared
2018-05-11

(431 days)

Product Code
Regulation Number
868.5915
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Single patient use manual resuscitator for use hospital, transport, emergency, and post hospital care to temporary ventilate a patient for the given body mass ranges of: Infant: less than or equal to 10Kg. Child: less than or equal to 23 Kg. Adult: greater than 23 Kg

This manual resuscitator may be supplied with a single patient use positive end expiratory pressure (PEEP) valve and / or disposable Airway Pressure Manometer.

The PEEP Valve is a single patient use positive end expiratory pressure (PEEP) valve for use hospital. transport, emergency, and post hospital care to evaluate end lung pressure above atmospheric at the end of exhalation in constant and intermittent gas flow conditions. Intended for patients that the clinician has determined need PEEP.

The Disposable Manometer is a single patient use manometer intended to be used for monitoring the patient's airway pressure during ventilation. The manometer is to be used with resuscitation systems.

Device Description

Foremount Disposable PVC Resuscitators are portable medical devices used to temporarily augment ventilation in patients during ventilatory insufficiency or ventilatory failure. They consist of Ventilation Bag, Patient Valve, Intake Valves, Reservoir Bag, Oxygen Tubing, Cushion Mask, and optional Diverter Ring, PEEP Valve and Manometer. Foremount Disposable PVC Resuscitators come in three sizes along with a ventilation bag:

  • . Infant - Less than or equal to 10 kg
  • Child less than or equal to 23 kg ●
  • Adult Greater than 23 kg. ●

The ventilation bags are available in three sizes based upon the intended patient population. They are provided with masks in three sizes (#1 - Infant, #3 - Child, and #5 - Adult). The patient valve includes a duck-bill valve to prevent rebreathing and incorporates a 40 cmH2O pop off valve for Child and Infant models. The patient valve includes a duck-bill valve to prevent rebreathing and incorporates a 60 cmH2O pop off valve for Adult models.

AI/ML Overview

This is a 510(k) premarket notification for the Foremount Disposable PVC Resuscitator Model A1, A2, B1, B2 and Accessories, including a PEEP valve and a disposable manometer. The submission aims to demonstrate substantial equivalence to previously cleared devices.

Here's an analysis of the acceptance criteria and study information provided:

1. Table of Acceptance Criteria and Reported Device Performance

The document doesn't explicitly define "acceptance criteria" in a dedicated table for each device component with specific numerical targets. Instead, it demonstrates performance by comparing the Foremount devices to their respective predicates or by stating that the device meets relevant ISO standards.

We can infer the acceptance criteria are met if the Foremount device's specifications and performance fall within acceptable limits defined by the predicate devices and applicable ISO standards.

Inferred Acceptance Criteria and Reported Device Performance (Resuscitator)

CharacteristicAcceptance Criteria (Inferred from Predicate/Standard)Reported Device Performance (Foremount)
General
Indications for UseSingle patient use manual resuscitator for hospital, transport, emergency, and post-hospital care to temporarily ventilate in given body mass ranges.Same as predicate: Infant ≤ 10Kg, Child ≤ 23 Kg, Adult > 23 Kg (for Models A1, B1); Adult > 23 Kg (for Models A2, B2)
Patient PopulationInfant ≤ 10Kg, Child ≤ 23 Kg, Adult > 23 KgSame as predicate
Environment of UseHospital, transport, emergency, post-hospital careSame as predicate
Duration of UseSingle patient, disposable < 24 hoursSingle patient, disposable < 24 hours
Operational Temperature-18°C ~ 50°C-18°C ~ 50°C
Storage Temperature-40°C ~ 60°C-40°C ~ 60°C
Functional Performance (ISO 10651-4 and others)
Dimensions (Size)Adult: 570x190 mm, Child: 510x190 mm, Infant: 430x180 mm (Predicate)Adult: 445x190 mm, Child: 350x177 mm, Infant: 325x165 mm. Note: Differences in dimensions were assessed and did not raise new safety/effectiveness concerns.
Ventilation Bag VolumeAdult 1500 ml, Child 600ml, Infant 280 ml (Predicate)Adult 1700 ml, Child 500 ml, Infant 320 ml. Note: Differences in volume were assessed and did not raise new safety/effectiveness concerns, as ISO 10651-4 specifies minimum requirements.
Oxygen Collection Bag VolumeAdult 1000 ml, Child 1000 ml, Infant 500 ml (Predicate)Adult 1000 ml, Child 1000 ml, Infant 600 ml
Max Stroke Volume (single hand)Adult 700 ml, Child 360 ml, Infant 150 ml (Predicate)Adult 650 ml, Child 370 ml, Infant 180 ml. Note: Differences in volume were assessed and did not raise new safety/effectiveness concerns as ISO 10651-4 specifies minimum requirements.
Dead Space4.5 ml (measured) / 6.8 ml (labeled) for all sizes (Predicate)~ 3.8 ml for all sizes
Inspiratory ResistanceInfant: ≤ 0.5cm H2O, Child: ≤ 0.7cm H2O, Adult: ≤ 3.3cm H2O (Predicate)Infant: 0.5cm H2O, Child: 0.5cm H2O, Adult: 3cm H2O
Expiratory ResistanceInfant: ≤ 0.8cm H2O, Child: ≤ 0.8cm H2O, Adult: ≤ 2.6cm H2O (Predicate)Infant: 0.5cm H2O, Child: 0.5cm H2O, Adult: 2.8 cmH2O
Supplemental Oxygen %Comparable to predicate at various flow rates and tidal volumes (e.g., Infant Vt-70ml x 20bpm: 91/97/99% at 2/5/10 lpm)Infant Vt-70ml x 20bpm: 90/98/98% at 2/5/10 lpm (and other comparable values).
Pop-Off or Pressure Limiting40 cm H2O (Infant/Child), 60 cm H2O (Adult)40 cm H2O (Infant/Child), 60 cm H2O (Adult)
Ability to add PEEP valvesYes, 0-20 cm H2OYes, 0-20 cm H2O
Patient connectors15 / 22 mm (standard)15 / 22 mm
PEEP valve fittings22 / 30 mm (standard)22 / 30 mm
BiocompatibilityNon-cytotoxic, non-sensitizers, non-irritants; meets ISO 10993-5, -10, -11; ISO 18562-2 for PM2.5 and inorganic gases.Found to be non-cytotoxic, non-sensitizers, non-irritants. Tests for Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Gas emission (VOC), PM2.5, Inorganic gases (Ozone, CO, CO2) were performed.
MaterialsPVC, Polycarbonate, Silicone (Predicate)PVC, Polycarbonate, Silicone
Conical connectorsMeets ISO 5356-1Meets ISO 5356-1
Shelf-life3 years (Predicate)5 years. Note: This is a difference, but the document concludes no new safety or effectiveness concerns are raised.

Inferred Acceptance Criteria and Reported Device Performance (PEEP Valve)

CharacteristicAcceptance Criteria (Inferred from Predicate/Standard)Reported Device Performance (Foremount)
Indications for UseSingle patient use PEEP valve for hospital, transport, emergency, and post hospital care to evaluate end lung pressure above atmospheric.Same as predicate
Pressure range2.5 to 10 cmH2O; 5 to 20 cmH2O (Predicate)0 to 10 cmH2O; 5 to 20 cmH2O. Note: Broader lower range, but within similar operational function.
AdjustableYes (Predicate also had fixed options)Yes
Accuracy+/- 2 cmH2O+/- 2 cmH2O
Connectors22 mm / 30 mm (standard)22 mm / 30 mm
BiocompatibilityMeets biocompatibility standards for externally communicating, limited duration use.Externally communicating, limited duration of use (<24 hours). Materials are Polycarbonate, Silicone, Stainless Steel spring; passed biocompatibility testing.
MaterialsPolycarbonate, Silicone, Stainless Steel spring (Predicate)Polycarbonate, Silicone, Stainless Steel spring
Performance TestingAccuracy, Repeatability, Effects of Aging, Drop Test, Conical fittingsAll listed tests were performed. Results implicitly met 'acceptance criteria' by demonstrating substantial equivalence, though specific numerical results beyond accuracy were not detailed in the comparison table.

Inferred Acceptance Criteria and Reported Device Performance (Disposable Manometer)

CharacteristicAcceptance Criteria (Inferred from Predicate/Standard)Reported Device Performance (Foremount)
Indications for UseSingle patient use manometer for monitoring patient's airway pressure during ventilation, used with resuscitation systems.Same as predicate
Pressure range0 to 60 cmH2O0 to 60 cmH2O
MarkingsIncrements of 10 cmH2OIncrements of 10 cmH2O
Accuracy±2 cmH2O @ 0-10 cmH2O, ±4 cmH2O @ 20 cmH2O, ±5 cmH2O @ 30-40 cmH2O, ±7 cmH2O @ 50-60 cmH2O (Predicate)±1 cmH2O @ 0-10 cmH2O, ±2 cmH2O @ 20-30 cmH2O, ±3 cmH2O @ 40-60 cmH2O. Note: The Foremount device reports better accuracy than the predicate. This difference is considered acceptable as it is a performance improvement.
Connectors15 / 22 mm (standard)15 / 22 mm
BiocompatibilityMeets biocompatibility standards for externally communicating, limited duration use.Externally communicating, limited duration of use (<24 hours). Materials are Polycarbonate, Silicone, Stainless Steel Springs; passed biocompatibility testing.
MaterialsPolycarbonate, Silicone, Stainless Steel Springs (Predicate)Polycarbonate, Silicone, Stainless Steel Springs
Performance TestingAccuracy, Repeatability, Effects of Aging, Drop Test, Conical fittingsAll listed tests were performed. Results implicitly met 'acceptance criteria' by demonstrating substantial equivalence, though specific numerical results beyond accuracy were not detailed in the comparison table.

2. Sample Size Used for the Test Set and Data Provenance

The document describes non-clinical bench testing. It does not specify a "test set" in terms of patient data. The testing involves physical samples of the devices.

  • Sample Size: Not explicitly stated for each individual bench test. The nature of the tests (e.g., stroke volume, resistance, oxygen concentration, drop test, aging) implies a sufficient number of devices were tested to draw statistically valid conclusions for engineering verification.
  • Data Provenance: This is bench testing data, originating from the manufacturer's testing facilities (Foremount Enterprise Co., Ltd. in Taichung City, Taiwan). It is prospective in that the testing was performed specifically for this 510(k) submission. No patient data (retrospective or prospective) is mentioned as this is a physical device rather than an AI/software device.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This information is not applicable as this submission concerns the hardware and functional performance of medical devices (resuscitators, PEEP valves, manometers) and involves bench testing against engineering standards and predicate device performance, not clinical studies involving expert interpretation of data or images.

4. Adjudication Method for the Test Set

This information is not applicable. Adjudication typically refers to the process of resolving disagreements between experts in clinical studies or image review. Bench testing results are objective measurements against specified standards or comparative measurements against predicate devices.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a MRMC comparative effectiveness study was not done. This type of study is relevant for AI or imaging software where human readers' performance with and without AI assistance is evaluated on a set of cases. This submission is for a physical medical device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. This submission is for a physical medical device. There is no "algorithm" or standalone AI component. Performance is assessed through comprehensive bench testing of the device's physical and functional attributes.

7. The Type of Ground Truth Used

The "ground truth" for this submission is established by:

  • Engineering Standards: Primarily ISO 10651-4:2002 (Lung ventilators - Particular requirements for operator-powered resuscitators) and ISO 5356-1:2004 (Anaesthetic and respiratory equipment - Conical connectors).
  • Predicate Device Performance: The performance characteristics (e.g., stroke volume, resistance, oxygen concentration, accuracy for PEEP and manometer) of the legally marketed predicate devices (K082092 - GaleMed Dispo-Bag Manual Resuscitator and K122077 - Intersurgical - Airway Pressure Manometer) serve as a comparative benchmark.
  • Biocompatibility Standards: ISO 10993 series and ISO 18562-2 for material safety.

8. The Sample Size for the Training Set

Not applicable. This submission is for hardware medical devices. There is no concept of a "training set" as there is no AI algorithm being developed or submitted. The devices are manufactured based on design specifications and then tested.

9. How the Ground Truth for the Training Set Was Established

Not applicable. Since there is no "training set," the concept of establishing ground truth for it does not apply.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

May 11, 2018

Foremount Enterprise Co., Ltd. % Paul Dryden President ProMedic LLC. 131 Bay point Dr, NE St. Petersburg, Fl 33704

Re: K170663

Trade/Device Name: Foremount Disposable PVC Resuscitator Model A1, A2, B1, B2 and Accessories Regulation Number: 21 CFR 868.5915 Regulation Name: Manual emergency ventilator Regulatory Class: Class II Product Code: BTM Dated: April 6, 2018 Received: April 9, 2018

Dear Paul Dryden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Geeta K. Pamidimukkala -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K170663

Device Name

Foremount Disposable PVC Resuscitator Model A1, A2, B1, B2 and Accessories

Indications for Use (Describe)

Single patient use manual resuscitator for use hospital, transport, emergency, and post hospital care to temporary ventilate a patient for the given body mass ranges of: Infant: less than or equal to 10Kg. Child: less than or equal to 23 Kg. Adult: greater than 23 Kg

This manual resuscitator may be supplied with a single patient use positive end expiratory pressure (PEEP) valve and / or disposable Airway Pressure Manometer.

The PEEP Valve is a single patient use positive end expiratory pressure (PEEP) valve for use hospital. transport, emergency, and post hospital care to evaluate end lung pressure above atmospheric at the end of exhalation in constant and intermittent gas flow conditions. Intended for patients that the clinician has determined need PEEP.

The Disposable Manometer is a single patient use manometer intended to be used for monitoring the patient's airway pressure during ventilation. The manometer is to be used with resuscitation systems.

Type of Use (Select one or both, as applicable)

XX Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

This 510(k) summary is being submitted by Foremount Enterprise in accordance with 21 CFR 807.92.

Date: May 8, 2018

Submitter

Submitter's name:Foremount Enterprise Co., Ltd.
Submitter's address:No. 17, Alley 15, Lane 5, Shenan Street, Shengang Dist., Taichung City429, Taiwan
Contact person:Tyson Hsu / President
Phone Number:+886-4-2561-8788
Fax Number:+886-4-2561-8798

Name of the Device(s) and Predicate(s)

Trade Name:Foremount Disposable PVC Resuscitator Model A1, A2, B1, B2 andAccessories
Device Name:Ventilator, Emergency, Manual (Resuscitator)
Product Code:BTM
Regulation #:868.5915
Device Class:2

Predicate: K082092 - GaleMed Dispo-Bag Manual Resuscitator and accessories including Ventilation Bag, Patient Valve, Intake Valves, Reservoir Bag, Oxygen Tubing, Cushion Mask and PEEP Valve

Disposable Manometer

Predicate: K122077 - Intersurgical - Airway Pressure Manometer

Device Description:

Foremount Disposable PVC Resuscitator Model A1, A2, B1, B2 and Accessories

Foremount Disposable PVC Resuscitators are portable medical devices used to temporarily augment ventilation in patients during ventilatory insufficiency or ventilatory failure. They consist of Ventilation Bag, Patient Valve, Intake Valves, Reservoir Bag, Oxygen Tubing, Cushion Mask, and optional Diverter Ring, PEEP Valve and Manometer. Foremount Disposable PVC Resuscitators come in three sizes along with a ventilation bag:

  • . Infant - Less than or equal to 10 kg
  • Child less than or equal to 23 kg ●
  • Adult Greater than 23 kg. ●

The ventilation bags are available in three sizes based upon the intended patient population. They are provided with masks in three sizes (#1 - Infant, #3 - Child, and #5 - Adult). The patient valve includes a duck-bill valve to prevent rebreathing and incorporates a 40 cmH2O pop off valve for Child and Infant models. The patient valve includes a duck-bill valve to prevent rebreathing and incorporates a 60 cmH2O pop off valve for Adult models.

Indications for Use:

Single patient use manual resuscitator for use hospital, transport, emergency, and post hospital care to temporary ventilate a patient for the given body mass ranges of:

{4}------------------------------------------------

Infant: less than or equal to 10Kg. Child: less than or equal to 23 Kg. Adult: greater than 23 Kg

This manual resuscitator may be supplied with a single patient use positive end expiratory pressure (PEEP) valve and / or disposable Airway Pressure Manometer.

The PEEP Valve is a single patient use positive end expiratory pressure (PEEP) valve for use hospital, transport, emergency, and post hospital care to evaluate end lung pressure above atmospheric at the end of exhalation in constant and intermittent gas flow conditions. Intended for patients that the clinician has determined need PEEP.

The Disposable Manometer is a single patient use manometer intended to be used for monitoring the patient's airway pressure during ventilation. The manometer is to be used with resuscitation systems.

Substantially Equivalent Information (Predicate Device)

Foremount devices are equivalent to the following predicate devices:

DescriptionForemountPredicate
Disposable resuscitatorWithout Pop-off valveGaleMed – K082092
With Pop-off valve
With Manometer and PEEP valve
With or without face mask and oxygen collection bag
PEEP valve
Pressure manometerIntended to monitor airway pressureIntersurgical – K122077Note other resuscitators havebeen cleared with the ability toadd a manometer.

Table 1 outlines the predicate selection for each.

Table 1 - Predicate Selection

{5}------------------------------------------------

Comparison to Predicate Devices Discussion of Substantial Equivalence for the Resuscitator

Subject DevicePredicate / K082092
Models A1Model A2Models B1Model B2GaleMed
Adult, Child,AdultAdult, Child,AdultDispo-Bag Manual
and Infantand InfantResuscitator
ClassificationBTMBTM
CFR 868.5915CFR 868.5915
Ventilator, Emergency, Manual (Resuscitator)Ventilator, Emergency,Manual (Resuscitator)
Indications for UseSingle patient use manualresuscitator for use hospital,transport, emergency, andpost hospital care totemporary ventilate a patientfor the given body massranges of:Single patient usemanual resuscitator foruse hospital, transport,emergency, and posthospital care totemporary ventilate apatient for the givenbody mass ranges of:Single patient use manualresuscitator for use hospital,transport, emergency, andpost hospital care totemporary ventilate a patientfor the given body massranges of:Single patient usemanual resuscitator foruse hospital, transport,emergency, and posthospital care totemporary ventilate apatient for the givenbody mass ranges of:Single patient use manualresuscitator for usehospital, transport,emergency, and posthospital care to temporaryventilate a patient for thegiven body mass ranges of:
PopulationInfant $≤$ 10KgChild $≤$ 23 KgAdult > 23 KgAdult > 23 KgInfant $≤$ 10KgChild $≤$ 23 KgAdult > 23 KgAdult > 23 KgInfant $≤$ 10KgChild $≤$ 23 KgAdult > 23 Kg
Environment of Usehospital, transport, emergency, andpost hospital carehospital, transport,emergency, andpost hospital care
ContraindicationsInfant > 10kgChild > 23 kgAdult < 23 kgAdult < 23 kgInfant > 10kgChild > 23 kgAdult < 23 kgAdult < 23 kgInfant > 10kgChild > 23kgAdult < 23 kg
ComponentsSelf-inflating bagSelf-inflating bag
Intake valvesIntake valves
Oxygen collection bagOxygen collection bag
Oxygen tubingOxygen tubing
Patient connectorPatient connector
Face maskFace mask
Options - Pop-off, PEEP valve, Pressure manometerOptions - Pop-off, PEEPvalve
Table 2 - Comparison Disposable Resuscitator Devices
--------------------------------------------------------

{6}------------------------------------------------

510(k) Summary

Subject DevicePredicate / K082092
Models A1Adult, Child,And InfantModel A2AdultModels B1Adult, Child,and InfantModel B2AdultGaleMedDispo-Bag ManualResuscitator
Principle of operationThe patient valve contains a duckbill valve that directs air from compression of the ventilation bag through a patient connector into the patient airway during inspiration and directs the patient expired air out to the atmosphere when the ventilation bag is released during exhalation. If the patient valve incorporates a pop off valve (40 cmH2O for infant and child and 60 cmH2O for adult), excessive pressure will be exhausted to atmosphere to prevent pressure trauma.Similar
Specifications
Duration of useSingle patient, disposable <24 hoursSingle patient, disposable <24 hours
Operationaltemperature-18°C ~50°C-18°C ~50°C
Storage temperature-40°C ~60°C-40°C ~60°C
DimensionsAdult: 445x190 mmChild: 350x177 mmInfant: 325x165 mmAdult: 445x190 mmAdult: 445x190 mmChild: 350x177 mmInfant: 325x165 mmAdult: 445x190 mmAdult: 570x190 mmChild: 510x190 mmInfant: 430x180 mm
Intake valvesExternal 2 valve designIntegrated design2 valvesAll in one design
Can providesupplemental oxygenYesYes
Ventilation Bag VolumeAdult 1700 mlChild 500 mlInfant 320 mlAdult 1700 mlAdult 1700 mlChild 500 mlInfant 320 mlAdult 1700 mlAdult 1500 mlChild 600mlInfant 280 ml
Oxygen collection BagVolumeAdult 1000 mlChild 1000 mlInfant 600 mlAdult 1000 mlAdult 1000 mlChild 1000 mlInfant 600 mlAdult 1000 mlAdult 1000 mlChild 1000 mlInfant 500 ml
Max Stroke Volume(single hand)Adult 650 mlChild 370 mlInfant 180 mlAdult 650 mlAdult 650 mlChild 370 mlInfant 180 mlAdult 650 mlAdult 700 mlChild 360 mlInfant 150 ml
Dead Space~ 3.8 ml for all sizes4.5 ml for all sizes astested6.8 ml from labelingNote: All values of thepredicate were measuredvalues
Subject DevicePredicate / K082092
Models A1Model A2Models B1Model B2GaleMed
Adult, Child,AdultAdult, Child,AdultDispo-Bag Manual
and Infantand InfantResuscitator
Inspiratory ResistanceInfant - 0.5cm H2OInfant - 0.5cm H2O
MaximumChild - 0.5cm H2OChild - 0.7cm H2O
Infant@5lpmAdult - 3cm H2OAdult - 3.3cm H2O
Child@5lpm
Adult@501pm
Expiratory ResistanceInfant - 0.5cm H2OInfant -0.8cm H2O
MaximumChild - 0.5cm H2OChild - 0.8cm H2O
Adult@50 lpmAdult - 2.8 cmH2OAdult - 2.6 cmH2O
Child & Infant@5 lpm
Supplemental Oxygen%Infant2 lpm5 lpm10 lpm2/5/10 lpm
at different flow ratesVt - 70 ml x 20 bpm90%98%98%91/97/99%
and Tidal Volumes (VT)Vt – 70 ml x 30 bpm87%99%98%87/96/98%
Supplemental Oxygen%Child2 lpm5 lpm10 lpm2/5/10 lpm
at different flow ratesVt – 200 ml x 20 bpm57%99%98%56/99/99%
and Tidal Volumes (VT)Vt - 300 ml x 30 bpm39%66%98%42/63/99%
Supplemental Oxygen%Adult5 lpm10 lpm15 lpm5/10/15 lpm
at different flow ratesVt - 600 ml x 12 bpm83%99%99%84/98/99%
and Tidal Volumes (VT)Vt - 750 ml x 12 bpm57%99%99%76/98/99%
Vt - 1000 ml x 20 bpm40%60%70%40/61/71%
Pop-Off or Pressure40 cm H2O40 cm H2O
Limiting60 cm H2O60 cm H2O
Ability to add PEEPYesYes
valves0-20 cm H2O0-20 cm H2O
Patient connectors15 / 22 mm15 / 22 mm
PEEP valve fittings22 / 30 mm22 / 30 mm
Face mask#1 - Infant#2 - Child#3 - Adult#3 - Adult#1 - Infant#2 – Child#3 - Adult#3 - Adult#1 - Infant#2 - Child#3 - AdultNote: All values of thepredicate were measuredvalues
Subject DevicePredicate / K082092
Models A1Model A2Models B1Model B2
Adult, Child,and InfantAdultAdult, Child,and InfantAdult
Pressure manometerOptional0-60 cm H2O0-60 cm H2OReference IntersurgicalK122077
BiocompatibilityExternally communicating, tissue and Surface Contact, skinLimited duration of use (<24 hours)Testing – Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Gas emission VOC, PM2.5, Inorganicgases (Ozone, CO, CO2)Same patient exposure andduration
MaterialsPVC, Polycarbonate, SiliconeSame
ISO 5356-1Conical connectorsSame
ISO 10651-4Lung ventilators - operator poweredSame
Shelf-life5 years3 years

{7}------------------------------------------------

K170663

510(k) Summary

{8}------------------------------------------------

Discussion of Substantial Equivalence and Differences for the Resuscitator

The Foremount Disposable Resuscitators are viewed as substantially equivalent to the predicate device because:

Indications for Use -

The proposed indications for use are similar to the predicate. There are no differences in the Indications for use.

Patient Population -

The patient population is similar to the predicate. There is no difference in the population compared to the predicate.

Environment of Use -

The proposed environments of use are similar to the predicate. There are no differences in the environment of use.

Technology -

The design and principle of operation is similar to the configuration and functionality is similar The differences in specification, e.g. maximum stroke volume and ventilation bag volume do not raise new safety or effectiveness concerns related to substantial equivalence as ISO 10651-4 specifies the minimum requirements and the predicate both meet the minimum requirements as listed in the standard.

{9}------------------------------------------------

Substantial Equivalence for the PEEP Valves

The subject device includes the use of PEEP Valves as part of the optional accessories.

Table 3 provides a detailed comparison of the subject PEEP valves to the predicate PEEP valves which are part of the predicate K082092 GaleMed.

Subject PEEP ValvePredicate
GaleMed – K082092
ClassificationBYECFR 868.5965Attachment, Breathing, Positive EndExpiratory PressureBYECFR 868.5965Attachment, Breathing, Positive EndExpiratory Pressure
Indications for UseSingle patient use positive endexpiratory pressure (PEEP) valve foruse hospital, transport, emergency,and post hospital care to evaluateend lung pressure above atmosphericat the end of exhalation in constantand intermittent gas flow conditions.Single patient use positive endexpiratory pressure (PEEP) valve foruse hospital, transport, emergency, andpost hospital care to evaluate end lungpressure above atmospheric at the endof exhalation in constant andintermittent gas flow conditions.
Patient PopulationIntended for patients that theclinician has determined need PEEP.Not specified
Environment of UseHospital, transport, emergency, andpost hospital careHospital, transport, emergency, andpost hospital care
ContraindicationsContraindicated in a patient whodoes not require elevated endexpiratory pressure therapy.Contraindicated in a patient who doesnot require elevated end expiratorypressure therapy.
Principle of OperationExhaling exerts pressure on a springwhich has tension that is calibratedto a range of pressures. Adjustmentof the spring tension changes thePEEP pressure.Exhaling exerts pressure on a springwhich has tension that is calibrated to arange of pressures. Adjustment of thespring tension changes the PEEPpressure.
Pressure range0 to 10 cmH2O5 to 20 cmH2O2.5 to 10 cmH2O5 to 20 cmH2O
AdjustableYesYes and fixed
Accuracy+/- 2 cmH2O+/- 2 cmH2O
Connectors22 mm / 30 mm22 mm / 30 mm
BiocompatibilityExternally communicating, tissueLimited duration of use (<24 hours)Same
MaterialsPolycarbonate, Silicone, Stainless Steel springSame
Performance TestingAccuracyRepeatabilityEffects of AgingDrop TestConical fittings

Table 3 – Substantial Equivalence for PEEP Valves

{10}------------------------------------------------

510(k) Summary

Discussion of Substantial Equivalence and Differences for the PEEP Valves

The Disposable PEEP Valves are viewed as substantially equivalent to the predicate device because:

Indications for Use –

The proposed indications for use are similar to the predicate. There is no difference in the indications for use compared to the predicate.

Patient Population -

There is no specific patient population for PEEP valves. There is no difference in the population compared to the predicate.

Environment of Use –

The proposed environments of use are similar to the predicate. There is no difference in the environment of use compared to the predicate.

Technology –

The design and principle of operation is similar to the predicate. There is no difference in the technology, design or principle of operation compared to the predicate

{11}------------------------------------------------

Substantial Equivalence for the Manometer

The subject device includes the use of a disposable pressure manometer as part of the optional accessories.

Table 4 provides a detailed comparison of the subject Manometer to the Intersurgical K122077.

Table 4 – Substantial Equivalence for Manometer

SubjectPredicate
ManometerIntersurgical – K122077
ClassificationCAPCFR 868.2600Monitor, Airway PressureCAPCFR 868.2600Monitor, Airway Pressure
Indications for UseThe Disposable Manometer is a single patient use manometer is intended to be used for monitoring the patient's airway pressure during ventilation. The manometer is to be used with resuscitation systems.The Disposable Manometer is a single patient use manometer is intended to be used for monitoring the patient's airway pressure during ventilation. The manometer is to be used with resuscitation systems.
Patient PopulationIntended for use with a resuscitatorIntended for use with a resuscitator
Environment of UseHospital, transport, emergency, and post hospital careHospital, transport, emergency, and post hospital care
Principle of OperationThe technology is the use of a calibrated spring that when pressure rises the springs is compressed raising an indicator and showing the pressure via markings on the manometer cylinder has calibrated marking at 10 cmH2O intervals between 0 through 60 cmH2O.Same
Pressure range0 to 60 cmH2O0 to 60 cmH2O
MarkingsIncrements of 10 cmH2OIncrements of 10 cmH2O
Accuracy±1 cmH2O @ 0-10 cmH2O±2 cmH2O @ 20-30 cmH2O±3 cmH2O @ 40-60 cmH2O±2 cmH2O @ 0-10 cmH2O±4 cmH2O @ 20 cmH2O±5 cmH2O @ 30-40 cmH2O±7 cmH2O @ 50-60 cmH2O
Connectors15 / 22 mm15 / 22 mm
BiocompatibilityExternally communicating, tissueLimited duration of use (<24 hours)Same
MaterialsPolycarbonate, Silicone, Stainless Steel SpringsSame

{12}------------------------------------------------

Performance TestingAccuracy
Repeatability
Effects of Aging
Drop Test
Conical fittings

Discussion of Substantial Equivalence and Differences for the Manometer

The Disposable Pressure Manometer is viewed as substantially equivalent to the predicate device because:

Indications for Use -

The proposed indications for use are similar to the predicate. There is no difference in the indications for use compared to the predicate.

Patient Population -

There is no specific patient population for Manometers. There is no difference in the population compared to the predicate.

Environment of Use –

The proposed environments of use are similar to the predicate. There is no difference in the environment of use compared to the predicate.

Technology -

The design and principle of operation is similar to the predicate. There is no difference in the technology, design or principle of operation compared to the predicate.

Summary of Performance - Non-clinical for All Devices

The following tests and standards were considered in determining the safety and performance of the subject devices as compared to the predicates.

Biocompatibility of Patient Contacting Materials -

The materials of the Resuscitator bag, PEEP valve, manometer, and face mask have the following patient contact:

  • . External Communicating (Indirect gas pathway)
  • . Tissue / Bone / Dentin communicating
  • . Duration of Use - limited (<24 hours)

The face mask are also

{13}------------------------------------------------

  • Surface Contact
  • Intact skin ●
  • Duration of Use limited (<24 hours) ●

Discussion -

We performed the applicable tests to support biocompatibility. These tests included:

  • . Cytotoxicity - ISO 10993-5 (2009) - Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
  • Sensitization and Irritation ISO 10993-10 (2010) Biological evaluation of medical devices Part 10: Tests . for irritation and skin sensitization
  • Acute Systemic Toxicity ISO 10993-11 (2010) Biological evaluation of medical devices Part 11: Tests for ● systemic toxicity
  • . Gas Emission VOC testing
  • PM25 ISO 18562-2:2017 Biocompatibility Evaluation of Breathing Gas Pathways Part 2: Tests for ● Emissions of Particulate Matter
  • Inorganic gases - CO, CO2, and Ozone

The materials were found to be non-cytotoxic, non-sensitizers, and non-irritants.

Bench Testing

Applicable standards for bench testing:

  • ISO 10651-4:2002 Lung ventilators Part 4: Particular requirements for operator-powered resuscitators. .
  • . ISO 5356-1:2004 - Anaesthetic and respiratory equipment - Conical connectors - Part 1: Cones and sockets.

Tests performed included:

  • . Storage condition performance
  • Operating condition performance ●
    • o Delivered stroke volume
    • Supplementary oxygen delivered concentration at different Tidal Volumes O
    • Measurement of Inspiratory and Expiratory Resistance O
    • Patient Valve malfunction O
    • Pressure limits of Pop-off о
    • o Dead space
  • Function after Contamination with Vomitus ●
  • Drop Test ●
  • Immersion in Water ●
  • Evaluation and testing of conical fittings ●

{14}------------------------------------------------

  • Effects of aging pre- and post-conditioning performance ●
  • Accuracy and Repeatability .

Discussion of Differences and Substantial Equivalence Conclusion

There are no differences between the proposed device and the predicates which raise different safety or effectiveness concerns. We can conclude that the proposed device and accessory components can be considered substantially equivalent. Based upon the testing the sponsor has demonstrated the equivalence of the subject device compared to the predicate.

§ 868.5915 Manual emergency ventilator.

(a)
Identification. A manual emergency ventilator is a device, usually incorporating a bag and valve, intended to provide emergency respiratory support by means of a face mask or a tube inserted into a patient's airway.(b)
Classification. Class II (performance standards).